Operating Instructions
82 Pages
Preview
Page 1
ROTOFIX 32 A
DE
Bedienungsanleitung... 5
EN
Operating Instructions... 20
FR
Mode d'emploi ... 35
IT
Istruzioni per l'uso ... 50
Rev. 00 / 12.09
Andreas Hettich GmbH & Co. KG
AB1206DEENFRIT
A
Fig. 1
Fig. 2
2/82
ROTOFIX 32 A
IMPULS
START
RCF
STOP
EG-Konformitätserklärung EC Conformity Declaration Déclaration de conformité CE Dichiarazione di conformità alle norme CEE Andreas Hettich GmbH & Co. KG • Föhrenstraße 12 • D-78532 Tuttlingen • Germany
Das bezeichnete Gerät, inklusive Zubehör entspricht den aufgeführten EG-Richtlinien und Normen. The denoted device, including accessories corresponds to the listed EC guidelines and standards. L'appareil désigné, y compris les accessoires, correspond aux directives CE et aux normes énumérées. L'apparecchio designato, compresi gli accessori, è conforme alle direttive CE e alle norme citate.
Geräteart, Type of device, Type d'appareil, Tipo di apparecchio:
Laborzentrifuge mit Zubehör, Laboratory centrifuge with accessories, Centrifugeuse de laboratoire avec des accessoires, Centrifuga da laboratorio con accessori
Typenbezeichnung, Type designation, Désignation de modèle, Contrassegno tipo:
ROTOFIX 32 A
EG-Richtlinien/Normen, EC guidelines/standards, Directives CE/Normes, Direttive/Norme CEE:
2006/95/EG, EN 61010-1:2001, EN 61010-2-020:2006 2004/108/EG, EN 61326-1:2006 2006/42/EG, EN ISO 12100-1:2004, EN ISO 12100-2:2004 98/79/EG, EN 61010-2-101:2003
Tuttlingen, 28.09.2009
H. Eberle Geschäftsleiter, Manager, Directeur, Gerente
3/82
Andreas Hettich GmbH & Co. KG Föhrenstraße 12, D-78532 Tuttlingen / Germany Phone +49 (0)7461 / 705-0 Fax
+49 (0)7461 / 705-125
[email protected], [email protected] www.hettichlab.com
© 2006 by Andreas Hettich GmbH & Co. KG All rights reserved. No part of this publication may be reproduced without the prior written permission of the copyright owner. Änderungen vorbehalten! , Modifications reserved! , Sous réserve de modifications ! , Con riserva di modifiche! AB1206DEENFRIT / Rev. 00 / 12.09
4/82
EN
Contents 1
Use according to specification ... 21
2
Residual risks... 21
3
Technical specifications ... 21
4
Notes on safety ... 22
5
Symbol meanings... 23
6
Delivery checklist ... 23
7
Unpacking the centrifuge ... 24
8
Initial operation... 24
9
Opening and closing the lid... 24 9.1
Opening the lid ... 24
9.2
Closing the lid... 24
10
Installation and removal of the rotor ... 25
11
Loading the rotor ... 25
12
Control and display elements ... 26 12.1
Symbols on the control panel... 26
12.2
Keys and setting options... 26
13
Setting the brake step ... 27
14
Setting the centrifuging radius... 27
15
Centrifugation ... 27 15.1
Centrifugation with preselected time... 28
15.2
Continuous operation... 28
15.3
Short-time centrifugation... 28
15.4
Display of the relative centrifugal force (RCF) ... 28
16
Relative centrifugal force (RCF) ... 28
17
Centrifugation of materials or mixtures of materials with a density higher than 1.2 kg/dm ... 29
18
Rotor Identification ... 29
19
Emergency release ... 29
20
Maintenance and servicing ... 30
3
20.1
Centrifuge (housing, lid and centrifuging chamber) ... 30
20.1.1
Surface cleaning and care ... 30
20.1.2
Surface disinfection... 30
20.1.3
Removal of radioactive contaminants ... 30
20.2
Rotors and Attachments ... 31
20.2.1
Cleaning and care... 31
20.2.2
Disinfection ... 31
20.2.3
Removal of radioactive contaminants ... 31
20.2.4
Trunnions ... 31
20.2.5
Rotors and accessories with limited service lives ... 32
20.3 20.4
Autoclaving ... 32 Centrifuge containers... 32
21
Faults... 33
22
Change mains input fuse ... 34
23
Acceptance of the centrifuges for repair ... 34
24
Disposal... 34
25
Anhang / Appendix ... 65 25.1
20/82
Rotoren und Zubehör / Rotors and accessories ... 65
EN
1 Use according to specification The machine presented here is a medical product (laboratory centrifuge) according to the IVD guideline 98/79/EG. The centrifuge is used to separate substances or substance mixtures with a density of max. 1.2 kg/dm³. This also includes substances and substance mixtures of human origin. The centrifuge is only intended to be used for this purpose. A different use or application over and above this is deemed not in accordance with the specifications. The company Andreas Hettich GmbH & Co. KG undertakes no liability for damages resulting therefrom. Belonging to the application according to specification is also the observance of all references contained in the Instruction Manual and compliance with the inspection and maintenance works.
2 Residual risks The machine is constructed according to the state of the art and the recognized technical safety regulations. Improper use and handling can result in dangers to life and limb of the user or third parties and impairments to the machine or to other material assets. The machine is only to be used for the specified applications and only in an impeccable technical safety condition. Disturbances that can interfere with the safety are to be immediately rectified.
3
Technical specifications
Manufacturer Model Type Mains voltage (± 10%) Mains frequency Connected load Current consumption Max. capacity Allowed density Speed (RPM) Force (RCF) Kinetic energy Obligatory inspection (BGR 500) Ambient conditions (EN / IEC 61010-1)
Andreas Hettich GmbH & Co. KG D-78532 Tuttlingen ROTOFIX 32 A 1206 1206-01 208 – 240 V 1∼ 100 – 127 V 1∼ 50 – 60 Hz 50 – 60 Hz 300 VA 300 VA 1.4 A 3.0 A 4 x 100 ml / 32 x 15 ml 3 1.2 kg/dm 6000 4186 3160 Nm no
− Set-up site − Altitude
Indoors only Up to 2000 m above sea level
− Ambient temperature − Humidity
2°C to 40°C Maximum relative humidity 80% for temperatures up to 31°C, linearly decreasing to 50% relative humidity at 40°C.
− Excess-voltage category (IEC 60364-4-443) − Pollution degree Device protection class
ΙΙ 2 Ι Not suitable for use in explosion-endangered areas.
EMC − Emitted interference, Interference immunity Noise level (dependent on rotor) Dimensions
EN / IEC 61326-1, Class B
FCC Class B ≤ 57 dB(A)
− Width
366 mm
− Depth
430 mm
− Height Weight
257 mm 23 kg
21/82
EN
4
Notes on safety No claim of warranty will be considered by the manufacturer unless ALL instructions in this manual have been followed.
•
The centrifuge should be installed on a good, stable base.
•
Before using the centrifuge absolutely check the rotor for firm placement.
•
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous substances or objects may be within the safety margin of 300 mm around the centrifuge.
•
Rotors, suspensions and accessories that possess traces of corrosion or mechanical damage or if their term of use has expired may not be used any longer.
•
The centrifuge may no longer be put into operation when the centrifuging chamber has safetyrelated damages.
•
With swing-out rotors the trunnions must be regularly lubricated (Hettich Lubricating Grease No. 4051) in order to ensure consistent swinging out of the hangers.
•
Before the initial operation of your centrifuge you should read and pay attention to the operating instructions. Only personnel that has read and understood the operating instructions are allowed to operate the device.
•
Along with the operating instructions and the legal regulations on accident prevention, you should also follow the recognised professional regulations for working in a safe and professional manner. These operating instructions should be read in conjunction with any other instructions concerning accident prevention and environmental protection based on the national regulations of the country where the device is to be used.
•
This centrifuge is a state-of-the-art piece of equipment which is extremely safe to operate. However, it can lead to danger for users or others if used by untrained staff, in an inappropriate way or for a purpose other than that it was designed for.
•
The centrifuge must not be moved or knocked during operation.
•
In case of fault or emergency release, never touch the rotor before it has stopped turning.
•
To avoid damage due to condensate, when changing from a cold to a warm room the centrifuge must either heat up for at least 3 hours in the warm room before being connected to the mains, or run hot for 30 minutes in the cold room.
•
Only the rotors and accessories approved by the manufacturer for this device may be used (see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories").
•
The centrifuge rotor may only be loaded in accordance with the chapter "Loading the rotor".
•
When centrifuging with maxim revolutions per minute the density of the materials or the material mixtures may not 3 exceed 1.2 kg/dm .
•
The centrifuge may only be operated when the balance is within the bounds of acceptability.
•
The centrifuge may not be operated in explosion-endangered areas.
•
The centrifuge must not be used with: − inflammable or explosive materials − materials that react with one another producing a lot of energy.
•
If users have to centrifuge hazardous materials or compounds contaminated with toxic, radioactive or pathogenic micro-organisms, they must take appropriate measures. For hazardous substances centrifuge containers with special screw caps must strictly be used. In addition to the screw cap centrifuge containers, for materials in hazard category 3 and 4 a biosafety system must be used (see the World Health Organisation’s “Laboratory Biosafety Manual”). In a biosafety system, droplets and aerosols are prevented from escaping by a bioseal (packing ring). If the hanger of a biosafety system is used without the lid, the packing ring must be removed from the hanger in order to prevent the packing ring from being damaged during the centrifugation run. Damaged packing rings must not be used to seal the biosafety system. Without the use of a biosafety system the centrifuge is not microbiologically sealed in the sense of the EN / IEC 610101-2-020 standard.
22/82
EN
For further details of available biosafety systems see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories". If in doubt, you should obtain relevant information from the manufacturer. •
The centrifuge must not be operated with highly corrosive substances which could impair the mechanical integrity of rotors, hangers and accessories.
•
Repairs must only be carried out by personnel authorised to do so by the manufacturer.
•
Only original spare parts and original accessories licensed by the Andreas Hettich GmbH & Co. KG company are allowed to be utilised.
•
The following safety regulations apply: EN / IEC 61010-1 and EN / IEC 61010-2-020 as well as their national deviations.
•
The safe operation and reliability of the centrifuge can only be guaranteed if: − the centrifuge is operated in accordance with the operating instructions, − the electrical installation on the site where the centrifuge is installed conforms to the demands of EN / IEC stipulations, − prescribed tests to BGV A1, BGR 500 are carried out by an expert.
5
Symbol meanings Symbol on the machine: Attention, general hazard area. Before using the centrifuge implicitly read the operating instructions and pay attention to the safety relevant references! Symbol in this document: Attention, general hazard area. This symbol refers to safety relevant warnings and indicates possibly dangerous situations. The non-adherence to these warnings can lead to material damage and injury to personal. Symbol on the machine and in this document: Beware of biohazard. Symbol in this document: This symbol refers to important circumstances. Symbol on the machine and in this document: Symbol for the separate collection of electric and electronic devices according to the guideline 2002/96/EG (WEEE). The device belongs to Group 8 (medical devices). Applies in the countries of the European Union, as well as in Norway and Switzerland.
6 Delivery checklist The following items and accessories are delivered with the centrifuge: 1 2 1 1 1 1 1
Connecting cable Fuses Lubricating grease for trunnions Hex. pin driver Release pin Notes on moving the equipment safely Operating instructions
The rotor(s) and associated accessories are included in the delivery in the quantity.
23/82
EN
7
Unpacking the centrifuge
•
Lift the carton upward and remove the padding.
•
Do not lift by the handle rail. Observe the weight of the centrifuge, refer to chapter "Technical specifications". Lift the centrifuge on both sides with an appropriate number of helpers and place it on the laboratory table.
8
Initial operation
•
According to the laboratory instrument standards IEC 61010-2-020 an emergency switch to separate power supply in the event of a failure must be installed in the building electrical system. This switch has to be placed remote from the centrifuge, prefered outside of the room in which the centrifuge is installed or near by the exit of this room. Remove the transportation safety device from the bottom of the housing, see sheet ”Transportation safety device". Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
• •
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous substances or objects may be within the safety margin of 300 mm around the centrifuge. •
•
Do not place any object in front of the ventiduct. Keep a ventilation area of 300 mm around the ventiduct. Check whether the mains voltage tallies with the statement on the type plate. Connect the centrifuge with the connection cable to a standard mains socket. For connection ratings refer to Chapter "Technical specifications". Turn on the mains switch. Switch position "Ι". The last used centrifuge data will be displayed. Open the lid.
9
Opening and closing the lid
• • •
9.1
Opening the lid The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be opened under these circumstances, see the section on “Emergency release”.
• •
Swing handle rail on the lid upwards. The symbol " " (lid open) illuminates in the rotation indicator Open the lid.
9.2
.
Closing the lid
Do not bang the lid shut. •
Place the lid and swing handle rail on the lid downward. The symbol " " (lid closed) illuminates in the rotation indicator .
24/82
EN
10
Installation and removal of the rotor • A
• B C D
11
• • •
Clean the motor shaft (C) and the rotor drilling (A), and lightly grease the motor shaft afterwards. Dirt particles between the motor shaft and the rotor hinder a perfect seating of the rotor and cause an irregular operation. Place the rotor vertically on the motor shaft. The motor shaft dog (D) has to fit in the rotor slot (B). The alignment of the groove is labelled on the rotor. Tighten the rotor tension nut with the supplied wrench by turning in a clockwise direction. Check the rotor for firm seating. Loosening the rotor: Loosen the tension nut by turning in a counter clockwise direction, and turning until the working point for lifting. After passing the working point for lifting the rotor is loosened from the motor shaft cone. Turn the tension nut until the rotor is able to be lifted from the motor shaft.
Loading the rotor
Standard centrifuge containers of glass will not stand RCF values exceeding 4000 (DIN 58970, pg. 2). •
•
Check the rotor for firm seating. With swing-out rotors all rotor positions must be lined with identical hangers. Certain hangers are marked with the number of the rotor position. These hangers may only be used in the respective rotor position. Hangers that are marked with a set number (e.g. S001/4) may only be used in the set. The rotors and hangers may only be loaded symmetrically. The centrifuge containers have to be distributed evenly on all rotor positions. For authorised combinations see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories". In the case of angle rotors all possible rotor positions must be loaded, see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories".
Rotor is evenly loaded •
• • • • •
Not permitted! Rotor is not evenly loaded
On certain hangers, the weight of the maximum load or the weight of the maximum load and the maximum weight of the completely loaded hanger is specified. This weight may not be exceeded. In case of exception, see chapter " Centrifugation of materials or mixtures of materials with a density higher than 1.2 kg/dm3".The weight specified for the maximum loading includes the total weight of adapter, frame, centrifuging container and content. In containers with rubber inserts, the same number of rubber inserts must always be among the centrifuge containers. Always fill the centrifuge containers outside of the centrifuge. No liquid should be allowed to enter the centrifugal chamber during filling and swinging out of the hangers. The maximum filling quantity for the centrifuge containers specified by the manufacturer must not be exceeded. In order to maintain the weight differences within the centrifuge container as marginal as possible, a consistent fill level in the containers is to be heeded.
25/82
EN
12 Control and display elements See figure on page 2. Fig. 2:
Display and control panel
12.1
Symbols on the control panel Rotation indicator. The rotation indicator lights up and rotates anticlockwise while the rotor is turning. When the rotor is stationary, the status of the lid is displayed by symbols in the rotation indicator: Symbol : Lid open Symbol : Lid closed
Operator errors and occurring faults are indicated on the display (see Chapter "Faults").
12.2
Keys and setting options
RPM/RCF x 100 • Speed
A numeric value of 500 RPM up to the maximum rotor speed can be set. For maximum rotor speed, see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories". Preset in steps of 100 (RPM = displayed value x 100). If the key or is kept pressed, the value changes with increasing speed. • Display the brake step and the centrifuging radius. t
START
STOP
RCF
IMPULS
26/82
• Running time - Preset from 1 - 99 minutes, in 1 minute steps - Continuous operation "--" • Centrifuging radius. Input in centimeters. Preset from 5 - 16 centimeters, in 1 centimeter steps. For centrifuging radius, see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories". • Braking steps 0 or 1. Step 1 = short run-down time, Step 0 = long run-down time. If the key or is kept pressed, the value changes with increasing speed. • Start centrifugation run. • End centrifugation run. The rotor runs down with the preselected brake step. • Save the brake step and the centrifuging radius. • Display of the relative centrifugal force (RCF). The display of the relative centrifugal force (RCF) appears while the key RCF is kept pressed. • Short-time centrifugation. The centrifugation run occurs while the key IMPULS is kept pressed. • Display the brake step and the centrifuging radius.
EN
13
Setting the brake step
• • •
Switch off the mains switch. Keep the key beneath the speed indicator and the key IMPULS pressed simultaneously. Switch on the mains switch and release the keys again. The speed indicator shows the machine version and the time indicator shows the set brake step: e.g.:
IMPULS
START
IMPULS
START
RCF
STOP
RCF
STOP
•
If the machine version and brake step are not displayed, press the key under the speed indicator until they are displayed. The machine version is set by the manufacturer and cannot be changed. Set the desired brake step with the keys beneath the time indicator. Step 1 = short run-down time, Step 0 = long run-down time. For run-down times, see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories". Press the key STOP to save the setting.
14
Setting the centrifuging radius
•
The centrifuging radius must be entered in centimeters. • • • •
Switch off the mains switch. Keep the key beneath the speed indicator and the key IMPULS pressed simultaneously. Switch on the mains switch and release the keys again. Press the key beneath the speed indicator until the following display appears:
IMPULS
START
RCF
STOP
•
The set centrifuging radius is displayed in the speed indicator. Set the desired centrifuging radius with the keys beneath the time indicator. For centrifuging radius, see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories". Press the key STOP to save the setting.
15
Centrifugation
•
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous substances or objects may be within the safety margin of 300 mm around the centrifuge. If the permissible weight difference is exceeded within the rotor loading, the drive switches off during the runup time, and error -3- is displayed (see chapter "Faults"). The centrifugation run can be interrupted at any time by pressing the key STOP . . The time and speed can be changed during the centrifugation run, with the keys If the key or is kept pressed, the value changes with increasing speed. After a centrifugation run, the display flashes until the cover is opened or a key is pressed. If the symbol " " (lid closed) and " " (lid open) flashes alternately in the rotation indicator , operation of the centrifuge can only be continued after opening the lid. If rot xx is displayed, no centrifugation run has taken place because the rotor has been changed, see chapter "Rotor Identification". • •
Switch on the mains switch (switch position "Ι"). Load the rotor and close the centrifuge cover.
27/82
EN
15.1 • • •
Centrifugation with preselected time Set the desired speed with the keys beneath the speed indicator. Set the desired time with the keys beneath the time indicator. Press the key START . The rotation indicator appears while the rotor is turning. The time is displayed in minutes. The last minute is counted down in seconds. When the time is displayed in minutes, a point flashes next to the number.
•
After expiry of the time or if the centrifugation run is interrupted by pressing the key STOP , the rotor runs down with the set brake step. During the centrifugation run, the rotor speed or the resulting RCF value and the remaining time are displayed.
15.2 • • •
Continuous operation Set the desired speed with the keys beneath the speed indicator. Set the time to zero with the key beneath the time indicator. "--" is displayed. Press the key START . The rotation indicator appears while the rotor is turning. The time count starts from 0. The first minute is counted up in seconds, and then the time is displayed in minutes. When the time is displayed in minutes, a point flashes next to the number.
• Press the key STOP to end the centrifugation run. The rotor runs down with the set brake step. During the centrifugation run, the rotor speed or the resulting RCF value and the expired time are displayed.
15.3 • •
Short-time centrifugation beneath the speed indicator. Set the desired speed with the keys Keep the key IMPULS pressed. The rotation indicator appears while the rotor is turning. The time count starts from 0. The first minute is counted up in seconds, and then the time is displayed in minutes. When the time is displayed in minutes, a point flashes next to the number.
• Release the key IMPULS again to end the centrifugation run. The rotor runs down with the set brake step. During the centrifugation run, the rotor speed and the expired time are displayed.
15.4 Display of the relative centrifugal force (RCF) The relative centrifugal force (RCF) can be displayed during the centrifugation run. If the relative centrifugal force (RCF) is used, the centrifuging radius must be entered. • •
Keep the key RCF pressed during the centrifugation run. The relative centrifugal force (RCF) appears in the speed indicator (RCF = displayed value x 100). Release the key RCF again. The speed is displayed.
16 Relative centrifugal force (RCF) The relative centrifugal force (RCF) is given as a multiple of the acceleration of gravity (g). It is a unit-free value and serves to compare the separation and sedimentation performance. These values are calculated using the formula below:
⎛ RPM ⎞ RCF = ⎜ ⎟ ⎝ 1000 ⎠
2 × r × 1,118
⇒
RPM =
RCF r × 1,118
× 1000
RCF = relative centrifugal force RPM = rotational speed (revolutions per minute) r = centrifugal radius in mm = distance from the centre of the turning axis to the bottom of the centrifuge. For more on the centrifugal radius see the chapter ” Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories”. The relative centrifugal force (RCF) stands in relation to the revolutions per minute and the centrifugal radius.
28/82
EN
17 Centrifugation of materials or mixtures of materials with a density higher than 1.2 kg/dm3 When centrifuging with maxim revolutions per minute the density of the materials or the material mixtures may not exceed 1.2 kg/dm3. The speed must be reduced for materials or mixtures of materials with a higher density. The permissible speed can be calculated using the following formula:
Reduced speed (nred) =
1.2 x maximum speed [RPM] Greater density [kg/dm³]
e.g.: maximum speed RPM 4000, density 1.6 kg/dm3 nred =
1.2 kg/dm³ x 4000 RPM = 3464 RPM 1.6 kg/dm³
In the exceptional case that the maximum loading indicated on the hanger is exceeded, the speed must also be reduced. The permissible speed can be calculated using the following formula: Reduced speed (nred) =
maximum load [g] x maximum speed [RPM] actual load [g]
e.g.: maximum speed RPM 4000, maximum load 300 g, actual load 350 g nred =
300 g x 4000 RPM = 3703 RPM 350 g
If in doubt you should obtain clarification from the manufacturer.
18 Rotor Identification After every start of a centrifugation run the rotor utilised is identified. After a rotor change, the drive switches off and the rotor code (rot xx) is displayed. •
Press the key START . The last used centrifuge data will be displayed.. A further operation of the centrifuge is only possible after a single opening of the lid. If, following a rotor change, the maximum speed of the rotor is less than the set speed, the speed is limited to the maximum speed of the rotor.
19 Emergency release The lid cannot be opened during power failure. An emergency release has to be executed by hand.
For emergency release disconnect the centrifuge from the mains. Open the lid only during rotor standstill. Only the plastic release pin provided may be used for emergency release. See figure on page 2. • • • •
Switch off the mains switch (switch position "0"). Look through the window in the lid to be sure that the rotor has come to a standstill. Insert the release pin (see scope of supply) horizontally into the hole (Fig. 1, A). Push the unlocking pin in until the handle can be lifted when the pin is pressed down. Open the lid.
29/82
EN
20
Maintenance and servicing
The device can be contaminated.
Pull the mains plug before cleaning. Before any other cleaning or decontamination process other than that recommended by the manufacturer is applied, the user has to check with the manufacturer that the planned process does not damage the device. • • • • •
Centrifuges, rotors and accessories must not be cleaned in rinsing machines. They may only be cleaned by hand and disinfected with liquids. The water temperature must be between 20 – 25°C. Only detergents/disinfectants may be used which: − have a pH between 5 - 8 − do not contain caustic alkalis, peroxides, chlorine compounds, acids and alkaline solutions In order to prevent appearances of corrosion through cleaning agents or disinfectants, the application guide from the manufacturer of the cleaning agent or disinfectant are absolutely to be heeded.
20.1
Centrifuge (housing, lid and centrifuging chamber)
20.1.1 • • • • • • •
Surface cleaning and care
Clean the centrifuge housing and the centrifuging chamber regularly, using soap or a mild detergent and a damp cloth if required. For one thing, this services purposes of hygiene, and it also prevents corrosion through adhering impurities. Ingredients of suitable detergents: soap, anionic tensides, non-ionic tensides. After using detergents, remove the detergent residue by wiping with a damp cloth. The surfaces must be dried immediately after cleaning. In the event of condensation water formation, dry the centrifugal chamber by wiping out with an absorbent cloth. Lightly rub the rubber seal of the centrifuge chamber with talcum powder or a rubber care product after each cleaning. The centrifuging chamber is to be checked for damage once a year. If damage is found which is relevant to safety, the centrifuge may no longer be put into operation. In this case, notify Customer Service.
20.1.2 • • • •
20.1.3 • • • •
Surface disinfection
If infectious materials penetrates into the centrifugal chamber this is to be disinfected immediately. Ingredients of suitable disinfectants: ethanol, n-propanol, isopropyl alcohol, glutardialdehyde, quaternary ammonium compounds. After using disinfectants, remove the disinfectant residue by wiping with a damp cloth. The surfaces must be dried immediately after disinfecting.
Removal of radioactive contaminants
The agent must be specifically labelled as being an agent for removing radioactive contaminants. Ingredients of suitable agents for removing radioactive contaminants: anionic tensides, non-ionic tensides, polyhydrated ethanol. After removing the radioactive contaminants, remove the agent residue by wiping with a damp cloth. The surfaces must be dried directly after removing the radioactive contaminants.
30/82
EN
20.2
Rotors and Attachments
20.2.1 • • • • • •
• •
Cleaning and care
In order to prevent corrosion and material changes, rotors and accessories must be cleaned regularly with soap or a mild detergent and a damp cloth. Cleaning is recommended at least once a week. Contaminants must be removed immediately. Ingredients of suitable detergents: soap, anionic tensides, non-ionic tensides. After using detergents, remove detergent residue by rinsing with water (only outside of the centrifuge) or wipe off with a damp cloth. The rotors and accessories must be dried directly after cleaning. Angle rotors, container and hanger made of aluminium are to be lightly greased after drying using acid-free grease, e.g. vaseline. In the case of biosafety systems (for further details of available biosafety systems see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories"), the packing rings must be checked and cleaned regularly (weekly).The sealing ring is to be replaced immediately upon indication of crack formation, embrittlement or abrasive wear. In order to prevent the packing ring from twisting when opening and closing the cover, the packing ring must be lightly rubbed with talcum powder or a rubber care product. In order to prevent corrosion as a result of moisture between the rotor and the motor shaft, the rotor should be disassembled and cleaned at least once a month, and the motor shaft should be lightly greased. The rotors and the accessory parts have to be checked on a monthly basis for wear and corrosion. For swing-out rotors, it is important to check the area of the lifting lugs, for hangers, the grooves and the base should be checked for cracks. Rotors and attachments may no longer be utilised upon indication of wear and tear or corrosion.
•
Check the firm seating of the rotor on a weekly basis.
20.2.2 • • • •
If infectious material should get on the rotors or accessories, they must be appropriately disinfected. Ingredients of suitable disinfectants: glutaraldehyde, propanol, ethyl hexanol, anionic tensides, corrosion inhibitors. After using disinfectants, remove disinfectant residue by rinsing with water (only outside of the centrifuge) or wipe off with a damp cloth. The rotors and accessories must be dried directly after disinfection.
20.2.3 • • • •
Disinfection
Removal of radioactive contaminants
The agent must be specifically labelled as being an agent for the removal of radioactive contaminants. Ingredients of suitable agents for removing radioactive contaminants: anionic tensides, non-ionic tensides, polyhydrated ethanol. After removing the radioactive contaminants, remove agent residue by rinsing with water (only outside of the centrifuge) or wipe off with a damp cloth. The rotors and accessories must be dried directly after removing the radioactive contaminants.
20.2.4 Trunnions With swing-out rotors the trunnions must be regularly lubricated (Hettich Lubricating Grease No. 4051) in order to ensure consistent swinging out of the hangers.
31/82
EN
20.2.5 Rotors and accessories with limited service lives The use of certain rotors, hangers and accessory parts is limited by time. These are marked with the maximum permitted number of operating cycles or with an expiration date and the maximum permitted number of operating cycles or just with the expiration date; e.g.: - "einsetzbar bis Ende: ΙV. Quartal 2011 / usable until end of: ΙV. Quarter 2011" or "einsetzbar bis Ende Monat/Jahr: 10/2011 / usable until end of month/year: 10/2011" - "Max. Lauf Zyklen / max. cycles: 40000".
For safety reasons, rotors, hangers and accessory parts may no longer be used if either the indicated maximum number of operating cycles or the indicated expiration date has been reached.
20.3 Autoclaving Swing-out rotors, angle rotors made of aluminium, suspension made of metal, lids with bio-containment as well as stands and reductions can be autoclaved at 121° C / 250°F (20 mins). Otherwise you must ask the manufacturer. No statement can be made about the degree of sterility.
The lids of the rotors and containers must be removed prior to autoclaving. Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics. We recommend that the packing rings of the bio-safety system be replaced after autoclaving.
20.4 • •
Centrifuge containers
With leakiness or after the breakage of centrifuging containers broken container parts and leaked centrifugation material are to be completely removed. The rubber inserts as well as the plastic sleeves of the rotors are to be replaced after a glass breakage. Remaining glass splitters cause further glass breakage!
•
If this concerns infectious material, a disinfection process is to be executed immediately.
32/82
EN
21 Faults If the fault cannot be eliminated with the help of the fault table, please inform Customer Service. Please specify the type of centrifuge and the serial number. Both numbers can be found on the name plate of the centrifuge.
Perform a MAINS RESET: − Switch off the mains switch (switch position "0"). − Wait at least 10 seconds and then switch on the mains switch again (switch position "Ι").
Fault No display
Display ---
Tacho error
-1-
Cause of fault No voltage Mains input fuses defective.
Failure of speed impulses during operation.
-
-
-
System reset
-2-
Power failure during the centrifugation run. (The centrifugation run was not finished.)
-
-
-
Balance error
-3-
The rotor is unevenly loaded.
Communication
-4-
Fault in control unit or power unit.
-
Overload
-5-
Motor or motor control defective.
-
Overvoltage Undervoltage
-6-8-
Supply voltage outside tolerance (see Technical Data).
-
Overspeed
-7-
Fault in the power unit.
Excess temperature
-9-
Excess temperature switch in motor has triggered.
Version error
A number is Incorrect machine version set, control displayed in unit jumps into Setting menu. the time indicator.
-
-
Controller watchdog
-C-
Fault in control unit.
-
Lid error
-d-
Error in lid locking or lid closure.
-
Short circuit
-E-
Short circuit in control unit / power unit.
-
No rotor code
-F-
-
New rotor identified
rot...
No rotor recognition at start. No rotor fitted or defective tacho. see section rotor identification.
-
Remedy Check distribution voltage. Check mains power input fuse, refer to Chapter "Change mains input fuse". Mains switch ON. Due to safety reasons the lid can only be opened after approx. 120 seconds. After this time has expired perform a MAINS-RESET. When stationary, open lid and press START key. Repeat the centrifugation run if necessary. Open lid when rotor is stationary. Check the loading of the rotor, see chapter "Loading the rotor". Repeat the centrifugation run. Perform a MAINS RESET when the rotor has been stationary. Perform a MAINS RESET when the rotor has been stationary. Perform a MAINS RESET when the rotor has been stationary. Check supply voltage. Perform a MAINS RESET when the rotor has been stationary When rotor is stationary, open lid using emergency unlocking (see Emergency Unlocking chapter). Allow motor to cool down. Set the number 4 using the keys underneath the time indicator. Press the key STOP to save the setting. Perform a MAINS RESET. Perform a MAINS RESET when the rotor has been stationary. Perform a MAINS RESET when the rotor has been stationary. Perform a MAINS RESET when the rotor has been stationary. Perform a MAINS RESET when the rotor has been stationary. Press the key START .
33/82
EN
22
Change mains input fuse
Switch off the mains switch and separate the centrifuge from the mains!
The fuse holder (A) with the mains input fuses is located next to the mains switch. • • •
A
Remove the connecting cable from the machine plug socket. Press the snap-fit (B) against the fuse holder (A) and remove. Exchange defective mains input fuses. Only use fuses with the rating defined for the type. See the following table.
B • •
Reinsert the fuse holder until the snap-fit clicks shut. Reconnect the centrifuge to the mains supply.
Model ROTOFIX 32 A ROTOFIX 32 A
Type 1206 1206-01
Fuse T 3.15 AH/250V T 5 AH/250V
Order no. E997 E914
23 Acceptance of the centrifuges for repair If the centrifuge is returned to the manufacturer for repair, it must be decontaminated and cleaned to protect persons, environment and material.
We reserve the right to accept contaminated centrifuges. Costs incurred for cleaning and disinfection are to be charged to the customer. We ask for your understanding in this matter.
24 Disposal Before disposal, the device must be decontaminated and cleaned to protect people, the environment and property. When you are disposing of the device, the respective statutory rules must be observed. Pursuant to guideline 2002/96/EC (WEEE), all devices supplied after August 13, 2005 may not be disposed as part of domestic waste. The device belongs to group 8 (medical devices) and is categorized in the business-to-business field. The icon of the crossed-out trash can shows that the device may not be disposed as part of domestic waste. The waste disposal guidelines of the individual EC countries might vary. If necessary, contact your supplier.
34/82
DE
EN
1324
1398
Ausschwingrotor 4-fach / Swing out rotor 4-times
1482
2x
2x 0500
0507
9 14 x 100 16 4000 2522 141
15 17 x 100 16 4000 2522 141
90°
Kapazität / capacity ml mm Maße / dimensions ∅ x L Anzahl p. Rotor / number p. rotor Drehzahl / speed RZB / RCF Radius / radius 9 (97%) 9 Probenerwärmung/Sample temp. rise
RPM mm sec sec
2,6 –2,9 13 x 65 16 4000 2272 127
4 – 4,5 15 x 75 16 4000 2272 127
9 – 10 16 x 92 16 4000 2522 141
10 15 x 102 16 4000 2522 141
12 17 x 100 16 4000 2522 141
4-7 16 x 75 16 4000 2397 134
27 30
K 1)
9
1324
1398
Ausschwingrotor 4-fach / Swing out rotor 4-times
1483
1484
1484
1482
ohne / without E2110
0509
Falcon ®
0513
0518
15 17 x 120 16 4000 2612 146
50 29 x 115 4 4000 2576 144
50 29 x 115 4 4000 2576 144
15 17 x 100 16 4000 2522 141
90°
Kapazität / capacity ml mm Maße / dimensions ∅ x L Anzahl p. Rotor / number p. rotor Drehzahl / speed RZB / RCF Radius / radius 9 (97%) 9 Probenerwärmung/Sample temp. rise
RPM mm sec sec K 1)
1) Probenerwärmung bei maximaler Drehzahl und 1 Stunde Laufzeit (nur bei Zentrifuge ohne Kühlung)
68/82
27 30 9 1) Sample temp. rise during maximum speed and 1 hour running time (only with centrifuges without cooling)