8 Pages

Fabius Family Breathing System Cleaning Guide   02 I  Fabius Family Breathing System Cleaning Guide  CAUTION To reduce the risk of infection to both hospital staff and patients, clean and disinfect medical devices after each use... rotective clothing, eye protection etc... ust be worn... Warning! This booklet is not a replacement for the Instructions for Use... or full information concerning the performance characteristics of the Dräger devices described in the booklet, each user must carefully read and fully comprehend the Instructions For Use before operating the specfic device...
File Type: PDF
File Size: 845KB
File Name: fabius_family_breathing_system_cleaning_guide_2012.pdf
Download

If your hospital with access, please login via your intranet.

Contact Us to Register Your InterestAlready a member? Log in hereAbout Us

fabius_family_breathing_system_cleaning_guide_2012.pdf

Page 1

Fabius Family

Breathing System Cleaning Guide

Page 2

02 I

Fabius Family Breathing System Cleaning Guide

CAUTION

To reduce the risk of infection to both hospital staff and patients, clean and disinfect

medical devices after each use. Protective clothing, eye protection etc. must be

worn.

Warning! This booklet is not a replacement for the Instructions for Use. For full

information concerning the performance characteristics of the Dräger devices

described in the booklet, each user must carefully read and fully comprehend the

Instructions For Use before operating the specfic device.

Page 3

Fabius Family Breathing System Cleaning Guide

FABIUS FAMILY BREATHING SYSTEM CLEANING

COSY Breathing System

CLEANING - Qty 1

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

•

Hot Steam sterilization is required to dry

the breathing system

CLIC Adapter

CLEANING - Qty 1 each

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

•

NB. Place black disc into clic adapter prior to

cleaning

COMPONENTS

03 I

Page 4

04 I

Fabius Family Breathing System Cleaning Guide

FABIUS FAMILY BREATHING SYSTEM CLEANING

Inspection Caps

CLEANING - Qty 2 pieces

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

Retaining Nuts

CLEANING - Qty 2

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

Unidirectional Valve Discs

CLEANING - Qty 2

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

COMPONENTS

Page 5

Fabius Family Breathing System Cleaning Guide

FABIUS FAMILY BREATHING SYSTEM CLEANING

Unidirectional Valve ‘O’ rings

CLEANING - Qty 2

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

Expiratory Port

CLEANING - Qty 1

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

APL Valve

CLEANING - Qty 1

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

COMPONENTS

05 I

Page 6

06 I

Fabius Family Breathing System Cleaning Guide

FABIUS FAMILY BREATHING SYSTEM CLEANING

Ventilator Diaphragm

CLEANING - Qty 1

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

Ventilator Lid

CLEANING - Qty 1

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

Flow Sensor

CLEANING - Qty 1

•

•

•

•

Disinfect the flow sensor by immersion.

Sufficiently rinse the flow sensor under water

(preferably drinking-water quality) until no

disinfectant residues can be recognized.

Inspect flow sensor for visible soiling and

damage (measuring wires and corresponding

pins). Repeat manual disinfection if necessary.

Remove remaining water and allow the flow

sensor to air for at least 30 minutes.

COMPONENTS

Page 7

Fabius Family Breathing System Cleaning Guide

FABIUS FAMILY BREATHING SYSTEM CLEANING

Ventilator Hose and Water Trap

CLEANING - Qty 3 pieces

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

Scavenge Hose

CLEANING - Qty 1

•

Machine clean and thermal disinfection

(preferably anaesthesia program)

COMPONENTS

07 I

Page 8

Fabius Family Breathing System cleaning Guide / V1.1 / 20.07.12 / Printed in Australia / Subject to modifications / © 2012 Draeger Medical Australia Pty Ltd

Draeger Medical Australia Pty Ltd

Unit 97, Axxess Corporate Park

45 Gilby Road

Mt Waverley VIC 3149

Tel: 1800 800 327

Fax: 1800 010 327

E-mail: med@draeger.com

www.draeger.com

Previous Page Next Page
© 2021 BioClinical Services Ptd Ltd
Terms of Use
Privacy Statement