Dräger Zeus® Breathing System Disassembly/Assembly  Disassembly 90 67 069 | 05... -1 | Marketing Communications | BM | PR | LE | Printed in Germany | Chlorfrei – umweltfreundlich | Änderungen vorbehalten | © 2014 Drägerwerk AG & Co... GaA  1  Screw the CLIC adapter (optional)... 4  5  3  6  Check the inspiration grill... Insert the flow sensors  8  Loosen and pull out the expiration port... Check the expiration grill... Insert and tighten the expiration port... 7  Remove the flow sensors... 5  Lift off the upper part of the breathing system... Set the upper part on the lower part of the breathing system... Insert cleaning disc into the adapter... lose adapter until it snaps... Remove cleaning disc from the adapter... lose adapter until it snaps... 8  Loosen and pull out the inspiration port... Insert and tighten the inspiration port... 7  2  Unscrew the CLIC adapter (optional)... Turn the screws using the key counterclockwise about 120°... Turn the screws using the key clockwise about 120°... 4  Screw off the APL valve... 2  Screw in the APL valve... 6  1  Assembly * If the grill is damaged, contact DrägerService... Reprocessing instruction Sterilization 134 °C (273... °F)  Machine cleaning and disinfection  Immersion desinfection  Please note: SpiroLife® flow sensors are not suitable for plasma or radiation sterilization... Recommendation:  Recommendation:  Position components so that all interior spaces and surfaces are completely flushed and water can drain off freely... Please use the disinfectants listed in the Instructions for use of the Dräger flow sensors... 3
File Type: PDF
File Size: 1MB
File Name: Drager Zeus Breathing System Disassembly and Assembly Guide Aug 2014.pdf
Download

If your hospital with access, please login via your intranet.

Contact Us to Register Your InterestAlready a member? Log in hereAbout Us

Drager Zeus Breathing System Disassembly and Assembly Guide Aug 2014.pdf

Page 1

Dräger Zeus® Breathing System

Disassembly/Assembly

Disassembly

90 67 069 | 05.12-1 | Marketing Communications | BM | PR | LE | Printed in Germany | Chlorfrei – umweltfreundlich | Änderungen vorbehalten | © 2014 Drägerwerk AG & Co. KGaA

1

Screw the CLIC adapter (optional).

4

5

3

6

Check the inspiration grill.*

Insert the flow sensors

8

Loosen and pull out the expiration port.

Check the expiration grill.*

Insert and tighten the expiration port.

7

Remove the flow sensors.

5

Lift off the upper part of the breathing system.

Set the upper part on the lower

part of the breathing system.

Insert cleaning disc into the adapter.

Close adapter until it snaps.

Remove cleaning disc from the adapter.

Close adapter until it snaps.

8

Loosen and pull out the inspiration port.

Insert and tighten the inspiration port.

7

2

Unscrew the CLIC adapter (optional).

Turn the screws using the key

counterclockwise about 120°.

Turn the screws using the key

clockwise about 120°.

4

Screw off the APL valve.

2

Screw in the APL valve.

6

1

Assembly

* If the grill is damaged, contact DrägerService.

Reprocessing instruction

Sterilization 134 °C (273.2 °F)

Machine cleaning and disinfection

Immersion desinfection

Please note:

SpiroLife® flow sensors are not suitable for plasma or

radiation sterilization.

Recommendation:

Recommendation:

Position components so that all interior spaces and surfaces

are completely flushed and water can drain off freely.

Please use the disinfectants listed in the

Instructions for use of the Dräger flow sensors.