FSN Medical Technologies
FSN Medical Monitor FHD Model FS Series Instructions for Use Aug 2021
Instructions for Use
44 Pages
Preview
Page 1
FHD Monitor
Instructions for Use FS-Y1901D FS-E2101D FS-E2101DT FS-L2403D FS-P2404D FS-P2603D FS-P2604D FS-L2702D FS-L2702DT FS-L3202D
Before connecting, operating or adjusting this product, please read this instruction booklet carefully and completely. English
Table of Contents Product Description / Intended Use... 3 Symbol Definitions... 4 Safety Instructions... 5 Warnings and Precautions... 7 Electromagnetic Compatibility... 10 Accessories... 15, 16 Connections and Mechanical Drawings... 17-20 Controls... 21 On Screen Display (OSD)... 22-26 Standard Signal Tables... 27-30 General Specifications... 31-41 Cleaning Instructions... 42
The specifications and information in this document are subject to change without notice.
eIFU indicator
2
Instructions for Use for this product are also available in electronic form (eIFU). Choose from several languages. Use Adobe Acrobat software to view eIFUs. Access the eIFUs online at fsnmed.com/support/eifu/ 8/2021
Product Description / Intended Use
This product from FSN Medical Technologies is a high-end surgical display monitor designed for advanced digital OR applications. This medical display is uniquely equipped to handle tasks in the demanding operating room environment. Performance characteristics include: • • • • •
Rapid signal detection, robust mode tables Artifact-free images Fanless - sterile field compatible Calibrated to clinical color Image pan, zoom, freeze, picture-in-picture
Intended Purpose This device is intended to be connected to other medical equipment, and to display images or videos from endoscopic cameras, room cameras, and patient information such as ultrasound, cardiology, and anesthesiology. This device is not intended for diagnosis. This device is intended to be compatible with other highly specialized surgical and diagnostic equipment used in surgical suites, operating rooms, emergency rooms, and procedural facilities.
Intended Use Environment This device is intended to be used by a trained medical professional in a healthcare facility setting where contact with a patient is unlikely (no applied part). This device is designed to meet the medical safety requirements for a patient vicinity device. Warning: This device may not be used in connection with life support equipment.
Indications for Use This device is to be used by a trained medical professional to display images from procedures, such as endoscopy, ultrasound, cardiology, and anesthesiology. This device connects to medical imaging equipment to display images, videos or patient information during surgical procedures. This device is not intended for diagnosis.
8/2021
3
Symbol Definitions The following symbols appear on the product, its labeling, or the product packing. Each symbol carries a special definition, as defined below: Dangerous : High Voltage
Power adapter
Consult accompanying documents
Direct current
lndicates equipotential earth ground
Unique Device Identifier
Indicates protective earth ground
Indicates top-bottom direction
Korea Certification
DC Power control switch
Fragile
Approved according to the CCC regulations
Do not get wet
Maximum Stacking
China RoHS labels
Consult the operating instructions
Indicates the manufacturer
Catalog Number
Indicates the manufacturing date
Authorized representative in the European community
Medical Device
Serial Number
Humidity limitation eIFU indicator
Temperature limitation
Atmospheric pressure limitation
Consult the operating instructions - electronic Importer Entity
UK Conformity Assessed Indicates proof of conformity to EU 2017/745 Medical Devices Regulation and applicable standards. Medical Equipment is in accordance with ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012) and CAN/ CSA-C22.2 No. 60601-1 (2014) in regards to electric shock, fire hazards, and mechanical hazard. Tested to comply with FCC Class B standard (USA). Waste electrical and electronic equipment (WEEE Directive 2012/19/EU). This symbol indicates that the waste of electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact the manufacturer or other authorized disposal company to decommission your equipment. Note: A printed copy of the manual in English is provided with the product. Users within EU member states, please contact local distributor for other languages. This applies to EU member states where the product has been purchased through authorized channels.
4
8/2021
Warnings and Precautions Caution Information CAUTION RISK OF ELECTRICAL SHOCK DO NOT OPEN
This symbol alerts the user that important literature concerning the operation of this unit has been included. Therefore, it should be read carefully in order to avoid potential problems.
This symbol warns users that un-insulated voltage within the unit may have sufficient magnitude to cause electrical shock. Therefore, it is dangerous to make contact with any part inside the unit. To reduce the risk of electrical shock, DO NOT remove cover (or back). There are no user-serviceable parts inside. Refer servicing to qualified service personnel. To prevent fire or shock hazards, do not expose this unit to rain or moisture. Do not use this unit’s polarized plug with an extension cord receptacle or other outlets unless the prongs can be fully inserted.
Underwriters Laboratories (UL) Classification: UL safety Compliance: This medical LCD monitor is U.L. Classified WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1/CAN/CSA C22.2 NO. 601.1
EU Conformity and EMC Compliance: This medical LCD monitor unit meets the requirements of EN60601-1 and EN60601-1-2 so as to conform to the EU Medical Devices Regulation (MDR 2017/745). CE class I medical device accessory. This medical LCD monitor complies to the above standards only when used with the supplied medical grade power supply. Use 120V rating 5-15P type plug only in the U.S. JMW190KB1200F04
ATM065T-P120
ATM090-P240
BM120S24F02
BPM150S24F10
ATM160T-P240
FS-Y1901D
FS-E2101D FS-E2101DT
FS-L2403D
FS-P2404D
FS-P2603D FS-L3202D
FS-P2604D FS-L2702D FS-L2702DT
Caution: Make sure the power cord is the correct type that is required in your geographic area. This medical LCD monitor has a universal power supply that allows operation in either 100-120V AC or 200-240V AC voltage areas (no user adjustment is required).
8/2021
5
Use the proper power cord with correct attachment plug type. If the power source is 120 V AC, use a power cord which is a Hospital Grade Power Cord with NEMA 5-15 style plug, labeled for 125 volts AC with UL and C-UL approvals. If the power source is a 240 V AC supply, use the tandem (T blade) type attachment plug with ground conductor power cord that meets the respective European country’s safety regulations. A ground post, located on the back of the display, may be used for the purpose of grounding the display’s chassis. Any such ground must be installed in accordance with applicable electrical codes. The ground post is shown on the mechanical drawing found in this instructions for use.
Recycling (WEEE Directive 2012/19/EU) Follow local governing ordinances and recycling plans regarding the recycling or disposal of this equipment. Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation. Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of this medical LCD monitor, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Warning: Using this equipment in the X-ray or magnetic resonance environment could result in degradation of the performance of this equipment, interference with other equipment or interference with radio services. Warning: The use of cables and/or other accessories with this device, other than those specified, may result in increased emissions or decreased immunity of this device. Warning: This product is not considered physically to connect to HF (High Frequency) electrosurgical equipment. Warning: Not suitable for use in the presence of a flammable anesthetics mixture with oxygen or with nitrous oxide.
6
8/2021
Safety Instructions On Safety 1. Before connecting the AC power cord to the DC adapter outlet make sure the voltage designation of the DC adapter corresponds to the local electrical supply. 2. Never insert anything metallic into the cabinet openings of the medical LCD monitor. Doing so may create the danger of electric shock. 3. To reduce the risk of electric shock, do not remove cover. No user-serviceable parts inside. Only a qualified technician should open the case of the medical LCD monitor. 4. Never use your medical LCD monitor if the power cord has been damaged. Do not allow anything to rest on the power cord, and keep the cord away from areas where people can trip over it. 5. Be sure to hold the plug, not the cord,when disconnecting the medical LCD monitor power cord from an electric socket. 6. Unplug your medical LCD monitor power cord when it is going to be left unused for an extended period of time. 7. Unplug your medical LCD monitor power cord from the AC outlet before any service. 8. If your medical LCD monitor does not operate normally, in particular, if there are any unusual sounds or smells coming from it, unplug it immediately and contact an authorized dealer or service center. 9. Please contact the manufacturer if the set should be installed in an inaccessible area. Warning: Do not touch input or output connectors and the patient simultaneously. Warning: This medical LCD monitor is intended for connection to input/output signals and other connectors that comply with relevant IEC standard (e.g.,IEC60950 for IT equipment and IEC60601 series for medical electrical equipment). In addition, all such combination-system shall comply with the standard IEC 60601-1-1 or clause 16 of the 3 Ed. of IEC 60601-1, respectively, safety requirements for medical electrical systems. Any person who has formed a combination-system is responsible for the system to comply with the requirements of IEC 60601-1-1 or clause 16 of the 3 Ed. of IEC 60601-1, respectively. If in doubt, contact qualified technician or your local representative. Warning: To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth. Power supply (AC/DC Adapter) is specified as a part of the LCD Color Display. Do not position equipment in such a way that it is difficult to disconnect the power cord plug from the appliance inlet. Warning: Do not modify this equipment without authorization of the manufacturer. Product fuse has a lower breaking capacity. Do not install at the building power system, prospective short-circuit current exceeding 35 A. 8/2021
7
Environmental Conditions for Operation and Storage Temperature range within 0°C to 40°C(operation), -20°C to 60°C (storage) Relative humidity range 10% to 85% Atmospheric pressure range within 500 to 1060hPa.
On Installation 1. Openings in the medical LCD monitor cabinet are provided for ventilation. To prevent overheating, these openings should not be blocked or covered. If you put the medical LCD monitor in a bookcase or some other enclosed space, be sure to provide adequate ventilation. 2. Do not expose the medical LCD monitor to rain or use it near water. If the medical LCD monitor accidentally gets wet, unplug it and contact an authorized dealer immediately. You can clean the medical LCD monitor with a damp cloth if necessary, but be sure to unplug the medical LCD monitor first. 3. Place your medical LCD monitor near an easily accessible AC outlet. 4. High temperature can cause problems. Max operating temperature is 40°C. Don’t use your medical LCD monitor in direct sunlight and keep it away from heaters, stoves, fireplaces,and sources of heat. 5. Don’t place your medical LCD Monitor on an unstable stand, Medical LCD monitor may malfunction or fall. 6. This medical LCD monitor should not topple over when tilted at a 5° angle, in any position, during NORMAL USE, excluding transport. 7. In the position specified for transport, medical LCD monitor shall not overbalance when tilted at a 10 degree angle. 8. When carrying this product, please use both handles (if included) on the left and right side of the product, and carry using two people. If you want the product to be installed in another place, please call your service center. 9. Always use only the original cables and accessories with the device. 10. Do not lay this monitor on other equipment.
8
8/2021
Repair Do not attempt to service the medical LCD monitor yourself, as opening or removing covers may expose you to dangerous voltages or other hazards, and will void the warranty. Refer all servicing to qualified service personnel. Unplug the medical LCD monitor from its power source and refer servicing to qualified personnel under the following conditions: - If the power cord or plug is damaged or frayed. - If liquid has been spilled into the medical LCD monitor. - If objects have fallen into the medical LCD monitor. - If the medical LCD monitor has been exposed to rain or moisture. - If the medical LCD monitor has been subjected to excessive shock by being dropped. - If the cabinet has been damaged. - If the medical LCD monitor seems to be overheated. - If the medical LCD monitor emits smoke or abnormal odor. - If the medical LCD monitor fails to operate in accordance with the operating instructions.
Biohazards To prevent spreading of infections, this device should only be used in environments where biological decontamination can be successfully performed.
Returned Product After troubleshooting, if problems persist, disinfect the monitor and return it to FSN using the original packaging. Include the accessories that came with the monitor in the return shipment. Please enclose a brief explanation of the malfunction. Contact FSN Medical Technologies for a Return Authorization Number and instructions, prior to returning the device.
Accessories Use only accessories specified by the manufacturer, or sold with the medical LCD monitor.
Classification for Safety Compliance - Protection against electric shock : Class I including AC/DC adapter. This medical equipment is in accordance with ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012) and CAN/CSA-C22.2 No. 60601-1 (2014) in regards to electric shock, fire hazards, and mechanical hazard. - Applied Parts : No Applied Parts. - Degree of safety in the presence of flammable anesthetics mixture with air or with oxygen or with nitrous oxide. Not suitable for use in the presence of a flammable anesthetics mixture with oxygen or with nitrous oxide. - For critical applications, it is recommended to have a replacement monitor available. - Mode of operation : Continuous.
Notice to the user: Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established. Contact your local FSN Medical Technologies sales representative for information on changes and new products.
8/2021
9
Electromagnetic Compatibility This medical monitor unit has been designed and tested to comply with IEC 60601-1-2:2014/ AMD1:2020 requirements for EMC with other devices. To ensure electromagnetic compatibility (EMC), the monitor must be installed and operated according to the EMC information provided in this Instructions for Use. This medical monitor unit has been tested and found to comply with the limits of a Class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against interference. This monitor can radiate radio frequency energy and, if not installed and used in accordance with the instructions, it may interfere with other radio communications equipment. There is no guarantee that interference will not occur in a particular installation. If this equipment is found to cause harmful interference to radio or television reception, the user is encouraged to try to correct the interference by carrying out one or more of the following measures: 1. Reorient or relocate the receiving antenna. 2. Increase the distance between the medical LCD monitor and the subject of interference. 3. Plug the monitor into an outlet on a different electrical circuit than that to which the subject of interference is connected. 4. Consult the dealer or an experienced radio/TV technician for help.
NOTICES TO USER This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
FCC WARNING This medical LCD monitor generates or uses radio frequency energy. Changes or modifications to this medical LCD monitor may cause harmful interference unless the modifications are expressly approved in the instruction manual. The user could lose authority to operate this equipment if an unauthorized change or modification is made.
PRODUCT LIFETIME The performance of LCD panels may deteriorate over long periods of time. Periodically check that this monitor is operating correctly. The expected service life of the device is four years. Keep the monitor clean to prolong its operational lifetime.
10
8/2021
1. Guidance and manufacturer’s declaration - electromagnetic emission The medical LCD monitor is intended for use in the electromagnetic environment specified below. The user of the device should make sure that the medical LCD monitor is operated in such an environment. Interference emission measurements
Conformity level
Electromagnetic environment -guidance
RF emissions acc. to CISPR 11
Complies with Group 1
RF emissions acc. to CISPR 11
Complies with Class B
Emission of harmonic oscillations acc. to IEC 61000-3-2
Complies with Class A
Voltage fluctuations/flicker emissions acc. to IEC 61000-3-3
Complies
The characteristics of this device determined by broadcasting permit its industrial and hospital use (CISPR 11, Class A). When used in a living area (for which CISPR 11 usually requires Class B), this device may not provide adequate protection of radio services. The user must, if necessary, take remedial action such as implementation or reorientation of the device.
2. For the use of ME devices in professional healthcare facilities. Guidance and manufacturer’s declaration - electromagnetic immunity The medical LCD monitor is intended for use in the electromagnetic environment specified below. The user of the medical LCD monitor should make sure that it is used in such an environment. Interference immunity test
IEC 60601-1-2:2014 conformity level
Electrostatic discharge (ESD) acc. to IEC 61000-4-2
Complies ±2 kV, ±4 kV, ±6 kV, ± 8 kV contact discharge ±2 kV, ±4 kV, ±8 kV, ± 15 kV air discharge
Floors should be made of wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity must be at least 30%
Rapid transient electric interferences/ bursts acc. to IEC 61000-4-4
Complies ± 2 kV for mains lines ± 1 kV for input/output lines
The quality of the supply voltage should correspond to that of a typical business or hospital environment.
Surge acc. to IEC 61000-4-5
Complies ± 1 kV push-pull voltage ± 2 kV common-mode voltage
The quality of the supply voltage should correspond to that of a typical business or hospital environment.
Voltage dips, short interruptions and fluctuations of the supply acc. to IEC 61000-4-11
0 % UT*; 0.5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°
Mains power quality should be that of a typical commercial or hospital environment.
0% UT; 1 cycle and 70% UT; 25/30 cycles Single phase: at 0°
If the user of the device requests continued functioning even when interruptions of the power supply occur, it is recommended that the device be supplied from a power supply that is free of interruptions.
0% UT; 250/300 cycle
Electromagnetic environment-guidance
*Note: UT is the mains alternating voltage before applying the test levels.
8/2021
11
3. For the use of ME devices in professional healthcare facilities. Test specification for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment (according to IEC 60601-1-2:2014) The medical LCD monitor is intended for use in the electromagnetic environment specified below. The user of the medical LCD monitor should make sure that it is used in such an environment. Test frequency MHz
Band MHz
Service
Modulation
Maximum power W
Distance m
IMMUNITY TEST LEVEL V/m
385
380 to 390
TETRA 400
Pulse modulation 18 Hz
1.8
1.0
27
450
430 to 470
GMRS 460, FRS 460
FM ± 5 kHz stroke ± 1 kHz sine wave
2
1.0
28
704 to 787
Band 13, 17
Pulse modulation 217 Hz
0.2
1.0
9
800 to 960
GSM 800/900 TETRA 800, iDEN 820, CDMA 850, LTE Band 5
Pulse modulation 18 Hz
2
1.0
28
Pulse modulation 217 Hz
2
1.0
28
Pulse modulation 217 Hz
2
1.0
28
Pulse modulation 217 Hz
0.2
1.0
9
710 745 780 810 870 930 1720 1845
1700 to 1990
1970
2450
2400 to 2570
GSM 1800, CDMA 1900, GSM 1900, DECT, LTE Band 1,3, 4, 25 UMTS Bluetooth, WLAN 802.11 b/g/n, RFID 2450, LTE Band 7
5240 5500
5100 to 5800
WLAN 802.11 a/n
5785 *Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the medical LCD monitor may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
12
8/2021
4. Guidance and manufacturer’s declaration – electromagnetic immunity – for equipment and systems that are not life-supporting The medical LCD monitor is intended for use in the electromagnetic environment specified below. The user of the medical LCD monitor should make sure that it is used in such an environment. Interference immunity tests
IEC 60601-12:2014 test level
Conformity level
Electromagnetic environment – guidelines Portable and mobile RF communications equipment should be used no closer to any part of the medical LCD monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance: d = 1.2
Conducted RF disturbances acc. to IEC 61000-4-6 Radiated RF disturbances according to IEC 61 000·4·3
3 V rms 150 kHz to < 80 MHz 3 V/m 80 MHz to 2.5 GHz
3 V eff 3 V/m
P
Where P is the nominal power of the transmitter in watts [W] according to the information provided by the manufacturer of the transmitter and d is the recommended separation distance in meters [m]. The field strength of stationary transmitters at all frequencies on site a should be, according to a study, less than the conformity level b. d = 1.2 P 80 MHz to < 800 MHz d = 2.3 P 800 MHz to 2.5 GHz Interference may occur in the vicinity of equipment marked with the following symbol:
Note: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of buildings, objects, and persons.
a Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment of the stationary transmitters, a site survey should be considered. If the measured field strength in the location at which the device is used exceeds the above conformity levels, the device should be observed to verify normal operation. If unusual performance characteristics are observed, additional measures may be necessary, such as a modified orientation or a different location for the device. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
8/2021
13
5. Recommended separation distances between portable and mobile RF communications equipment and the medical LCD monitor The medical LCD monitor is intended for use in the electromagnetic environment in which the RF disturbances are controlled. The user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device – as a function of the output power of the communication device, as shown below. Nominal power of transmitter [W]
Separation distanced [m] according to frequency of transmitter 150kHz to< 80 MHz d = 1.2 P
80 MHz to< 800 MHz d = 1.2 P
800 MHz to 2.5 GHz d = 2.3 P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Connecting the Power Supply
AC power.
AC-DC adaptor.
Monitor
Power input on back of monitor.
Maximum DC Extension Cable Length* (feet)
FS-P2404D, FS-P2603D, FS-P2604D, FS-L2702D, FS-L2702DT, FS-L3202D
75
FS-L2403D
33
FS-Y1901D, FS-E2101D, FS-E2101DT
25
* If longer extension is used, there is a risk of abnormal operation of the product.
14
8/2021
Accessories Item
IFU
AC-DC Adaptor 6.23ft/1.9m
AC Power Cord 6ft/1.8m*
DVI-D Cable 6ft/1.8m
BNC Cable 6ft/1.8m
Mounting Screws
FS-Y1901D 1, 2
FS-E2101D 7
FS-E2101DT 6, 7
FS-L2403D 2, 3
FS-P2404D 1, 4 * US,UK,EU, China. Hospital grade. 1 D-SUB cable available for purchase. 2 S-Video cable available for purchase. 3 RS-232C cable available for purchase.
8/2021
4 DC cable terminal male/female available for purchase. 5 Remote control. 6 USB cable, 8.2ft/2.5m. 7 DVI to HDMI cable.
15
Accessories Item
IFU
AC-DC Adaptor 6.23ft/1.9m
AC Power Cord 6ft/1.8m
DVI-D Cable 6ft/1.8m
BNC Cable 6ft/1.8m
Mounting Screws
FS-P2604D 1, 4
FS-P2603D 4
FS-L2702D 4
FS-L2702DT 4, 6
FS-L3202D 4, 5 * US,UK,EU, China. Hospital grade. 1 D-SUB cable available for purchase. 2 S-Video cable available for purchase. 3 RS-232C cable available for purchase.
16
4 DC cable terminal male/female available for purchase. 5 Remote control. 6 USB cable, 8.2ft/2.5m. 7 DVI to HDMI cable.
8/2021
FS-L2403D
565.9 mm / 22.28 in. 85 / 3.35 520.4 / 20.49
100 / 3.94 326 12.83 100 3.94 378.4 14.90
GROUND
3G SDI VGA DVI S-VIDEO IN IN IN 3G SDI C-VIDEO RS-232C OUT
DC IN
FS-P2404D
82 / 3.23 575.4 mm / 22.66 in.
35.7 1.41
392 15.43
4-M4 100 / 3.94 VESA MOUNT HOLES
100 3.94
171.5 6.75
DVI 2 IN (fiber optional)
18
8/2021
FS-P2603D 88.8 / 3.5 668 mm / 26.3 in.
39.5 1.56
421 16.57
100 / 3.94
FS-P2604D
78 / 3.07 640 mm / 25.20 in.
31.7 1.25
402 15.83
8-M4 VESA MOUNT HOLES
100 / 3.94 200 / 7.87
100 3.94
171.2 6.74
DVI 2 IN (fiber optional)
8/2021
19
Controls On Screen Display (OSD) QUICK ADJUST. Press to decrease the display contrast. No need to enter the OSD main menu.
QUICK ADJUST. Press to decrease the display brightness. No need to enter the OSD main menu.
When the OSD menu is activated, press to decrease the adjustment of the selected function.
When the OSD menu is activated, press to move the menu selection downward.
Press to show the input selection menu and to change the display signal source.
QUICK ADJUST. Press to increase the display contrast. No need to enter the OSD main menu.
Press 1 second for DSUB Analog auto adjustment.
When the OSD menu is activated, press to enter sub menu, or increase the adjustment of the selected function.
Press to enable PIP (Picture in Picture) function.
Press to turn power on/ off to the display’s front LCD screen. If this icon is not illuminated, then the power switch on the back of the display has been turned off.
QUICK ADJUST. Press to increase the display brightness. No need to enter the OSD main menu. When the OSD menu is activated, press to move the menu selection upward.
Press to activate the OSD menu. When the OSD menu is active, press to exit from the main menu or submenu.
On Screen Display (OSD) Lock For models with key lock function, press Up + Down buttons simultaneously to enable or disable.
8/2021
21
On Screen Display (OSD) Menus FSN display monitors come equipped with a rich set of features for system set-up, image adjustments, and screen layout control. These features are managed through the On Screen Display, or OSD. Some options presented in the OSD are contextual and vary depending on the active input signal. See the Controls section for a complete description of each OSD button.
1. Enter the OSD To activate the OSD menu, press the MENU button on the front of the display monitor. To close the OSD menu, press the menu button to exit from the main menu or a sub menu.
2. Pick a Main Menu Category After entering the OSD, use the UP and DOWN buttons on the front of the display monitor to navigate to a main menu category: ADJUST, COLOR TEMP, IMAGE, SETUP, PIP.
3. Pick a Submenu Category
+ + -
After entering the desired main menu category, press the button to enter the submenus and DOWN buttons to navigate associated with the selected main menu. Use the UP to the desired submenu, then adjust as needed with the and buttons. Select the MENU button to exit from the submenu or main menu.
On Screen Display (OSD) Menus FS-Y1901D Sub Menu Categories Main Menu Categories
Sub Menu Value or Setting
Video Source Information
Submenus under the ADJUST menu (variable by signal type) 1. BRIGHTNESS Increases or decreases the brightness. (Range : 0~100) 2. CONTRAST Increases or decreases the contrast. (Range : 0~100) 3. CLOCK Increases or decreases the sampling frequency. (Range : 0~100) 4. SHARPNESS Adjusts the sharpness of video image (Range : 0~100) 5. PHASE Increases or decreases the phase level. (Range : 0~100) 6. SATURATION Changes the tone of color. (Range : 0~100) 7. COLOR Changes the richness of color. (Range : green 0~50, red 0~50) 8. BACKLIGHT Adjusts backlight dimming level. (Range : 0~100) 9. AUTO ADJUST fits to the most appropriate screen on the D-SUB Analog / RGBs signal.
22
8/2021