Evita V800-V600-V500-V300 Flow Reduction InService Guide.pdf
Page 2
Flow reduction
– Anti-air shower
– Manual disconnection
Footnote: Please insert appropriate company
Page 3
Manual
disconnection
August 2019, Lübeck
Drägerwerk AG & Co KGaA
Footnote: Please insert appropriate company
Page 4
Manual disconnection
Evita V800 / Evita V600
Evita V800, Evita V600
When the Manual disconnection function is activated, the device reduces
flow delivery in case of disconnection.
The patient can be disconnected within the next 10 seconds.
When the patient is reconnected, the device resumes ventilation in the
set ventilation mode.
1.
To activate manual disconnection touch the Procedures button
2.
Touch the Manual disconnection button
1
2
3. Confirm by pressing the rotary knob
3
Drägerwerk AG & Co KGaA
4
Page 5
Manual disconnection
Evita Infinity V500
Evita Infinity V500
When the Manual disconnection function is activated, Evita V500 reduces
flow delivery in case of disconnection.
The patient can be disconnected within the next 10 seconds.
When the patient is reconnected, the device resumes ventilation in the
set ventilation mode.
1. Touch the Special procedures... button in the main menu bar.
The device opens the Special procedures dialog window
1
2
2. Touch the Functions button if the page is not already preset
3. Touch the Manual discon. button (B) and
4
4. Confirm by pressing the rotary knob
The Manual disconnection maneuver can be configured as a Manual
discon. button in the main menu bar to enable direct access.
Drägerwerk AG & Co KGaA
3
5
Page 6
Manual disconnection
Evita V300
Evita V300
When the Manual disconnection function is activated, Evita V300 reduces
flow delivery in case of disconnection.
The patient can be disconnected within the next 10 seconds.
When the patient is reconnected, the device resumes ventilation in the
set ventilation mode.
1. To activate touch the Special maneuvers button in the main menu bar.
Evita V300 opens the Special maneuvers dialog window
1
2. To activating manual disconnection touch the Manual discon. button
3. Confirm by pressing the rotary knob
The Manual disconnection maneuver can be configured as a Manual
discon. button in the main menu bar to enable direct access.
Drägerwerk AG & Co KGaA
2
4
6
Page 7
Anti-Air Shower
August 2019, Lübeck
Drägerwerk AG & Co KGaA
Footnote: Please insert appropriate company
Page 8
Anti-air shower
Evita V800 / Evita V600
Evita V800, Evita V600
Flow reduction – Anti-air shower
When the Anti-air shower function is activated and a disconnection is
detected during ventilation, the flow is reduced until reconnection is
detected.
Simultaneously, the “Disconnection detected” alarm is displayed.
1
2
For non-invasive ventilation, the time before the alarm is triggered can be
delayed with the setting Tdiscon. The minute volume may be reduced
due to the already reduced flow.
The Anti-air shower function can be configured on the dialog page
System setup > Ventilation > General settings
The password must be entered.
The settings can be activated or deactivated. Ventilation starts with the
selected settings.
Drägerwerk AG & Co KGaA
3
8
Page 9
Anti Air Shower
Evita Infinity V500 / Evita V300
Flow reduction Anti Air Shower
If a disconnection is detected, flow delivery is reduced until after
reconnection.
1
1. To activate manual disconnection touch System setup …
2. Touch the Ventilation page button (A)
3. Touch the General settings tab (B)
2
4. Switch Anti Air Shower (H) on or off
3
5. Confirm by pressing the rotary knob
5
Evita V500 starts with the selected settings.
4
Drägerwerk AG & Co KGaA
9
Page 10
Many thanks
Footnote: Please insert appropriate company