How to clean the Dräger Caleo®  !  Please reprocess the device according to your hospital’s guidelines... his document is for information purposes only... lease refer to the device instructions for use for complete disassembly, assembly, cleaning and surface disinfection... nly use disinfectants as listed in the instructions for use... nsure that no liquid enters the device... Prerequisites: There must be no patient in the device... llow device to cool for 30 minutes prior to disassembly... 1  Surface disinfection for: • • • •  Disassembly ►  Basic device/trolley All disassembled parts Hood and side walls (interior and exterior) X-ray valve air channels: Pull a cloth soaked in disinfectant through each air channel  Set the incubator to the lowest position... 2  C  lose and remove the transfer set... Machine cleaning and thermal disinfection: • Grommets • Fan impeller • Water tank (option)  Before starting use, ensure that all surfaces are dry and free from disinfectant residues...   Remove sterile water bag or water tank (option)... 16 Connect water tank  or sterile water bag... 4  Remove upper double wall and continuous seal... 15 Apply continuous seal and  refit upper double wall... 18 Set the incubator to the  desired working height... 5  Remove hood... 14 Refit hood... 17 Connect and open  the transfer set... 6  Cleaning mode... 13   ◄ Assembly   Disassembly ► Open side walls... 8  Remove grommets... 9  Open head and foot access ports... 10 Remove mattress... 12 Close side walls... 11 Insert grommets... 10 Close head and foot  9  11 Remove mattress tray... 12 Remove hose holder arm  13 Unlock and remove  14 Remove trough... 8  Insert mattress tray... 15 Remove fan impeller  and connecting piece... 4  Insert fan impeller and connecting piece... (option)... 7  Insert hose holder arm (option)... 16 Remove drawer... 3  Insert drawer... access ports... x-ray tray... 6  Insert and lock x-ray tray... 17 Clean and disinfect  basic device in full... 2    5  Insert mattress... Insert trough... 18 Remove display unit rotary  knob and clean and disinfect display unit... 1  Insert display unit rotary knob... ◄ Assembly  91 08 695 | 2004-1 | NW | GT | Subject to modifications | © 2020 Drägerwerk AG & Co... GaA  7
File Type: PDF
File Size: 2MB
File Name: Caleo How to Clean Guide .pdf
Download

If your hospital with access, please login via your intranet.

Contact Us to Register Your InterestAlready a member? Log in hereAbout Us

Caleo How to Clean Guide .pdf

Page 1

How to clean the Dräger Caleo®

!

Please reprocess the device according to your hospital’s guidelines. This document is for information

purposes only. Please refer to the device instructions for use for complete disassembly, assembly,

cleaning and surface disinfection. Only use disinfectants as listed in the instructions for use.

Ensure that no liquid enters the device.

Prerequisites: There must be no patient in the device.

Allow device to cool for 30 minutes prior to disassembly.

1

Surface disinfection for:

•

•

•

•

Disassembly ►

Basic device/trolley

All disassembled parts

Hood and side walls (interior and exterior)

X-ray valve air channels: Pull a cloth soaked

in disinfectant through each air channel

Set the incubator to

the lowest position.

2

C

 lose and remove

the transfer set.

Machine cleaning and thermal disinfection:

• Grommets

• Fan impeller

• Water tank (option)

Before starting use, ensure that all surfaces

are dry and free from disinfectant residues.

3

Remove sterile water bag

or water tank (option).

16 Connect water tank

or sterile water bag.

4

Remove upper double wall

and continuous seal.

15 Apply continuous seal and

refit upper double wall.

18 Set the incubator to the

desired working height.

5

Remove hood.

14 Refit hood.

17 Connect and open

the transfer set.

6

Cleaning mode.

13 

◄ Assembly

Page 2

Disassembly ►

Open side walls.

8

Remove grommets.

9

Open head and foot

access ports.

10 Remove mattress.

12 Close side walls.

11 Insert grommets.

10 Close head and foot

9

11 Remove mattress tray.

12 Remove hose holder arm

13 Unlock and remove

14 Remove trough.

8

Insert mattress tray.

15 Remove fan impeller

and connecting piece.

4

Insert fan impeller and

connecting piece.

(option).

7

Insert hose holder arm

(option).

16 Remove drawer.

3

Insert drawer.

access ports.

x-ray tray.

6

Insert and lock x-ray tray.

17 Clean and disinfect

basic device in full.

2



5

Insert mattress.

Insert trough.

18 Remove display unit rotary

knob and clean and

disinfect display unit.

1

Insert display unit

rotary knob.

◄ Assembly

91 08 695 | 2004-1 | NW | GT | Subject to modifications | © 2020 Drägerwerk AG & Co. KGaA