VISION Enteral Feeding Pump Instructions for use
Contents
GB
Operating instructions for APPLIX Vision Please read the following instructions before using the pump.
Information Intended purpose Indication Contraindications Complications Important notes
Description Illustration of the pump Pump holder and pump Attaching the pump to the pump holder Nurse call and data interface
Giving set Priming of the giving set Installing the giving set in the pump
4 4 4 4 4 5
6 6 7 8 9
10 10 11
Keypad
12
Starting up the pump
18
Mains operation Battery operation Switch on Insert giving set Automatic priming using the pump Program selection Start Stop Switch off 2
18 18 18 19 19 19 19 19 19 4150-2_nu_Applix_Vision_GB
GB
Contents
Administration program
20
Continuous feed administration Volume/time feed administration Bolus feed administration
21 23 25
Alarms
27
Cleaning the pump
32
Pump and pump holder Pump door Pump mechanism and sensor area
Technical information Technical data Operating conditions Warranty Inspection Repair Contact Explanation of symbols Guidance and manufacturer’s declaration
Ordering information
4150-2_nu_Applix_Vision_GB
32 32 33
34 34 35 36 36 36 37 38 39
43
3
GB
INFORMATION Intended purpose The APPLIX Vision is intended exclusively for enteral feeding. It can be used with both home patients and hospital patients. It is an easy-to-use feeding pump with the application programs: Continuous, Volume/time, Bolus. The alarm functions ensure optimum reliability.
Indication Only suitable for enteral feeding. The user should refer to the operating conditions (see p. 35), in order to check that the use of the APPLIX pump is compatible with the clinical condition of the patient, particularly on infants and neonates. Due to the diversity of the use conditions with a wide range of solution characteristics (density, viscosity, quantity of dissolved gasses, rheology, particles, temperature, individualized nutrition preparation, ...), the performance of the occlusion detection may not be granted under all circumstances. Therefore Fresenius Kabi recommends not using the APPLIX device without a timely and regular surveillance especially when an interruption of the nutrition delivery may lead to a patient hazard (e.g. comatose situation).
Contraindications Do not use for the intravenous administration of infusions. Do not use if enteral feeding is contraindicated (e.g. ileus, uncontrolled diarrhoea, severe acute pancreatitis or intestinal atony) or if the patient needs to be fed parenterally.
Complications Even with pump assisted feeding, problems such as diarrhoea or a bloated feeling may arise. The speed of administration must be adapted individually to the patient. Regular checks are required.
4
4150-2_nu_Applix_Vision_GB
GB
Important notes The APPLIX Vision is intended solely for enteral feeding. When using the pump, please note the operating conditions (see p. 35). The function of the pump is guaranteed only if it is operated with the appropriate Fresenius Kabi giving sets, (see ordering information on p. 43), and if the giving set is correctly installed (see Installing the giving set in the pump p. 11). Otherwise patients may be put at risk under certain circumstances. The giving sets should be single use only for no longer than 24 hours as depicted by Bristish Standard EN 1615. The feeding tube should be flushed through before and after every administration of feed and before, between and after the administration of individual drugs. The APPLIX Vision may not be operated in areas where there is a risk of explosion. The feeding pump has a BF-classification. To conform with CF requirements in a cardiac environment where the patient has direct cardiac access, keep the pump attached on its holder. When changing the nutrition bag/bottle/container without changing the giving set, it is recommended to open the door of the pump and check the right position of the set before starting the pump again. In case of prolonged storage of the device, check the general aspect of this one in order to make sure it is in good condition.
4150-2_nu_Applix_Vision_GB
5
GB
DESCRIPTION Illustration of the pump
Giving set
Giving set
Serial number label (Unit number)
Pump door
Guide rail
Door level
View from the front
6
Contact pins for connecting to pump holder
View from the rear
4150-2_nu_Applix_Vision_GB
GB
Pump holder and pump The pump can be attached universally, vertically and horizontally to both round and square rails. E.g.: drip stands wheelchairs hospital beds table-top pump stand
Pump
Pump holder
Insert point for the pole stand to suspend the administration system
The clamping device can be rotated 180°
Screw to lock the clamping device
Clamping device
4150-2_nu_Applix_Vision_GB
7
GB
Attaching the pump to the pump holder
Depress button to release pump
8
4150-2_nu_Applix_Vision_GB
GB
Nurse call and data interface The pump holder provides the opportunity of connecting to a nurse call system. Further information about connection is available from the Technical Service department. The function is available if the pump is properly engaged in the holder and the holder is connected to the mains. Plug the connecting cable for the nurse call function into the connector on the holder and guide through the lateral cable fixing. Then connect to the nurse call system. Before using the pump, check that the transfer line for the nurse call system is working properly by initiating an alarm (e.g. start without giving set) at the pump.
Cable fixing
Mains socket
Nurse call connection and data interface
4150-2_nu_Applix_Vision_GB
9
GB
GIVING SET Priming of the giving set for example: APPLIX Vision Pump Set VarioLine
APPLIX Vision - Pump Sets WARNING! Operate the pump only with the appropriate Fresenius Kabi APPLIX giving sets (see ordering information on p. 43). Otherwise administration problems cannot be ruled out. NOTE: When opening the pump door, the tube clamp is closed automatically (free-flow prevention device). 1. Close tube clamp. 2. Connect food container to giving set and hang up. 3. Fill drip chamber half full by pressing gently. 4. Use preferably the automated priming process: Ensure that the set is correctly installed in the pump prior to closing the door, (see page 11 "Installing the giving set in the pump“). 5. Close the door by pushing firmly on the grey door-lever. 6. Press the automated priming key: keep the priming key depressed. Prime to the end of the giving set. 7. Release the key once priming is completed. Verify the flow by checking the liquid flow in the drip chamber after starting the pump.
VarioLine adaptor
EasyBag connector T-piece
Tube clamp
NOTE: When priming the set manually, the steps following step n° 3 above are: 4. Open tube clamp Prime to the end of the giving set. 5. Close tube clamp 6. Install the set in the pump to start nutrition administration (see "Installing the giving set in the pump“, page 11 and “starting up the pump”, page 18). Verify the flow by checking the liquid flow in the drip chamber after starting the pump. NOTE: For giving sets without drip chamber, use only the automatic priming.
10
4150-2_nu_Applix_Vision_GB
GB
Installing the giving set in the pump 1. Unlock and open the pump door with the door lever. 2. Insert the tube clamp on the giving set into the pump in accordance with the arrow marking, with the clamp lever pointing upwards. 3. Guide the giving set over the pump mechanism and fix it in the lower tube guide ensuring that it is in a straight line and slightly stretched but not elongated. 4. Push the pump door shut and ensure that it engages properly.
Pump door
Clamp lever
Door lever Giving set correctly installed
Pump mechanism
Lower tube guide (Sensor area)
Giving set incorrectly installed
4150-2_nu_Applix_Vision_GB
11
GB
KEYPAD Keys
Symbol
Operation
Function
On/Off
Press key for at least 1 second
Pump switches on (self-test) or pump switches off.
Automatic priming
Keep key depressed
Initially fill giving set with feed (600 ml/h). Release key once priming is complete.
Select application programs
Press key (several times if appropriate
Choose between programs: continuous volume/time bolus
Set rate
Press key
Up arrow key increases (rate or dose). Down arrow key decreases.
Set time
Press key
Up arrow key increases (feed administration time or time interval). Down arrow key decreases.
Target volume On/Off
Press keys simultaneously
Activates target volume function or switches off target volume function.
Press keys individually Sets target volume.
12
4150-2_nu_Applix_Vision_GB
GB
Keys
Symbol
Start/Stop
Operation
Function
Press key
Starts administration or stops administration: STOP appears in the bottom display cell for 3 seconds. (During which time parameters cannot be changed). The following functions are available only if the pump is stopped:
Reset
Press keys simultaneously
Pump returns to original factory settings of continuous feeding at a rate of 100 ml/h.
Selection of a feeding program and blocking of the two other feeding programs
Press key
Selection of application program to be used.
Press key
Key symbol and symbol of program selected flash in display.
Press key
Confirmation of feeding program selected. The appropriate program is shown permanently in the display. The key symbol continues to flash.
Press key
Key symbol disappears from the display. The feeding pump can be operated in the application program selected. The two other programs are blocked. When the MODE key is pressed, an alarm signal sounds and the key symbol appears in the display.
4150-2_nu_Applix_Vision_GB
13
GB
Keys Cancel blocking of feeding programs
Symbol
Operation
Function
Press key
Key symbol in display flashes.
Press key
The key symbol and the symbol for the feeding program selected flash in the display.
Press key
The symbol for the feeding program selected is shown permanently in the display. The key symbol is extinguished. The MODE key can be used to select other feeding programs.
Keypad lockout
All key functions except ON/OFF and START/STOP are locked or released. Press key
Key symbol in display flashes
Set code 7 using top set of arrow keys
Alarm volume
14
Press key again
Key symbol in display: appears: keypad is locked disappears: keypad is released
Press key (repeatedly to alter alarm volume)
Set at different volumes. The selected setting can be both seen and heard.
In alarm mode: Press key (once to snooze the alarm)
The alarm remains silent for 2 minutes. The Backlight and the error message keep on flashing on the display
4150-2_nu_Applix_Vision_GB
GB
Keys
Symbol
Operation
Function
Information display
Press key briefly
The following information is automatically displayed if the information key is actuated in the sequence described.
when option Target volume countdown is not activated
When administering feed
1. Volume of feed administered since the pump was last reset. 2. Remaining volume "-" of feed to be delivered before the target volume is reached (if target volume is activated).
When pump is stopped 1. Volume of feed administered since the pump was last reset. 2. The icons corresponding to the last alarm triggered are shown on the display. when option Target volume countdown is activated
When administering feed
1. Volume of feed administered since the pump was last reset. 2. Programed target volume to be delivered for this application (if target volume is activated).
When pump is stopped 1. Volume of feed administered since the pump was last reset. 2. Programed target volume to be delivered for this application (if target volume is activated). 3. The icons corresponding to the last alarm triggered are shown on the display.
4150-2_nu_Applix_Vision_GB
15
GB
Keys
Symbol
To reset Information display
Operation
Function
Keep key depressed for 3 seconds
Reset volume display to zero.
Time set up between 2 alarm signals
The user can select the time between two alarm signals (except for reminder and battery pre-alarm). Press key
Key symbol in display flashes.
Set code 15 using top set of arrow keys Set time (from 5 to 300 seconds) using bottom set of arrow keys
Time in display appears. Note: This option is only available if there is no administration in progress. To cancel administration, press the 'i' key 3 seconds.
Press key again
16
4150-2_nu_Applix_Vision_GB
GB
Keys
Symbol
Operation
Target volume countdown option
Function When programing the administration with a target volume, the user can choose to display the countdown of the programed target volume during the administration, instead of displaying the target volume programed.
Press key
Key symbol in display flashes.
Set code 26 using top set of arrow keys
Symbol "-" appears in display, before the target volume to be administered. In this option, there is no automatic reset of the volume infused if the administration has been interrupted and/or the pump has been switched off (to reset the remaining volume to be infused, press the 'i' button during 3 seconds. A reset of the volume will also occur in case of reset to return to original factory settings). To select the displaying of the target volume instead of the countdown, repeat the operation and select "off" with the bottom set of arrow keys.
Select "on" using bottom set of arrow keys Press key again
Please note that the pumps are delivered with the configuration displaying the target volume programed per default. Note: this option is only available if there is no administration in progress. To cancel administration, press the 'i' key 3 seconds.
4150-2_nu_Applix_Vision_GB
17
GB
STARTING UP THE PUMP Mains operation With pump holder and mains lead
1. Push the pump onto the guide rail of the holder until it clicks. 2. Plug the mains lead into the mains connector on the holder (see page 9). 3. Connect the mains lead to the mains (the green "Mains" indicator on the holder comes on). The "Plug" symbol appears in the display.
Battery operation Battery life: 24 hrs Rate: 125 ml ⁄ h Battery is full
Battery is empty
Before starting for the first time, charge the batteries for approx. 5 hours! If the pump is connected to the mains, the batteries also charge during operation. When the pump is disconnected from the mains, it automatically switches to battery mode. The maximum life of the batteries is only achieved after several charge/discharge cycles. In the event of frequent mains operation, battery life may be less. Battery indicator shows 3 bars when fully charged. The "battery pre-alarm" will be activated when there is at least 30 minutes of battery life (or charge) remaining. If the pump is not connected to an electrical power source to recharge within 30 minutes, the pump will stand by and the "battery alarm" will activate for 10 minutes. If at this time the battery is not recharged, it will shut off.
Switch on Press key for approx 1 second
18
An acoustic signal sounds. The pump performs a self-test. The numbers 1-4 and then all the display symbols appear. The user should have a look at the display during the self test, in order to detect potential failures. The last feeding program activated including settings is displayed.
4150-2_nu_Applix_Vision_GB
GB
Insert giving set (See page 11)
Automatic priming using the pump Keep key depressed
Prime to the end of the giving set (Release key once priming is complete).
Program selection Press key
For selection of application programs see section "Application Programs". continuous volume/time bolus
Start Press key
Flashing bars and drops indicate the start of feed/fluid administration (*).
Stop Press key
STOP appears in the display.
Switch off Keep depressed for approx. 1 second
In mains mode, the "Plug" symbol is visible in the display. The pump retains all information when switched off. This information is saved for 1 month provided the battery was fully charged at switch off.
(*) In order to detect an upstream occlusion occurred before starting the pump, the APPLIX Vision delivers a small initial volume at a higher rate, right after START, if the door has been previously opened or an upstream occlusion alarm has been triggered (see Operating conditions p. 35).
4150-2_nu_Applix_Vision_GB
19
GB
ADMINISTRATION PROGRAM The APPLIX Vision offers three different administration programs:
Continuous:
Volume/time:
Bolus:
Feed is administered continuously at the selected speed
The Volume/time program is a continuous program that calculates the feed rate according to the specified feed quantity (target volume) and administration time. The rate calculated is shown in the display.
Feed is administered portionwise at the time intervals selected. The feed speed is fixed at 600 ml/h.
20