SH870 Instructions For Use Rev A.pdf
Page 1
SURGICAL
HUMIDIFICATION
SYSTEM
SH870
INSTRUCTIONS FOR USE
REF SH870
www.fphcare.com/surgical
NON
STERILE
STERILE
R
Page 2
en
ENGLISH
A1
Page 3
Fisher & Paykel Healthcare
SH870 SURGICAL HUMIDIFICATION SYSTEM
SH870AEA
SH870AEK
SH870AFU
01
en
Page 4
02
INSTRUCTIONS FOR USE
Surgical Humidification System
BEFORE YOU START
Please read these user instructions carefully before use and save for future reference.
If the device or any accessories are not operating correctly, please contact your
Fisher & Paykel Healthcare office (as detailed on the back cover) or applicable distributor.
INTENDED USE
For use in warming and humidifying surgical insufflation gas.
Laparoscopic Procedures
Intended for patients of all ages undergoing laparoscopic procedures where carbon
dioxide (CO2) insufflation gas is used.
Open surgery Procedures
Intended for patients of all ages undergoing open abdominal or cardiothoracic surgical
procedures where CO2 insufflation gas is used.
Operator Profile
To be operated by a healthcare provider with experience in the set up of surgical
procedures, including handling of sterile items. The user must read and understand
these instructions for use and the instructions for use of the F&P HumiGard™ Surgical
Humidification System (SH870) before use.
General Precautions
• Only use accessories approved by Fisher & Paykel Healthcare. Use of
non-approved accessories could impair performance or compromise safety.
• Monitor patient temperature throughout the procedure and adjust patient warming
protocols to prevent hyperthermia or hypothermia.
TO SET-UP
I. Remove the humidifier from its packaging. Plug the power cord into the power supply.
II. E
nsure that the humidifier is level and securely mounted. A range of mounting
solutions is available from Fisher & Paykel Healthcare. Please contact your
Fisher & Paykel Healthcare representative or distributor for assistance in
selecting the most suitable mounting bracket.
Page 5
Fisher & Paykel Healthcare
CONTROLS AND INDICATORS
01
02 04
03
05
01.
Power / Standby
Pressing this button turns the humidifier on or off (Stand-by).
02. O
n Indicator
Solid light indicates that the humidifier is ON.
03. H
eaterwire Adaptor Plug Indicator
Flashing light indicates that the heater wire plug is disconnected.
04. F
low Out / Water Out Indicator
Flashing light indicates that the chamber requires refilling and/or the gas supply is
running low. Check both the chamber water level and gas supply.
05. H
ardware Fault Indicator
A flashing or solid light indicates a fault with the humidifier requiring servicing.
Please refer to SH870 Humidification Technical Manual (Ref 185048422) Humidified
gas will not be delivered when this indicator is on.
03
Page 6
04
INSTRUCTIONS FOR USE
Surgical Humidification System
INSTRUCTIONS FOR USE
GENERAL CLEANING INSTRUCTIONS
Surgical Humidifier
01. Disconnect the humidifier from any electrical outlet
02. Using a damp cloth, clean the humidifier with one of the following solutions:
• Isopropyl alcohol
• Dishwashing detergent
03. Wipe the humidifier clean of any cleaning residues before use
Warning: To avoid electric shock do not immerse the humidifier into any liquid.
ROUTINE MAINTENANCE, SERVICING AND CALIBRATION
In order to keep your humidifier in good working order, it is necessary for qualified service personnel to
perform maintenance at regular intervals. A full technical description including maintenance and service
data is contained in the SH870 Humidifier Technical Manual (REF 185048422) available from your
Fisher & Paykel Healthcare office or distributor.
Device
Regular Maintenance Period
Maintenance Aspects*
SH870 Humidifier
12 months
Check for damage.
Carry out a performance test
* Refer to SH870 Humidification Technical Manual for detailed instructions (REF 185048422)
SPECIFICATIONS
General Specifications
Specifications for F&P HumiGard™ Surgical Humidifier
Dimensions (without chamber fitted)
94 x 154 x 135mm
Weight (without chamber fitted)
1.7kg
Supply Voltage
230 V~(SH870A--) 115 V~ (SH870J--)
Supply Frequency
50-60 Hz
Power Input
200 VA
Heater-wire
22 V ~ at 30 W
Heater-plate
150 W
Heater plate high temperature
93 ± 6 °C cut-out
Operating environment
18 °C to 26°C
Humidity performance (0 to 10 L/min)
> 33 mg/L
Transport and storage conditions
-10°C to 50°C
Specifications for Laparascopic Procedures
with ST320 Surgical Humidification Insufflation Kit
Specifications for Open Procedures
with ST320 Surgical Humidification Insufflation Kit
Input set flow rate: ≤45 L/min
Recommended flow rate: 10 L/min
Refer to ST320 user instructions (REF 185048421) for set up. Refer to product technical manual (REF 185048422)
for electro-magnetic compatibility information.
NOTE: Not made with natural rubber latex.
Page 7
Fisher & Paykel Healthcare
SYMBOLS
Caution - Hot Surface
STANDBY/ON
Button
Resistant to vertical falling drips
On Indicator
Fragile.
Handle With Care
Heaterwire Adaptor Plug Indicator
This Way Up
Flow Out / Water Out Indicator
C-Tick for EMC
Hardware Fault Indicator
DEVICE DISPOSAL INSTRUCTIONS
The SH870 Humidifier is intended for repeat use. If disposal is required please do not discard as
regular rubbish. Dispose of electronics according to local guidelines.
APPROVED ACCESSORIES
Using non-approved accessories or spare parts could be unsafe.
Use only the following approved products from your Fisher & Paykel Healthcare supplier:
General Accessories Laparoscopic and Open Surgery
• ST320 Surgical Humidification Insufflation Kit
Open Surgery Accessories
• VITA-diffuser™ manufactured by Cardia Innovation AB
NOTE: Accessories may not be available in all countries. Contact your local Fisher & Paykel Healthcare office
or authorised distributor for more information.
05
Page 8
06
INSTRUCTIONS FOR USE
Surgical Humidification System
For more information please contact
your local Fisher & Paykel Healthcare representative
Manufacturer
Fisher & Paykel Healthcare Ltd
15 Maurice Paykel Place
East Tamaki, Auckland 2013
PO Box 14 348, Panmure
Auckland 1741
New Zealand
Tel:
Fax:
Email:
Web:
+64 9 574 0100
+64 9 574 0158
info@fphcare.co.nz
www.fphcare.com
Australia
Fisher & Paykel Healthcare
Pty Ltd 36-40 New Street,
Ringwood, Melbourne,
Victoria 3134, Australia
Tel:
Fax:
+61 3 9879 5022
+61 3 9879 5232
Austria
Tel: 0800 29 31 23
Fax: 0800 29 31 22
Benelux
Tel: +31 40 216 3555
Fax: +31 40 216 3554
China
Tel: +86 20 3205 3486
Fax: +86 20 3205 2132
France
Tel: +33 1 6446 5201
Fax: +33 1 6446 5221
Germany
Tel: +49 7181 98599 0
Fax: +49 7181 98599 66
Switzerland
Tel: 0800 83 47 63
Fax: 0800 83 47 54
India
Tel: +91 80 4284 4000
Fax: +91 80 4123 6044
Taiwan
Tel: +886 2 8751 1739
Fax: +886 2 8751 5625
Irish Republic
Tel: 1800 409 011
Turkey
Tel: +90 312 354 34 12
Fax: +90 312 354 31 01
Italy
Tel: +39 06 7839 2939
Fax: +39 06 7814 7709
Japan
Tel: +81 3 5117 7110
Fax: +81 3 5117 7115
Korea
Tel: +82 2 6205 6900
Fax: +82 2 6309 6901
Northern Ireland
Tel: 0800 132 189
Spain
Tel: +34 902 013 346
Fax: +34 902 013 379
185048419 REV A © 2015 Fisher & Paykel Healthcare Limited
Sweden
Tel: +46 8 564 76 680
Fax: +46 8 36 63 10
www.fphcare.com/surgical
UK
(EU Authorised
Representative)
Fisher & Paykel Healthcare Ltd
Unit 16, Cordwallis Park,
Clivemont Road, Maidenhead,
Berkshire SL6 7BU, UK
Tel:
Fax:
+44 1628 626 136
+44 1628 626 146
USA/Canada
Tel: 1 800 446 3908
or
+1 949 453 4000
Fax: +1 949 453 4001