CU Medical Systems
CU-SPR Instructions for Use Edition 1
Instructions for Use
70 Pages

Preview
Page 1
Instructions for Use CU-SPR The information in these Instructions for Use applies to the CU-SPR. This information is subject to change. Please contact CU Medical Systems, Inc. or its authorized representatives for information on revisions.
Revision History Edition 1 Publication Date: July 1st, 2021 Document No.: SPR-OPM-E-01 Published by: CU Medical Systems, Inc. Printed in the Republic of Korea
Copyright © 2020 CU Medical Systems, Inc. No part of these Instructions for Use may be reproduced without the permission of CU Medical Systems, Inc.
Medical Device Directive The CU-SPR complies with the requirements of the Medical Device Directive 2007/47/EC and its revisions.
Important: Quick defibrillation is needed if sudden cardiac arrest occurs. Since the chance of success is reduced by 7% to 10% for every minute that defibrillation is delayed, defibrillation must be performed promptly.
1
Contact Us Product and Order Inquiries Overseas Sales Team CU Medical Systems, Inc. 130-1, Donghwagongdan-ro, Munmak-eup, Wonju-si, Gangwon-do, Republic of Korea Tel : +82 31 421 9700/Fax: +82 31 421 9911 E-mail addresses : [email protected]
German Branch Office CU Medical Germany GmbH, Berliner Straße 44, 10713 Berlin, Germany Tel: +49 30 6781 7804 Fax: +49 30 6782 0901 E-mail addresses : [email protected]
Service and Technical Support Customer Service Team CU Medical Systems, Inc. 130-1, Donghwagongdan-ro, Munmak-eup, Wonju-si, Gangwon-do, Republic of Korea Tel : +82 31 421 9700/Fax: +82 31 421 9911 E-mail addresses : [email protected]
2
Table of contents INTRODUCTION ... 6 OVERVIEW ... 7 1.
INTRODUCTION ... 8
1.1 DEVICE DESCRIPTION ... 8 1.2 INDICATED USE ... 8 1.3 INTENDED USERS ... 9 1.4 LOCAL PROTOCOL ... 9 1.5 ADDITIONAL INFORMATION ... 9 2.
DEVICE FEATURES ... 10
3.
PREPARATION FOR USE... 13
3.1 STANDARD PACKAGE CONTENTS ... 13 3.2 SETTING UP THE CU-SPR... 14 4.
HOW TO USE THE CU-SPR ... 16
4.1 CHAIN OF SURVIVAL ... 16 4.2 PREPARATION FOR DEFIBRILLATION ... 17 4.3 DEFIBRILLATION PROCEDURES FOR ADULT VICTIM... 20 4.4 DEFIBRILLATION PROCEDURES FOR PEDIATRIC VICTIM ... 25 5.
AFTER USING THE CU-SPR ... 27
5.1 MAINTENANCE AFTER EACH USE ... 27 5.2 SAVING AND TRANSFERRING TREATMENT DATA... 27 5.2.1
Device Usage... 27
5.2.2
Transferring Treatment Data ... 28
5.3 DEVICE SETTING ... 29
6.
5.3.1
CPR Guide Setting ... 29
5.3.2
Setting the CPR Guide... 29
MAINTENANCE ... 32
6.1 DEVICE STORAGE ... 32 6.2 MAINTENANCE ... 33 6.2.1
Device Inspection ... 33
6.2.2
Replacing Supplies... 33
4
6.2.3
Cleaning the CU-SPR ... 36
7.
DISPOSAL ... 37
8.
TROUBLESHOOTING ... 38
8.1 SELF-TESTS ... 38 8.2 DEVICE STATUS ... 39 8.2.1 Status LCD... 39 8.2.2 Status LED ... 40 8.2.3 Other indicators... 40 8.3 TROUBLESHOOTING ... 41 8.3.1 Troubleshooting While the Device is Operating... 41 8.3.2 Troubleshooting While the Device is not Operating(Standby Mode) ... 42 9.
DEVICE SERVICE... 44
APPENDIX ... 47
5
Introduction These Instructions for Use contain information necessary for the correct use of this device. Please contact us regarding any questions or issues on the use of the device arising from information found in these Instructions for Use [Chapter 9: Device Service]. The company or its authorized distributor is not responsible for any injury incurred by the user or patient due to any apparent negligence or improper use by the user. Here in after, "Device" refers to [CU-SPR] "We" or “Us" refers to CU Medical Systems, Inc. "Pads" refers to defibrillation electrode pads, "Battery Pack" refers to a disposable battery pack. These Instructions for Use emphasize the safety procedures and precautions for the device use by using the terms below. Please acquaint yourself with the warnings, cautions and references stated in these Instructions for Use in order to safely use the device.
Conditions, hazards, or unsafe practices that can result in serious personal injury or loss of life.
Conditions, hazards, or unsafe practices that can result in minor or moderate personal injury, damage to the device, or loss of treatment data stored in the device, particularly if precautionary steps are not taken.
Used to denote items that are important during installation, operation, or maintenance of the device.
6
Overview Thank you for purchasing the CU-SPR. This device can be effectively and safely used for a long period if you familiarize yourself with the instructions, warnings, precautions, and notices contained in these Instructions for Use prior to its use.
You must follow the instructions, warnings, cautions, and notices in these Instructions for Use when using this device. The manufacturer will not be responsible for any problems involving the device that are caused by the user's negligence. This device shall be serviced only by the manufacturer or its authorized service centers. If the Device is intended to be connected to equipment other than those stated in these Instructions for Use, contact the manufacturer. If this Device does not operate properly, contact the manufacturer or its authorized service center.
A defibrillator discharges electric shocks with high voltage and current. You must be wellacquainted with the instructions, warnings, and precautions contained in these Instructions for Use.
7
1. Introduction 1.1 Device Description CU-SPR is an easy-to-use Semi-Automated External Defibrillator (AED) that is small, light, portable, and uses a battery. The AED automatically reads the patient's electrocardiogram (ECG) and determines if a cardiac arrest that requires defibrillation has occurred, so that both medical professionals and the general public can easily operate it. Cardiac arrest can occur anytime to anyone at any place and may threaten the patient's life if the appropriate CPR and/or electric shock with a defibrillator are not applied within a few minutes.
The CU-SPR is a semi-automated external defibrillator (AED). If connected to a patient, the CUSPR automatically acquires and analyzes the electrocardiogram (ECG) of the patient for the presence of Ventricular Fibrillation or Ventricular Tachycardia (also known as shockable rhythms). If a shockable rhythm is detected, the device automatically charges itself. Defibrillating shock is delivered when you press the SHOCK button.
The CU-SPR is easy to use. It guides you throughout a rescue operation using voice prompts and indicators (LED and graphical indicators).
The CU-SPR is small, light, highly portable, and battery powered. It is highly suitable for use in public, out-of-hospital settings.
1.2 Indicated Use The CU-SPR is indicated for use on patients that are exhibiting the symptoms of sudden cardiac arrest (SCA) with all of the following signs:
a) No movement and no response when shaken b) No normal breathing
Do not use the CU-SPR on patients who show either of the following signs:
a) Movement or response when shaken b) Presence of normal breathing
8
1.3 Intended Users The CU-SPR is intended for use in or out of the hospital by emergency care personnel or healthcare professionals or laypersons. The manufacturer recommends that users train on the use of the device.
1.4 Local Protocol Please contact your local health authority for information on the requirements of ownership and use of defibrillators.
1.5 Additional Information Please contact CU Medical Systems, Inc. or its local distributors for any additional information on the CU-SPR.
9
2. Device Features Adult/Pediatric Mode
Status LED Status LCD
Selection Button Defibrillator Pads Connector
Pads Storage Compartment
Do-Not-Touch-
CPR Step
Patient Indicator
Indicator
Early Access
Defibrillation
CPR Detection
Instruction
Indicator
Indicator
Indicator Pad Position Indicators Power Button Shock Button
IrDA Port Battery Pack
USB Port
※ Pictures shown on this document may be different from the actual device and accessories.
10
Power Button
Turns the device on or off (When the device is on, a green LED is lit)
Status LED
Displays the current status of the device.
Status LCD
Displays the current status of the device.
Shock Button
Delivers a defibrillating shock when pressed while flashing in orange.
Adult/Pediatric mode
Selects adult or pediatric mode.
selection button (i-Button)
Defibrillator Pads
Connects with the connectors of the pads.
Connector
Pads Position
Indicates the pads position on the patient.
Indicators
Early Access
Instructs immediate recognition of cardiac arrest and
Instruction Indicator
activation of the emergency response system.
Do-Not-Touch-Patient
Warns when not to touch the patient.
Indicator
Defibrillation
When the defibrillation shock is needed, instructs to deliver
Indicator
electrical shock energy.
CPR Step Indicator
Indicates the step of CPR cycle.
CPR Detection
Indicates performance of CPR on the patient.
Indicator
(The indicator is lit if CPR is performed, and flashes if CPR is not performed)
Battery Pack
The disposable power source of the device.
11
USB Port
Port for copying device records to a USB.
Pads Storage
Stores pads with removable magnetic attached cover.
Compartment
12
3. Preparation for Use 3.1 Standard Package Contents The following are the standard package contents of this device
CU-SPR Semi-automated External
Instructions for Use
Defibrillator
1 Battery Pack (Disposable)
1 Pack of Adult/Pediatric Pads (Disposable)
※ Pictures shown on this document may be different from the actual device and accessories.
Please contact the manufacturer for replacement supplies (refer to [Appendix B: Parts and Accessories] of these Instructions for Use).
13
Only parts and accessories recommended and approved by CU Medical Systems, Inc. may be used with the CU-SPR. Using parts or accessories manufactured by a third party may compromise device safety and performance. Any damage, loss, or incident resulting from the use of parts or accessories that were supplied by unauthorized distributors will not be covered by CU Medical Systems, Inc.
Extra battery packs and pads are recommended.
3.2 Setting up the CU-SPR Do the following to set up the CU-SPR ① Open the package and verify that it contains all the items listed in the packing list. ② Familiarize yourself with the device features by referring to [Chapter 2: Device Features] of these Instructions for Use. ③ Insert the battery pack into the battery compartment of the device as shown in the figure below.
As the battery pack is inserted, a beep sounds for 1 second and the power button flashes in green, the device starts a self-test according to following process.
• Battery Level Check • Pads Connector Connection Check • Pads Status Check • Button Check
14
To check the result of the self-test, please refer to [Chapter 8: Troubleshooting] of these Instructions for Use. ④ If you have a carrying case, please safely store the Device in the carrying case. If you want to purchase the carrying case, please contact us by referring to [Appendix B: Parts and Accessories] of these Instructions for Use. ⑤ Storage and maintenance considerations:
• Refer to [Section 6.1: Device Storage] for proper device storage instructions. • When the device is in storage, check the Status LCD and LED periodically to ensure that the device is in normal condition.
• Store the CU-SPR in accordance with your local emergency first aid protocol. • Store the device in an easy-to-access location where its status LCD and LED can be checked periodically and its technical alarms can be easily heard (e.g. alarm on low battery or other device problems).
• It is also recommended to place an emergency use telephone near the device’s storage area so that emergency medical services can be easily called during emergencies.
• Store the accessories along with the device in the device’s carrying case for easy and quick access.
Electromagnetic interference may affect the performance of the device. While the device is in use, it should be kept away from devices that cause electromagnetic interference. Devices that may cause such interference include motors, X-ray equipment, radio transmitters, and cell phones. Refer to [Appendix F: Electromagnetic Compatibility] of these Instructions for Use for more information. The use of accessories or cables other than those referred to in these Instructions for Use may increase electromagnetic radiation from the device or reduce the device’s electromagnetic immunity. Only accessories and cables that are authorized by the manufacturer should be used with the CU-SPR.
15
4. How to Use the CU-SPR 4.1 Chain of Survival If you think that you are witnessing someone go down in sudden cardiac arrest, perform the chain of actions recommended by the American Heart Association (AHA) in its Chain of Survival emergency response to sudden cardiac arrest.
1
2
3
4
5
6
1. Activation of Emergency Response
• Check for a response by tapping the victim on the shoulder and shouting at the victim. • Activate the community emergency response system (e.g. call 911 or the equivalent service in your locality) 2. High-Quality CPR
• Perform CPR. 3. Defibrillation.
• Use this device (CU-SPR). Using this device can be summarized in 3 steps: After pressing the Power Button, Step 1: Place pads on the patient. Step 2: Press the Shock Button if instructed by the device. Step 3: Perform CPR. 4. Advanced Resuscitation. 5. Post-Cardiac Arrest Care. 6. Recovery.
If finding and/or operating the defibrillator takes time, monitor the patient's status until the defibrillator is available, perform CPR if necessary.
16
4.2 Preparation for Defibrillation ①Turn the device on by pressing the Power Button. When the power turns ON, the following occurs in sequence: The beeper will beep for 1 seconds. The power button has a green light. Self-tests are performed. Voice instruction: “Call emergency medical services, now”
As the power is on, self-tests following status of device are performed.
• Pads Connector Connection Check • Adult/Pediatric Mode Check • Check whether the pads are used • Pads expiration date Check
To check the result of self-tests, please refer to the [chapter 8: Troubleshooting].
When the device is turned on, if X is displayed on the status LCD and the device is not working, get a replacement defibrillator immediately. If there is no replacement device, do not delay the time to care for the patient and perform CPR as soon as possible.
② Select patient mode: Adult/Pediatric Mode When the pads are connected to the device, the following occurs: Voice instruction: “Adult Mode.”
If it is need to use the device with the pediatric mode, please press the adult/pediatric mode selection button to change the patient mode.
If the Adult/Pediatric Mode Selection Button is pressed during the operation of the device, the defibrillation time might be delayed because the ECG analysis starts again. Therefore, it is necessary to carefully select the Adult/Pediatric Mode before using the device.
17
When the pediatric pads are connected to the device, the Adult/Pediatric Mode selection button will not function.
③ Remove clothes from patient's chest.
Time is essential for the cardiac arrest patient. Tear or cut clothes if removing them will take time. Dry the patient's skin such that pads can adhere well on the chest. Shave hair on the chest if necessary.
④ Take out the pads package from the Pads Storage Compartment at the bottom of the device.
18
⑤ Open the pads package.
⑥ Take pads out of the pads package.
⑦ Refer to the pictures on both pads. Adult/Pediatric Pads
The adhesive material on the pads starts to dry out as soon as the package is opened. Use immediately after opening. Refer to [Section 6.2: Maintenance] of these Instructions for Use for procedures on how to check the expiration date of the pads and pads maintenance.
19
4.3 Defibrillation Procedures for Adult victim Step 1: Place pads on the patient. ① Peel pad 1 off the liner, and stick the pad to the patient’s upper chest as shown below.
② Peel pad 2 off the liner, and stick the pad to the patient’s side torso as shown below.
③If the device detects the connection with the patient after placing the pads, the Do-Not-TouchPatient Indicator lights. Please follow the voice instruction of the device.
If the device does not detect the patient, the device repeats the voice instructions for attaching the pads to the patient until the device detects the connection with the patient after placing the pads.
In an emergency of a pediatric patient (1 to 8 years old, 10kg to 25kg), follow the Defibrillation Procedures for pediatric patients in this Instructions for Use. Defibrillation can be done even if the pads are reversed. If the locations of pads are switched, follow the next voice instruction without changing the directions of pads. It is more important to begin defibrillation as soon as possible. In the event that the pad is not adhering well, check if the adhesive side of the pads is dry. Each pad has an adhesive gel. If the gel does not adhere well, replace it with new pads.
20
Ensure the patient is not on a wet surface when performing defibrillation. If the patient's skin is wet, dry the skin prior to using the device.
Step 2: Press the Shock Button if instructed. The device acquires and analyzes the patient's ECG immediately after being connected. The device will instruct you not to touch the patient by flashing the Do-Not-Touch-Patient Indicator and by issuing the voice prompt: “Do not touch the patient, analyzing heart rhythm”. After analyzing the ECG, the device will determine whether or not the patient needs defibrillation.
Do not move or touch the patient during ECG analysis. Contact with the patient while analyzing the patient’s ECG may result in errors in the ECG analysis.
If the patient needs defibrillation, the device will do the following: The device announces that a defibrillation shock is needed, and instructs you to keep away from the patient.
While the device is charging after a shockable rhythm is detected, the ECG of the patient is continuously acquired and analyzed. The device disarms itself if the ECG rhythm changes to a non-shockable rhythm before shock delivery.
When it is charged, the device activates the following indicators in sequence: The device informs you that an electric shock is needed and instructs you to stay away from the patient. Continuously beep while the Shock Button flashes in orange. The device instructs you to press the flashing orange Shock button.
You should press the Shock Button within 15 seconds.
21