Dräger Zeus® Breathing System
Disassembly/Assembly
Disassembly
90 67 069 | 05.12-1 | Marketing Communications | BM | PR | LE | Printed in Germany | Chlorfrei – umweltfreundlich | Änderungen vorbehalten | © 2014 Drägerwerk AG & Co. KGaA
1
Screw the CLIC adapter (optional).
4
5
3
6
Check the inspiration grill.*
Insert the flow sensors
8
Loosen and pull out the expiration port.
Check the expiration grill.*
Insert and tighten the expiration port.
7
Remove the flow sensors.
5
Lift off the upper part of the breathing system.
Set the upper part on the lower
part of the breathing system.
Insert cleaning disc into the adapter.
Close adapter until it snaps.
Remove cleaning disc from the adapter.
Close adapter until it snaps.
8
Loosen and pull out the inspiration port.
Insert and tighten the inspiration port.
7
2
Unscrew the CLIC adapter (optional).
Turn the screws using the key
counterclockwise about 120°.
Turn the screws using the key
clockwise about 120°.
4
Screw off the APL valve.
2
Screw in the APL valve.
6
1
Assembly
* If the grill is damaged, contact DrägerService.
Reprocessing instruction
Sterilization 134 °C (273.2 °F)
Machine cleaning and disinfection
Immersion desinfection
Please note:
SpiroLife® flow sensors are not suitable for plasma or
radiation sterilization.
Recommendation:
Recommendation:
Position components so that all interior spaces and surfaces
are completely flushed and water can drain off freely.
Please use the disinfectants listed in the
Instructions for use of the Dräger flow sensors.