Covidien

iDrive Ultra sterilization tray

iDrive Stapling System IDRVULTRAx Cleaning and Sterilization Guide

2 Pages

CLEANING AND STERILIZATION GUIDE†  iDrive™ Ultra Powered Stapling System  MANUAL 1.  Wipe down the components with a lint-free cloth soaked in a pH-neutral detergent solution.  The recommended automatic cleaning cycle is as follows1,‡: Treatment  Time Temperature (MM:SS)  Chemical  00:45  Cold Tap Water  N/A  2.  Scrub all exposed surfaces with a soft brush to remove surgical debris, as required.  Pre-Wash  3.  Immerse the components in a warm enzymatic bath diluted per the manufacturer’s specifications and soak for ten minutes.  Enzyme Wash  04:00  Hot Tap Water  Enzymatic Cleaner (Prolystica™* 2X concentrate enzymatic cleaner or equivalent)  Rinse  00:15  Hot Tap Water  N/A  4.  Inspect the components for cleanliness.  5.  Manually agitate the devices in the enzymatic bath for a minimum of one minute.  6.  Rinse under warm running tap water for a minimum of one minute.  7.  Perform a final rinse under purified water for a minimum of one minute.  8.  Dry with a clean, soft, lint-free cloth.  AUTOMATIC 1.  Wipe down the components with a lint-free cloth soaked in a pH-neutral detergent solution.  2.  Scrub all exposed surfaces with a soft brush to remove surgical debris, as required.  3.  Rinse under warm running tap water for a minimum of one minute.  Wash  03:00  Hot Tap Water  pH Neutral Detergent (Prolystica™* 2X concentrate neutral detergent or equivalent pH neutral detergent)§  Rinse  00:15  Hot Tap Water  N/A  Thermal Rinse 01:00  Hot Purified Water heated to 180˚ F N/A (82˚ C)  Dry  High Setting 203˚ F N/A (95˚ C)  05:00  †Refer to user’s guide for complete instructions. ‡Tested to allow exposure to 203° F (95° C) for up to 5 minutes during the automated washer’s thermal rinse cycle to allow for variations in local requirements and regulations. (See note, Automatic Cleaning section, iDrive™ Ultra IFU). §Tested for material compatibility with cleaners that have pH ranges from neutral to 10.8 (such as Dr. Weigert neodisher™* Mediclean forte). Please refer to the cleaner’s manufacturer for information on the microbiological effectiveness of the cleaner. (See note, Automatic Cleaning section, iDrive™ Ultra IFU).
File Type: PDF
File Size: 862 KB
File Name: medtronic - 2281033 - iDrive Stapling System IDRVULTRAx Cleaning and Sterilization Guide - 2018-03.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions