MAC 2000 ECG Analysis System Quick Guide Skin Preparation and Resting 12-lead Placement  The quality of your ECG tracing is a direct result of skin prep and lead placement. 1. Ensure that skin is dry, clean and remove excessive hair. 2. Mark each electrode site with a felt tip pen. 3. Degrease each site with a skin preparation cream. 4. Use a mild abrasion to remove the mark left by the felt tip pen. 5. Apply the electrodes to the prepared sites. 6. Look at the hook-up advisor and on-screen messages for indication of lead problems.  Label V1 V2 V3 V4 V5 V6 RA and LA RL and LL  Electrode Placement Fourth intercostal space next to the sternum on the right side. Fourth intercostal space next to the sternum on the left side. Midway between locations V2 and V4. At the mid-clavicular line in the fifth intercostal space. At the anterior axillary line on the same horizontal level as V4. At the mid-axillary line on the same horizontal level as V4 and V5. Place on the corresponding right and left wrist. An alternate placement to reduce muscle artifact is on the right and left deltoid respectively. Place on the corresponding right and left ankle. An alternate placement to reduce muscle artifact is on the upper leg as close to the torso as possible.