User Manual
52 Pages
Preview
Page 1
MARK II Manometric Perfusion Pump Horizontal Deck
Dentsleeve International Ltd. Manufactured by Mui Scientific 145 Traders Blvd. East, Unit 33-34 Mississauga, Ontario, CANADA L4Z 3L3 Tel: (905) 712-8828 Fax: (905) 890-3523 Toll Free: 866 751-6345 Email: [email protected] Website: www.dentsleeve.com
Handbook November, 2004
In
Index
Section In
Index
Section A
Overview
A-1 A-2
Schematic of pump Major pump components
Section B
Precautions & Warnings
Section C
Normal Use
C-1 C-2
Assembly & filling of perfusate reservoir Check, connect & set air supply
C-3 C-4 C-5 C-6 C-7
Connection & flushing of water manifold Check flow values of each hydraulic resistor Turn on water perfusion to fill each transducer Transducer calibration (if required) Connect catheter to transducers
C-8 C-9 C - 10 C - 11
Water injection procedure - each channel Check flow value for air perfusion manometry Observation & refilling of perfusate reservoir Maintain correct perfusate reservoir pressure
C - 12
Prevention, recognition and correction of hydraulic resistor blockage
C - 13
Compressor
Section D D-1 D-2 D-3 Section E
Steps on completion of measurements Perfusate reservoir Water manifold Air supply Set Up Procedures & Connection of Components
E-1 E-2 E-3
Air supply/compressor Perfusate reservoir prior to first use Installation of pressure transducers
E-4
Water perfusion manifold – removal
E-5 E-6 E-7
Standard hydraulic resistors Compact resistors Air perfusion circuit
2
In
Index
Section F
Cleaning & Disinfection
F-1 F-2
Cleaning of plastic parts Sterilization of perfusate reservoir
F-3 F-4 F-5
Sterilization of manifolds Hydraulic resistors Cleaning of compressor
Section G
Regular Maintenance
G-1
Every 3 months (or as needed) : perfusate water filter
G-2
Every year (or as needed): gas filters
G-5 G-6
Service of control box by Dentsleeve Compressor dessicant
Section H
Problem Solving
H-1
Diagnosis of abnormal Air consumption
H-2
Abnormal Air consumption - perfusate reservoir leakage
Section I
○
○
○
Specifications, Support, Spart Parts & Accessories
I-1
Specifications
I-2
Support
I-3 I-4 I-5
Spare parts Parts List Regulatory information
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
Notes on the use of this manual • References within the manual are shown in brackets eg (C – 4.2) = section C, part 4, instruction 2 • Part numbers (#) given in the text are unique for Dentsleeve • The technical information and illustrations in this manual reflect specifications and operating procedures at the time of drafting. Some specifications and operating procedures differ from earlier manuals for Mark II perfusion pump models. Dentsleeve reserves the right to vary specifications and operating procedures as part of its continuous product improvement process.
3
May 2003
○
○
Overview
Schematic of pump
100 kpa or 15 psi
A A–1
Regulated Air Supply to perfusate reservoir
Water Perfusate
Air Supply
Pump control box Perfusate Manifold
Manifold Purge Clamp
Manifold side-arm pinch valves or pinch clamps
Hydraulic resistors
3 Way stop cocks Pressure transducers Catheter connector tubes
Catheter
Note: Schematic diagram of water perfusion circuit. Only critical components are shown.
May 2003
4
A
Overview
Major pump components
A–2
Perfusate reservoir Pump control box Reservoir pressurization line
Reservoir outflow control device
Air supply gauge Manifold cover Air inlet
Pinch valves
Compressor connection plug
Compressor on/off switch
Main deck
Pinch valve or pinch clamp deck
Pressure transducer rack
Manifold flush pinch clamp Standard hydraulic resistors Transducer rack slide
Note: Pump length, channel numbers and spacings vary according to individual specifications. Compact deck version shown. Transducer types vary.
5
May 2003
!
B
Precautions & Warnings
B–1
!
Only pressurise reservoir with this line (C – 3.1).
! Air perfusion channel. Only use for pharyngeal swallow monitoring.
!
!
Never drop. Do not use if visible damage.
!
Air Pressures never more than 300 kpa (45psi).
Water level never less than 2cm (C – 10).
May 2003
6
C
Normal Use
Assembly & filling of perfusate reservoir C – 1
Note: See E – 2 for Dentsleeve part #’s Fill with particle free, degassed, distilled H20
1
2
Add water, then place perfusion float barrier on top
Blow-off valve
Perfusate outlet hole
‘O’ ring
Reservoir outflow control device
! !
Always use diffusion float barrier
○
○
○
3
7 May 2003
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
Tighten firmly to seal
May 2003
○
C
Normal Use
For set-up and first use see E – 1 to E – 9
C–2
Check, connect & set air supply
1
Push together
Pull back to release
Click!
Air ○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
2
Air Set to 100 kPa - 15 psi
3
At least 1000 kpa (150 psi) in high pressure Air supply cylinder
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
May 2003 8
Normal Use
Connection & flushing of water manifold
1
Push to seal onto reservoir
Part #: ZP1FIT/SW/B-QM2-B-200ZN
2
Connect to water manifold
Reservoir support posts
3
Air line from control box
Open outflow control
4
Open to vent bubbles
Manifold venting pinch clamp
9
May 2003
C C–3
C C–4
Normal Use
Check flow values of each hydraulic resistor
1
Flow ml/min
1
Standard resistor shown: flow is colour coded
Tube Colour
Screw Colour
0.01
Black
White
R01HRE/ST/1(2)
0.02
Black
Black
R01HRE/ST/2(2)
0.04
Red
Red
R01HRE/ST/4(2)
0.08
Red
White
R01HRE/ST/8(2)
0.15
Red
Black
R01HRE/ST/15(2)
0.3
Green
White
R01HRE/ST/30(2)
0.45
Green
Red
R01HRE/ST/45(2)
0.6
Green
Black
R01HRE/ST/60(2)
Alternative: Compact hydraulic resistor
Part #
Water flow on case
2
The flow value is...
• Correct, go to
C–5
• Changed, go to
E–5 or 6 May 2003
10
Normal Use
C
Turn on water perfusion to fill each transducer
1
Open pinch valve or clamp
2
Fill transducers
2
Fill transducers
Prefered position for 3 way stopcock - other than PVB DPT-6100 Note: PVB DPT-6100 Transducer shown
11 May 2003
Note: Abbott Transpac 42582-10 Transducer Shown
C–5
C
Normal Use
C–6
Transducer calibration (if required)
1
Turn perfusion off - (C – 5)
2
Set tranducer gain by applying standard external pressure
Shows AP1CMA/TR/2
Connect to transducers Note: Shows Dentsleeve calibration manifold; made according to channel numbers and spacings Part #’s AP1CMA/TR/1to5
Pressurise with syringe
C–7
Connect catheter to transducers
1
Select correct channel
2
Twist & push on firmly
Note: Catheter detail is for Dentsleeve product
May 2003
12
Normal Use
C
Water injection procedure - each channel
1
H2O - no bubbles
3
Inject 5ml of H2O
13 May 2003
2
Open side-port
C–8
C C–8
Normal Use
Water injection procedure - each channel (cont)
4
Close side-port then remove syringe
Note: Arrangement for PVB DPT-6100 transducer shown
May 2003 14
C
Normal Use
Check flow value for air perfusion manometry
1
C–9
Identify air perfusion channel
Red Band
! Only use air perfusion for detection of swallowing in pharynx
2
Ensure resistor for air perfusion channel is correct
!
Air flow rate through hydraulic resistor is x100 water flow rate.
Water flow ml/min
Part #
0.02
Black
Black
R01HRE/ST/2(2)
0.04
Red
Red
R01HRE/ST/4(2)
0.08
Red
White
R01HRE/ST/8(2)
Suitable compact resistors give airflow on case
! Must be installed between manifold and transducer to limit air flow into catheter to less than 10ml/min
15
May 2003
Air flow ml/min
Water Flow ml/min
Part #
2
0.02
R01HRE/CO/2
4
0.04
R01HRE/CO/4
8
0.0
R01HRE/CO/8 8
C
Normal Use
C – 10
Observation & refilling of perfusate reservoir
If perfusate exhausted, large volumes of gas may be perfused down catheter
1
Refill when at 2cm
2
Release pressure
Pull Back
3
Refill
Particle free, degassed, distilled H 20
4
Restart perfusion
Tighten
Repressurise
May 200316
Normal Use
C
Maintain correct perfusate reservoir pressure
C – 11
Monitor air pressure. Maintain at 100kPa/15psi.
1
150
100 15
20 25
10 50
0
2
200
30 35
5 psi kPa
250
Check if air pressure drops
3
Check that air supply is adequate
300 200
400 40 60 20
100
500 80
0
psi kPa
600
OR At least 130 kPa/ 20psi
At least 1000 kPa/ 150psi
17 May 2003
C C – 12
Normal Use
Prevention, recognition and correction of hydraulic resistor blockage
1
Prevention of blockages - Always use perfusate filter
2
Minimise perfusate particle load
Double-distilled or de-ionised water
No particles in reservoir (E-2) Perfusate filter Part # AP1FIL/WA/1
3
4
Replace water perfusate filter regularly - (G-1)
Check perfusate flow rates - assemble 1ml syringe barrel on transducer
1ml syringe
5
Start perfusion - (C-5)
6
Measure time to deliver water
Female/Female adapter Part # ZP1FIT/VP/FTLLC-9
Transducer on rack
Note: for resistor flow rates < 0.15ml/min use 0.25 - 0.50 ml syringe barrels
May 2003
18
C
Normal Use
Prevention, recognition & correction of hydraulic resistor blockage (cont.)
7
Flushing of blocked resistors - remove resistor from circuit
Transducer end
8
Backflush forcibly
9
Observe water flow
OR 11
Remove resistor from circuit
12
Insert flush support tool - (E – 6)
13 Same steps as C – 12.7 to 12.10 above
19 May 2003
10
Re-install - (E – 5)
C – 12
C
Normal Use
C – 13
Compressor
1
Always earth
2
Turn on power switch. Power is on when switch is illuminated
3
Connect compressor air outlet to pump control box
4
Plug in compressor to pump control
5
Press compressor on/off switch. Functioning normally when switch is illuminated
May 2003 20