DANFLOW
DANFLOW 1100 Quick Guide
Quick Guide
2 Pages

Preview
Page 1
DANFLOW 1100 QUICK QUIDE STEP 1 – SWITCHING ON Turn ON the printer unit with the ON/OFF button on the backside of the unit. After a selftest of approx 10 seconds, the POWER LED is ON and the unit is ready. Check that the flowmeter is ON too. STEP 2 – ZEROING MODE Place an empty container on the flowmeter. During the zeriong LED MEASURE is blinking approx. 5 seconds. Now the unit is ready to record. NOTE! Do not touch the flowmeter or container during the zeroing. STEP 3 – TESTING MODE When LED MEASURE has a constant light you ask the patient to void. Normally the unit stays in waiting mode for 3 minutes but this time can be ajusted. The micturition is finished if there is no additional flow during 50 seconds. You can also manually end the recording by pressing START. STEP 4 – PRINTING THE REPORT When the recording is finished LED MEASURE turns off and LED PRINT lightening up while the report is automatically being printed. For the flowmetry report to be generated you need a volume > 10 ml and a flow > 2 ml/s. When the report is printed, the flowmeter returns to sleeping mode.
STEP 5 - PRINTING AN EXTRA COPY If you want to print an extra copy of the report, just press COPY. Remember that if the unit is turned OFF, the last recording will disappear from the internal memory. STEP 6 - FEED PAPPER You can at any time press FEED to move the paper. You can use this function to create extra space for adding notes on the report. STEP 7 - TRANSFER DATA VIA SD MEMORY CARD TO COMPUTER The unit is equipped with a 2 GB memory card where all the examinations are stored. Press the SD card so it bounces out approx 5 mm and pull it out. Place the SD card in a card reader in your computer and copy the recordings you wish to save or print on an archive resistant A4 paper. Medkonsult medical technology s. r. o., Pasteurova 67/15, 779 00, Czech Republic www.mmtsystems.com
TROUBLESHOOTING Error
Possible cause
Solution
POWER LED does not stay lit after switching ON
Disconnected from mains.
Connect device to the mains.
Wrong fuses T1A.
Change fuses in fuse box.
Wrong control unit.
Contact service center.
All LED indicators are lit after switching ON.
Error of initialization of control unit's electronics.
Switch OFF the device. Wait at least 1 minute and switch ON again. If the fault persists, contact the service center.
The instrument is not measuring or no sound signal after placing of container on wireless uroflowmeter.
Wireless uroflowmeter is switched off.
Switch ON wireless uroflowmeter by switch on its bottom side.
Battery of wireless uroflowmeter is flat.
Switch OFF and again switch ON wireless uroflowmeter by the switch on its bottom side. If there is no sound signal contact service centre.
Wrong sensor of uroflowmeter.
Contact service centre.
Results of measurement are not satisfactory.
Error during automatic setting of Repeat the test. Note that urination can zero level. not start immediately after placing the urine container on the measuring plate. Respect user manual.
Results of measurement are not printed.
Methodical mistake of test procedure.
Respect procedures according the user manual.
Conditions for valid measurement were not fulfilled
Repeat measuring.
Missing paper in printer.
Insert paper to printer. Repeat print by pressing COPY button.
Locking lever of printer is in closed position.
Check, if locking lever is properly closed. Repeat print by pressing COPY button.
SOUND SIGNALS Beep
Meaning
– •••••••••••
Initialization of sensor after switch ON.
•
Start of measurement (test begins).
-
End of measurement (test finished, data are transmitted to control unit and print of test protocol).
•••
Transit to sleep mode (device is in mode of minimal energy consumption and could be activated by changing the load on measuring surface).
••••••
Error status – attempt to start the test, although data are not sent to the control unit. The measurement can not continue in this state. It is necessary to check the state of the control unit, or the state of the radio link. It may be necessary to restart (turn OFF and turn ON) the control unit.
••
Error status - low battery (beep repeats every 60 seconds). It is possible to continue in testing, but it is necessary to order immediately a replacement of battery at service centre.