Page 1
Bubble CPAP System
Page 3
Bubble CPAP System Setup
1 LPM
10
9
8
7
6
5
10
9
8
7
6
5
6-8 LPM
(<15LPM)
10
9
8
7
6
5
INTERNATIONAL Fisher & Paykel Healthcare Limited, 15 Maurice Paykel Place, East Tamaki, Auckland 2013; P O Box 14 348, Panmure, Auckland 1741, New Zealand Tel: +64 9 574 0100 Fax: +64 9 574 0158 Email: info@fphcare.com Web Site: www.fphcare.com
AUSTRALIA Tel:+61 3 9879 5022 Fax:+61 3 9879 5232 AUSTRIA Tel: 0800 29 31 23 Fax: 0800 29 31 22 BENELUX Tel:+31 40 216 3555 Fax:+31 40 216 3554 CHINA Tel:+86 20 3205 3486 Fax:+86 20 3205 2132 FRANCE Tel:+33 1 6446 5201 Fax:+33 1 6446 5221
GERMANY Tel:+49 7182 93777 0 Fax:+49 7182 93777 99 INDIA Tel:+91 80 4123 6041 Fax:+91 80 4123 6044 IRISH REPUBLIC Tel: 1800 409 011 ITALY Tel:+39 06 7839 2939 Fax:+39 06 7814 7709 JAPAN Tel:+81 3 3661 7205 Fax:+81 3 3661 7206
MIDDLE EAST Tel: +9 71 4 360 0849 Fax: +9 71 4 368 8110 NORTHERN IRELAND Tel: 0800 132 189 SPAIN Tel:+34 902 013 346 Fax:+34 902 013 379 SWEDEN Tel: +46 8 564 76 680 Fax: +46 8 36 63 10 SWITZERLAND Tel: 0800 83 47 63 Fax: 0800 83 47 54
UK (EU Authorised Representative) Fisher & Paykel Healthcare Ltd, Unit 16, Cordwallis Park, Clivemont Road, Maidenhead, Berkshire SL6 7BU, United Kingdom Tel:+44 1628 626 136 Fax:+44 1628 626 146 USA Tel: +1 800 446 3908 or +1 949 453 4000 Fax:+1 949 453 4001
Page 4
Bubble CPAP System Setup
10
9
8
7
6
5
MR850: 37°
MR730: 40°, -3
REF 185042850 Rev E 2008-06
en
Page 5
Bubble CPAP Setup (Flow Driver)
10
9
8
7
6
5
1 LPM
10
9
8
7
6
5
6-8 LPM
(<15LPM)
10
9
8
7
6
5
Page 6
10
9
8
7
6
5
REF 185042849 Rev B 2008-06 en
Bubble CPAP Setup (Flow Driver)
INTERNATIONAL Fisher & Paykel Healthcare Limited, 15 Maurice Paykel Place, East Tamaki, Auckland 2013; P O Box 14 348, Panmure, Auckland 1741, New Zealand Tel: +64 9 574 0100 Fax: +64 9 574 0158 Email: info@fphcare.com Web Site: www.fphcare.com
AUSTRALIA Tel:+61 3 9879 5022 Fax:+61 3 9879 5232 AUSTRIA Tel: 0800 29 31 23 Fax: 0800 29 31 22 BENELUX Tel:+31 40 216 3555 Fax:+31 40 216 3554 CHINA Tel:+86 20 3205 3486 Fax:+86 20 3205 2132 FRANCE Tel:+33 1 6446 5201 Fax:+33 1 6446 5221
GERMANY Tel:+49 7182 93777 0 Fax:+49 7182 93777 99 INDIA Tel:+91 80 4123 6041 Fax:+91 80 4123 6044 IRISH REPUBLIC Tel: 1800 409 011 ITALY Tel:+39 06 7839 2939 Fax:+39 06 7814 7709 JAPAN Tel:+81 3 3661 7205 Fax:+81 3 3661 7206
MIDDLE EAST Tel: +9 71 4 360 0849 Fax: +9 71 4 368 8110 NORTHERN IRELAND Tel: 0800 132 189 SPAIN Tel:+34 902 013 346 Fax:+34 902 013 379 SWEDEN Tel: +46 8 564 76 680 Fax: +46 8 36 63 10 SWITZERLAND Tel: 0800 83 47 63 Fax: 0800 83 47 54
UK (EU Authorised Representative) Fisher & Paykel Healthcare Ltd, Unit 16, Cordwallis Park, Clivemont Road, Maidenhead, Berkshire SL6 7BU, United Kingdom Tel:+44 1628 626 136 Fax:+44 1628 626 146 USA Tel: +1 800 446 3908 or +1 949 453 4000 Fax:+1 949 453 4001
Page 7
Patient Interface Setup Guide
360px
Measure the baby’s head circumference in centimeters.
Choose the correct sized bonnet. The bonnet should be snug fitting.
Stretch the bonnet with your hands initially for ease of placement.
Slip the bonnet onto the baby’s head completely covering the ears and with the back
edge of the bonnet at the base of the neck. The front edge of the bonnet should be
just above or on the eyebrows.
360px
Use the prong-sizing guide to choose the suitable prongs. Prongs should fill the
nares completely without stretching the skin. Use the biggest possible size.
Choose the correct length of the nasal tubing. Use the shortest length possible. The
following may be used as a guide:
50mm ≤ 1.5kg, 70mm ≤ 2.5kg, 100mm > 2.5kg
INTERNATIONAL: Fisher & Paykel Healthcare Limited, 15 Maurice Paykel Place, East Tamaki, Auckland 2013, P O Box 14 348, Panmure, Auckland 1741, New Zealand
Tel:+64 9 574 0100 Fax:+64 9 574 0158 Email: info@fphcare.com Web Site: www.fphcare.com USA Tel: +1 800 446 3908 or +1 949 453 4000 Fax:+1 949 453 4001
AUSTRALIA Tel:+61 3 9879 5022 Fax:+61 3 9879 5232 AUSTRIA Tel: 0800 29 31 23 Fax: 0800 29 31 22 BENELUX Tel:+31 40 216 3555 Fax:+31 40 216 3554
CHINA Tel:+86 20 3205 3486 Fax:+86 20 3205 2132 FRANCE Tel:+33 1 6446 5201 Fax:+33 1 6446 5221 GERMANY Tel:+49 7182 93777 0 Fax:+49 7182 93777 99
INDIA Tel:+91 80 4123 6041 Fax:+91-80-4123 6044 IRISH REPUBLIC Tel: 1800 409 011 ITALY Tel:+39 06 7839 2939 Fax:+39 06 7814 7709 JAPAN Tel:+81 3 3661 7205
Fax:+81 3 3661 7206 MIDDLE EAST Tel: +9 71 4 360 0849 Fax: +9 71 4 368 8110 NORTHERN IRELAND Tel: 0800 132 189 SPAIN Tel:+34 902 013 346 Fax:+34 902 013 379
SWEDEN Tel: +46 8 564 76 680 Fax: +46 8 36 63 10 SWITZERLAND Tel: 0800 83 47 63 Fax: 0800 83 47 54 UK (EU Authorised Representative) Fisher & Paykel Healthcare Ltd,
Unit 16, Cordwallis Park, Clivemont Road, Maidenhead, Berkshire SL6 7BU, United Kingdom Tel:+44 1628 626 136 Fax:+44 1628 626 146
Check the height of the nasal tubing on the baby’s forehead. The nasal tubing
should be parallel to the infant’s face not slanted upwards or downwards. If needed,
remove the tear-off strips (there are 2, remove as needed) at the bottom of the
foam block along the tri-lumen tube. Note: The tear-off strip may be replaced if
removed inadvertently.
REF 185042883 Rev E 2008-06 en
Page 8
Attach the prongs to the nasal tubing. Squeeze the prongs to open up the slots at
the sides. Insert one end of the prongs into the triangular plastic block at the end of
the nasal tubing. Work the prongs along the rest of the block to lock the prongs in
place. The rectangular portion of the nasal prongs should sit within the base of the
triangular plastic block.
Attach the circuits to the nasal tubing. Blue inspiratory limb to the blue collapsible
extension and white expiratory limb to white collapsible extension. Ensure that the
gas flow has been set and turned on before connecting to the baby. Operational Input
flow range is 4 - 15 L/min (max). Recommended Input flow range is 6 - 8 L/min.
Ensure the nasal prongs are positioned at least 2 mm (0.08 inches) from the
septum to avoid pressure necrosis. Re-adjust as necessary.
The circuit should be kept below the level of the nasal tubing to minimize condensate
at the baby. The use of angel frames, x-mas trees or similar devices are not
recommended as they cause the circuit position to be elevated and restrict the
movement of the baby’s head.
The foam block on the nasal tubing should sit at the middle of the blue velcro strap of
the bonnet just above the infant’s forehead. Place the strap over the foam block and
tighten it to hold the tubing firmly in place.
Clear nasal secretions before inserting the nasal prongs. Insert the nasal prongs
curve down into the infant’s nose then rotate the nasal tubing until the foam block
rests on the velcro strap on the baby’s forehead. Ensure that the nasal prongs sit well
inside the nose following the anatomical curve.
Hook the clips from the side straps of the bonnet to the glider. Pull both straps at the
same time for central positioning and until there is a slight bend on the glider. Affix
the ends of the straps to the velcro at the sides of the bonnet.
In an incubator setting, it is recommended that the circuits exit the incubator
through the side corner hole.
Should there be a need to raise the head of the bed, the baby should be supported
to prevent him/her from sliding down and dislodging the prongs.
The key to a successful setup is keeping the nasal tubing in place. Ensure the weight
of the circuit is supported to reduce tension on the nasal tubing. The circuit must have
free movement to allow the baby’s head to move without restriction.
en
Page 9
BOOKLET REF 043042435 Rev C 2008-06
en
INTERNATIONAL Fisher & Paykel Healthcare Limited, 15 Maurice Paykel Place, East Tamaki, Auckland 2013; P O Box 14 348, Panmure, Auckland 1741, New Zealand Tel: +64 9 574 0100 Fax: +64 9 574 0158 Email: info@fphcare.com Web Site: www.fphcare.com
AUSTRALIA Tel:+61 3 9879 5022 Fax:+61 3 9879 5232 AUSTRIA Tel: 0800 29 31 23 Fax: 0800 29 31 22 BENELUX Tel:+31 40 216 3555 Fax:+31 40 216 3554 CHINA Tel:+86 20 3205 3486 Fax:+86 20 3205 2132 FRANCE Tel:+33 1 6446 5201 Fax:+33 1 6446 5221
GERMANY Tel:+49 7182 93777 0 Fax:+49 7182 93777 99 INDIA Tel:+91 80 4123 6041 Fax:+91 80 4123 6044 IRISH REPUBLIC Tel: 1800 409 011 ITALY Tel:+39 06 7839 2939 Fax:+39 06 7814 7709 JAPAN Tel:+81 3 3661 7205 Fax:+81 3 3661 7206
MIDDLE EAST Tel: +9 71 4 360 0849 Fax: +9 71 4 368 8110 NORTHERN IRELAND Tel: 0800 132 189 SPAIN Tel:+34 902 013 346 Fax:+34 902 013 379 SWEDEN Tel: +46 8 564 76 680 Fax: +46 8 36 63 10 SWITZERLAND Tel: 0800 83 47 63 Fax: 0800 83 47 54
UK (EU Authorised Representative) Fisher & Paykel Healthcare Ltd, Unit 16, Cordwallis Park, Clivemont Road, Maidenhead, Berkshire SL6 7BU, United Kingdom Tel:+44 1628 626 136 Fax:+44 1628 626 146 USA Tel: +1 800 446 3908 or +1 949 453 4000 Fax:+1 949 453 4001