Prismaflex Spare Parts Instruction Replacing the Loader Stepper Motor Ver 2.0 Rev 4 Sept 2006
Spare Parts Instruction
2 Pages
Preview
Page 1
Loader Stepper Motor G5008001
Spare Parts Instruction Replacing the Loader Stepper Motor Tools needed: • •
PhiIips head screwdriver Smm Hex wrench
Note: • • • •
The installer is required to use an ESD (electro-static discharge) Grounding Wrist strap during this procedure to avoid unintentional damage to the electronic devices in the machine. Do not remove any cards or IC chips from their antistatic containers until you are ready to install them. When removing cards or chips from a system, immediately place them in an antistatic bag or container. When handling cards or IC's, hold them by their edges. Avoid touching the components and connector leads on the card. Avoid touching the leads on the IC. Do not slide cards or IC's over any surface. Avoid plastic, vinyl, and Styrofoam in your work area
Setup instructions 1. 2.
3.
4.
5.
ne is switched off. a Smm Hex wrench!
Disconnect the cable connected to the loader board and the earth connecting cable connected to the stepper motor. Unscrew the screws holding the stepper motor.
Turn the stepper motor a little bit counter clockwise and pull it out towards you.
6.
Separate the lead screw from the stepper motor Gently pull or screw it out from the motor. 7. Unscrew the two screws holding the lead screw and the lead screw fixing plate. S. Replace the lead screw and the lead screw fixing plate with the new one. Put back the screws and tighten them.
,.,
Template·2235
GAMBRO Renal Products
Page 1/2
Revision: 4 Reference: POP·2279
MDREC-100091
Vers ion: 2.0
Effective date:2006-09-04
Spare Parts Instruction
Loader Stepper Motor G5008001 9.
Thread the stepper motor on to the lead screw and gently push it on till the metal plate on the stepper motor passes the micro switch. 10. Put back the screws holding the stepper motor and tighten them. 11. Reconnect the earth connecting cable to the stepper motor and connect the cable from the stepper motor to the loader board. 12. Close the cabinet
13. To continuously improve our customer support, please send the following information to the below address. • Running Time: • Serial number: • Reason for change: • Changed component: • Country/Hospital .com E-mail - ric .customer
Final check: •
Enter SERVICE/DIAGNOSE and run Load/Unload tool, verify function.
Working time necessary: Replacement: Final check:
10 minutes. 5 minutes.
,..,
Template-2235
GAMBRO Rena l Products
Page 2/2
Revision: 4 Reference: POP-2279
MDREC-100091
Version: 2.0
Effective date:2006-09-04