Operating Instructions
13 Pages
Preview
Page 1
Operating Instructions
Anaesthetic Gas Evacuation System
AGFS
Operating Instructions AGFS
1. Manufacturer's information ... 3 2. Foreword... 4 3. Function and mode of operation... 5 4. Assembly dimensions and parts specification... 6 4.1
Front view with front plate...6
4.2
Front view without front plate...7
4.3
Side view...7
4.4
Hose plan...9
5. Technical data... 11 6. List of spare parts ... 12 7. Warranty... 13
2
Operating Instructions AGFS
1. Manufacturer's information
Product
Anaesthetic Gas Evacuation System AGFS
Manufacturer and Distributor
Heyer Aerotech GmbH Nieverner Straße 30 56132 Nievern Tel.: (0 26 03) 20 04 Fax.: (0 26 03) 40 20 E-Mail: [email protected]
3
Operating Instructions AGFS
2. Foreword Dear customer! Thank you for choosing a HEYER Aerotech product! HEYER Aerotech products represent stateof-the-art technology and are manufactured according to the highest quality standards. You have chosen a professional, durable, easy-to-maintain product, the result of many years of experience and know-how from a competitive supplier of medical technology. The anaesthetic gas evacuation system AGFS is in accordance with all standards and specifications that currently apply to the production and operation of this product. In the unlikely event of malfunction please contact our Customer Service. For details refer to page 3 of these Operating Instructions.
Before using the unit, please read these Operating Instructions carefully and particularly observe the warnings.
4
Operating Instructions AGFS
3. Function and mode of operation Whenever anaesthesia is carried out in half-closed or half-open systems, excess anaesthetic gas escapes through the pressure control valve of the circulation system and is inhaled by users and staff. This is avoided by installing an anaesthetic gas evacuation system. The anaesthetic gas evacuation system AGFS is installed in the vicinity of the anaesthesia working area. A transvector integrated in the AGFS generates the required underpressure. The suction power is adjusted separately on every evacuation box to a standard value of 45-50 l/min, but it can be tailored to the customer's needs. The anaesthetic gas evacuation system is started up by connecting the anaesthetic gas suction plug, supplying the transvector with driving air (5 bar). The underpresssure generated by the transvector is indicated by the green colour of the visual signal. The anaesthetic gas drawn off, including the driving air of the transvector, is passed into the open air through a separate pipe or duct. When the plug is disconnected the driving gas inflow to the transvector is interrupted, which ceases to generate a low pressure, and the visual signal turns white. Simultaneously the integrated safety valve closes and prevents the anaesthetic gas in the exhaust pipe from flowing back.
5
Operating Instructions AGFS
4. Assembly dimensions and parts specification 4.1 Front view with front plate
1 2
3
Fig. 1 Front view with front plate
Legend: 1 2
Front plate Connection AGFS gas suction plug
3
Visual signal
6
Operating Instructions AGFS
4.2 Front view without front plate 7
8
4 5 9
6
Fig. 2 Front view without front plate
Legend: 4 Bend 15 5001A 5 Hose clip 6 Connection AGFS suction plug
7 VACUFLEX hose 8 Flush socket AGFS 9 Transvector complete
Ensure that no foreign substance can enter the open connections.
7
Operating Instructions AGFS
4.3 Side view
10 11 12
19
13 14 15
16
17
18
Fig. 3 Side view
Legend: 10 11 12 13 14 15
O-ring 14 x 1 Valve disk Spring for valve disk O-ring AGFS, upper part Flat head screw
16 17 18 19
Valve housing AGFS Seat valve O-ring 7 x 1,5 Headless pin
8
Operating Instructions AGFS
4.4 Hose plan
20
21
22 24
23
Fig. 4 View of hose plan
Legend: 20 PE hose 150 mm 21 PE hose 150 mm 22 PE hose 240 mm
23 Choke valve 24 Swing fixture
9
Operating Instructions AGFS
10
Operating Instructions AGFS
5. Technical data Performance: Suction power: 45-50 l/min Air consumption: 40-60 l/min maximum Suction volume: 80 l/min maximum
Connections: Compressed air: Copper pipe 8 x 1 mm Exhaust air: Copper pipe 15 x 1 mm
Operating pressure: 5 bar
Dimensions: Front plate: 160 x 160 mm Flush socket: 150 x 150 x 60 mm Plaster equalisation: stepless up to approx. 20 mm
11
Operating Instructions AGFS
6. List of spare parts Pos.
Order number
Description
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 not shown not shown not shown
117-1955 --117-1440 131-1810 046-4080 047-4102 006-3900 117-1600 117-2010 049-3102 117-1370 117-1200 117-1360 117-1300 117-1150 047-3606 117-1155 117-1165 049-3081 047-4172 017-1550 017-1550 017-1550 017-3091 017-3076 017-3095
Front plate AGFS Connection AGFS plug Green visual signal Bend 15 5001A Hose clip Cap screw (Allen) VACUFLEX hose Flush socket AGFS Transvector complete O-ring 14 x 1 Valve disk Spring for valve disk Ball ∅ 4 Lip seal gasket AGFS, upper part Flat head screw Valve housing AGFS Seat valve O-Ring 7 x 1,5 Headless pin PE hose 150 mm PE hose150 mm PE hose 240 mm Choke valve Swing fixture T-piece Kuhnke
017-3100
Spacer
047-3622
Flat head screw
12
Operating Instructions AGFS
7. Warranty Warranty for the HEYER Aerotech anaesthetic gas evacuation system AGFS is twelve (12) months from the date of purchase, according to the following conditions: 1. Within the guarantee period we will remedy free of charge any product which proves to be defective or non-conforming in materials or workmanship, if we are notified without delay as soon as the defect or non-conformity is detected. The warranty does not extend to fragile parts, such as glass, or perishable parts and consumables. 2. In the case of warranty we will at our discretion repair or replace the product. Neither does such claim cause the guarantee period to be extended, nor will a new guarantee be started. There will be no independent guarantee period for any spare parts installed. 3. This warranty does not apply to defects arising out of improper or inexpert treatment, faulty operation, mechanical damage or disregard of our Operating Instructions, as well as any such damage caused by acts of God or exceptional environmental conditions. 4. These guarantee conditions are invalidated in the case of third-party intervention, modification and/or repair of the product by persons not authorised by HEYER Aerotech, or if the product is operated with auxiliary equipment or spare parts of from another source. Any and all further claims as well as so-called incidental or consequential damages are excluded, insofar as not otherwise compelling stated by law.
Subject to technical changes! Rev. Nr. : 1.1