infinity_acute_care_system_models_c700_and_m540__in.pdf
Page 1
.
Instructions for use
Infinity Acute Care System
WARNING
To properly use this medical device,
read and comply with these instructions
for use.
Monitoring Applications
Software VG2
Page 2
Typographical conventions
1
Consecutive numbers indicate steps of action,
with the numbering restarting with "1" for each
new sequence of actions.
z Bullet points indicate individual actions or different options for action.
–
Dashes indicate the listing of data, options, or
objects.
(A) Letters in parentheses refer to elements in the
related illustration.
A Letters in illustrations denote elements referred
to in the text.
Any text shown on the screen and any labeling on
the device are printed in bold and italics, for example, Alarms, or Trends/ Data....
The "greater than" symbol > indicates the navigation path in a dialog window, for example, System
setup... > Screen setup > General settings. In
this example, System setup... represents the dialog window title, Screen setup represents a horizontally aligned tab, and General settings a vertically aligned tab.
Screen images
Schematic renderings of screen images are used,
which may differ in appearance or in configuration
from the actual screen images.
2
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 3
Trademarks
–
Acute Care SystemTM
–
OxiMaxTM
–
Babylog®
–
SatSecondsTM
–
DrägerService®
–
Evita®
are trademarks of Nellcor Puritan Bennett, LLC, a
Covidien company
–
Hemo4®
–
–
Hemo2®
–
Infinity®
is a registered trademark of ViewSonic Corporation
in the U.S. and other countries.
–
Innovian®
–
MCableTM
–
Medical Cockpit®
–
MegaCareTM
–
MPod®
–
PatientWatchTM
–
TruST®
–
MonoLead®
–
Perseus A500
are trademarks of Dräger.
–
Masimo®
–
Masimo® SET® (Signal Extraction Technology)
–
Masimo® rainbow® SET®
–
PVI®
–
SET®
–
SpCO®
–
SpHb®
–
SpMet®
–
SpOCTM
–
Pulse CO-Oximeter TM
–
M-LNCS TM
ViewSonic®
–
Edwards Vigilance®
–
Edwards Vigileo®
are trademarks of Edwards Lifesciences, LLC.
–
ComplianceTM
is a trademark of Cardinal Health, Inc.
–
Dismozun®pur
is a trademark of BODE Chemie GmbH.
–
Sporox II®
is a trademark of Sultan Healthcare, Inc.
–
SERVO-i®
is a trademark of Maquet, Inc.
are trademarks of the Masimo corporation.
–
NellcorTM
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
3
Page 4
Safety information definitions
WARNING
A WARNING statement provides important information about a potentially hazardous
situation which, if not avoided, could result in
death or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially hazardous situation
which, if not avoided, may result in minor or moderate injury to the user or patient or in damage to
the medical device or other property.
NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.
Abbreviations and symbols
For explanations, refer to the sections
""Abbreviations" on page 27" and "Device symbols" on page 25".
4
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 5
Contents
Contents
For your safety and that of your patients. . .
9
General safety information . . . . . . . . . . . . . . . . 14
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Intended use – Infinity Acute Care System –
Monitoring Applications . . . . . . . . . . . . . . . . . . 18
System overview . . . . . . . . . . . . . . . . . . . . . . 19
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Infinity Medical Cockpit (Cockpit) . . . . . . . . . . .
Infinity PS250 Comm Hub (PS250) power
supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Infinity P2500 power supply (P2500) . . . . . . . .
Infinity M540 patient monitor (M540) . . . . . . . .
Infinity M500 docking station (M500) . . . . . . . .
Additional hardware . . . . . . . . . . . . . . . . . . . . .
Device symbols . . . . . . . . . . . . . . . . . . . . . . . .
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . .
20
20
21
21
22
22
23
25
27
Operating concept . . . . . . . . . . . . . . . . . . . . . 33
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The IACS components . . . . . . . . . . . . . . . . . . .
M540 and Cockpit communication . . . . . . . . . .
Communicating with the Infinity network . . . . .
Remote control and remote view . . . . . . . . . . .
Communication management. . . . . . . . . . . . . .
Loss of power . . . . . . . . . . . . . . . . . . . . . . . . . .
Locked options . . . . . . . . . . . . . . . . . . . . . . . . .
Secondary display . . . . . . . . . . . . . . . . . . . . . .
Export protocol . . . . . . . . . . . . . . . . . . . . . . . . .
User interface . . . . . . . . . . . . . . . . . . . . . . . . . .
Header bar . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Monitoring area. . . . . . . . . . . . . . . . . . . . . . . . .
Supported banners . . . . . . . . . . . . . . . . . . . . . .
Main menu bar and quick access toolbar . . . . .
Filtering the parameter content. . . . . . . . . . . . .
Auto and manual display modes . . . . . . . . . . .
Auto view setup toolbar . . . . . . . . . . . . . . . . . .
Customizing the display . . . . . . . . . . . . . . . . . .
Parameter priority . . . . . . . . . . . . . . . . . . . . . .
Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Standby mode. . . . . . . . . . . . . . . . . . . . . . . . . .
34
35
36
38
40
42
44
44
45
45
46
47
48
51
53
55
55
56
56
60
61
62
63
Privacy mode. . . . . . . . . . . . . . . . . . . . . . . . . .
64
Assembly and preparation . . . . . . . . . . . . . .
65
Assembly overview . . . . . . . . . . . . . . . . . . . . . 66
Docking/undocking the M540 . . . . . . . . . . . . . 67
Locking/unlocking the M540 . . . . . . . . . . . . . . 68
Connecting the system cables . . . . . . . . . . . . 69
Mounting the Infinity MCable – Masimo
SET/Masimo rainbow SET/ Nellcor OxiMax . . 71
Getting started . . . . . . . . . . . . . . . . . . . . . . . .
73
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Turning the IACS on/off . . . . . . . . . . . . . . . . . . 74
Viewing demographic data . . . . . . . . . . . . . . . 75
Admitting a patient. . . . . . . . . . . . . . . . . . . . . . 76
Discharging a patient. . . . . . . . . . . . . . . . . . . . 77
Patient categories . . . . . . . . . . . . . . . . . . . . . . 78
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
81
Overview of alarms . . . . . . . . . . . . . . . . . . . . .
Alarm priorities . . . . . . . . . . . . . . . . . . . . . . . .
Alarm processing. . . . . . . . . . . . . . . . . . . . . . .
Activating or deactivating alarm validation . . .
Visual alarm signals . . . . . . . . . . . . . . . . . . . .
Acoustic alarm signals. . . . . . . . . . . . . . . . . . .
Testing visual and acoustic alarm signals . . . .
Viewing current alarm messages . . . . . . . . . .
Special alarm behavior . . . . . . . . . . . . . . . . . .
Pre-silencing alarms . . . . . . . . . . . . . . . . . . . .
Pausing acoustic alarm signals . . . . . . . . . . . .
Activating or deactivating acoustic alarm
signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pausing alarm monitoring temporarily. . . . . . .
Activating or deactivating alarm monitoring . .
Configuring the alarm settings for a patient. . .
Configuring the alarm setup for an individual
parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuring the alarm setup for multiple
parameters . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuring the arrhythmia alarm setup . . . . .
Alarm setup for ST . . . . . . . . . . . . . . . . . . . . .
Auto setting all alarm limits . . . . . . . . . . . . . . .
Alarm history and stored events . . . . . . . . . . .
Viewing the alarm history . . . . . . . . . . . . . . . .
83
84
84
86
87
89
90
91
91
93
94
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
95
95
96
97
98
100
102
104
106
107
108
5
Page 6
Contents
Viewing a snapshot of a single event . . . . . . . . 109
Remote alarm control . . . . . . . . . . . . . . . . . . . . 110
External device disconnection alarm . . . . . . . . 111
The Code button. . . . . . . . . . . . . . . . . . . . . . . . 111
Alarm ranges and defaults . . . . . . . . . . . . . . . . 112
Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Accessing the calculation functions . . . . . . . . . 123
Viewing the calculation results . . . . . . . . . . . . . 125
Lab data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Calculation equations . . . . . . . . . . . . . . . . . . . . 127
Drug calculations . . . . . . . . . . . . . . . . . . . . . . . 131
Accessing the drug calculation functions . . . . . 131
Customized drug list. . . . . . . . . . . . . . . . . . . . . 132
Drug calculator equations. . . . . . . . . . . . . . . . . 134
Trends/data dialog windows . . . . . . . . . . . . . 137
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Trending behavior. . . . . . . . . . . . . . . . . . . . . . . 138
Trend graphs . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Interacting with the trend graphs pages . . . . . . 142
Recruitment page . . . . . . . . . . . . . . . . . . . . . . . 145
Interacting with the Recruitment page . . . . . . . 146
Trend table . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Interacting with the trend table . . . . . . . . . . . . . 150
Mini-trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Data review pages . . . . . . . . . . . . . . . . . . . . . . 153
Reports tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
ECG, arrhythmia, and ST segment . . . . . . . . 157
Overview of ECG and heart rate monitoring . . 159
ECG precautions . . . . . . . . . . . . . . . . . . . . . . . 160
Connecting the 3-, 5-, 6-lead wire sets for
ECG monitoring . . . . . . . . . . . . . . . . . . . . . . . . 160
Connecting the lead wire set for 12-lead
monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Connecting the lead wires for neonatal
monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
Patient preparation for ECG monitoring . . . . . . 163
ECG display . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
ECG colors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Electrode placement. . . . . . . . . . . . . . . . . . . . . 167
12-lead monitoring . . . . . . . . . . . . . . . . . . . . . . 170
Accessing the ECG functions. . . . . . . . . . . . . . 170
ECG parameter setup functions . . . . . . . . . . . . 171
Monitoring paced patients . . . . . . . . . . . . . . . . 174
6
Pacemaker precautions. . . . . . . . . . . . . . . . . .
Optimizing pacer processing . . . . . . . . . . . . . .
Arrhythmia monitoring overview . . . . . . . . . . .
Selecting arrhythmia leads . . . . . . . . . . . . . . .
Arrhythmia modes . . . . . . . . . . . . . . . . . . . . . .
Arrhythmia display. . . . . . . . . . . . . . . . . . . . . .
Accessing the arrhythmia functions. . . . . . . .
Arrhythmia parameter setup functions . . . . . .
Monitoring ST overview. . . . . . . . . . . . . . . . . .
Standard ST monitoring. . . . . . . . . . . . . . . . . .
TruST 12-lead monitoring . . . . . . . . . . . . . . . .
12-lead ST monitoring . . . . . . . . . . . . . . . . . . .
Connecting lead wire sets for ST monitoring. .
ST display . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reviewing ST complexes . . . . . . . . . . . . . . . .
ST measuring points . . . . . . . . . . . . . . . . . . . .
ST reference . . . . . . . . . . . . . . . . . . . . . . . . . .
ST alarm settings . . . . . . . . . . . . . . . . . . . . . .
Accessing the ST settings . . . . . . . . . . . . . . .
ST setup functions. . . . . . . . . . . . . . . . . . . . . .
Learning/relearning QRS pattern . . . . . . . . . .
175
177
177
178
178
180
181
182
182
183
183
183
184
184
185
187
188
189
189
190
192
Impedance respiration (RRi) . . . . . . . . . . . . 193
Overview of respiration monitoring . . . . . . . . .
Respiration precautions. . . . . . . . . . . . . . . . . .
Connecting the 3-, 5-, 6-lead wire sets for
respiration monitoring . . . . . . . . . . . . . . . . . . .
Connecting the lead wire set for 12-lead
monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connecting the lead wires for neonatal
monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Patient preparation for respiration monitoring .
Respiration display . . . . . . . . . . . . . . . . . . . . .
Adjusting the detection threshold and
activating the respiration marker . . . . . . . . . . .
Respiration measuring modes. . . . . . . . . . . . .
Accessing the respiration settings. . . . . . . . . .
Respiration parameter setup functions . . . . . .
194
194
195
196
197
197
199
200
200
201
201
SpO2 and Pulse CO-Ox monitoring with
Masimo SET MCables . . . . . . . . . . . . . . . . . . 203
Overview of SpO2 and Pulse CO-Ox
monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SpO2 and Pulse CO-Ox precautions . . . . . . . .
Connecting the Masimo SET MCable . . . . . . .
Connecting the Masimo rainbow SET
MCable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Patient preparation . . . . . . . . . . . . . . . . . . . . .
204
206
208
209
210
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 7
Contents
SpO2 and Pulse CO-Ox display . . . . . . . . . . . . 211
Reviewing the SpO2 and Pulse CO-Ox
parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214
Accessing the SpO2 settings . . . . . . . . . . . . . . 215
SpO2 parameter setup functions . . . . . . . . . . . 216
Masimo rainbow SET Pulse CO-Ox
parameter setup functions . . . . . . . . . . . . . . . . 218
Password-protected Masimo rainbow SET
setup functions . . . . . . . . . . . . . . . . . . . . . . . . . 221
SpO2 and pulse rate with Nellcor OxiMax
MCable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
Overview of SpO2 monitoring . . . . . . . . . . . . . . 224
SpO2 precautions . . . . . . . . . . . . . . . . . . . . . . . 225
Connecting the Nellcor OxiMax MCable. . . . . . 226
Patient preparation for SpO2 monitoring. . . . . . 227
SpO2 display. . . . . . . . . . . . . . . . . . . . . . . . . . . 228
Accessing the SpO2 settings . . . . . . . . . . . . . . 229
SpO2 parameter setup functions . . . . . . . . . . . 229
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 231
Overview of temperature monitoring . . . . . . . . 232
Connecting the temperature sensors . . . . . . . . 232
Temperature display . . . . . . . . . . . . . . . . . . . . . 235
Accessing the temperature settings . . . . . . . . . 236
Temperature parameter setup functions. . . . . . 236
Non-invasive blood pressure (NIBP) . . . . . . 237
Overview of non-invasive blood pressure
monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238
Non-invasive blood pressure precautions . . . . 239
Connecting the non-invasive blood pressure
hose and cuff . . . . . . . . . . . . . . . . . . . . . . . . . . 240
Patient preparation for non-invasive blood
pressure monitoring . . . . . . . . . . . . . . . . . . . . . 241
Non-invasive blood pressure display . . . . . . . . 242
Non-invasive blood pressure measurement
modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 243
Venous stasis . . . . . . . . . . . . . . . . . . . . . . . . . . 246
Accessing the non-invasive blood pressure
settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246
Non-invasive blood pressure parameter setup
functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 247
Invasive blood pressure (IBP) . . . . . . . . . . . . 249
Overview of invasive blood pressure
monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250
IBP precautions . . . . . . . . . . . . . . . . . . . . . . . .
Connecting the Hemo4 pod and Hemo2 pod .
Connecting the MPod – QuadHemo . . . . . . . .
Connecting the Dual Hemo MCable . . . . . . . .
Patient preparation for invasive blood
pressure monitoring. . . . . . . . . . . . . . . . . . . . .
Invasive blood pressure display . . . . . . . . . . .
Labeling invasive blood pressure channels. . .
Standard labels . . . . . . . . . . . . . . . . . . . . . . . .
Pressure label conflicts . . . . . . . . . . . . . . . . . .
Zeroing an invasive blood pressure
transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pulmonary wedge pressure. . . . . . . . . . . . . . .
Starting wedge measurements from the pods .
Accessing the invasive blood pressure
settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Invasive blood pressure parameter setup
functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
252
253
254
255
256
256
258
259
260
260
263
264
266
267
Cardiac output (C.O.) . . . . . . . . . . . . . . . . . . 269
Overview of cardiac output monitoring . . . . . .
C.O. Precautions . . . . . . . . . . . . . . . . . . . . . . .
Connecting the cardiac output hardware. . . . .
Patient preparation for cardiac output
monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cardiac output display. . . . . . . . . . . . . . . . . . .
Cardiac output computation constant . . . . . . .
Cardiac output measuring modes . . . . . . . . . .
Saving the cardiac output value . . . . . . . . . . .
Reviewing the cardiac output averages . . . . .
Accessing the cardiac output settings . . . . . . .
Cardiac output parameter setup functions . . .
270
270
271
273
274
275
277
280
281
282
282
Carbon dioxide concentrations (CO2). . . . . 285
Overview of CO2 monitoring . . . . . . . . . . . . . .
CO2 precautions . . . . . . . . . . . . . . . . . . . . . . .
Connecting the sensor . . . . . . . . . . . . . . . . . .
Patient preparation for CO2 monitoring. . . . . .
CO2 display. . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessing the CO2 settings . . . . . . . . . . . . . .
CO2 parameter setup functions . . . . . . . . . . .
286
287
288
289
290
293
293
External device – continuous cardiac
output (CCO) . . . . . . . . . . . . . . . . . . . . . . . . . 295
Overview of CCO monitoring. . . . . . . . . . . . . . 296
CCO precautions. . . . . . . . . . . . . . . . . . . . . . . 298
CCO/SvO2 Display . . . . . . . . . . . . . . . . . . . . . 299
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
7
Page 8
Contents
Viewing the CCO/SvO2 parameters. . . . . . . . . 299
Accessing the CCO/SvO2 settings. . . . . . . . . . 300
SvO2 parameter setup functions . . . . . . . . . . . 300
External device – ventilation / anesthesia . . 301
Overview of ventilation monitoring . . . . . . . . . . 302
Ventilator precautions . . . . . . . . . . . . . . . . . . . . 308
Ventilator display . . . . . . . . . . . . . . . . . . . . . . . 309
Loops . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 309
Viewing all ventilation parameters . . . . . . . . . . 310
Accessing the ventilator settings . . . . . . . . . . . 310
Ventilator Paw setup functions . . . . . . . . . . . . . 311
Ventilator parameter setup functions . . . . . . . . 312
CO2 parameter setup functions . . . . . . . . . . . . 313
Overview of anesthesia monitoring . . . . . . . . . 313
Perseus A500 precautions . . . . . . . . . . . . . . . . 316
Perseus A500 display. . . . . . . . . . . . . . . . . . . . 316
System configuration. . . . . . . . . . . . . . . . . . . 317
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 318
Screen setup . . . . . . . . . . . . . . . . . . . . . . . . . . 318
Screen setup – general settings . . . . . . . . . . . . 319
Screen setup – auto view functions . . . . . . . . . 320
Configuring parameters for display . . . . . . . . . 322
Screen setup – configuring views. . . . . . . . . . . 325
Screen setup – configuring function keys. . . . . 326
Screen setup – the View editor. . . . . . . . . . . . . 328
Configuring the alarm setup . . . . . . . . . . . . . . . 330
Configuring the recording and report settings . 335
Biomed setup . . . . . . . . . . . . . . . . . . . . . . . . . . 338
Biomed IT setup . . . . . . . . . . . . . . . . . . . . . . . . 346
Profile setup . . . . . . . . . . . . . . . . . . . . . . . . . . . 351
Saving profiles . . . . . . . . . . . . . . . . . . . . . . . . . 352
Configuring profiles . . . . . . . . . . . . . . . . . . . . . 354
Transferring profiles . . . . . . . . . . . . . . . . . . . . . 356
Reports/recordings . . . . . . . . . . . . . . . . . . . . 359
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 360
R50N recorder . . . . . . . . . . . . . . . . . . . . . . . . . 360
Timed recordings . . . . . . . . . . . . . . . . . . . . . . . 361
Continuous recordings . . . . . . . . . . . . . . . . . . . 363
Requesting recordings . . . . . . . . . . . . . . . . . . . 363
Available reports. . . . . . . . . . . . . . . . . . . . . . . . 364
Printing reports . . . . . . . . . . . . . . . . . . . . . . . . . 364
Configuring a case summary report . . . . . . . . . 368
8
IT applications (options). . . . . . . . . . . . . . . . 369
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuring IT tabs . . . . . . . . . . . . . . . . . . . . .
Accessing an IT tab. . . . . . . . . . . . . . . . . . . . .
Supported IT applications . . . . . . . . . . . . . . . .
370
370
371
372
Problem solving . . . . . . . . . . . . . . . . . . . . . . 373
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device communication messages. . . . . . . . . .
ECG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ARR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Respiration (RRi). . . . . . . . . . . . . . . . . . . . . . .
SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . .
IBP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cardiac Output (C.O.) . . . . . . . . . . . . . . . . . . .
CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Recording status messages . . . . . . . . . . . . . .
374
374
376
379
380
382
385
390
393
394
398
401
405
Cleaning and disinfection. . . . . . . . . . . . . . . 407
Overview of general precautions. . . . . . . . . . . 408
Approved agents . . . . . . . . . . . . . . . . . . . . . . . 408
Cockpit components . . . . . . . . . . . . . . . . . . . . 409
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . 411
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 412
Maintenance of the IACS components . . . . . . 412
Safety inspections . . . . . . . . . . . . . . . . . . . . . . 414
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417
Technical data . . . . . . . . . . . . . . . . . . . . . . . . 419
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Infinity PS250 Comm Hub . . . . . . . . . . . . . . . .
Infinity P2500. . . . . . . . . . . . . . . . . . . . . . . . . .
Infinity MCable – Nurse Call . . . . . . . . . . . . . .
Infinity R50N . . . . . . . . . . . . . . . . . . . . . . . . . .
Secondary display . . . . . . . . . . . . . . . . . . . . . .
Electromagnetic compatibility . . . . . . . . . . . . .
420
420
422
423
424
425
425
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 429
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 9
For your safety and that of your patients
For your safety and that of your patients
Strictly follow these instructions for use . . . . . .
Training. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety checks . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installing accessories . . . . . . . . . . . . . . . . . . . .
Sterile accessories . . . . . . . . . . . . . . . . . . . . . .
Restrictions for use. . . . . . . . . . . . . . . . . . . . . .
Connected devices . . . . . . . . . . . . . . . . . . . . . .
Safe connection with other electrical
equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . .
Networking and connection to other devices . .
Connection to hospital network . . . . . . . . . . . .
Patient safety . . . . . . . . . . . . . . . . . . . . . . . . . .
Patient monitoring. . . . . . . . . . . . . . . . . . . . . . .
10
10
10
10
11
11
11
11
11
11
12
12
13
13
13
General safety information . . . . . . . . . . . . . . 14
Not for use in areas of explosion hazard . . . . .
Site of operation . . . . . . . . . . . . . . . . . . . . . . . .
Defibrillator precautions . . . . . . . . . . . . . . . . . .
Electrosurgery . . . . . . . . . . . . . . . . . . . . . . . . .
Virus protection. . . . . . . . . . . . . . . . . . . . . . . . .
14
15
15
16
16
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
9
Page 10
For your safety and that of your patients
Strictly follow these instructions for use
WARNING
Any use of the medical device requires full understanding and strict observation of all sections of these instructions for use. The medical device must only be used for the purpose
specified under "Intended Use" and in conjunction with appropriate patient monitoring.
Strictly observe all WARNING and CAUTION
statements throughout these instructions for
use and all statements on medical device labels. Failure to observe these safety information statements constitutes a use of the medical device that is inconsistent with its intended use.
Training
Maintenance
WARNING
The medical device must be inspected and
serviced regularly by professionals who possess the required qualifications due to their
training and their experience. Repair of the
medical device must also be performed by
trained personnel with additional productspecific DrägerService training. Dräger recommends that a service contract is obtained
with DrägerService and that all repairs are performed by DrägerService. Dräger further recommends that only authentic Dräger repair
parts are used for maintenance.
If the above are not complied with, the correct
functioning of the medical device may be
compromised.
See chapter "Maintenance."
Training for users is available from the responsible
Dräger organization. See www.draeger.com for
more information.
Safety checks
The medical device must be subject to regular safety inspections. See chapter "Maintenance".
10
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 11
For your safety and that of your patients
Accessories
WARNING
Only the accessories indicated in the Infinity
Acute Care System – Monitoring Accessories
instructions for use (latest edition) have been
tested and approved for use with the medical
device.
Therefore, it is strongly recommended that
only these accessories are used in conjunction with the medical device. Otherwise, the
correct functioning of the medical device may
be compromised.
Installing accessories
CAUTION
Install accessories to the basic device in accordance with the instructions for use of the basic device. Make sure that there is a safe connection to
the basic device system.
Strictly observe assembly instructions and instructions for use.
Sterile accessories
CAUTION
Do not use sterile-packaged accessories if the
packaging has been opened, is damaged or there
are other signs of non-sterility. Disposable articles
must not be reprocessed and resterilized. Reuse,
reprocessing, or resterilization can lead to a failure of the medical device and cause injury to the
patient.
Restrictions for use
CAUTION
Device for use in health care facilities only and exclusively by persons with specific training and experience in its use.
Connected devices
CAUTION
Federal (U.S.) law restricts this device to sale by
or on the order of a physician.
WARNING
Risk of electric shock and of device malfunction
Any connected devices or device combinations not complying with the requirements
mentioned in these instructions for use may
compromise the correct functioning of the
medical device. Before operating any combination of devices, refer to and strictly comply
with the instructions for use for all connected
devices and device combinations.
Safe connection with other electrical
equipment
CAUTION
Risk of patient injury
Electrical connections to equipment not listed in
these instructions for use or assembly instructions
must only be made when approved by each respective manufacturer.
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
11
Page 12
For your safety and that of your patients
WARNING
The leakage current increases when multiple
medical devices are connected to a patient.
Make sure that the electrical isolation of each
device is suitable for the intended application.
Connect only equipment that is set up and
tested according to IEC standards to the analog and digital signal inputs and outputs.
Connect only passive USB devices to the
IACS (Infinity Acute Care System) Cockpit.
To protect the patient from possible injury due
to electrical shock, peripheral devices should
only be connected to a monitor within the
same room. The installer or service provider
should verify that the leakage current of the
interconnected system meets the electrical
safety requirements of IEC 60601-1 and
IEC 60601-1-1.
Electrical safety
WARNING
Because of the danger of electric shock, never
remove the cover of a device while it is in use
or plugged into a power socket.
CAUTION
Connect the PS250 (Infinity PS250 Comm Hub)
or the P2500 with an attached power cord only to
hospital-grade electrical power sockets to make
sure that it is properly grounded.
CAUTION
To avoid injuring the patient, do not touch any
connector or mounting screw on the device when
you are touching the patient. Do not allow the conductive parts of electrodes and cables to contact
other conductive parts or ground, either.
12
Networking and connection to other
devices
When combining Dräger devices with other electrical devices, the owner must ensure that the resulting system meets the requirements of the following
standards:
– IEC 60601-1 (EN 60601-1)
Medical electrical equipment
Part 1: General requirements for safety
– IEC 60601-1-1 (EN 60601-1-1)
Medical electrical equipment
Part 1-1: General requirements for safety
Collateral standard: Safety requirements for
medical electrical systems
– IEC 60601-1-2 (EN 60601-1-2)
Medical electrical equipment
Part 1-2: General requirements for safety
Collateral standard: Electromagnetic compatibility; Requirements and tests
– IEC 60601-1-4 (EN 60601-1-4)
Medical electrical equipment
Part 1-4: General requirements for safety
Collateral standard: Programmable electrical
medical systems
CAUTION
Combinations of Dräger devices and third-party
devices that are not approved by Dräger may adversely affect operation of those devices and may
put the patient at greater risk of injury.
CAUTION
The medical device must only be used with software tested and approved by Dräger. Any modifications of the operating system settings can impair operating safety. Responsibility for any such
modifications lies with the owner.
Strictly follow the assembly instructions and instructions for use for each connected device.
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 13
For your safety and that of your patients
Connection to hospital network
Patient monitoring
Many medical devices manufactured by Dräger use
networks to transmit patient data in real-time and to
notify clinical users of alarm conditions. Hospitals
should refer to IEC 80000-1 before attempting to
connect such medical devices to their IT networks.
The technical documentation that IEC 80000-1 requires manufacturers such as Dräger to make
available in support of such network connections
can be requested. Contact your Dräger representative for that information or to facilitate negotiation of
an IEC 80000-1 Responsibility Agreement for additional support from Dräger.
Patient safety may be achieved by a wide variety of
means ranging from electronic surveillance of medical device performance and patient condition to
simple, direct observation of clinical signs.
The responsibility for selecting the best level of patient monitoring lies solely with the user of the medical device.
Patient safety
The design of the medical device, the accompanying documentation, and the labeling on the medical
device are based on the assumption that the purchase and the use of the medical device are restricted to trained professionals, and that certain inherent characteristics of the medical device are
known to a trained user. Instructions and WARNING and CAUTION statements are therefore largely limited to the specifics of the Dräger medical device.
These instructions for use do not contain references to various hazards which are obvious to a professional who operates this medical device as well
as references to the consequences of medical device misuse, and to potentially adverse effects in
patients with different underlying diseases. Medical
device modification or misuse can be dangerous.
CAUTION
Risk of patient injury
Do not make therapeutic decisions based solely
on individual measured values and monitoring parameters.
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
13
Page 14
For your safety and that of your patients
General safety information
The following WARNING and CAUTION statements apply to general operation of the medical device.
WARNING and CAUTION statements specific to
subsystems or particular features of the medical
device appear in the respective sections of these
instructions for use or in the instructions for use of
another product being used with this device.
WARNING
Follow local regulations for safe disposal of
batteries. To prevent fire or explosion, never
dispose of batteries in fire.
WARNING
Follow local regulations for safe disposal of
batteries. To prevent fire or explosion, never
dispose of batteries in fire.
WARNING
To avoid electric shock, inspect all cables before use. Never use cables that appear
cracked, worn, or damaged in any way (doing
so may compromise performance or put the
patient at risk).
WARNING
Do not cover the device with blankets or bed
sheets. To prevent burns to the patient, avoid
direct contact between external surfaces and
the patient.
CAUTION
To avoid injuring the patient, disconnect all sensors that will not be used during transport, before
moving the patient.
14
CAUTION
Read all cleaning instructions (for example, originating from the disinfectant manufacturer and the
hospital) carefully before cleaning the device. Refer to the chapter entitled "Cleaning and disinfection" on page 407 for device-specific cleaning instructions. Moisture may damage the circuits,
compromise critical performance and present a
safety risk.
WARNING
Dräger recommends using the Infinity Acute
Care System or the M540 (if on wireless transport) for primary diagnosis and the (ICS) Infinity CentralStation for patient viewing only.
For countries subject to the EU directive
2002/96/EC
This device is subject to EU Directive 2002/96/EC
(WEEE). In order to comply with its registration according to this directive, this device may not be disposed of at municipal collection points for waste
electrical and electronic equipment. Dräger has authorized a company to collect and dispose of this
device.
To initiate collection or for further information, visit
Dräger on the Internet at www.draeger.com. Use
the Search function with the keyword "WEEE" to
find the relevant information. If access to Dräger's
website is not possible, contact the local Dräger organization.
Not for use in areas of explosion hazard
WARNING
When placing the device, make sure that adequate airflow exists. To prevent overheating,
position the device with at least 5 cm (2 in) of
space all around.
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 15
For your safety and that of your patients
Information on electromagnetic compatibility
General information on electromagnetic compatibility (EMC) according to international EMC standard
IEC 60601-1-2:
WARNING
Do not connect connectors with an
ESD warning symbol and do not
touch the pins of such connectors
without implementing ESD protective measures. Such protective
measures may include antistatic clothing and
shoes, touching a ground stud before and
during connection of the pins, or using electrically insulating and antistatic gloves. All relevant personnel must be instructed in these
ESD protective measures.
Site of operation
Only use devices (monitor, MPod, MCable, and accessories) in areas that meet the environmental requirements outlined in the technical data section.
WARNING
To avoid interfering with device operation, do
not operate devices (monitor, MPod, MCable,
and accessories) within 10 m (33 feet) of
equipment that emits microwave or other
high-frequency emissions.
WARNING
Make sure that the device is properly mounted
and secured to prevent injury. Make sure the
requirements for maximum load and slope of
floor are met. Consult the documentation of
the mounting manufacturer for detailed information.
WARNING
To minimize the risk of patient strangulation,
carefully position and secure sensor cables.
Also position the sensor cables to minimize
inductive loops.
CAUTION
To prevent overheating, do not place the device in
direct sunlight or near radiant heaters.
CAUTION
After extended exposure in a cold environment,
acclimate the device carefully so that condensation does not form on the electronic parts and
damage the device.
CAUTION
To avoid damaging the touch-sensitive screen, do
not allow sharp instruments to touch the front panel of the devices.
CAUTION
To avoid short-circuiting and otherwise damaging
the device, Dräger recommends that no fluids
come in contact with the IACS devices when they
are connected to a power socket. If fluids are accidentally spilled on the equipment, remove the
affected device from service as soon as possible
and have technical personnel verify that patient
safety is not compromised.
Defibrillator precautions
The IACS and the peripheral devices are protected
against high-frequency interference from defibrillators and electrosurgical units and against 50-Hz
and 60-Hz power line interference.
CAUTION
To prevent burns and electric shock due to rerouting of electrical current through electrodes, do not
position the defibrillator pads near any electrodes
or sensors.
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
15
Page 16
For your safety and that of your patients
CAUTION
Only defibrillate across the chest.
CAUTION
Using ECG electrodes and cables specified by
Dräger protects the device from damage during
defibrillation and reduces noise and other interference on the ECG waveform.
WARNING
To reduce the hazard of burns during electrosurgery, keep the sensor or transducer (ECG,
pressure, SpO2) and their associated cables
away from the surgical site, the ESU return
electrode, and earth ground.
NOTE
Cover internally placed reusable temperature
sensors with temperature probe sheaths.
Electrosurgery
Observe the following precautions during electrosurgery to reduce electrosurgical unit (ESU) interference and improve operator and patient safety.
WARNING
For better performance and to reduce the hazard of burns during surgery, always use accessories designed for ESU environments. Do
not use skin temperature sensors.
16
Virus protection
CAUTION
The IACS does not have virus protection software
and relies therefore on the firewall of your institution to prevent access to infected files. While setting up IT applications to access web sites, evaluate each web site with regard to possible virus infection.
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 17
Application
Application
Intended use – Infinity Acute Care
System – Monitoring Applications . . . . . . . . 18
Indications for use . . . . . . . . . . . . . . . . . . . . . . 18
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
17
Page 18
Application
Intended use – Infinity Acute Care System – Monitoring Applications
The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric, and neonatal patients in environments where patient care is
provided by trained healthcare professionals.
The IACS obtains the physiologic, multi-parameter
data from the connection to the M540 monitor and
optional medical devices and displays. The transfer
of this data is accomplished by the Infinity network.
The IACS and any connected optional hardware
are not intended for use in the following hospital environments:
–
Hyperbaric chambers
–
Environments containing MRI equipment
18
Indications for use
The M540 monitors the following parameters:
–
Heart rate
–
Arrhythmia (adult and pediatric patients only)
–
12-lead ECG monitoring
–
12-lead ECG monitoring including TruST (adult
and pediatric patients only)
–
ST segment analysis (adult and pediatric patients only)
–
12-lead ST segment analysis (adult and pediatric patients only)
–
Apnea
–
Respiration rate
–
Invasive blood pressure
–
Non-invasive blood pressure
–
Temperature
–
Cardiac output (adult and pediatric patients
only)
–
Arterial oxygen saturation – SpO2
–
Pulse rate
–
Perfusion index – PI
–
Total hemoglobin – SpHb, (adult and pediatric
patients only)
–
Total oxygen content – SpOC, (adult and pediatric patients only)
–
Carboxyhemoglobin saturation – SpCO
–
Methemoglobin saturation – SpMet
–
Pleth variability index – PVI
–
Mainstream etCO2
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 19
System overview
System overview
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Infinity Medical Cockpit (Cockpit). . . . . . . . . 20
Infinity PS250 Comm Hub (PS250) power
supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Infinity P2500 power supply (P2500) . . . . . . 21
Infinity M540 patient monitor (M540) . . . . . . 22
Infinity M500 docking station (M500) . . . . . . 22
M500 front panel. . . . . . . . . . . . . . . . . . . . . . . . 22
M500 back panel . . . . . . . . . . . . . . . . . . . . . . . 23
Additional hardware . . . . . . . . . . . . . . . . . . . . 23
Device symbols . . . . . . . . . . . . . . . . . . . . . . . 25
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . 27
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
19
Page 20
System overview
Overview
These instructions for use describe the Cockpit
(Medical Cockpit), the primary display and user interface of the Infinity Acute Care System – Monitoring Applications – M540 Patient Monitor (IACS).
Specifically, these instructions for use describe the
setup tasks and features available on the Cockpit.
For detailed information on the M540 patient monitor, refer to the instructions for use Infinity Acute
Care System – Infinity M540.
–
Undocking the M540 – refers to removing the
M540 from the M500 for patient transport.
The following diagram shows the basic components of the IACS. In addition, you can connect various hardware to expand the viewing and monitoring capabilities (see "Additional hardware"
on page 23).
Some terms used in these instructions for use:
Cockpit – refers to the Infinity C700 Medical
Cockpit or the Infinity C500 Medical Cockpit
–
M540 – refers to the Infinity M540 transport
component and patient connection point of the
IACS.
–
M500 – refers to the Infinity M500 Docking Station that secures the M540, provides communication between the M540 and the Cockpit, and
charges the battery in the M540.
A
B
D
C
001
–
–
PS250 – refers to the Infinity PS250 Comm Hub
A C500 / C700
–
P2500 – refers to the Infinity P2500 Communications Hub
B PS250 or P2500
–
Docking the M540 – refers to placing the M540
on the M500.
D M540
C M500
Infinity Medical Cockpit (Cockpit)
The Cockpit is the primary display and user interface for the IACS and is available in the following
sizes:
–
C500 – 43 cm (17 in) wide screen
–
C700 – 50 cm (20.1 in) wide screen
For detailed description regarding the front and
back panel of the Cockpit, refer to the instructions
for use Infinity Acute Care System – Infinity Medical
Cockpit.
20
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 21
System overview
Infinity PS250 Comm Hub (PS250) power supply
The following diagram shows the bottom of the
PS250.
A Infinity network connectors
B Nurse call connector
C Export protocol connector
A
BC
D
D Power cord connection
E Two interchangeable system cable connectors
– one for the M540, one for the Cockpit
F
F
F Network connection LEDs
The front panel of the PS250 has the following two
LEDs:
–
Power mains – lights up green when the
device is connected to AC power.
–
Battery indicator – yellow LED that
lights up briefly during startup or fault
conditions such as a faulty battery.
004
E
Infinity P2500 power supply (P2500)
The following diagram shows the bottom of the
P2500.
A Two interchangeable system cable connectors
– one for the M540, one for the Cockpit
B Power cord connection
A
C Infinity network connector
D Nurse call connector
E Export protocol connector
The front panel of the P2500 has the following two
LEDs:
C B
357
ED
–
Power mains – lights up green when the
device is connected to AC power.
–
battery indicator – yellow LED that lights
up briefly during startup or fault conditions such as a faulty battery.
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
21
Page 22
System overview
Infinity M540 patient monitor (M540)
The following diagram shows the M540 when it is
docked in the M500 docking station.
A M540 patient monitor
B M500 docking station
The M540 acquires patient signals, processes
them, and relays them to the Cockpit for display.
The M540 also provides patient monitoring when it
is undocked during patient transport. For more detailed information on the M540, refer to the instructions for use Infinity Acute Care System – Infinity
M540.
A
300
B
Infinity M500 docking station (M500)
The M500 is the mechanical device that secures
and powers the M540. It also charges the battery,
and controls the communication between the M540
and the Cockpit through an optical Ethernet link.
A Locking mechanism – secures the M540 (for
more detailed information, see "Locking/unlocking the M540" on page 68)
B Release buttons for undocking the M540 (you
only have to press one button to release the
M540)
M500 front panel
B
C Optical Ethernet links
A
D Pins for charging the battery of the M540 and
providing power to the M540 when docked
B
C
301
D
22
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 23
System overview
M500 back panel
E Nurse call connector
F Network LED – lights up green when connected
to the network
G
E
F
302
G System cable connector
Additional hardware
The following table lists the additional devices that
can be connected to the IACS.
Device
Description
Infinity MCable – Masimo SET
Measures the percentage of func- Connects directly to the SpO2
tional hemoglobin saturated with connector of the M540
oxygen (%SpO2) and reports the (see page 208 and page 226).
perfusion index (PI) and the pulse
rate (PLS).
Connection
Infinity MCable – Masimo SET
rainbow
Measures the percentage of functional hemoglobin saturated with
oxygen (%SpO2) and reports the
perfusion index (PI) and the pulse
rate (PLS). In addition, it measures total hemoglobin (SpHb),
total oxygen content (SpOC),
pleth variability index (PVI), Carboxyhemoglobin saturation
(SpCO), methemoglobin saturation (SpMet).
Infinity MCable – Nellcor OxiMax Measures the percentage of functional hemoglobin saturated with
oxygen (%SpO2) and the pulse
rate (PLS).
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
23
Page 24
System overview
Hemo4 pod
Infinity MPod – QuadHemo
Measures up to 4 pressures, car- Connects directly to the Hemo
diac output, core and body tem- connector of the M540 (see inforperature.
mation starting on page 253).
Hemo2 pod
Measures up to 2 pressures, cardiac output, core and body temperature.
Infinity MCable – Dual Hemo
Measures up to 2 pressures.
Infinity MCable – Mainstream
CO2
Measures mainstream CO2.
Connects directly to the CO2
connector of the M540
(see page 288).
Infinity MCable – Nurse Call
Provides remote notification of
medium and high priority alarm
conditions.
Connects to the PS250 / P2500
(see page 21) or to the M500
(see page 22).
Infinity MCable – Analog/Sync
Provides a sync pulse to synchronize defibrillators to the heart beat
of the patient during cardioversion. The cable’s analog out function provides an ECG and arterial
blood pressure signal to a device
such as intra-aortic balloon pump.
Connects to the Temp/Aux connector of the M540
(see page 232) or to the CO2
connector with a Y-cable.
Secondary video display
Extends the viewing capabilities
of a Cockpit to an additional video
display. Secondary displays mirror the content of the Cockpit.
Connects to a Cockpit using the
DVI 1 connector located on the
back panel (see the instructions
for use Infinity Acute Care System – Medical Cockpit).
R50N recorder
Produces timed and continuous
recordings.
Connects to the Infinity network
or the PS250 / P2500.
Laser printer
Prints various reports and Cockpit Connects to the Infinity network.
print screens.
24
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
Page 25
System overview
Device symbols
Consult instructions for use
Lower alarm limits
Directive 93/42/EEC
concerning Medical Devices
Upper alarm limits
Access to trend pages
Autoset alarm limits
The button next to this symbol accesses special procedure pages
Alarm monitoring deactivated temporarily
Access to alarm functions
Alarm monitoring deactivated permanently
Access to the standby and privacy
modes, and access to patient discharge
Acoustic alarm tone paused temporarily
Access to pre-configured Views and
layouts
Acoustic alarm tone turned off permanently
Access to parameter pages
Change clinical password
Adult patient category
Lung symbol that pulsates with each detected breath
Pediatric patient category
Heart blip that flashes with each detected pulse
Neonatal patient category
Pacer detection is activated; the heart
symbol flashes with each detected
paced pulse
Battery status LED
Scrolls to additional tabs and pages
Battery charging error
Power on/off
Instructions for Use Infinity Acute Care System – Monitoring Applications VG2
25