Infinity Cockpit C700 and C500 Quick Reference Guide sw VG4.pdf
Page 1
Infinity Acute Care System
Quick Reference Guide
Infinity Cockpit
C700 / C500
This Infinity Quick Reference
Guide is not a replacement or
substitute for the Instructions for
Use (IFU). It is for informational
purposes only.
Software Level: VG4
Page 2
INFINITY COCKPIT QUICK REFERENCE GUIDE
|2
Content
Electrode & Lead Placement
Docking, Transport, Power and Keys
Blue Header Bar, Quick Access & Function Keys
Patient Category and Admit
Discharge, Standby and Privacy
Alarm Priorities
Alarm Pause & Alarm Setup
Screen Setup Restore Profile and Views
Parameter Setup
ECG
RRi/Impedance Respirations
SpO2
NIBP
Mainstream etCO2 Monitoring
12 lead ECG
Invasive Pressures
PAWP
Cardiac Output
Calculations
3-6
7-9
10-12
13
14
15
16-18
19-20
21
21-22
22-23
23-24
24
25-26
26
27
28
28-29
30
Warning! This “Quick Reference Guide” is not intended to replace the Instructions for Use manual.
For full information concerning the performance characteristics of the Dräger devices described in
this Quick Reference Guide, each user must carefully read and fully comprehend the Instructions for
Use before operating the respective device.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 3
INFINITY COCKPIT QUICK REFERENCE GUIDE
|3
Electrode & Lead Placement
DO’S & DO NOT’S
9 Do place the electrode on a clean and dry
area of skin.
9 Do utilize soap, water, and dry skin thoroughly.
9 Do remove hair if necessary for proper electrode placement.
9 Do change electrodes Q 24-48 hours, or per
hospital policy.
U Do Not utilize alcohol to dry and clean area
of skin for electrode placement.
U Do Not place electrodes over bony prominences or area of increased muscle movement/activity.
U Do Not assume electrodes are “good” if the
tape portion is attached to the patient’s skin.
U Do Not braid, tie or knot leads.
Note: Always follow hospital-approved practices or the recommendations of the electrode manufacturer.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 4
INFINITY COCKPIT QUICK REFERENCE GUIDE
|4
Electrode Colors*
IEC
AHA/US
Electrode Location
L
Yellow
LA
Black
Left arm
F
Green
LL
Red
Left leg
R
Red
RA
White
Right arm
N
Black
RL
Green
Right leg (neutral)
C
White
V
Brown
Moveable chest
C+
Gray & White V+
Gray & Brown 2nd Moveable chest
Standard 6 Lead Configuration
IEC
AHA
Standard 5 Lead Configuration
IEC
AHA
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 5
INFINITY COCKPIT QUICK REFERENCE GUIDE
Standard 3 Lead Configuration
IEC
|5
AHA
Pacemaker Configuration 5 electrodes
IEC
AHA
Warning: Make sure the pacemaker detection is turned OFF for patients
without pacemakers and that it is ON for patients with pacemakers. Always keep pacemaker patients under close surveillance.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 6
INFINITY COCKPIT QUICK REFERENCE GUIDE
|6
12-Lead Configuration (ten electrodes)
AHA
IEC
Respiratory lead placement for infants
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 7
INFINITY COCKPIT QUICK REFERENCE GUIDE
|7
Docking and Transport, Power &
Frequently Used Keys
M500 DOCKING STATION WITH M540 AND COCKPIT
1. M500 secures and powers the M540 monitor
2. M500 connects M540 to the Cockpit and the Infinity network
that may include an Infinity Central Station (ICS), a network recorder, and a network printer.
3. M540 facilitates transport of patient when undocked, by maintaining all patient connections and alarm settings.
4. M540 when configured as wireless can send information to the
ICS over a Wi-Fi network.
TO UNDOCK M540 (release for transport)
1. Push locking tab toward the back of M500
2. While holding the M540 firmly with one hand press one of the
release buttons on M500 and remove from M500
TO DOCK M540
1. Align curved portions of M540 and M500
2. Press M540 into M500 until it clicks in place.
3. Push the locking tab toward the front of M500.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 8
INFINITY COCKPIT QUICK REFERENCE GUIDE
|8
COCKPIT
Refers to the C500/C700 medical cockpit. The main user interface
for the IACS
TO POWER ON/OFF
1. Press the Power On/Off key
2. Select Shutdown inside the dialog window. The message
3. System is shutting down appears on the screen while the
4. Infinity Medical Cockpit is being powered down.
ROTARY KNOB
Navigates through menus and tables to define settings
1. Touch a Parameter box or Function key to display a menu.
Turn the Rotary knob to the left or right to increase or decrease a numeric value, to scroll down a list, or navigate
through a menu.
2. Press the Rotary knob to accept a selected setting.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 9
INFINITY COCKPIT QUICK REFERENCE GUIDE
|9
AUDIO PAUSE KEY
This key is located to the right of the rotary knob.
Pauses acoustic alarm signals for two minutes.
Note: Quiet Mode is Activated – If a new alarm condition with a priority
higher than the currently paused alarm occurs, a truncated alarm tone
sounds. In addition, the alarm is represented by visual alarm signals corresponding to the alarm priority. If the new alarm is of lower priority than the
paused alarm, the new alarm condition is only represented by a visual alarm
signal. No acoustic alarm tones sound.
If Quiet Mode is Deactivated - Any new alarm condition breaks through the
Audio Pause period with full acoustic and visual alarm annunciation.
TO CALIBRATE THE C700 / 500 SCREEN
1. Press the rotary knob for 30 sec. until the Calibrate Touch
Screen popup appears. (Requires several seconds)
2. Select the Calibrate button in the popup or press the rotary
knob again to access the calibration screen.
3. Follow the onscreen prompts
4. Depress the circles until they turn Dark Blue.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 10
INFINITY COCKPIT QUICK REFERENCE GUIDE
|10
Blue Header Bar, Quick Access Icons, &
Function Keys
BLUE HEADER BAR
Appears along the top of the screen and displays the following:
1.
Patient category field (adult pediatric, neonate)
2. System data field
3. Patient name field
4. Time and date field
5. Alarm message field
6. Alarm banner field
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 11
INFINITY COCKPIT QUICK REFERENCE GUIDE
|11
QUICK- ACCESS SYMBOLS
Symbols that offer access to frequently used functions
Quick access symbol next to:
Associate tool bar
All alarms off, or All alarms paused
Auto set all
ALARMS…
VIEWS…
Show all ECG
Remote view
Print screen
TRENDS AND
DATA…
Trend Table Report, ECG Reports, Timed
Waveform Report, Timed Recording and
Continuous Recording, etc.
SENSOR
PARAMETERS…
Zero all
NIBP continuous
Venous stasis
Standby
Discharge
START / STANDBY… Privacy
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 12
INFINITY COCKPIT QUICK REFERENCE GUIDE
|12
FUNCTION KEYS
Preconfigured and Configurable quick access keys
Key
Function
ALARMS
Opens the Alarms dialog window.
MARK EVENT
Stores an event in the alarm history.
CODE
Executes pre-configured functions associated with a code
RELEARN ARR
Establishes a new QRS template
VIEWS…
Opens the dialog window for setting up
screen view and Remote view.
PRINT SCREEN
Prints the current screen view on a connected laser printer
FREEZE WAVEFORMS Stops all waveforms for 60 seconds.
TRENDS AND DATA…
Opens the Trends/Data dialog window.
PROCEDURES
Opens the dialog window for setting
up Wedge and C.O.
SENSOR PARAMETERS…
Opens the dialog window for setting up each
parameter.
NIBP START / STOP
When selected starts an NIBP measurement.
To cancel the BP, select the button again.
ZERO
Zeroes all pressures (not available on C500)
SYSTEM SETUP
Opens the System setup dialog window.
START / STANDBY
Opens the Start/Standby dialog window.
HOME
Returns to the main screen and closes any
dialog window.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 13
INFINITY COCKPIT QUICK REFERENCE GUIDE
|13
Patient Category, Admit, Discharge,
Standby, and Privacy
PATIENT CATEGORY
Each patient category for Adult, Pediatric, and Neonate have specific defaults associated with it.
in left most field of the
1. Touch patient category symbol
blue header bar on Cockpit.
2. Select desired Patient category.
3. Press rotary knob to confirm setting.
o The message “Changing category will change alarm settings and algorithmic processing” appears.
o Selected patient category icon
will appear in blue
header bar.
Note: If you dock an M540 with a patient category that differs from the one
selected on the Cockpit, the patient category setting of the Cockpit changes
to match the M540.
ENTER PATIENTS DEMOGRAPHIC DATA FOR ADMISSION
1. Touch patient category symbol
in left most field of the
blue header bar on Cockpit.
2. Select Name, ID, Birth date, Admit date, Physician, height
and weight, and use key board to enter data.
TO ADMIT A PATIENT USING “GET HIS” FUNCTION
1. Touch patient category symbol
in left most field of the
blue header bar on Cockpit.
2. Enter the patient ID number.
3. Select Get HIS button on the demographics page.
o If button is grayed out and is not selectable, the Cockpit is
not connected to the hospital information system.
4. Demographics will populate automatically. If demographics do
not appear, confirm correct ID #.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 14
INFINITY COCKPIT QUICK REFERENCE GUIDE
|14
TO DISCHARGE A PATIENT
1. Select Discharge function key on the main menu bar.
2. The message Caution discharge will delete patient data appears.
3. Select Discharge to confirm, or Cancel to resume monitoring
same patient.
4. The message Touch Screen to initiate monitoring appears.
Note: All patient data is deleted upon discharge, and either factory or user
default limit settings are restored.
TO PLACE MONITOR IN STANDBY
You can temporarily interrupt patient monitoring by placing the
Cockpit and the M540 in standby mode.
1. Select the Start/Standby... button on the main menu bar to
display the Standby dialog window.
2. Select the Start tab if it is not already selected.
3. Select the Standby button. The banner Standby-Touch Screen
to resume monitoring appears in the center of the Cockpit
screen.
4. To take the Cockpit out of Standby mode, touch the screen to
resume monitoring the vital signs
TO PLACE MONITOR IN PRIVACY
All patient data are removed from the Cockpit and the M540 displays, but continue to be displayed and alarm signals produced at
the Infinity Central Station.
1. Select the Start/Standby... button on the main menu bar to
display the Standby dialog window.
2. Select the Start tab if it is not already selected.
3. Select the Privacy button. The banner Privacy Touch Screen to
resume monitoring is displayed in the center of the monitor.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 15
INFINITY COCKPIT QUICK REFERENCE GUIDE
|15
Alarm Priorities
HIGH-PRIORITY ALARMS
- Report conditions that can be life threatening and require immediate intervention; for example, asystole.
- The alarming parameter’s box and alarm message field have red
background, and audible alarms*sound when active.
- Are latched, (audio / visual alarms continue until the alarm is
acknowledged, even if the alarm condition has ceased to exist). To acknowledge press
.
MEDIUM-PRIORITY ALARMS
- Report conditions that require prompt attention but may not be
life-threatening; for example, low SpO2.
- The alarming parameter’s box and alarm message field have a
yellow background, and audible alarms* sound when alarms are
active.
- Are non-latched (audio / visual flashing alarms stop when situation corrects self), but alarm status message will remain in the
header bar. To remove message press
.
LOW -PRIORITY ALARMS
- Alerts to technical issues that may compromise the ability of the
system to monitor the patient; for example, SpO2 probe off. Also
may be set to alert to lower risk arrhythmias, such as couplets.
- The alarming parameter’s box and alarm message field have cyan
background, and audible alarms* sound when active.
- Are also non-latched alarms, except status bar does not remain
active when situation resolved.
Note: Audible alarm signals are distinct for each alarm priority. Specific
characteristics including break through alarms will vary depending on the
default alarm pattern (Infinity, IEC fast or IEC slow) & software level.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 16
INFINITY COCKPIT QUICK REFERENCE GUIDE
|16
Alarm Pause & Alarm Setup
TO PAUSE AUDIO ALARMS FOR 2 MINUTES
1. Press the yellow key
on the Cockpit.
2. Pressing the key again turns alarms back on
Note: Quiet Mode is Activated – If a new alarm condition with a priority
higher than the currently paused alarm occurs, a truncated alarm tone
sounds. In addition, the alarm is represented by visual alarm signals corresponding to the alarm priority. If the new alarm is of lower priority than the
paused alarm, the new alarm condition is only represented by a visual alarm
signal. No acoustic alarm tones sound.
If Quiet Mode is Deactivated - Any new alarm condition breaks through the
Audio Pause period with full acoustic and visual alarm annunciation.
TO PRE-SILENCE ALARMS FOR 2 MINUTES
This function allows you to pre-silence (audio pause in advance)
potential alarm conditions before they occur.
1. Press the yellow key
on the Cockpit
o Alarm conditions are reported visually by a corresponding alarm message and a blinking parameter box.
o A single alarm tone sequence is generated for the first
occurrence of an alarm condition of low, medium or high
priority.
o The Audio paused % banner appears in the far right
field of the header bar along with a countdown timer
and the following symbol:
o If multiple alarm conditions arise during an active presilence period, the Cockpit triggers a single alarm tone
sequence for the highest grade.
Note: Pre-Silence functions only if Quiet Mode is activated.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 17
INFINITY COCKPIT QUICK REFERENCE GUIDE
|17
ARCHIVE FUNCTION
In response to an alarm limit violation for each Parameter or Arrhythmia the following occurs depending on the Archive setting:
o Off – no event is stored and no recording is generated.
o Store – stores the event for later review.
o Record – generates a timed recording
o Str/Rec – generates a timed recording and stores the event.
Note: All Arrhythmia Alarms if set to Str or Str/Rec will hold the event in the
alarm history for later review even when the alarm is turned Off.
CONFIGURING INDIVIDUAL PARAMETER ALARM SETTINGS
1. Select desired parameter box to access the parameter setup
page.
2. Select button above the current patient value to adjust the
High limit. Select the button below the current patient value to
adjust the Low limit.
3. When button is highlighted yellow use Rotary knob to adjust
the parameter limit higher or lower.
4. Press Rotary knob to confirm.
5. Select button below low limit to select how an alarm violation
for this parameter will be Archived.
6. To turn alarm Off or On for the parameter
o Select the Alarm on/off button, a crossed-out triangle
appears in the parameter box when alarm monitoring is
deactivated.
WARNING: The banner HR alarms off appears in the right most
field in the header bar whenever you deactivate heart rate alarms.
Setting alarm limits to extreme values may prevent certain alarm
conditions from being detected and from being annunciated with
acoustic and visual alarm signals.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 18
INFINITY COCKPIT QUICK REFERENCE GUIDE
|18
CONFIGURING ARR (ARRYTHMIA) ALARM SETTINGS
1. Select the Alarms button on the main menu bar.
2. Select the Limits tab (if not already selected).
3. Select the ARR tab along the right side.
4. Select the corresponding setup button for the Alarm priority
column to select the alarm priority. A crossed-out triangle appears when alarm monitoring is deactivated. The priority for
asystole and ventricular fibrillation events cannot be changed.
The alarm priority ’high’ is always assigned to these categories.
5. Select the corresponding setup button in the Rate column to
set the rate.
6. Select the corresponding setup button in the Count column to
set the count.
7. Select the corresponding setup button in the Archive column
to set how it will be archived with alarm violation.
8. Select the Arrhythmia mode button to change Arrhythmia
monitoring to Advanced, Basic or Off (if available).
9. Select the Relearn button to initiate the relearn process of the
ECG leads. A new reference template is created.
TO ADJUST ALARM VOLUME
1. Touch the Alarm Volume Indicator to open the Alarm Volume
Menu
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 19
INFINITY COCKPIT QUICK REFERENCE GUIDE
|19
Screen Setup, Restore Profile & Views
DISPLAY MODE SETUP/AUTO VIEW TAB
The Mainscreen of the Cockpit will show parameters and waveforms based on two settings, Auto or Manual
1. Select Setup button on far right of Auto display toolbar if
available
o OR, select System Setup from menu bar key, Screen Setup
tab (on top), and Auto view tab (on right).
2. Select Auto display mode selection button to have the Main
screen display only those parameters with cables connected
to the monitor
OR
3. Select Manual display mode to have the Main screen show all
parameters even those not connected.
4. Adjust position and number of waveforms and parameter
boxes, turn on Split screen, and setup and turn on Mini trends
display.
AUTO VIEW TOOLBAR
1.
2.
On Auto view toolbar at bottom of screen, touch the desired
parameter icon to move.
With rotary knob, dial left to move to higher priority, or dial
right to move parameter to a lower priority.
o
Parameter is displayed as a waveform with a parameter
box.
o
Parameter is displayed only as a parameter box.
o
Parameter is actively in use but not displayed on the
screen.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 20
INFINITY COCKPIT QUICK REFERENCE GUIDE
|20
TO SELECT OR RESTORE A PROFILE
A profile consists of customizable pre-defined settings for each
patient category, and may include settings unique for each parameter, the alarm defaults, and configurable menu bar keys. Default
Profile is restored when selected or upon discharge.
1. Select System setup…
2. Select Profiles
3. At drop down box, select down arrow and select desired profile name. Default profile has an asterisk*.
o Profile may already be highlighted, but select, profile has
now been restored.
Note: Patient category does not change automatically.
TO SELECT A DIFFERENT VIEW
A View refers to the layout and content of the screen. Each profile
has a pre-defined view associated with it. Each profile can also
have alternate views associated with it.
1. Select Views… key on main menu bar.
2. Select name of desired view.
“Auto” in the name indicates a plug-and-play concept where the
content of the main screen depends on the connected parameter
signals.
REMOTE VIEW
Cockpit allows you to view data from other infinity monitors in the
same monitoring unit.
1. Select the Views button on the main menu bar.
2. Select the Remote view tab. This dialog lists all of the beds in
the monitoring unit.
3. Select a bed from the list in the Views... dialog to access the
remote view of an individual patient.
4. Select the Connect button.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 21
INFINITY COCKPIT QUICK REFERENCE GUIDE
|21
Parameters
TO ACCESS ANY PARAMETER SETTINGS
1. Select the desired parameter box.
OR, if desired parameter is not visible
2. Select Sensor parameters… on Main menu bar.
3. Select desired parameter tab from tabs across top of dialog box.
Alarm settings
Touch desired
parameter box
Or
Touch Sensor
parameter menu
bar key and then
top tab for desired
parameter
ECG Settings 1
tab dialog box
Additional ECG settings tabs
ECG
TO ADJUST PULSE TONE ON/OFF
1. Select Heart rate parameter box.
2. At Pulse tone volume %, select button.
3. Use rotary knob to dial and select desired setting.
4. Click rotary knob to confirm.
TO CHANGE LEADS
1. Select Heart rate parameter box.
2. At Leads select 1st, 2nd, or 3rd button to choose which displayed channel on screen to change.
3. Make selection from available leads
4. Click rotary knob to confirm.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 22
INFINITY COCKPIT QUICK REFERENCE GUIDE
|22
TO TURN PACEMAKER DETECTION ON / OFF
1. Select Heart rate parameter box.
2. Select Settings 2 tab.
3. At Pacer Detection select On or Off.
o If buttons are grayed out, make sure choice of Filter (located directly above Pacer detection) is not set to ESU
TO ADJUST QRS DETECTION THRESHOLD
1. Select the Heart rate parameter box.
2. Select Settings 2 tab.
3. Select Normal or Low QRS Threshold
4. Click rotary knob to confirm.
Note – Low detects QRS complexes of low amplitude.
TO SHOW ALL LEADS AND / OR PRINT ECG REPORT
5. Select Heart rate parameter box.
6. Select Show all tab.
7. Select ECG report Print for non-diagnostic report
8. Select Rest ECG report Print for diagnostic report.
o Diagnostic report requires specific software.
TO SELECT NUMBER OF LEADS FOR ARR PROCESSING
1. Select Heart rate parameter box.
2. Select Settings 2 tab.
3. At ARR processing, select ECG 1 or ECG 1 & 2 button
TO RELEARN ECG
1. Select Heart rate parameter box.
2. Select Settings tab > ARR tab > Relearn.
Note: Not available in Neonate Mode
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 23
INFINITY COCKPIT QUICK REFERENCE GUIDE
|23
RRi - Impedance Respiration
TO RELEARN RESP PROCESSING
1. Select RRi parameter box.
2. Select Relearn.
3. RRi waveform is re-analyzed and saved as a reference for
counting the Resp waveform.
TO SELECT AUTO OR MANUAL MODE
1. Select RRi parameter box.
2. At Mode, select Auto or Manual to change.
o Auto – Optimal breath-detection threshold calculated at
the beginning of RESP monitoring, or with RESP Relearn.
Intended for patients with regular breathing patterns. Relearn the RRi waveform regularly to reduce artifact and
optimize breath count.
o Manual –The monitor detects breath based on the size of
the waveform. After turning on Manual mode adjust
waveform size. Manual mode is intended for adult or pediatric patients whose breathing patterns show excessive
variation; or for neonates whose breathing rhythms tend
to be irregular, and whose signals may not be reliably detected in Auto mode.
TO ADJUST RRi WAVEFORM SIZE
1. Select RRi parameter box.
2. Select Size % and select desired amplitude.
3. Dial rotary knob to select desired setting.
4. Click rotary knob to confirm.
TO DISPLAY RESPIRATORY MARKERS
1. Select RRi parameter box.
2. At Resp. marker select On or Off.
Note: Respiration markers indicate the time of breath detection. A white
vertical line displays on the Resp waveform where each breath is detected.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 24
INFINITY COCKPIT QUICK REFERENCE GUIDE
|24
SpO2 Monitoring
The accuracy of SpO2 monitoring depends largely on the strength
and quality of the SpO2 signal
TO CHANGE PULSE TONE VOLUME
1. Select SpO2 parameter box.
2. At Pulse tone volume %, select button.
3. Dial rotary knob to select desired setting.
4. Click rotary knob to confirm.
TO CHANGE PULSE TONE SOURCE
1. Select SpO2 parameter box.
2. Select Tone Source and select ECG or SpO2.
Note: Refer to M540 Instructions for Use manual for further device specific
information, Masimo or Nellcor.
NIBP - Non-Invasive Blood Pressure
TO START A NIBP SINGLE MEASUREMENT
1.
Press the NIBP Start / Stop fixed key
.
o Pressing the key again stops the measurement.
o NIBP inflation modes- Adult, pediatric, and neonatal- can be
selected in NIBP parameter box and Settings tab.
TO SET NIBP INTERVAL TIME
1. Select NIBP parameter box.
2. Select Interval time (min).
3. Dial rotary knob to select desired setting.
4. Click rotary knob to confirm.
5.
6.
Press NIBP Start / Stop fixed key
to initiate cycling.
An interval timer bar appears in the parameter box between
interval measurements. This indicates that the interval time
option is in use.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.
Page 25
INFINITY COCKPIT QUICK REFERENCE GUIDE
|25
Note: Pressing the NIBP start/stop button longer than two seconds suspends
interval mode and sets the Interval time [min] to Off. To resume interval
times, must re-set to desired time interval and initiate by pressing NIBP Start
/ Stop.
TO STOP A NIBP MEASUREMENT
1.
2.
Press the NIBP Start / Stop fixed key
l
To resume interval times, must re-set to desired time interval
and initiate by pressing NIBP Start / Stop.
Mainstream etCO2 Monitoring
CONNECTING THE CO2 SENSOR
1. Connect yellow end of CO2 sensor cable to yellow CO2 connector on M540.
2. Insert correct airway adapter for patient between endotracheal tube adapter and the ventilator Y-piece.
3. Snap the CO2 mainstream sensor firmly onto the airway
adapter and make sure the cable is directed away from the patient.
Note: Always position the sensor windows of the airway adapter vertically
to prevent patient secretions from obscuring the adapter window.
ADJUST GAS COMPENSATION
Compensates for supplemental O2 greater than 50% N2O or Heliox.
Failure to adjust compensation can cause inaccurate measurements.
1. Select etCO2 parameter box.
2. Select Gas compensation.
3. Dial rotary knob to correct compensation.
4. Click rotary knob to confirm.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
© 2015 Draeger Medical, Inc.