EndoWrist Stapler Stapler Release Kit Addendum
Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086 USA intuitive.com
Intuitive Surgical SAS 11 avenue de Canteranne 33600 Pessac, France
PN 553872-02 REV. A 2021.04
DRAFT/PRE-RELEASE/CONFIDENTIAL
2
Copyright ©2021 Intuitive Surgical, Inc. All rights reserved.
Trademarks Intuitive, Intuitive Surgical, Beyond the Limits of the Human Hand, da Vinci, da Vinci S, da Vinci Si, da Vinci Xi, ProGrasp, HotShears, EndoWrist, EndoPass, ClearField, CardioVac, Vacuum Source, TilePro and InSite are trademarks or registered trademarks of Intuitive Surgical, Inc. Other parties’ trademarks are the property of their respective owners and should be treated as such.
Rx only
da Vinci Xi EndoWrist Stapler
English
English Introduction
• Proper use of the da Vinci Xi EndoWrist Stapler Release Kit (PN 381215) Keep this addendum with your da Vinci Xi System User Documentation.
Overview The Stapler Release Kit (Figure 1) is intended to be used in situations when the da Vinci Xi EndoWrist Stapler jaws will not open. Opening the Stapler jaws using the Stapler Release Kit requires rotating the provided tool in up to three release holes on the instrument, per the instructions on the integrated Stapler Release Kit instruction tag. See the da Vinci Xi EndoWrist Stapler Instruments and Accessories User Manual Addendum for full instructions.
4/27/21
DRAFT/PRE-RELEASE/CONFIDENTIAL
This addendum to the da Vinci Xi EndoWrist Stapler Instruments and Accessories User Manual Addendum describes information specific to:
Figure 1 Stapler Release Kit
da Vinci Xi EndoWrist Stapler
3
4
English
Before Using the Stapler Release Kit
Release hole 2 covered by white flag
Figure 2 Release hole 2 covered by white flag If release hole 2 is not covered by the white flag (Figure 3), rotating the tool in release hole 1 is not required. Attempting to rotate the tool in release hole 1 may result in breaking the tool.
4/27/21
DRAFT/PRE-RELEASE/CONFIDENTIAL
Before using the Stapler Release Kit, examine release hole 2 on the instrument housing. If release hole 2 is covered by a white flag (Figure 2) and is not accessible, the tool must be rotated in release hole 1 first, then release holes 2 and 3.
Release hole 2 not covered by white flag
Figure 3 Release hole 2 not covered by white flag
da Vinci Xi EndoWrist Stapler
English
Guidelines for Using the Stapler Release Kit To avoid breaking the Stapler Release Kit tool: • Follow the manual stapler release instructions found in the da Vinci Xi EndoWrist Stapler Instruments and Accessories User Manual Addendum.
If the Stapler Release Kit tool breaks: • Release hole 2 and/or release hole 3 may be rotated using the tool included in the da Vinci Xi Instrument Release Kit (PN 381216). • If the tool has broken in a release hole, that hole has been fully rotated and it does not require further actuation with the tool. -End-
4/27/21
DRAFT/PRE-RELEASE/CONFIDENTIAL
• Do not continue to rotate the tool in release hole 2 after resistance is felt (up to 12 full turns).
da Vinci Xi EndoWrist Stapler