airvo_2_user_manual_rev_g_may_2016.pdf
Page 1
AIRVO 2
TM
User Manual
Page 3
• This User Manual is intended for healthcare professionals.
• This User Manual applies to AIRVO 2 units with manufacturing date after 2016-06-06.
• Read this User Manual including all warnings. Failure to do so may result in injury. In addition, watch the
AIRVO 2 Video Guide. Keep them both in a safe place for future reference.
• Before the AIRVO 2 is used for the first time, it must be set up according to the instructions in the
AIRVO 2 Technical Manual.
• Some accessories may not be available in certain countries. Please contact your local Fisher & Paykel
Healthcare representative for more information.
OTHER REFERENCES
•
•
•
•
•
Refer to the AIRVO 2 User Manual for detailed instructions for use.
Refer to all relevant accessory User Instructions.
Watch the training videos on the AIRVO website www.fphcare.com/airvo
For troubleshooting information, please refer to the AIRVO 2 Technical Manual.
Download the AIRVO 2 Simulator App to learn how to use the AIRVO 2.
You can change settings, simulate faults and test your skills.
Available from the Apple, Google Play and Windows App stores.
• Visit the Fisher & Paykel education & resources website at www.fphcare.co.nz/education
to find self-paced online courses and local training events.
• If the unit is ever used by multiple patients, the unit must be cleaned and disinfected between patients
according to instructions in the Disinfection Kit Manual (900PT600).
• For further assistance, please contact your Fisher & Paykel Healthcare representative.
TABLE OF CONTENTS
1. Overview ...................................................................................................................................................... A - 2
Intended Use ................................................................................................................................................................................ A - 2
Warnings ........................................................................................................................................................................................ A - 2
AIRVO 2 and Accessories ........................................................................................................................................................ A - 3
2. Setting up AIRVO 2 .................................................................................................................................. A - 4
3. Using AIRVO 2 ........................................................................................................................................... A - 6
Target dew-point temperature............................................................................................................................................... A - 7
Target flow...................................................................................................................................................................................... A - 7
Oxygen .......................................................................................................................................................................................... A - 8
Alarms
.......................................................................................................................................................................................... A - 10
4. Reprocessing .............................................................................................................................................. A - 12
Schedule for changing accessories....................................................................................................................................... A - 12
Filter replacement....................................................................................................................................................................... A - 12
Servicing.......................................................................................................................................................................................... A - 12
5. Technical Information .............................................................................................................................. A - 13
A–1
English
BEFORE YOU START
Page 4
1. OVERVIEW
The AIRVO 2 is a humidifier with integrated flow generator that delivers high flow warmed and humidified
respiratory gases to spontaneously breathing patients through a variety of patient interfaces.
INTENDED USE
The AIRVO 2 is for the treatment of spontaneously breathing patients who would benefit from receiving
high flow warmed and humidified respiratory gases. This includes patients who have had upper airways
bypassed. The flow may be from 2 - 60L/min depending on the patient interface. The AIRVO 2 is for patients
in hospitals and long-term care facilities.
The AIRVO 2 provides respiratory support through a nasal cannula interface, promoting slower, deeper
breathing and increasing alveolar ventilation. To patients with either nasal cannula or tracheostomy interfaces,
the AIRVO 2 provides airway hydration and may be used with supplemental oxygen, as required.
!
WARNINGS
• The unit is not intended for life support.
• Appropriate patient monitoring must be used at all times. Loss of therapy will occur if power is lost.
• Nasal delivery of respiratory gases generates flow-dependent positive airway pressure (PAP). This must
be taken into account where PAP could have adverse effects on a patient.
To avoid burns:
• The unit should only be used with interfaces, water chambers and breathing tubes specified in this user
manual.
• Using the breathing tube or interface for longer than the specified time can result in serious injury
including infection.
• Before using oxygen with the unit, read all warnings in the “Oxygen” section of this manual.
• Never operate the unit if:
• the heated breathing tube has been damaged with holes, tears or kinks,
• it is not working properly,
• the case screws have ever been loosened.
• Do not block the flow of the air through the unit and breathing tube.
• The unit should be located in a position where ventilation around the unit is not restricted.
• Never block the air openings of the unit or place it on a soft surface such as a bed or couch/sofa, where
the filter area may be blocked. Keep the air openings free of lint, hair etc.
To avoid electric shock:
• Do not store or use the unit where it can fall or be pulled into water. If water has entered the unit enclosure,
disconnect the power cord and discontinue use.
• Never operate the unit if:
• it has been dropped or damaged,
• it has a damaged power cord or plug,
• it has been dropped into water.
• Avoid unnecessary removal of the power cord from the rear of the device. If removal is necessary, hold the
connector during removal. Avoid pulling on the power cord.
• Return the unit to an authorized service center for examination and repair, except as outlined in this
manual.
To avoid choking, or inhalation of a foreign object:
• Ensure an air filter is fitted when operating your unit.
• Never drop or insert any object into any opening or tube.
Miscellaneous:
• Do not use the unit when the room temperature exceeds 30°C (86°F) or is below 10°C (50°F). Humidity
output will be compromised below 18°C (64°F) and above 28°C (82°F).
• The unit is not suitable for use in the presence of a flammable, anesthetic mixture with air or oxygen or
nitrous oxide.
• Prior to each patient use, ensure that the auditory alarm signal is audible by conducting the alarm system
functionality check described in the Alarms section.
A–2
Page 5
AUDIO PAUSE
ON/OFF (STANDBY)
Heated
breathing
tube
UP
Patient
interface
DISPLAY
DOWN
MODE
HEATED BREATHING TUBE
CONNECTION PORT
HOSPITAL
STAND
POLE
MOUNTING
TRAY
OXYGEN
INLET PORT
MEASUREMENT POINT OF
DISPLAYED DEW POINT
TEMPERATURE
CHAMBER PORTS
SERIAL PORT
POWER CORD
and
CONNECTOR
AIRVO 2
FILTER COVER
PT101xx
HEATER
PLATE
FINGER
GUARD
AUTO-FILL WATER
CHAMBER (MR290)
(with adapter fitted)
Water chamber
AIR FILTER
Nasal Cannula (20-pack)
Tube &
Chamber kits
(10-pack)
RT013
(Mask Adapter)
OPT870
(Direct Trache)
OPT846
(large)
OPT844
(medium)
900PT501
900PT531
AirSpiral™
900PT551*
900PT561*
Cleaning and Disinfection
900PT600
900PT601
900PT602
900PT603
OPT842
(small)
Optiflow™
OPT980
(Mask Adapter)
OPT970
(Direct Trache)
OPT946
(large)
OPT944
(medium)
Optiflow™+
OPT942
(small)
OPT318
(pediatric)
OPT316
(infant)
Optiflow™
Junior
Miscellaneous
Disinfection Kit
Disinfection Filter (2-Pack)
Cleaning Sponge-Stick (20-Pack)
Clean Storage Cover (20-Pack)
900PT405
900PT411
900PT420
900PT421
900PT422
900PT426
900PT427
900PT427L
900PT428
900PT912
900PT913
OPT012
OPT014
A–3
Pole mounting tray
UPS mounting kit
Mobile Pole Stand (extendable)
Mobile Pole Stand
Oxygen inlet extension kit
Plastic Basket
Oxygen bottle holder
Oxygen bottle holder (large)
Pole Clamp
Filter holder
Air filter (2-Pack)
Wigglepads (OPT316/OPT318) (20-pack)
Oxygen Tubing (Optiflow Junior)
English
AIRVO 2 AND ACCESSORIES
Page 6
2. SETTING UP AIRVO 2
1. BEFORE YOU BEGIN
The AIRVO 2 should be fixed on a pole mounting tray (900PT405) below
patient head height.
Open the packaging of the tube & chamber kit (heated breathing tube,
MR290 auto-fill chamber and adapter).
2. INSTALL WATER CHAMBER
Remove the blue port caps from the chamber by pulling the tear tab
upwards then remove the bracket holding the water supply tube.
Fit the supplied adapter over the two vertical ports on the chamber and
push on fully then clip the water supply tube into position.
Fit the water chamber to the unit by pressing down the finger guard and
sliding the chamber on, carefully aligning with the blue chamber port
ends.
Push the chamber on firmly until the finger guard clicks into place.
WARNINGS
To avoid burns:
• Do not start the unit without the water chamber in place.
• The water in the chamber becomes hot during use. Exercise caution when
removing and emptying the chamber.
• Do not touch the heater plate, water chamber or chamber base during use.
To avoid electric shock:
• When handling the unit with the water chamber in place, avoid tilting the
machine to prevent any chance of water entering the unit enclosure.
• Empty all the water from the water chamber before transporting the unit.
CAUTIONS
To ensure optimal therapy (MR290 only):
• Do not use the auto-fill MR290 chamber if it has been dropped or been allowed
to run dry this could lead to the chamber over filling.
• Do not use the MR290 chamber if the water level rises above the maximum water
level line as this may lead to water entering the patient’s airway.
3. CONNECT WATER BAG
H2O
Attach the sterile water bag to the hanging bracket 20cm (8”) above the
unit, and push the bag spike into the fitting at the bottom of the bag.
Open the vent cap on the side of the bag spike. The chamber will now
automatically fill to the required level and maintain that level until the
water bag is empty.
To ensure continual humidification, always ensure that the water chamber
and/or water bag are not allowed to run out of water.
Check that water flows into the chamber and is maintained below the
fill line. If the water level rises above the fill line, replace the chamber
immediately.
MR290: Flow setting vs usage time (2-litre sterile water bag)
L/min
2
5
10
15
20
25
30
35
40
45
50
55
60
hrs
378
151
75
50
37
30
25
21
18
16
15
13
12
A–4
Page 7
One end of the heated breathing tube has a blue plastic sleeve. Lift the
sleeve and slide the connector onto the unit. Push the sleeve down to
lock.
WARNINGS
To avoid burns:
• Do not modify the breathing tube or interface in any way.
• Do not allow the breathing tube to remain in direct contact with skin for
prolonged periods of time.
• Adding heat, above ambient levels, to any part of the breathing tube or interface
e.g. covering with a blanket, or heating it in an incubator or overhead heater for a
neonate, could result in serious injury.
• Do not use an insulating sleeve or any similar accessories which are not
recommended by Fisher & Paykel Healthcare.
CAUTIONS
• Position the heated breathing tube away from any electrical monitoring leads
(EEG, ECG/EKG, EMG, etc), to minimize any possible interference with the
monitored signal.
5. SELECT PATIENT INTERFACE
The AIRVO 2 can be used with a variety of patient interfaces. Read the separate user instructions for the
patient interface that will be used, including all warnings.
Nasal cannula
Tracheostomy
interface
2
VO
AIR
Mask interface
adapter
TM
Optiflow™+
OPT942
OPT944
OPT946
Optiflow™
OPT842
OPT844
OPT846
Optiflow™ Junior
OPT316, OPT318
(Refer to “Using AIRVO 2” ”Junior Mode”)
OPT970 /
OPT870
OPT980 / RT013 (with mask)
Note that the RT013 Mask Interface Adapter
is designed to be used with vented masks
only. Do not use sealed masks.
The following table shows the target dew-point temperature settings and target flow settings able to be
used with these interfaces.
Patient Interface
OPT316
OPT942
2
OPT318
OPT944
2
10
OPT942
OPT946
OPT944
OPT870
10
OPT946
RT013
10
OPT970
10
OPT980
10
OPT842
10
OPT844
10
OPT846
10
OPT870
10
RT013
10
Low temperature ambient conditions may prevent the unit from reaching a 37 °C target temperature
setting at high target flow settings. In these cases, consider decreasing the target flow setting.
WARNINGS
To avoid burns:
• Do not modify the breathing tube or interface in any way.
• Do not use any patient interfaces not listed here.
A–5
English
4. INSTALL HEATED BREATHING TUBE
Page 8
3. USING AIRVO 2
1. SWITCH ON UNIT
Plug the unit’s power cord into the mains power supply. The connector
at the other end of the power cord should be well secured to the rear of
the unit.
WARNINGS
To avoid electric shock:
• Ensure that the unit is dry before plugging into the power socket.
Switch on the unit by pressing the On/Off button.
2. CHECK DISINFECTION STATUS
The unit will show you whether it is safe for use on a new patient.
Last
Disinfection:
#16
This AIRVO 2 is safe for use on a new patient.
This AIRVO 2 has not been cleaned and disinfected since last use.
This AIRVO 2 is NOT safe for use on a new patient.
3. WARM-UP
The unit will begin to warm up. You will see numbers showing the
current output dew-point temperature, flow and oxygen values. These
numbers will pulse until they approach their target settings.
This screen is called the “Summary screen”.
4. JUNIOR MODE
If the patient will be using an Optiflow Junior nasal cannula (OPT316/
OPT318), you must activate Junior Mode.
Junior Mode limits the target settings to: 34 °C and 2 - 25 L/min, in
increments of 1 L/min.
To activate Junior Mode:
Hold the Mode button for 5 seconds.
New target settings
The target settings for dew-point temperature and
flow will be changed automatically. The colorful icons
in the corners of the screen indicate that this unit is in
Junior Mode.
New target
settings
To deactivate Junior Mode, follow the same
procedure: hold the Mode button for 5 seconds.
A–6
Page 9
Press the Mode button to view target settings.
These settings are locked by default.
TARGET DEW-POINT TEMPERATURE
You can set the AIRVO 2 to three target dew-point temperature
settings:
•
37°C (98.6°F)
•
34°C (93°F) [if compliance at 37°C is a problem]
•
31°C (88°F) [for face masks only].
You may not have access to all settings, if:
•
the unit is in Junior Mode (limited to 34 °C),
•
the unit was initially set up with tighter limits.
The AIRVO 2 will return to its default setting (37°C) after every
disinfection cycle.
To change the target
dew-point temperature setting:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
The lock will disappear and be replaced by an arrow
showing the minimum and maximum accessible
settings. Press the Up and Down buttons to choose
the new setting.
When you have finished, press the Mode button to
'lock' the setting again.
The lock will reappear.
Press the Mode button to move on to the next screen.
TARGET FLOW
You can set the AIRVO 2 to flows between 10 L/min and 60 L/min, in
increments of 1 L/min (10-25 L/min) and 5 L/min (25-60 L/min).
You may not have access to all settings, if:
•
the unit is in Junior Mode (limited to 2 - 25 L/min,
in increments of 1 L/min),
•
the unit was initially set up with tighter limits.
The AIRVO 2 will remember its target flow setting when you switch it
off.
To change the target flow setting:
Follow the same sequence of steps as above in “To change the
target dew-point temperature setting”.
A–7
English
5. CONFIGURE TARGET SETTINGS
Page 10
Press the Mode button to move on to the next screen.
OXYGEN
You can connect supplementary oxygen to the AIRVO 2 (up to
60 L/min). The AIRVO 2 contains an oxygen analyzer to help you
determine the oxygen fraction you are delivering to the patient. Your
unit may have been initially set up with tighter limits.
Use continuous oxygen monitoring on patients who would desaturate
significantly in the event of disruption to their oxygen supply.
WARNINGS
Before using the AIRVO 2 with oxygen, read all of the following warnings:
• The use of oxygen requires that special care be taken to reduce the risk of fire.
Accordingly, for safety it is necessary that all sources of ignition be kept away
from the unit and preferably out of the room in which it is being used. Oxygen
should not be used while smoking or in the presence of an open flame. The
unit should be located in a position where ventilation around the unit is not
restricted.
• A spontaneous and violent ignition may occur if oil, grease or greasy
substances come in contact with oxygen under pressure. These substances
must be kept away from all oxygen equipment.
• Ensure that the AIRVO 2 is switched on before connecting oxygen.
• Oxygen must only be added through the special oxygen inlet port on the back
of the unit. To ensure that oxygen enters the unit correctly, the oxygen inlet port
must be fitted properly to the filter holder and the filter holder must be fitted
properly to the unit. The power cord connector should also be well secured.
• Do not connect supplementary oxygen to the AIRVO 2 at flow rates higher
than the AIRVO 2 target flow rate, as excess oxygen will be vented into the
surroundings, or 60 L/min.
• The oxygen concentration delivered to the patient can be affected by changes
to the flow setting, oxygen setting, patient interface or if the airpath is
obstructed.
• When finished, turn off the oxygen source. Remove the output of the oxygen
source from the oxygen inlet port on the back of the unit. The oxygen flow must
be turned off when the unit is not operating, so that oxygen does not build up
inside the device.
• The oxygen analyzer within the AIRVO 2 uses ultrasonic measurement
technology. It does not require in-field calibration. It is designed for use with
pure oxygen - connecting any other gases or mixtures of gases will cause it to
function incorrectly.
CONNECT OXYGEN
Connect the output from the oxygen source to the oxygen inlet port on
the back of the unit. Make sure you push the oxygen tube firmly onto
this connection port.
ADJUST OXYGEN
Adjust the level of oxygen from the oxygen source, until the desired
oxygen fraction is displayed onscreen. It may take the reading several
minutes to settle. You can set the oxygen fraction between the
maximum and minimum values displayed above and below the arrow.
Real-time O2 measurement is displayed when O2 > 25% and O2 < 95%.
However, note that oxygen fractions below 25% and above 95% will be
displayed as 21% and 100% respectively.
If the oxygen fraction exceeds 95%, the oxygen reading will pulse red
and the device will beep.
WARNINGS
• Note that if the patient’s peak inspiratory demand exceeds the flow delivered
by the unit, the fraction of oxygen inspired by the patient will be lower than the
value shown onscreen, due to the additional entrainment of ambient air.
• Check that suitable blood saturation levels are achieved at the prescribed flow.
Press the Mode button to return to the Summary screen.
A–8
Page 11
6. CONNECT YOUR PATIENT
“Ready for use” symbol
Connect the patient interface to the heated breathing tube.
Monitor the flow and oxygen values displayed on the Summary screen.
Adjust the level of oxygen from the oxygen source as necessary.
When the patient first uses the unit, the air will feel warm. This is
normal. The patient should continue to breathe normally through the
nose and/or mouth, or tracheostomy.
7. DURING USE
If the “Ready for use” symbol has been displayed for 2 minutes and no
button has been pushed in this time, a screensaver will be launched.
CONDENSATE MANAGEMENT
The unit must be placed below head height and flat, this allows
condensate to drain towards the water chamber, away from the patient.
If excess condensate accumulates in the heated breathing tube,
disconnect the patient interface from the heated breathing tube, drain
the condensate by lifting the patient end of the tube, allowing the
condensate to run into the water chamber.
At higher target flow rates, it may be necessary to first reduce the
target flow rate to 30 L/min or below, to ensure the condensate drains
into the water chamber.
Minimize local sources of cooling acting on the heated breathing tube,
such as a fan to cool the patient, or an air-conditioning unit/vent.
If condensate persists, consider turning the target temperature down.
Note, a lower target temperature will decrease the humidity output of
the unit, decreasing the level of condensation.
Note: The temperature and humidity level delivered to the patient will
also be reduced.
8. AFTER USE
Switch off the unit by pressing the On/Off button.
A–9
English
Wait until the “Ready for use” symbol is displayed on the Summary
screen.
Page 12
ALARMS
The AIRVO 2 has visual and auditory alarms to warn you about interruptions to your patient’s treatment.
These alarms are generated by an intelligent alarm system, which processes information from the sensors
and target settings of the unit and compares this information to pre-programmed limits.
ALARM SIGNALS
Symbols
Meaning
Visual alarm signal
Alarm condition.
(message)
Audio paused.
Auditory alarm signal
Press this button to mute the auditory alarm for 115 seconds.
The auditory alarm can be reactivated by pressing this
button again.
3 beeps in 3 seconds.
Repeated every 5 seconds.
ALARM CONDITIONS
All of the alarms listed below have been assessed as “Medium Priority”. These priorities have been
allocated for an operator’s position within 1 meter of the device. The unit also uses an internal priorityranking system. If multiple alarm conditions occur simultaneously, the unit will display the highest-priority
alarm.
The following table lists all of the alarm conditions from highest-priority to lowest priority, their causes,
possible solutions and delays. Alarm conditions that affect oxygen delivery require an immediate response
to assess the patient’s saturation levels. Alarm conditions that affect humidity delivery require a prompt
response to assess potential drying of mucus and associated blockages.
The following alarm delays assume operation in ‘Ready for use’ mode.
Message
Fault
(E###)
Check tube
Check for
leaks
Check for
blockages
Meaning
Affects
delivery of:
Delays
The unit has detected an internal fault and has shut itself down.
Switch the unit off and then restart. If the problem persists, note the fault
code and contact your Fisher & Paykel Healthcare representative.
Oxygen,
humidity.
< 5 seconds
The unit cannot detect the heated breathing tube.
Check that the heated breathing tube is not damaged and that it is
plugged in correctly. If the problem persists, then change the heated
breathing tube.
Oxygen,
humidity.
< 5 seconds
The unit has detected a leak in the system.
The most likely cause is that the water chamber has been removed or has
not been pushed into place correctly.
Check that the heated breathing tube is not damaged and that it is
plugged in correctly.
Check that the nasal interface is fitted.
Check that the filter is fitted.
Oxygen,
humidity.
< 120 seconds
The unit has detected a blockage in the system.
Check the heated breathing tube or patient interface for blockage.
Check the air filter and filter holder for blockage.
Check whether the unit should be in Junior Mode. If the patient will be
using an Optiflow Junior nasal cannula (OPT316/OPT318), you must
activate Junior Mode.
Oxygen,
humidity.
< 10 seconds
O2 too low
The measured oxygen level has fallen below the allowed limit.
Check that the oxygen source is still correctly connected.
Adjust the level of oxygen from the oxygen source as necessary.
Oxygen
< 20 seconds
O2 too high
The measured oxygen level has exceeded the allowed limit.
Adjust the level of oxygen from the oxygen source as necessary.
Oxygen
< 20 seconds
A – 10
Page 13
(continued)
Meaning
Affects
delivery
of:
Delays
The unit cannot reach the target flow setting.
Cannot
reach target Check the heated breathing tube or patient interface for blockage.
Check whether the target flow setting is too high for the patient interface
flow
being used (refer to “Setting up AIRVO 2” - ”Select Patient Interface”).
You will be prompted for acknowledgement.
WARNINGS
Oxygen
< 120
seconds
• The oxygen concentration delivered to the patient can be affected by
changes to the flow setting. Adjust the level of oxygen from the oxygen
source as necessary.
Check water The chamber has run out of water.
When a chamber runs dry, the chamber float may be damaged. Replace the
chamber and water bag.
To ensure continual humidification, always ensure that the water chamber
and/or water bag are not allowed to run out of water.
Humidity
< 30
minutes
The unit cannot reach the target temperature setting.
Cannot
reach target You will be prompted for acknowledgement.
temperature The most likely cause for this is that the unit is operating at a high flow rate
in low ambient conditions. Consider decreasing the target flow setting.
WARNINGS
Humidity
30 +/- 3
minutes
Humidity
60 +/- 6
seconds
Oxygen,
humidity.
< 5 seconds
• The oxygen concentration delivered to the patient can be affected by
changes to the flow setting. Adjust the level of oxygen from the oxygen
source as necessary.
Check
operating
conditions
The unit has detected that it is operating in unsuitable ambient conditions.
This alarm may be caused by a sudden change in ambient conditions.
Leave the unit running for 30 minutes. Switch the unit off and then restart.
[Power out] The unit has been disconnected from the mains power supply.
No visual alarm. The auditory alarm will sound for at least 120 seconds. If
power is reconnected in this time, the unit will automatically restart, unless
the “Audio Pause” button has been pressed.
WARNINGS
• Appropriate patient monitoring must be used at all times. Loss of therapy
will occur if power is lost.
ALARM LIMITS
Most alarm limits are pre-programmed. The exceptions are listed below. These alarm limits may be
changed to other values by authorized personnel. Changes will be preserved during or after any power
loss.
Alarm
condition
Factory-set
alarm limit
Possible preset
values
O2 too low
21% O2
21 or 25% O2
O2 too high
95% O2
30 – 100% O2,in
5% increments
WARNINGS
• A hazard can exist if different alarm presets are used on different units within any single area, eg. an intensive care unit.
• Alarm limits set to extreme values can render the alarm system useless.
CHECKING ALARM SYSTEM FUNCTIONALITY
The functionality of the alarm system can be checked at any time when the unit is turned on.
Remove the heated breathing tube. You should see the “Check tube” visual alarm signal and hear the
auditory alarm signal. If either alarm signal is absent, do not use the unit and refer to the AIRVO 2 Technical
Manual for a guide on troubleshooting. If problems persist, contact your Fisher & Paykel Healthcare
representative.
AUDITORY INFORMATION SIGNALS
In addition to auditory alarm signals, auditory information signals are provided. These are described below.
Melody
Meaning
Ascending sequence of 5 tones
The “Ready for use” symbol has appeared
Ascending sequence of 3 tones
Activation/deactivation of Junior Mode
Single tone every 5 seconds
Measured oxygen level ≥ 33% at turn-off
Single tone every 30 seconds
Measured oxygen level > 95%
A – 11
English
Message
Page 14
4. REPROCESSING
The AIRVO 2 must be cleaned and disinfected between patients according to the instructions in the
Disinfection Kit Manual (900PT600).
This should take place as soon as possible after use. The unit utilizes warmed water and can pose a risk of
bacterial colonization and patient infection if cleaning, disinfection and replacement procedures are not
followed.
Standard aseptic techniques to minimize contamination should be followed when handling the unit
and accessories. This includes proper hand-washing, avoiding hand contact with connection ports, safe
disposal of the used consumables and suitable storage of the unit after cleaning and disinfection.
SCHEDULE FOR CHANGING ACCESSORIES
The accessories for the unit must be changed frequently to avoid the risk of infection. Parts should be
replaced immediately if they are damaged or discolored; otherwise they must be replaced within the
periods shown in the following table.
Maximum
period of use
Part number and description
Patient interfaces excluding Optiflow™+
1 week
(single-patient
use)
OPT316
Nasal Cannula - Infant
OPT318
Nasal Cannula - Pediatric
OPT842
Optiflow™ Nasal Cannula - Small
OPT844
Optiflow™ Nasal Cannula - Medium
OPT846
Optiflow™ Nasal Cannula - Large
OPT870
Tracheostomy Interface
RT013
Mask Interface Adapter - 22mm
Optiflow™+ patient interfaces
2 weeks
(single-patient
use)
3 months
or
1000 hours
OPT942
Optiflow™+ Nasal Cannula - Small
OPT944
Optiflow™+ Nasal Cannula - Medium
OPT946
Optiflow™+ Nasal Cannula - Large
OPT970
Optiflow™+ Tracheostomy Interface
OPT980
Optiflow™+ Mask Interface Adaptor
All tube & chamber kits
900PT551 /
900PT561
AirSpiral™ Heated breathing tube, MR290 auto-fill chamber and adapter
900PT501
Heated breathing tube, MR290 auto-fill chamber and adapter
900PT531
Junior heated breathing tube, MR290 auto-fill chamber and adapter
(for use with OPT316/318 only)
900PT913
Air filter
(or more often if significantly discolored)
FILTER REPLACEMENT
After the AIRVO 2 has been switched on for 1000 hours, a prompt will appear at the start of the next
disinfection cycle indicating that an air filter change is due. Follow the steps below if filter change is due:
Air filter change
due
Replace now?
Now
Later
1.
2.
Take the filter holder from the back of the unit and remove the filter.
Replace the old filter with a new filter (900PT913).
3.
Reattach the filter holder to the unit (clip the bottom of the filter
holder in first, then rotate it upwards until the top clips into place).
4.
5.
6.
Press the Mode button to move on to the “Replace now” screen.
Press the Up button to select “Now”.
Press the Mode button to confirm.
The hours counter will be reset to zero.
If you choose the “Later” option, the prompt will continue to appear at
the start of subsequent disinfection cycles.
SERVICING
This device contains no internal serviceable parts.
Refer to the AIRVO 2 Technical Manual for a list of external spare parts.
A – 12
Page 15
5. TECHNICAL INFORMATION
Caution
Hot Surfaces
Type BF
Applied Part
ATTENTION
Consult
accompanying
documents
Do not
throw away
Drip Proof
Alternating
Current
Class ll
Double
Insulated
Power
On/Off
(Standby)
93/42/EEC
Class IIa
PRODUCT SPECIFICATIONS
Dimensions
295 mm x 170 mm x 175 mm
(11.6” x 6.7” x 6.9”)
Weight
2.2 kg (4.8 lb) unit only,
3.4 kg (7.5 lb) packaged in bag
incl. accessories
Supply frequency
50-60 Hz
Target temperature
settings
37, 34, 31 °C
Humidity
>33 mg/L at 37 °C target
>10 mg/L at 34 °C target
>10 mg/L at 31 °C target
Maximum temperature 43 °C (109 °F)
(in accordance with ISO 8185:2007)
of delivered gas
100-115 V 2.2 A (2.4 A max)
Supply voltage/current
220-240 V 1.8 A (2.0 A max)
Flow range (default)
10-60 L/min
110 seconds
Flow range
(Junior Mode)
2-25 L/min
Serial port
The serial port is used for
downloading product data,
using F&P Infosmart™ software.
Maximum oxygen input 60 L/min
Warm-up time
10 minutes to 31 °C (88 °F),
30 minutes to 37 °C (98.6 °F)
using a MR290 chamber with
flow rate of 35 L/min and
starting temperature 23 ± 2 °C
(73 ± 3 °F)
Sound pressure level
Alarms exceed 45dbA @ 1 m
Auditory alarm pause
Oxygen analyzer
accuracy
<±4%
(within the range 25-95% O2)
Operating conditions:
18-28 °C (64-82 °F),
30-70% RH
Designed to conform to the requirements of:
IEC 60601-1
UL 60601-1
CSA C22.2/No. 601.1
AS 3200.1.0
EN 60601-1
ISO 8185
The unit complies with the electromagnetic compatibility requirements of IEC 60601-1-2. In
certain circumstances, the unit may affect or be affected by nearby equipment due to the
effects of electromagnetic interference. If this should happen, try moving the unit or the
location of the unit causing interference, or alternatively consult your healthcare provider.
Accessory equipment connected to the serial port of the device must be certified to either IEC 60601-1 or IEC
60950-1. Furthermore all configurations shall comply with the system standard IEC 60601-1-1. Anyone who connects
additional equipment to the signal input part or signal output part configures a medical system and is therefore
responsible for ensuring that the system complies with the requirements of the system standard IEC 60601-1-1. If in
doubt, consult the technical services department or your local representative.
OPERATING CONDITIONS
Ambient temperature
18 - 28 °C (64 - 82 °F)
Humidity
10 - 95% RH
Altitude
0 - 2000 m (6000 ft)
Mode of operation
Continuous operation
STORAGE AND TRANSPORT CONDITIONS
The unit should be stored and transported in environmental conditions of -10 °C to 60 °C (14 °F to 140 °F),
10 to 95% RH, non-condensing.
DISPOSAL INSTRUCTIONS
Unit Disposal Instructions
This unit contains electronics. Please do not discard with regular waste. Return to Fisher & Paykel
Healthcare or dispose according to local guidelines for disposing of electronics. Dispose according to Waste
Electrical and Electronic Equipment (WEEE) directive in European Union.
Consumables Disposal Instructions
Place the interface, breathing tube and chamber in a waste bag at the end of use. Hospitals should discard
according to their standard method for disposing of contaminated product.
A – 13
English
SYMBOL DEFINITIONS
Page 17
For more information please contact
your local Fisher & Paykel Healthcare representative
Manufacturer
Fisher & Paykel Healthcare Ltd
15 Maurice Paykel Place
East Tamaki, Auckland 2013
PO Box 14 348 Panmure
Auckland 1741
New Zealand
Tel:
Fax:
Email:
Web:
+64 9 574 0100
+64 9 574 0158
info@fphcare.co.nz
www.fphcare.com
Australia (Sponsor)
Fisher & Paykel Healthcare Pty Limited
36–40 New Street, PO Box 167
Ringwood, Melbourne
Victoria 3134, Australia
Tel: +61 3 9879 5022
Fax: +61 3 9879 5232
Austria
Tel: 0800 29 31 23
Fax: 0800 29 31 22
Benelux
Tel: +31 40 216 3555
Fax: +31 40 216 3554
Brazil
Fisher & Paykel do Brasil
Rua Sampaio Viana, 277 cj 21
Paraíso, 04004-000
São Paulo – SP, Brazilzil
REF 185045494 REV G 2016-May © 2016 Fisher & Paykel Healthcare Limited
Tel:
+55 11 2548 8002
China
代理人/售后服务机构:
费雪派克医疗保健(广州)有限公司,
广州高新技术产业开发区科学城科丰
路31号G12栋301号
电话: +86 20 32053486
传真: +86 20 32052132
Finland
Tel: +358 (0)405 406618
Fax: +46 (0)8 36 6310
France
Tel: +33 1 6446 5201
Fax: +33 1 6446 5221
Sweden
Tel: +46 8 564 76 680
Fax: +46 8 36 63 10
Germany
Tel: +49 7181 98599 0
Fax: +49 7181 98599 66
Switzerland
Tel: 0800 83 47 63
Fax: 0800 83 47 54
India
Tel: +91 80 4284 4000
Fax: +91 80 4123 6044
Taiwan
Tel: +886 2 8751 1739
Fax: +886 2 8751 5625
Irish Republic
Tel: 1800 409 011
Turkey
Fisher Paykel Sağlık Ürünleri
Ticaret Limited Şirketi,
Alinteri Bulvari 1161/1 Sokak
No. 12-14, P.O. Box 06371 Ostim,
Ankara, Turkey
Italy
Tel: +39 06 7839 2939
Fax: +39 06 7814 7709
Japan
Tel: +81 3 5117 7110
Fax: +81 3 5117 7115
Korea
Tel: +82 2 6205 6900
Fax: +82 2 6309 6901
Tel: +90 312 354 34 12
Fax: +90 312 354 31 01
Northern Ireland
Tel: 0800 132 189
UK
Fisher & Paykel Healthcare Ltd
Unit 16, Cordwallis Park
Clivemont Road, Maidenhead
Berkshire SL6 7BU, UK
Russia
Tel and Fax: +7 495 782 21 50
Tel: +44 1628 626 136
Fax: +44 1628 626 146
Spain
Tel: +34 902 013 346
Fax: +34 902 013 379
USA/Canada
Tel: +1 800 446 3908
or
+1 949 453 4000
Fax: +1 949 453 4001
0123
One or more products in this manual are covered by or for use under one or more of the following patents, design registrations or design patents in the corresponding countries listed below.
Australia: AU2006325613, AU2011245770, AU2009251939, AU2011245771, AU2006266564, AU2005273119, AU2003204474, AU756477, AU2004202274, AU2009243444, AU2004216567, AU2003244579,
AU2007214363, AU2003219624, AU355510, AU355511, AU345792.
Europe:
EP1973577 (DE, FR, GB), EP1481702 (DE, FR, GB, SE), EP1396277 (DE, FR, GB, IT), EP1484161 (DE, FR, GB, IT), EP1495855 (DE, FR, GB, IT), EP1471962 (DE, FR, GB, SE), EP2799103
(DE, ES, FR, GB, IT, NL, SE, TR), CRD no: 002372292-0001, CRD no:002372292-0002, CRD no:002452839-0001, CRD no:002452839-0002, CRD no:002554857-0001, CRD no:002554857-0002,
CRD no:002100289-0001, CRD no:002100289-0002, CRD no:002100289-0003, CRD no:002132464-0001, CRD no:002132464-0002.
USA:
US8741220, US8616202, US8453641, US8186345, US7111624, US8091547, US8550072, US6953354, US6598604, US7493902, US8267092, US8220463, US8997747, US7458615, US7468116,
USD747792.
One or more of the products in this manual are also covered by or for use under one or more pending patent applications and/or design applications.
www.fphcare.com