Addendum

20 Pages

Giraffe Display Repair Addendum For future reference, add to your Giraffe OmniBed or Incubator Service Manual. The existing electro-luminescent display used in the Giraffe Incubator and Giraffe OmniBed accounts for a significant portion of the material cost for the product. This old technology display is also becoming increasingly hard to source (lead times are 15+ weeks) and further cost increases are expected. The Giraffe Incubator and OmniBed have been redesigned to include a new display that has a lower cost and is widely available. This new design has replaced the previous design on all production units going forward. The following repair procedures and illustrated parts are to be used in place of the existing service manual content for the new display components. The contents of this addendum will be incorporated into the service manuals in the next revision.  Conventions in this Addendum A Warning statement is used when the possibility of injury exists.  SENSITIVE TO ELECTROSTATIC DISCHARGE CAUTION An Electrostatic Discharge (ESD) Susceptibility symbol is displayed to alert service personnel that the part(s) are sensitive to electrostatic discharge and that static control procedures must be used to prevent damage to the equipment.  © 2009 by General Electric Company. All rights reserved.  M1181029 003  i