Fabius Family Breathing System Cleaning Guide 2012.pdf
Page 1
Fabius Family
Breathing System Cleaning Guide
Page 2
02 I
Fabius Family Breathing System Cleaning Guide
CAUTION
To reduce the risk of infection to both hospital staff and patients, clean and disinfect
medical devices after each use. Protective clothing, eye protection etc. must be
worn.
Warning! This booklet is not a replacement for the Instructions for Use. For full
information concerning the performance characteristics of the Dräger devices
described in the booklet, each user must carefully read and fully comprehend the
Instructions For Use before operating the specfic device.
Page 3
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
COSY Breathing System
CLEANING - Qty 1
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
•
Hot Steam sterilization is required to dry
the breathing system
CLIC Adapter
CLEANING - Qty 1 each
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
•
NB. Place black disc into clic adapter prior to
cleaning
COMPONENTS
03 I
Page 4
04 I
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
Inspection Caps
CLEANING - Qty 2 pieces
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
Retaining Nuts
CLEANING - Qty 2
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
Unidirectional Valve Discs
CLEANING - Qty 2
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
COMPONENTS
Page 5
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
Unidirectional Valve ‘O’ rings
CLEANING - Qty 2
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
Expiratory Port
CLEANING - Qty 1
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
APL Valve
CLEANING - Qty 1
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
COMPONENTS
05 I
Page 6
06 I
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
Ventilator Diaphragm
CLEANING - Qty 1
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
Ventilator Lid
CLEANING - Qty 1
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
Flow Sensor
CLEANING - Qty 1
•
•
•
•
Disinfect the flow sensor by immersion.
Sufficiently rinse the flow sensor under water
(preferably drinking-water quality) until no
disinfectant residues can be recognized.
Inspect flow sensor for visible soiling and
damage (measuring wires and corresponding
pins). Repeat manual disinfection if necessary.
Remove remaining water and allow the flow
sensor to air for at least 30 minutes.
COMPONENTS
Page 7
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
Ventilator Hose and Water Trap
CLEANING - Qty 3 pieces
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
Scavenge Hose
CLEANING - Qty 1
•
Machine clean and thermal disinfection
(preferably anaesthesia program)
COMPONENTS
07 I
Page 8
Fabius Family Breathing System cleaning Guide / V1.1 / 20.07.12 / Printed in Australia / Subject to modifications / © 2012 Draeger Medical Australia Pty Ltd
Draeger Medical Australia Pty Ltd
Unit 97, Axxess Corporate Park
45 Gilby Road
Mt Waverley VIC 3149
Tel: 1800 800 327
Fax: 1800 010 327
E-mail: med@draeger.com