Drufomat scan 1. Description of unit The Drufomat scan is a universally usable pressure thermoforming unit for daily use in the dental laboratory. An integrated scanner allows a direct taking over of product names, blank thicknesses as well as individual heating and cooling times by means of a barcode that can be found on each blank package.
A variably adjustable forming pressure of 29-87 psi (2-6 bar) in connection with a vertically forming blank guarantees for very precise results. In combination with optionally available inserts the Drufomat scan can also be used as flask press (P insert) and as pressure pot (D insert).
2. Technical data Dimensions (H x W x D): Weight: Supply Voltage: Max. power consumption: Fuse of unit: Operating pressure: Min. operating pressure: Max. pressure in pipe:
30
460 x 290 x 290 mm 16 kg 100-240 V/50-60 Hz 115V-270 W, 230 V-315 W T 3,15 A 29-87 psi (2-6 bar) 29 psi (2 bar) 145 psi (10 bar)
english
deutsch
Applied harmonised standards: 2006/42/EG According to Machinary Directive EN ISO 12100:2010 Safety of machinery - General principles for design – Risk assessment and risk reduction (ISO 12100:2010); German version EN ISO 12100:2010. EN ISO 13732-1:2008 Ergonomics of the thermal enviroment - Methods for the assessment of human responses to contact with surfaces – Part 1: Hot surfaces (ISO 13732-1:2006); German version EN ISO 13732-1:2008-12. EN 60204-1:2006/A1:2009 Safety of machinery – Electrical equipment of machines – Part 1: Gen-eral requirements (IEC 60204-1:2005/A1 2008); German version EN 60204-1:2006/A1:2009. 2006/95/EG Low Voltage Directive EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General requirements (IEC 61010-1:2010 + Cor.: 2011); German version EN 61010-1:2010. 2004/108/EG EMC Directive EN 61000-3-2:2006 + A1:2009 + A2:2009 Electromagnetic compatibility (EMC) – Part 3-2: Limits – Limits for harmonic current emissions (equipment input current <= 16 A per phase) (IEC 61000-3-2 :2005 + A1:2008 + A2:2009); German version EN 61000-3-2:2006 + A1:2009 + A2:2009. EN 61000-3-3:2008 Electromagnetic compatibility (EMC) – Part 3-3: Limits – Limitation of voItage changes, voItage fluctuations and flicker in public Iow-voltage supply systems, for equipment with rated current <= 16 A per phase and not subject to conditinoal connection (IEC 61000-3-3:2008; German version EN 61000-3-3:2008. EN 61326-1:2006 Electrical equipment for measurement, control and laboratory use – EMC requirements; Part 1: General requirements (IEC 61326-1:2005); German version EN 61326-1:2006 Emission according to living area, business and industrial undertakings as well as small enterprises interference stability according to industrial area.
français
Designation of machine: Drufomat scan D3300 / D3300X1 / D3300X2 Type of machine: Pressure thermoforming unit Applicable EC-guidelines: 2006/42/EC Machinery Directive 2006/95/EC Low Voltage Directive 2004/108/EC EMC Directive
español
According to Machinary Directive (2006/42/EG) Herewith we declare that the machine described below is in its conception and design and in the shape delivered by us in accordance with the fundamental requirement for safety and health as prescribed in the applicable EC-guidelines. In the case of any change or modification of the machine not authorised by us this declaration becomes invalid.
italiano
3. Declaration of Conformity
Dreve Dentamid GmbH · Max-Planck-Straße 31 · 59423 Unna/Germany Tel.: +49 2303 8807-0 · Fax: +49 2303 8807-55 · [email protected] · www.dreve.com/dentamid
31
Drufomat scan 4. Safety advice Attention! Read the following notes before connecting and putting the unit into service. The operating service and the function of the unit can only be guaranteed for, if the general safety and accident prevention instructions of the legislation as well as the safety instructions of the working instructions are observed. 1. This unit must only be used according to the present working instructions. We cannot be held liable for damages which have been caused by improper treatment resp. faulty manners of operation. 2. Put the unit onto a stable and even basis which is suitable for the operation weight of approx. 20 kg of the unit. 3. In order to avoid the intrusion of water into the unit (e. g. splash water) the unit is to be placed in a dry surrounding. 4. Do not store any flammable substances in the close surrounding of the unit. 5. The indicated voltage on the type plate has to meet the voltage of the power source. 6. The unit must only be connected with a socket with protective conductor. Never touch the plug with wet hands. 7. Authorised operators: The operating company of the machine has to hold the working instructions at the operators’ disposal and has to make sure that authorised operators have read
32
and understood the manual. Only in that case the operators may operate the unit. 8. The unit has to be examined before operation in view of proper condition and operating safety. If the unit is not in proper condition, it must not be used and has to be marked accordingly. 9. The pressure of the compressed air line must not exceed 145 psi. 10. Do not intrude any objects into the unit. 11. Do not lock-up the operation elements of the unit. 12. Do not block the downwards-moving piston. 13. Do not touch the downwards-moving piston. 14. The swivelling housing of the radiator reaches high temperatures and must therefore not be touched - not even for short moments. 15. If the unit is not used, please disconnect the mains.
deutsch
20. Arbitrary rebuilding or modifications are not permitted due to safety reasons.
18. The opening of the unit and repairs are only to be carried out by accordingly trained experts.
21. The required operation and maintenance conditions mentioned in these working instructions are to be complied with imperatively.
italiano
19. Only those appliances and spare parts may be used which are released by
english
17. Before cleaning and maintenance of the unit or the exchange of parts the plug has to be disconnected.
français
the manufacturer. We are not responsible for damages caused by the use of foreign parts.
español
16. Plates and labels have to be maintained in a well legible condition and must not be removed.
5. Initiation Before initiation make sure that the voltage indicated on the type plate meets the voltage of the power source. Connect the unit to an air compressor. Push the pressure hose (included in the delivery) into the compressed air connection (9) until it is locked in place. The hose does not need any further fastening. To release the pressure hose push the blue ring upwards and draw the hose off by pulling downwards. The pressure of the compressed air line must not exceed 145 psi. Put the power cord into the female connector (33) and create a connection with the Euro/USA plug.
Important: Before the first initiation please read the working instructions thoroughly. In case of comprehension problems contact your responsible dealer or Dreve Dentamid GmbH, if necessary. Scope of delivery 1 x Drufomat scan 1 x drawer for granules 1 x F insert incl. insert bowl and nub plate 1 x tension ring with exhaust valve 1 x base range of blanks 1 x bowl for granules 1 x power cord 1 x air connection hose incl. hose clamps and fast coupling 1 x working instruction
33
Drufomat scan 6. Functional components 1 2 3
8 9
4
10
5 12
11 13
6 7
No. 1 2 3 4 5 6 7 8 9 10 11 12 13 33 34
Description scanner display keypad heating element handle heating F insert drawer for granules push button compressed air supply handle plate reception plate reception exhaust valve tension ring mains supply/main breaker
deutsch
a g
b
français
english
7. Display
h
d i
e f
j
Letter a b c d e f g h i j
italiano
español
c
Description indication of heating time indication of product name and material strength indication of cooling time change of menu items (backwards) change of time indications and pressure (upwards) confirmation key/start button heating indication of pressure inside piston control indication for individual time modifications change of time indications and pressure (downwards) change of menu items (forwards)
35
Drufomat scan 8. Operation Turn the unit on by means of the main switch (33) at the back of the unit. By this the scanner (1) and the display (2) at the front of the unit will be activated. An acoustic signal shows that the unit is ready for operation; the display will show the name of the machine Drufomat scan.
8.1 Scan modus Every outer package of Dreve thermoforming blanks is labelled with a barcode label.
This barcode contains the encrypted product name, material strength as well as the individual heating and cooling times. In order to scan the material data with the Drufomat scan the barcode is to be placed as centrally and evenly as possible onto the scan field (1).
36
There will be an acoustic signal for the confirmation of the data capture and the display will show all necessary data.
Info The light emanating from the scanner is harmless. Even in case of direct eye contact it has no negative impact! After 15 min. without use the scanner turns off automatically. By pressing any key it can be reactivated again. During an active thermoforming process the scanner is deactivated. After termination of the process it will be re-activated automatically.
english
deutsch
Info The minimum operating pressure is 29 psi (2 bars). If a thermoforming process is interrupted before having reached the 29 psi (2 bar) mark (indication of bar chart (g) in the display), the piston will be ventilated and will return to its original position.
français
The arrow to the left of the bar chart shows the currently adjusted pressure. The arrow can be moved along the bar upwards and downwards with the +/keys (e/i). The pressure can also be modified during an ongoing thermoforming process!
español
The pressure of the Drufomat scan is controlled digitally and can individually be adapted to the thermoforming blank in question between 29 and 87 psi infinitely variable. It is possible to control and to modify the adjusted pressure and the actual pressure during the thermoforming process via the bar chart (g) of the display.
italiano
8.2 Pressure adjustment
8.3 Start of programme In order to start the selected programme, put a blank that corresponds to the entered barcode resp. programming onto the plate reception (11) and fix it by means of the tension ring (13). The model can either be put onto the nub plate of the F insert (6) or fixed by taking out the nub plate and embedding it into granules directly in the F insert. Info For an optimum adaptation of the blank towards the model always place the models in a way that the anterior teeth are directed towards the centre of the blank. When using hard blanks it is recommended to embed the models into granules. Models that are placed onto the nub plate should not exceed a max. height of 23 mm.
Swivel the plate reception inwards at the handle (10) just up to the mechanical interlock in the centre of the unit. Also swivel in the heating element (4) at the corresponding handle (5) just up to the mechanical interlock in the centre of the unit. Close the ventilation valves of the tension ring (12) and the F insert (6). Press the OK key (f). There will be an acoustic signal as confirmation for the activation of the heating; the scanner will turn off and the adjusted heating time will elapse in the display. As safety and identifying feature for an active heating there will be four LED alight at the front bottom edge of the housing.
37
Drufomat scan
Info For safety reasons the quartz radiator is equipped with a final switch and only works in swivelled in position after having started a programme. If the quartz radiator is swivelled outwards after having started a programme, the heating process will be interrupted automatically. The indication in the display (a) will stop and show the remaining time. By swivelling the quartz radiator back in the programme will be re-activated.
Therefore it must not be touched – even not for short moments. After the adjusted time has elapsed the indicator (a) in the display will stop and count the duration after termination of the heating time. This process may be stopped by abort (long pressing of button 8) or by successive thermoforming.
Attention! The swivelling housing of the heating element reaches high temperatures.
8.4 Thermoforming After the heating time has elapsed there will be an acoustic signal; the heating time in the display will be 00:00. Heating and LED illumination will be switched off automatically.Pull the heating (4) with the right hand at the knob rod (5) as far as it will go to the front and hold it tight. Then push the push button (8) with the left hand at the left side of the housing. The piston will move downwards and close the pressure chamber; an acoustic signal will be heard. Info The closing pressure of the downwards moving piston is always higher than the forming pressure! 38
The downwards moving piston is equipped with a dual electronic safety feature and can only be activated, if the unit is operated with two hands at the same time (twohand control). Always make sure that no objects are in the contact area between piston and F insert while the piston is closing! Hold both buttons until the pressure bar in the display (g) shows the adjusted minimum of 29 psi (2 bars). Then release the push button (8) and finally the knob rod of the heating (5). After having reached the 29 psi (2 bar) mark the pressure will automatically rise up to the pre-adjustment.
français
english
deutsch
The adjusted pressure will be verified by the electronics and regulated, if necessary. The display shows the adjusted cooling time (c) which then counts down to 00:00.
español
will show the last stored values. In order to start the thermoforming process newly the process can be repeated immediately, as described in this chapter.
italiano
Info For safety reasons the process of an unwanted downwards movement of the piston or an interlock of the escapement can be stopped by the electronics: a) The process of moving the piston downwards is carried out too slowly (max. 3 seconds between action right hand and action left hand) or b) The two-hand control is aborted before reaching the minimum pressure. In both cases there will be 5 short acoustic warnings, the piston will be ventilated and remove to its original position. The display
8.5 Opening of the pressure chamber In order to attain a complete cool down of the thermoforming blank the valve at the tension ring (12) is to be opened slightly after 2/3 of the adjusted cooling time. Through this the warm air inside the piston will be released and replaced by cool air due to the pressure regulation of the electronics. After the cooling time has elapsed there will be an acoustic signal (indication in the display (c) will blink).
Directly before releasing the pressure inside the piston the ventilation valve at the F insert (6) should be opened. Push the push button (8) for approx. two seconds; the pressure inside the piston will be released. After complete ventilation the piston will remove to its original position. Swivel out the plate reception (11) at the corresponding handle (10) and remove tension ring and thermoformed result. The display (2) will show the last used programme; the scanning device (1) is ready for a new thermoforming process.
39
Drufomat scan 9. Deviation to the rule 9.1 Modification of given times Scanned times may be prolonged or shortened individually, if required. Info The performance of the heating might decrease slightly after 1000 operating hours. By prolonging the heating times by 5-10 seconds optimum thermoforming results can be attained again. The times given by the barcodes are average times and designed for general use. In case of extreme undercut areas or very high models a prolongation of the heating time can optimise the results. After having scanned the pre-adjusted material parameters (see paragraph 8.1) the heating time (1 x) or the cooling time (2 x) can be selected by pressing the menu key (j). There will be an arrow in the display in front of the chosen time. The heating time can be modified positively and negatively in steps of 5 seconds. For this press the menu keys (e/i) according to the key description. The max. deviation from the scanned time is 35 seconds.
40
With the same working steps the cooling time can be modified in steps of 10 seconds by max. 70 seconds.
For identification of the modified paraMeters the display will show a + resp. a - sign. The degree of deviation will be shown by a scale at the bottom of the display. Modified values have to be confirmed by pressing the OK key (f). All further working steps correspond to 8.2.
english
deutsch
If the created time programme is to be stored for repeated use, 3 memory capacities (F1-F3) are available. Do not confirm the created time programme by pushing OK, but press menu key (j) again. The display will show F1 in position b, the entered times will be taken over. The memory capacities F2 and F3 can be selected by pushing the menu key (j) 2 resp. 3 times. Store the times in the desired memory capacity by pushing the OK key (f). Then go on as described in point 8.2.
français
Turn on the unit – the initial display will be visible. When pushing the menu key (j) one time the display will show – – – – – at position b.
For single use confirm the two entered values by pushing OK (f). Then go on as described in point 8.2.
español
Up to three individual time programmes may be created and stored for thermoforming foils that are not equipped with a corresponding barcode.
italiano
9.2 Creation and storage of manual time programmes
As described in point 9.1, push menu key (j) and select the desired heating time by the menu keys +/- (e/i). Following push menu key (j) again and enter cooling time.
41
Drufomat scan 9.3 Calling of individual time programmes Turn on the unit – the initial display will be visible. By pushing the menu key (j) several times you may select one after another a) enter modus b) F1 c) F2 d) F3
Then go on as described in point 8.2.
By pressing the menu keys (d/j) you may navigate arbitrarily between the a. m. programme points. Confirm the selected programme point with the OK key (f).
9.4 Plasticizing of blanks with unknown heating times Create an individual time programme with high heating and cooling times (see point 9.2). Start the programme (see point 8.2). You may test the plasticization degree of the blank by means of an edgeless metal instrument (e. g. wax knife). Info The complete deformability of the blank can only be tested at the edges, as that is where warmth
will leak through the metal contacts and the necessary flexibility will be attained at last! After the blank has attained the necessary flexibility initiate the thermoforming process (see point 8.4). The remaining heating time will be erased from the display (a) and the adjusted cooling time will be counted down (c). The cooling time can be terminated at any time by pushing the push button (8).
9.5 Abort/termination of programme It is possible to abort / terminate all actions of the Drufomat scan by pushing the push button (8) at the left side
42
of the unit (keep pushed for aprox. 2 seconds!).
deutsch
9.7 Entering a wrong barcode If a barcode is scanned that does not correspond to the programming of the Drufomat scan, four acoustic warning signals will be heard and the display will show ERROR at position b.
By scanning the correct barcode you can proceed further, as described in point 8.0.
español
français
will be blinking in the display as control mechanism.
italiano
After elapse of times adjusted regularly (display position a/c) there will be an acoustic signal. The exceeded time
english
9.6 Exceeding of entered heating/cooling times
43
Drufomat scan 10 Optional additional functions 10.1 Pressure polymerisation In combination with the optionally available D insert the Drufomat scan can be used for pressure polymerisation of resin repairs or as compressor of plasters, embedding masses and silicones. Turn on the unit, swivel out plate reception (11) and heating element completely (4) by means of the operating knobs (10/5) and take the F insert (6) out of the bottom plate. The material to be compressed (e. g. cast flask, duplication etc.) is to be placed inside the D insert.
For the polymerisation of resins the D insert is to be filled before with hand warm water (corresponding to the indications of the resin manufacturer). Fix the D insert in the bottom plate. By moving the piston downwards the D insert will be closed and will build up the pre-adjusted pressure inside.
For this purpose go on as described in point 8.4.
44
deutsch
Place flasks onto the nub plate. By moving the piston with the P insert downwards the necessary pressure will be exerted upon the flasks.
italiano
español
Turn on the unit, swivel out plate reception (11) and heating element completely (4) by means of the operating knobs (10/5) and take the F insert (6) out of the bottom plate. Insert the P insert into the pressure chamber from underneath.
english
In combination with the optionally available P insert the Drufomat scan can also be used as flask press. One or two flasks may be pressed at the same time.
français
10.2 Flask press
For this purpose go on as described in point 8.4.
For one flask: Put back the F insert (6) into the bottom plate. For two flasks: Remove nub plate and insert bowl from F insert and insert them into the bottom plate.
Info In order to attain a homogeneous distribution of the material to be compressed in the flask we recommend an initial pressure of 29 psi (2 bars) which can be increased step by step up to 87 psi (6 bars) during the pressing procedure (see point 8.2).
45
Drufomat scan 10.3 Fabrication of cast form parts With the K-7 insert it is possible to thermoform up to 7 resin bases for the crown and bridge technique at the same time. Fix the single dies by means of a flexible blocking out material (e. g. Fillin) in the holes of the K-7 insert. Remove the nub plate from the F insert (6) and replace it by the K-7 insert. All further working steps as described from point 8.1 on.
11. Care and maintenance Generally disconnect the unit from mains supply before executing care and maintenance. Ideally cleaning is to be effected dryly with a soft piece of cloth, if necessary with a slightly humid sponge and a mild cleaning liquid. Water and cleaning liquids must not get into the unit. The O-rings should be greased regularly with silicone grease.
46
Removal and cleaning of the filter sieve Disconnect the power cable from the socket and shut down the compressed air supply. Push the blue ring of the compressed air connection upwards and draw the hose off the unit by pulling it downwards. Now unscrew the connection using a suitable tool. Then carefully remove the filter sieve with a slotted screwdriver by turning anti-clockwise. Clean with stream of compressed air and reassemble in reverse order.
deutsch english
12. Exchange of fuses
In order to exchange the fuse press the locking cap slightly inside and turn it to the left in accordance with the arrow.
Insert fuse carrier again and push it slightly to the right.
italiano
Attention: Only use the fuses indicated on the back of the unit.
español
français
Separate power cord from the mains. The fuse housing is to be found on the left back side of the unit.
Take out fuse carrier. Pull the defect fuse carefully out of the clamp and replace it by a new one.