CRYSTAL VISION Model 450-D Operating and Installation Manual Rev C July 2016.pdf
Page 1
I. C. MEDICAL, INC.
CRYSTAL VISION®
Model 450D
REF
ICM-450-0000
OPERATING AND INSTALLATION MANUAL
I. C. Medical, Inc.
2340 W. Shangri La Road
Phoenix, Arizona 85029 USA
Tel: +1(623) 780-0700 (USA)
Fax: +1(623) 780-0887 (USA)
MDSS GmbH
Schiffgraben 41
30175 Hannover, Germany
Made in USA
Copyright 1996-2005. All rights reserved.
Document#: ICM-450-9000 Rev: C Date: 2016-07
Page 2
Table of Contents
LIMITED WARRANTY: ........................................................................................................................................ III
IMPORTANT SAFEGUARDS AND NOTICES ................................................................................................... IV
GENERAL WARNINGS ........................................................................................................................................... V
GENERAL CAUTIONS............................................................................................................................................. V
SPECIFICATIONS ..................................................................................................................................................... 1
INSTALLATION/OPERATIONS INSTRUCTIONS .............................................................................................. 3
FILTERS AND POWER CORD INSTALLATION: .................................................................................................. 4
ESU SENSOR INSTALLATION: .............................................................................................................................. 4
LASER SENSOR INSTALLATION .......................................................................................................................... 5
CHECK PROPER OPERATION OF THE CRYSTAL VISION®.......................................................................... 7
DESCRIPTION OF SWITCHES, CONTROL BUTTONS & INDICATORS....................................................... 9
LAPAROSCOPIC OPERATION ............................................................................................................................ 13
NON-LAPAROSCOPIC GYNECOLOGY PROCEDURES: ................................................................................ 15
OTHER NON-LAPAROSCOPIC PROCEDURES: .............................................................................................. 16
THEORY OF OPERATION .................................................................................................................................... 18
PNEUMATIC CIRCUITS ........................................................................................................................................ 18
ELECTRONIC CIRCUITS ...................................................................................................................................... 18
START PUMP: ........................................................................................................................................................ 18
STOP PUMP: .......................................................................................................................................................... 18
TIME SETTING: ...................................................................................................................................................... 18
FLOW SETTING: .................................................................................................................................................... 19
OPEN MODE: ......................................................................................................................................................... 19
LAP MODE: ............................................................................................................................................................ 19
TROUBLE SHOOTING GUIDE ............................................................................................................................. 20
PREVENTIVE MAINTENANCE ............................................................................................................................ 22
CLEANING INSTRUCTIONS: ............................................................................................................................... 22
CRYSTAL VISION® PARTS/ORDER INFORMATION ..................................................................................... 23
HOSPITAL LEVEL CALIBRATION INSTRUCTIONS ...................................................................................... 24
i
Page 3
Table of Figures
Figure 1: Filters and Power Cord Installation .................................................................................................................4
Figure 2: ESU sensor installation when used with an electrosurgical unit (ESU) ............................................................4
Figure 3: Laser Sensor Installation .................................................................................................................................5
Figure 4: Laser Sensor Installation to the SIDE of the laser foot switch HOUSING........................................................5
Figure 5: Sensors should be positioned so that ANY movement of the pedal interrupts the beam. ..................................6
Figure 6: ON/OFF power switch description. .................................................................................................................7
Figure 7: Manual and Time Control locations. ...............................................................................................................7
Figure 8: Flow buttons. ..................................................................................................................................................8
Figure 9: ON/OFF power switch location.......................................................................................................................9
Figure 10: Manual button and sensor connectors. ...........................................................................................................9
Figure 11: Adjustment for time setting. ........................................................................................................................10
Figure 12: Setting the flow RANGE.............................................................................................................................10
Figure 13: Selecting the desired flow. ..........................................................................................................................11
Figure 14: Front Panel Indicators. ................................................................................................................................11
Figure 15: Input Filter connector. Charcoal Output Filter .............................................................................................12
Figure 16: Laparoscopy set up when using the CRYSTAL VISION®. .......................................................................13
Figure 17: Set up for gynecology procedure for either ESU or CO2 laser delivered through a colposcope or
Micromanipulator .................................................................................................................................................15
Figure 18: Set up for using CRYSTAL VISION® with an ESU system during an open procedure. ..........................16
Figure 19: Set-up for using a CRYSTAL VISION® with a laser hand piece. ..............................................................16
ii
Page 4
LIMITED WARRANTY:
For the periods and the conditions specified below, I. C. Medical, Inc. warrants to
the original purchaser that I. C. Medical, Inc.'s products will perform to our
published specifications when used and maintained in accordance with our written
instructions.
With respect to the Crystal Vision®, the warranty period is one (1) year from
delivery.
If due to a defect in materials or workmanship a Product fails to perform to our
published specification, or if a Consumable is not free from defects in materials and
workmanship when shipped from our factory, I. C. Medical, Inc. will, at its option,
repair or replace the defective Product or Consumable without charge, using new or
remanufactured parts. I. C. Medical, Inc. reserves the right to make a repair in its
factory, at any authorized repair facility. The purchaser shall pay factory return
shipping charges, if any.
The warranty is null and void if the purchaser attempts to service or repair the
product (other than the performance of routine maintenance as described in the
Operator's Manual), or if service is performed by persons not authorized by I. C.
Medical, Inc. In addition, the warranty is null and void if the Product or consumable
is used other than as specified in the Operators Manual. Without limitation, the
warranty does not cover damage caused by customer misuse of a Product.
THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTIES EXPRESSED
OR IMPLIED, AND ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE IS EXPRESSLY DISCLAIMED.
Purchaser's exclusive remedy for any failure of any Product or Consumable is as
provided in this Limited Warranty, and in no event shall I. C. Medical, Inc. be liable
for any special, incidental, consequential, indirect or other similar damages arising
from breach of warranty, breach of contract, negligence or any other legal theory.
iii
Page 5
IMPORTANT SAFEGUARDS AND NOTICES
The following pages provide important guidelines for operators and service personnel. Specific warnings and
cautions appear throughout the manual where they apply. Please read and follow this important information,
especially those instructions related to risk of electric shock or injury to patient or staff members.
Warning: Any instructions in this manual that require opening the equipment cover or enclosure are for
use by I. C. Medical, Inc. qualified service personnel only. To reduce the risk of electric shock, do not
perform any other service than that contained in the operating instructions unless you are determined by
IC. Medical, Inc. to be qualified to do so.
Caution risk of danger or Attention, consult ACCOMPANYING DOCUMENTS
The lightning flash with arrowhead symbol within an equilateral triangle alerts the user to the
presence of electrical shock risks within the smoke evacuator’s enclosure. Always take
precautions to protect users/qualified personnel from electrical shock risks.
The device is Class 1,
Type BF applied part
The fuse symbol indicates that the fuse referenced in the text must be replaced with one having the
ratings indicated.
The Thermometer represents that during shipping and storage, the smoke evacuator can be left in
an environment where temperatures range from -40ºC to +70ºC and pressures of 500 hPa to 1060 hPa
This symbol represents that during shipping and storage, the smoke evacuator can be left in an
environment where Relative Humility can range from 10% to 100%.
Federal (USA) law restricts this device to sale by or on the order of a physician.
iv
Page 6
GENERAL WARNINGS
A warning indicates a possible hazard to personnel, which may cause injury. Observe the following general
warnings when using or working on this equipment:
Heed all warnings on the unit and in the operating instructions.
Do not use this equipment in or near water.
This equipment is grounded through the grounding conductor of the power cord. To avoid electrical
shock, plug the power cord into a properly wired receptacle before operating the equipment.
Route power cords so they are not likely to be damaged.
Disconnect power before cleaning the equipment. Do not use aerosol cleaners, use a damp cloth.
Dangerous voltages may exist at several points in this equipment. To avoid injury, do not touch exposed
connections and components while power is on.
Do not wear rings or wristwatches when troubleshooting the equipment.
To avoid fire hazard, use only specified fuse(s) with the correct type number, voltage, and current ratings
as referenced on the equipment. Qualified service personnel should replace fuses.
To avoid explosion, do not operate this equipment in an explosive atmosphere.
Qualified service personnel should perform safety checks periodically and after any service.
If the equipment is modified, appropriate inspection and testing must be conducted to ensure continued
safe use of equipment.
Keep the back of the unit away from the patient vicinity (which is commonly defined as the space within
1.8m/6 feet of the patient/operating table), or otherwise be generally inaccessible to the patient.
Do not operate unit without Input Filter (ICM-000-0014; ICM-000-0434; ICM-000-0307).
Do not operate machine without Output Filter (ICM-000-0025). Turn OFF the unit when replacing the
filter. Replace the Output Filter as soon as odors become noticeable, or every three months, whichever
occurs first.
GENERAL CAUTIONS
A caution indicates a possible hazard to equipment that could result in equipment damage. Observe the
following cautions when operating or working on this equipment.
When installing this equipment, do not attach the power cord to building surfaces.
To prevent damage to equipment when replacing fuses, locate and correct the problem that caused the
fuse to blow before re-applying power.
Use only specified replacement parts.
Follow precautions for static sensitive devices when handling this equipment.
This product should only be powered as described in the manual. To prevent equipment damage, select
the proper voltage outlet.
To prevent damage to the equipment, read the instructions in the equipment manual for proper input
voltage.
Keep unit at operating environment for at least 6(six) hours before use, if unit was exposed to extreme
shipping and storage conditions.
Make sure the unit is in a safe and stable environment as to prevent falling or being dropped, which may
cause damage.
v
Page 7
Crystal Vision® Model 450D
SPECIFICATIONS
INTENDED USE:
The CRYSTAL VISION® 450D is intended to remove smoke created in any surgical
procedure. It is especially useful during laparoscopy and thoracoscopy procedures. The
Model 450D can be used to remove smoke produced by lasers, electrosurgical devices,
argon beam coagulators, LEEP devices, and other devices that create smoke during
surgical procedures.
During internal surgical procedures such as laparoscopy, it helps to maintain the desired
internal pressure (pneumoperitoneum). The smoke evacuator removes up to 20 liters-perminute of gases, vapors, or smoke that is produced, or introduced, by devices such as
insufflators, lasers, electrosurgical generators, argon beam coagulators and other devices.
The Model 450D automatically activates when active (smoke producing) devices that are
coupled to the Model 450D with special sensors are turned on. The Model 450D
automatically turns off, at a time predetermined by the operator, after the active device
turns off. The Model 450D also automatically activates when the high pressure limit is
exceeded in the pneumoperitoneum and it remains running to remove smoke, vapors, and
gases until the internal pressure returns to levels below the preset maximum.
The Model 450D can also be used to evacuate CO2 gas from the pneumoperitoneum at the
end of laparoscopic procedures.
PRODUCT DISPOSITION
At the end of service life, dispose of product in accordance with your institutional protocol
for capital equipment. I.C. Medical, Inc. has defined the service life for the Crystal Vision®
as 20 years from date of manufacture.
SIZE:
7.2"H x 14.07"W x 15.05"D (18.28cm H x 35.73cm W x 38.22cm D). Allow an additional
1.0" (2.5 cm) on both sides and 6.0" (15.2 cm) behind the device for the Charcoal Output
Filter and adequate cooling.
WEIGHT:
Approximately 17 pounds (7.7 kg).
SHIPPING/STORAGE
ENVIRONMENT
An ambient temperature range of -40ºC to +70ºC;
A relative humidity range of 10% to 100%, including condensation;
An atmospheric pressure range of 500 hPa to 1060 hPa.
OPERATING ENVIRONMENT
10o - 25o C, 30-75%RH, 700-1060hPa.
POWER REQUIREMENTS:
100-240 VAC, single phase, and 4.0 A, 47-63 Hz
LEAKAGE CURRENT:
<100 μamps, Conforms to IEC 60601-1
FUSE RATING:
F4AH 250V.
FLOW RATE:
OPEN RANGE:
LAP RANGE:
ACCURACY
Minimum: N/A Maximum: At least 90 Liters/Minute.
Minimum: 4 (±1) lpm Maximum: (At least 18 Liters/Minute)
±10%
MAXIMUM VACUUM:
Will not be more negative than -350mmHg.
MANUAL START SWITCH:
YES
1
Page 8
INDICATORS:
POWER ON
OPEN FLOW ON
LAP FLOW ON
OCCLUSION
CHANGE FILTER
NO PATIENT
OVER PRESSURE
OPEN FLOW RATE
LAP FLOW RATE
OPEN FLOW SET POINT
LAP FLOW SET POINT
TIME SETTING
FLOW READING
Visual Indicator
Visual Indicator
Visual Indicator
Visual & Audio Indicators
Visual Indicator
Visual Indicator
Visual & Audio Indicators
LED Meter
LED Meter
LED Display
LED Display
LED Meter
LED Meter
ULPA INPUT FILTER: WITH
WATER TRAP:
Multiple Use: Change when CHANGE FILTER illuminates on front panel; replace cap on
input connector when ULPA Input Filter with water trap not in use.
0.1 micron, ULPA rated filter
99.9975% Efficient
CHARCOAL OUTPUT FILTER:
RE-USABLE, REPLACE WHEN NOTICEABLE ODOR IS DETECTED, OR
EVERY THREE MONTHS, WHICHEVER OCCURS FIRST.
TURN OFF THE UNIT WHEN REPLACING THE FILTER.
INITIAL REMOVAL EFFICIENCY OF 99.99% FOR THE FOLLOWING:
OIL MIST, AEROSOLS, VAPORS
HYDROCARBON VAPORS
FAT VAPORS
MOISTURE DROPLETS
INITIAL REMOVAL EFFICIENCY TO 0.5 micron FOR THE FOLLOWING:
DIRT
ATMOSPHERIC DUST
FINE PARTICLES
RUST
SMOKE PARTICLES
BACTERIA
THE FOLLOWING ARE REMOVED UNTIL ODORS ARE NOTICEABLE:
ODORS
ORGANIC SOLVENT VAPORS
TRACE ORGANIC GASES
TOXIC VAPORS
2
Page 9
INSTALLATION/OPERATIONS
INSTRUCTIONS
WARNINGS:
IMPORTANT NOTICE:
Use only under the direction of a licensed
physician.
Do not exceed 27-mmHg intra-abdominal
pressure.
Do not use OPEN MODE in Laparoscopic
Procedure
Do not re-use, disposable Sterile Tubing Sets,
PenEvac1®, and Disposable ESU Shrouds
that are SINGLE USE ONLY.
During
laparoscopic
procedures,
the
®
CRYSTAL VISION is designed to remove
automatically smoke plume and water vapor
from the peritoneal cavity while maintaining
the pneumoperitoneal pressure that the
surgeon has selected on the insufflator.
Therefore, the volume of smoke that can be
removed by the CRYSTAL VISION® is directly
dependent on the flow rate of the insufflator.
Insufflators that cannot deliver a flow of at
least 6 Liters Per Minute will not permit results
that are satisfactory for most surgeons during
typical laparoscopic cases. They may be
suitable for procedures that generate small
amounts of plume.
Insufflators that deliver flow rates of 6 Liters
Per Minute will provide results that are
satisfactory for most surgeons during a large
majority of laparoscopic procedures. However,
there may be a few occasions when very large
quantities of smoke are generated over long
periods and results will be marginal, or even
unsatisfactory.
Insufflators that deliver flow rates ABOVE 6
Liters Per Minute are recommended to be used
by surgeons who frequently perform
procedures that generate very large amounts of
plume over extended amounts of time.
The CRYSTAL VISION® Model 450D is intended to
remove smoke created in any surgical procedure. It is
especially useful during laparoscopy and thoracoscopy
procedures. The Model 450D can be used to remove
smoke produced by lasers, electrosurgical devices, argon
beam coagulators, LEEP devices, and other devices that
create smoke during surgical procedures.
During internal surgical procedures such as laparoscopy, it
helps to maintain the desired internal pressure
(pneumoperitoneum). The smoke evacuator removes at
least 18 liters-per-minute of gases, vapors, or smoke that is
produced, or introduced, by devices such as insufflators,
lasers, electrosurgical generators, argon beam coagulators
and other devices.
The Model 450D automatically activates when active
(smoke producing) devices that are coupled to the Model
450D with special sensors are turned on. The Model 450D
automatically turns off, at a time predetermined by the
operator, after the active device turns off.
The Model 450D also automatically activates when the
high pressure limit is exceeded in the pneumoperitoneum
and it remains running to remove smoke, vapors, and gases
until the internal pressure returns to levels below the preset
maximum.
The Model 450D can also be used to evacuate CO2 gas
from the pneumoperitoneum at the end of laparoscopic
procedures.
3
Page 10
The following items will be needed and are available for
use with your CRYSTAL VISION®:
ESU Sensor Assembly ICM-000-0009
Laser Sensor Assembly with Cable (ICM-000-0018).
Hospital Grade Power Cord.
Charcoal Output Filter ICM-000-0025.
Ulpa Filter w/water trap ICM-000-0014 or ICM-000-0434.
Installation and Operating Instructions.
Be sure that you have inspected them for any sign of
damage.
FILTERS and POWER CORD
INSTALLATION:
1.
2.
3.
4.
Attach the Charcoal Output Filter to the connector on
the back of the CRYSTAL VISION®.
Attach the Ulpa filter w/water trap to the connector on
the front of the CRYSTAL VISION®..
Attach the power cord to the CRYSTAL VISION®.
Refer to Figure 1.
Figure 1: Filters and Power Cord Installation
ESU SENSOR INSTALLATION:
If you are going to use your CRYSTAL VISION® with an
electrosurgical unit (ESU), proceed with the following:
1.
Plug the ESU sensor connector into the ESU
connector on the CRYSTAL VISION® front panel.
(The Laser connector may also be used.)
2.
Place the ESU sensor on the sides of ESU Generator
as showed on Figure 2.
3.
The CRYSTAL VISION® will now turn ON any time
that the accessory is activated.
Figure 2: ESU sensor installation when used
with an electrosurgical unit (ESU)
4
Page 11
LASER SENSOR INSTALLATION
If you are going to use your CRYSTAL VISION® with
a laser, proceed with the following:
1.
Attach the Sensor Cable (Figure 3) to the LASER
connector on the CRYSTAL VISION® and to the
Sensor Connector Box.
2.
The Sensor Assembly has three parts: the Sensor
Connector Box, the Transmit Sensor, and the
Receive Sensor. Both Sensors have double-backed
tape on one side and an infrared lens on the side
opposite the tape. The Receive Sensor has a red
indicator lamp that will light when the sensor
assembly is plugged into the operating CRYSTAL
VISION®.
(NOTE: When the red light goes off, CRYSTAL
VISION® will start to operate).
Test the Sensor Assembly for proper operation:
Figure 3: Laser Sensor Installation
Plug the Sensor Cable into the Sensor
Connector Box and the LASER connector of
the CRYSTAL VISION®.
Plug the CRYSTAL VISION® into an AC outlet
and turn on the Power Switch on the Control Box
back panel and front panel.
Align the Transmit Sensor Lens and the
Receive Sensor Lens until the red light goes off
and the CRYSTAL VISION® starts.
Move the Sensor until the red light comes on.
The pump in the CRYSTAL VISION® will
stop operating sometime within 35 seconds
after the light goes out. (NOTE: the actual
amount of time that is required for the pump to
cease operation is determined by the TIME
adjustment on the CRYSTAL VISION® front
panel).
3.
Place the foot switch for the laser on an easily
accessible work surface (Figures 4 & 5).
4.
Inside the foot switch housing (Figure 4), position
the Transmit Sensor and the Receive Sensor on
opposite sides of the foot switch. Do not position
the sensors on the sides of the foot switch.
Figure 4: Laser Sensor Installation to the SIDE
of the laser foot switch HOUSING.
WITHOUT REMOVING THE PROTECTIVE
COVERING ON THE TAPE, position them so
that the red light comes on.
5
Page 12
5.
Keep the Sensors in the same position and press the
Laser Foot Pedal down. The red light should go out.
If it does not, reposition the Sensors until the red light
goes out when the Laser Foot Pedal is depressed and
it remains on when the pedal is not depressed.
THE LIGHT SHOULD GO OUT FOR THE
SLIGHTEST MOVEMENT OF THE FOOT
PEDAL. If it does not, move both sensors higher up
the wall of the protective housing of the laser pedal.
Carefully mark the location of both sensors.
HOUSING
FOOT PEDAL
6.
Remove the protective backing from one Sensor and
place it in the correct position on the side of the Laser
Foot Switch Assembly. (NOTE: It is usually very
helpful to only LIGHTLY position the sensors at first
and only after you are certain that they are in the exact
position press them firmly into place.)
BEAM
BEAM
WARNING- Repositioning either
sensor after it has been firmly set in
place can easily destroy it.
7.
Repeat the process for the other Sensor.
8.
Move the Sensor Connector Box to a convenient
location inside the Laser Foot Switch Assembly. Be
sure that the cables from the Sensors to the Sensor
Connector Box do not interfere with the operation of
the foot switch or with the surgeon's foot. Carefully
mark this location
9.
Remove the protective covering from the tape on the
Sensor Connector Box and attach it to the previously
marked location inside the Laser Foot Switch
Assembly.
(a)
FOOT PEDAL
TRANSMITTER
OR
RECEIVER
(b)
Figure 5: Sensors should be positioned so that
ANY movement of the pedal interrupts the beam.
6
Page 13
CHECK PROPER OPERATION OF
THE CRYSTAL VISION®
1.
The CRYSTAL VISION®. power switch (Figure 6) is
located on the back panel, next to the power cord.
Place this in the "ON" [I] position. When unit is
turned ON, the FLOW SET, indicators and display for
TIME and FLOW should illuminate.
2.
Adjust the TIME by pressing the push buttons (Figure
7) until the TIME display reads 2 SECONDS. The
pump should operate when the ESU Sensor or Laser
Foot Switch is activated and stops within
approximately 2 seconds after the foot switch is
released.
3.
The pump should operate when the MANUAL push
button (Figure 7) on front panel is depressed and stop
operating within approximately 2 seconds after the
button was released.
4.
Adjust TIME by pressing the push button arrow Up to
maximum, it should read 35 seconds. Press the
MANUAL button. The pump should start and then
stop approximately 35 (thirty five) seconds after the
button was released
5.
Figure 6: ON/OFF power switch description.
Adjust TIME by pressing the push button arrow
Down until the TIME display reads 2 seconds. Press
and release the MANUAL button. The pump should
start and then stop approximately 2 seconds after the
button was released.
NOTE:
The Up and Down arrow buttons adjust the
desired FLOW rate on the Flow Setting display.
The digital FLOW READING displays the
actual flow through the Crystal Vision®.
The OPEN (Orange) button and LAP (Green)
button changes the operating mode of the unit
The operating mode of Crystal Vision® is
indicated by the LED situated above OPEN and
LAP push buttons.
Figure 7: Manual and Time Control locations.
7
Page 14
6.
The FLOW SETTING buttons (Up and Down
arrows), (Figure 8), adjusts the desired Flow level on
the Flow Setting display. The Flow Reading digital
meter indicates the actual Flow through the unit. Some
RF noise may cause the Flow Set and Flow Reading
to flicker, however it will not affect the actual flow or
the functionality of the CRYSTAL VISION®.
7.
Push OPEN Button. Orange LED indicator should
light up. Adjust Flow setting to maximum (95 l/min)
by pushing Flow Setting Up arrow button. Press
Manual button, pump should start. Flow Reading
display should indicate at least 90 l/min.
8.
Push LAP button. Green LED indicator should lit up.
Set the Flow Setting to minimum (4 l/min) by pushing
Flow Setting arrow button down. Press Manual
button, pump should start. Flow reading display
should indicate 4 (±1) l/min. Repeat the process for
maximum flow. The Flow Reading display should
indicate at least 18 l/ min.
9.
Place a finger over the ULPA Input Filter with water
trap input and press the MANUAL button. The
CHANGE FILTER and OCCLUSION LEDs should
lit up. The OVER PRESSURE indicator requires
special test equipment and should be tested only by
trained qualified personnel.
Figure 8: Flow buttons.
10. For any problems, or if the CRYSTAL VISION® fails
to perform as indicated, contact I. C. Medical, Inc.
8
Page 15
DESCRIPTION OF SWITCHES,
CONTROL BUTTONS &
INDICATORS
The Sensor Assembly and the Charcoal Output Filter
should already be installed according to the
INSTALLATION INSTRUCTIONS.
1.
ON/OFF SWITCH is located on the back
panel (Figure 9) next to the power cord. This
switch controls the power to the CRYSTAL
VISION®. International symbols are used.
The [I] symbol indicates power "ON" and [O]
indicates "OFF." This switch also controls the
cooling fan.
2.
MANUAL button (Figure 10) used to turn the
CRYSTAL VISION® ON when the surgeon is
not activating a smoke-producing device. This
is particularly helpful when insufflators are
used that deliver 6 liters-per-minute or less.
The MANUAL button can be used to clear
residual plume during these situations in (LAP)
procedures. It can also be used to clear smoke
and plume if more than one device is used and
only one sensor is available. This frequently
occurs during laser cases when a sensor is
attached to the laser foot pedal and other
sensors are not available to be attached to an
Electrosurgical Unit (ESU).
3.
LASER CONNECTOR (Figure 10) is the
input for the LASER SENSOR CABLE. In
reality, both LASER & ESU connectors are
identical and either will accept all standard
I.C.Medical sensor assemblies.
4.
ESU CONNECTOR (Figure 10) is the input
for the ESU SENSOR CABLE. This sensor
can be used with a variety of devices to
activate the CRYSTAL VISION®.
Figure 9: ON/OFF power switch location.
Figure 10: Manual button and sensor connectors.
9
Page 16
5.
TIME SETTING Buttons (Up and Down
arrows) (Figure 11) varies the amount of time
that the CRYSTAL VISION® continues to
draw smoke, vapor, and gases from the
surgical site. Many insufflators deliver CO2 at
rates less than 6 liters-per-minute. These low
flows into the pneumoperitoneum limit the
amount of gas that can be evacuated from the
pneumoperitoneum without causing the
abdomen to deflate. In cases such as this, it is
very useful to limit the flow rate to a value that
can be supplied by the insufflator. To extend
the amount of time that the CRYSTAL
VISION® operates after the foot pedal (or hand
switch) is released. This allows for a greater
volume of gas to be eliminated from the
abdomen and therefore reduces the amount of
residual smoke that is left in the abdomen. The
ideal situation is to increase the amount of flow
into the abdomen. See the highlighted box
titled "FOR BEST RESULTS" in the
LAPAROSCOPIC OPERATION section that
follows.
Figure 11: Adjustment for time setting.
TIME DISPLAY (METER) (Figure 11)
indicates the amount of time that the
CRYSTAL VISION® will operate after the
foot switch, or hand switch, is deactivated (2
sec min- 35 sec max)
6.
OPEN and LAP buttons (Figure 12) selects the
flow range of the vacuum pump. The operator,
according to the type of procedure selects the
range. LAP MODE SHOULD BE
SELECTED FOR ALL LAPAROSCOPIC
PROCEDURES and the readouts are
GREEN. OPEN MODE should be used during
open and external cases when plume is being
eliminated through a handpiece. Using OPEN
MODE during laparoscopy will quickly deflate
the pneumoperitoneum. The OPEN and LAP
buttons also select the corresponding FLOW
SETTING.
7.
Only one of the LED mode indicators is
illuminated at a time. This eliminates
confusion of which mode is used.
Figure 12: Setting the flow RANGE.
10
Page 17
8.
FLOW SETTING control buttons (Figure 13)
sets the maximum flow that is desired by the
surgeon. This value is indicated on the FLOW
SETTING display.
9.
FLOW READING (Figure 13) is registering
the amount of gas and vapor that is actually
flowing at the present time. This value should
be zero, when the vacuum pump is not turned
on.
10.
CHANGE FILTER indicator is located above
filter (Fig. 14). It illuminates when there is a
reduced flow into the CRYSTAL VISION®.
The ULPA Input Filter with water trap should
be changed when this light first illuminates.
Do not attempt to clean, or re-use the ULPA
Input Filter with water trap. Dispose of the
ULPA Input Filter with water trap according
to your institution's biological waste protocol.
This indicator will also come ON when there
is a total occlusion. If the OCCLUSION
indicator is also lit, be sure to clear the
obstruction that caused it first and then check
the CHANGE FILTER indicator. If it is still
illuminated and the OCCLUSION indicator is
not, then the ULPA Input Filter with water
trap needs to be changed.
11.
Figure 13: Selecting the desired flow.
OCCLUSION indicates that flow into the
CRYSTAL VISION® has stopped (Figure 14).
The CHANGE FILTER light will also come
ON at this time. The operator should check for
kinked tubing, stopcocks that are turned OFF,
clogged insufflator output filters, or a
completely clogged ULPA Input Filter with
water trap. In order to prevent pump failure,
the CRYSTAL VISION® pump will not
activate when this lamp is illuminated.
Figure 14: Front Panel Indicators.
11
Page 18
12.
OVER PRESSURE indicates pressure
exceeding 27 mmHg in the pneumoperitoneum
(Figure 14). At pressures over 30 mmHg, the
vacuum pump starts in order to reduce the
pressure in the pneumoperitoneum. This
situation can be caused by many factors.
Frequently, the surgeon presses on the
abdomen and this causes intra-abdominal
pressure to rise. Correct the cause before
continuing.
13.
The "NO PATIENT" indicator (Figure 14)
turns on when the CRYSTAL VISION® is in
the LAP MODE, when zero pressure, or
atmospheric pressure, is present at the ULPA
Input Filter with water trap. The tubing set
may not be connected to the Trocar sheath or
to the ULPA Input Filter with water trap; the
valve on the Trocar sheath may be turned off;
or the ULPA Input Filter with water trap may
not be attached to the input filter connector.
14.
INPUT FILTER CONNECTOR (Figure 15)
holds the ULPA Input Filter with water trap.
The metal ring is pushed down to allow the
ULPA Input Filter with water trap to be
released. Replace cap on end of ULPA Input
Filter with water trap between uses. Change
ULPA Input Filter with water trap when
CHANGE FILTER indicator illuminates.
Charcoal
OUTPUT
FILTER
CONNECTOR (connector not shown)
(Figure 15). The Charcoal OUTPUT FILTER
CONNECTOR is located on the back panel of
the CRYSTAL VISION®. The Charcoal
Output filter- is reusable for several cases. The
Charcoal Output filter has a functional life of
three months. The Charcoal Output filter
provides odor control and reduces vacuum
pump noise.
15.
Figure 15: Input Filter connector. Charcoal
Output Filter
Do not operate the unit without a
Charcoal Output filter.
12
Page 19
LAPAROSCOPIC OPERATION
6.
Plug CRYSTAL VISION® into power outlet.
FOR BEST RESULTS:
7.
Use with insufflators that deliver more than 6
liter per minute.
Insufflators that deliver only 6 liters per
minute will provide satisfactory results
during most procedures.
Insufflators that provide less than 6 liters per
minute will not provide satisfactory results
in most cases.
8.
Insufflator should already be connected to the port
on the laparoscope. IT SHOULD NEVER BE
CONNECTED TO THE TROCAR SHEATH.
Insufflator should be set for a flow of AT LEAST 6
LITERS-PER-MINUTE as directed by the surgeon.
Connect Sterile Intra-Abdominal Plume Eliminator
Tubing (ICM-000-0003)* to Trocar sheath and
ULPA Input Filter with water trap assembly.
NEVER connect the CRYSTAL VISION® to the
laparoscope port.
*ICM-000-0448 Non-sterile Intra-Abdominal
Plume Eliminator Tubing Set is available.
1.
Installation should already be completed according
to the installation instructions.
2.
Be familiar with all operating controls as described
in description of switches, controls, and indicators.
3.
Refer to Figure 16 for set-up instructions.
10. Select LAP Mode on CRYSTAL VISION® by
pushing LAP button.
4.
Connect ULPA Input Filter with water trap.
11. Adjust TIME SETTING to maximum.
5.
Connect sensor LASER SENSOR and/or the ESU
SENSOR, as desired, to the CRYSTAL VISION®
front panel.
9.
Turn ON POWER SWITCH
CRYSTAL VISION®.
Figure 16: Laparoscopy set up when using the CRYSTAL VISION®.
13
on back of
Page 20
15.
OVER PRESSURE occurs when a pressure
more than 27 mmHg (±10% and 1 digit) is detected by
the CRYSTAL VISION. When this happens, the
OVER PRESSURE lamp and audio indicators turn ON.
If the pressure exceeds 30 mmHg (±10% and 1 digit)
the CRYSTAL VISION pump starts and attempts to
reduce the over pressure situation. DO NOT EXCEED
30 mmHg PRESSURE!
Set the FLOW SETTING for a flow slightly LESS than
the flow that the insufflator can provide.
Push the MANUAL button.
Observe the effect of a complete cycle by noting the
pneumoperitoneal
pressure
on
insufflator
gauge/indicator before the CRYSTAL VISION® starts
to operate and immediately after the pump stops. The
intra- abdominal pressure should not drop more than 1-2
mmHg. If it does, reduce the flow on the CRYSTAL
VISION® FLOW SETTING and repeat the process
until the pressure drops very slightly. It is important to
set the CRYSTAL VISION® flow as high as possible
without diminishing the pneumoperitoneal pressure.
16.
NO PRESSURE indicator illuminates when
the Trocar sheath is turned off, the tubing set is not
connected to the Trocar sheath, or the ULPA Input
Filter with water trap is not connected to the input filter
connector.
IF SMOKE PERSISTS WHEN
THE PUMP STOPS BE SURE
THAT
NOTE:
The FLOW SETTING indicates the
desired maximum Flow. The FLOW
READING display indicates the actual
Flow occurring at that moment. The
meter will read zero when the pump is
off. The meters are ORANGE in OPEN
Mode and GREEN in LAP Mode.
Leur lock on the cannula is open,
12.
The MANUAL button can be used to start the
vacuum pump if smoke still exists when the TIME
SETTING is set for the maximum amount of time. Be
sure to monitor the pneumoperitoneal pressure so that it
does not fall more than a few mmHg (adjust the FLOW
SETTING, if necessary).
Increase the TIME SETTING if there are no
leaks.
Tubing is not crimped,
ULPA Input Filter with water trap is clean,
may need to be replaced if it is clogged,
Inspect for leaks at the Trocar sheaths, tubing,
or instruments,
If TIME SETTING is set to maximum time,
use the MANUAL button until the abdominal
cavity is smoke free. Do not use it excessively,
or the pneumoperitoneum will collapse.
OPERATIONAL HINT:
PROCEDURES WITH LARGE
QUANTITIES OF FLUIDS
The TIME SETTING can be used to
conserve insufflator gas by reducing
the amount of time that the Crystal
Vision pumps after lasing stops. This
process requires more attention from
the circulating nurse, or the laser
operator.
Suctioning large quantities of fluids will shorten the life
of the ULPA Input Filter with water trap. To extend the
ULPA Input Filter with water trap life, use a suction
canister and non-sterile suction tubing set. Connect the
sterile ESU shroud to the canister and then connect the
non-sterile suction tubing set between the canister and
the ULPA Input Filter with water trap. The canister and
non-sterile suction tubing sets are available from I. C.
Medical.
13.
CHANGE FILTER illuminates when the
ULPA Input Filter with water trap needs to be changed.
14.
OCCLUSION indicator lights when flow
virtually ceases from the pneumoperitoneum. Check for
kinked tubing, closed Leur lock on the Trocar sheath,
severely clogged ULPA Input Filter with water trap,
heavy fluid accumulation in the water trap, or
obstructions in the Trocar sheath. The CHANGE
FILTER indicator always lights when the OCCLUSION
indicator comes on.
.
14
Page 21
NON-LAPAROSCOPIC
GYNECOLOGY PROCEDURES:
1.
Installation should already be completed according
to the installation instructions.
2.
Be familiar with all operating controls as described
in description of switches, controls, and indicators.
3.
Refer to Figure 17 for set-up .
4.
Connect ULPA Input Filter with water trap to the
CRYSTAL VISION®.
5.
Connect Disposable Sterile Tubing Set to laser
speculum port and to the ULPA Input Filter with
water trap.
6.
Connect ESU and/or LASER SENSOR CABLE to
SENSOR ASSEMBLY and the CRYSTAL
VISION®.
Figure 17: Set up for gynecology procedure for
either ESU or CO2 laser delivered through a
colposcope or Micromanipulator
®
7.
Plug CRYSTAL VISION into power outlet and
turn on POWER SWITCH on back panel.
8.
Push OPEN (Orange) button.
9.
Adjust FLOW SETTING for the desired flow rate
indicated on the display.
10. Push MANUAL button and observe the actual flow
indicated by FLOW READING display. Adjust
FLOW SETTING if necessary.
11. Adjust TIME SETTING for the desired amount of
time that the CRYSTAL VISION® runs after the
smoke producing equipment is shut off.
12. Increase the FLOW SETTING and/or the TIME
SETTING if smoke is not eliminated from the
uterus.
13. CHANGE FILTER light may indicate a partially
obstructed speculum port or smoke tubing. Check
and clean them, if necessary. The ULPA Input
Filter with Water Trap may also need to be
changed.
14. OCCLUSION alarm indicates obstructed speculum
port, kinked, or obstructed tubing. Check and clean
as necessary. It may also mean that the ULPA
Input Filter with Water Trap has become
extremely filled with smoke particles. Replace it,
if required.
15
Page 22
OTHER NON-LAPAROSCOPIC
PROCEDURES:
1.
Installation should already be completed according
to the installation instructions.
2.
Be familiar with all operating controls as described
in description of switches, controls, and indicators.
3.
Refer to Figure 18 or 19 for set-up.
4.
Connect ULPA Input Filter with Water Trap to
CRYSTAL VISION®.
5.
Connect Laser Hand piece Accessory or ESU
Handpiece Accessory to the ULPA Input Filter with
Water Trap.
6.
Connect ESU sensor to ESU port of the CRYSTAL
VISION® or Laser Sensor Cable to Sensor
Assembly and the CRYSTAL VISION®.
7.
Plug CRYSTAL VISION® into power outlet.
8.
Turn ON POWER switch on back panel.
9.
Push OPEN (Orange) button. Indicator should
illuminate.
Figure 18: Set up for using CRYSTAL VISION® with
an ESU system during an open procedure.
10. Adjust FLOW SETTING for the desired flow.
11. Push MANUAL button and observe that desired
flow is shown on the digital FLOW meter display
12. Adjust TIME SETTING for the desired amount of
time that the CRYSTAL VISION® runs after the
smoke producing equipment is deactivated.
IF SMOKE PERSISTS DURING THE
SURGICAL PROCEDURE, TRY THE
FOLLOWING.
13. Increase the FLOW SWTTING and/or the TIME
SETTING if smoke is not eliminated from the
surgical site.
14. CHANGE FILTER light may indicate a partially
obstructed speculum port or tubing. Check and
clean them, if necessary. The ULPA Input Filter
with Water Trap may also need to be changed.
Figure 19: Set-up for using a CRYSTAL VISION® with
a laser hand piece.
16
Page 23
15. OCCLUSION alarm indicates an obstructed smoke
collection nozzle, kinked or obstructed smoke
tubing. Check and clean as necessary. It may also
mean that the ULPA Input Filter with Water
Trap has become extremely filled with smoke
particles. Replace it, if required.
OTHER ACCESSORIES
IF SMOKE PERSISTS WHEN THE
PUMP STOPS BE SURE THAT:
Tubing is not crimped
ULPA Input Filter with Water Trap is clean,
may need to be replaced if it is clogged
Be sure there are no leaks in the tubing, or
instruments:
Increase the TIME SETTING if there are no
leaks.
Part number
Description
ICM-000-0021
Speculum Tubing, Non-Sterile
ICM-000-0027
Speculum Tubing, Sterile
ICM-000-0075
Smoke Evacuator Wand, Sterile
ICM-000-0098
Smoke Evacuator Tube, Sterile
ICM-000-0274
2’Tube with 10mm Swivel, NonSterile
ICM-000-0347
15 mm Smoke Tube, Sterile
Other options are available. Please refer to
http://www.icmedical.com for further information.
PROCEDURES WITH LARGE
QUANTITIES OF FLUIDS:
Suctioning large quantities of fluids will shorten the life
of the ULPA Input Filter with Water Trap. To extend
the ULPA Input Filter with Water Trap life, use a
suction canister and non-sterile suction tubing set.
Connect the sterile ESU shroud to the canister and then
connect the non-sterile suction tubing set between the
canister and the ULPA Input Filter with Water Trap.
The canister (ICM-000-0082) is available from I. C.
Medical.
PENEVAC1® ACCESSORIES:
PenEvac1® accessories may also be used with the
CRYSTAL VISION® Model 450D. The PenEvac1®
combines the function of an ESU pencil and smoke
evacuator into a single hand-held device.
The
PenEvac1® also has a telescoping tip that enables the
surgeon to change the length of the electrode without
actually replacing the electrode. Several styles of
electrodes are available. The PenEvac1® is available as
a disposable single-use.
Description
Part number
ICM-000-0450
PenEvac1®, SS Blade, Tube &
Holster, Sterile
ICM-000-0452
PenEvac1®, Non-Stick Blade, Tube
& Holster, Sterile
ICM-000-0458
PenEvac1®, Megadyne E-Z Clean
Blade, Tube & Holster, Sterile
Other options are available. Please refer to
http://www.icmedical.com for further information.
17
Page 24
THEORY OF OPERATION
The CRYSTAL VISION® Model 450D was designed to
effectively remove smoke during surgical procedures
and to eliminate problems associated with other types of
smoke evacuators.
The CRYSTAL VISION®
automatically turns on when the surgeon activates a
smoke-producing device and it turns itself off when it is
not needed to remove surgical smoke. The Model 450D
has two flow ranges to provide optimal performance
under a variety of surgical conditions. In addition, the
evacuator notifies the operator of conditions that may
limit the effective removal of smoke.
PNEUMATIC CIRCUITS
Surgical Smoke collection devices are used to remove
smoke from the surgical site and deliver it through the
fluid trap and into the ULPA rated 0.1 micron input
filter. The filtered air then travels from the input filter,
through either the OPEN or LAP pneumatic circuits,
into the pump, then through the output charcoal 0.5
micron filter and finally delivered back into the
operating room.
2.
The Manual button also starts the pump if it is
pushed. This push button is used during
procedures to activate the pump if a remote sensor
is not used with all of the active smoke producing
devices. The switch can also be used when very
low flow insufflators are used during laparoscopic
procedures. It is helpful in this circumstance
because increasing the flow rate would deflate the
pneumoperitoneal pressure. Instead, the flow
TIME can be set to 35 seconds and if additional
purging is required, the MANUAL button will
continue the evacuation.
3.
An OVER PRESSURE condition: When the
evacuator is in the LAP Mode, the pressure in the
pneumoperitoneum is monitored.
When the
pressure reaches approximately 27 mmHg, a light
alert is activated. If the pressure continues to
build to 30 mmHg, the pump is started and it
continues to operate until the pressure drops
below 30 mmHg. The pressure is not monitored
in OPEN Mode because this range is not intended
to be used during laparoscopic procedures.
STOP PUMP:
ELECTRONIC CIRCUITS
When the smoke producing device is deactivated,
MANUAL button not pressed, or the OVER
PRESSURE condition is not present, the pump will
continue to operate for the amount time shown on the
TIME SETTING on the front panel.
START PUMP:
There are three ways to start the pump: trigger a
remote sensor, push the MANUAL button on the front
panel, or an OVER PRESSURE condition.
1. Remote Sensors: Remote sensors are attached to
the front panel at the ESU or LASER plug. Both
plugs are electrically identical and can be used
interchangeably with either sensor.
a) The LASER sensor consists of an infrared
transmitter and receiver that are carefully
positioned on a foot switch that activates a
surgical laser. Placement is critical because it
is very important to position the infrared
beam so that it is interrupted at the first
movement of the foot pedal. In this manner,
the evacuator is started before the laser
actually activates and the air stream is already
moving when the laser beam first strikes
tissue.
Whenever an OCCLUSION condition is sensed, the
pump will stop. After a brief period of time, if any
start condition exists, it will attempt to restart to see if
the occlusion has been cleared. If it has not, the pump
will again stop. This will continue until the occlusion
is cleared, or until TIME expires.
TIME SETTING:
The TIME SETTING buttons are used to adjust the
amount of time that the evacuator continues to pump
after the active smoke producing device (laser, ESU,
etc.) is deactivated. This lag in the deactivation of the
evacuator is necessary in order to remove any residual
smoke from the surgical site. The control is variable
from approximately 2 to 35 seconds.
b) The ESU sensor is placed on, or near the
input connector of the electrocautery pencil
cable and the other end is plugged into the
evacuator front panel. Activation of the ESU
pencil is detected by the sensor. The remote
sensor (ESU or Laser) sends a trigger,
through the front panel board, to the Master
Board, and then to the motor controller.
18
Page 25
CHANGE FILTER light is illuminated. The pump is
not stopped unless the restrictions are allowed to
continue (filter gets fuller) and an occlusion is sensed.
FLOW SETTING:
The FLOW SETTING control buttons vary the speed
of the vacuum pump depending on the flow range in
use. OPEN Mode flow is adjustable to at least 90 lpm
and LAP Mode flow from 4 + 1 to at least 18 lpm.
LAP MODE:
LAP flow range is used during laparoscopic (“closed”)
procedures, they are characterized by a limited amount
of gas (air, carbon dioxide, etc.) being available for
evacuation. LAP MODE is used as many insufflators
can only provide 4 - 6 lpm to the cavity. If the FLOW
SETTING is set to evacuate much higher rates, the
pneumoperitoneum will collapse.
Therefore, the
operator needs to balance the FLOW and TIME
SETTINGS to optimize the amount of gas removed at
any one time.
The FLOW SETTING is used only as a general
indicator of what the flow will be when the pump is
started. The purpose of the display is to allow the
operator to preset the desired flow when the pump is
not running. The FLOW READING is meant to
provide a more accurate measure of the flow actually
occurring. There may be relatively large differences
between the FLOW SETTING and the FLOW
READING.
The FLOW READING display provides a good
measure of the actual flow when the pump is running.
The FLOW READING displays 00.0 (Zero) when the
pump is not running. Measuring flow is a very
complicated process. The actual flow through the
pump will vary significantly depending on the
accessories, and the measuring devices used. Be sure
to follow the procedures listed in the Calibration
section explicitly when trying to calibrate, or measure,
the flow through the evacuator.
The LAP MODE is also monitored for Occlusion and
Change Filter conditions. These are monitored by the
same vacuum sensor and uses parallel circuits to those
in the OPEN MODE. For more information, refer to
the OPEN MODE description.
The LAP MODE is also monitored for two other
conditions. The CRYSTAL VISION®, in LAP flow
only, monitors the pressure in the flow circuit when the
pump is NOT running. The NO PATIENT circuit
looks for the presence of a positive pressure to indicate
that the evacuator is attached to the pneumoperitoneum
correctly and that there are no occlusions (typically
closed stopcocks) in the circuit. If the circuit is not
attached to the patient, or a stopcock is closed, or the
filter is completely clogged, the positive pressure in
the pneumoperitoneum cannot be sensed by the
CRYSTAL VISION®. Under these circumstances, the
NO PATIENT lamp comes on. Whenever this light
comes on, it is important to correct the problem
because the evacuator will not be able to monitor the
pressure in the pneumoperitoneum for OVER
PRESSURE conditions nor will it be able to evacuate
smoke when it is generated.
OPEN MODE:
The OPEN MODE is used with the PenEvac1®, ESU
Pencil Shroud™, or other types of surgical smoke
collection 10mm or greater tubing sets.
These
procedures are considered “open” in that the flow of
air and vapor to the surgical site is not restricted like it
is during a laparoscopic or other “closed” procedure.
During the open procedures, higher flow rates are very
desirable to remove effectively all smoke from the site
and to keep it from permeating the operating room.
When the OPEN button is pushed, the high flow
solenoid is activated to select the high flow air circuit.
The flow circuit is monitored by a vacuum sensor in
order to detect an occluded air stream. If an occlusion
is detected, the pump immediately stops to prevent
tissue damage. Within a few seconds, the pump will
attempt to restart. If the occlusion has been cleared,
the pump will run. If the occlusion is still present the
pump will not start and will again attempt a restart in a
few seconds. The restart attempts will continue until
the occlusion is cleared, or until TIME SETTING
expires.
The pneumoperitoneal pressure is monitored (if the
circuit is connected properly and the NO PATIENT
light is out) when the pump is not running. If the
pressure sensed is approximately 27 mmHg, the
OVER PRESSURE light alert is activated until the
pressure drops to below 27 mmHg. If the pressure
exceeds approximately 30 mmHg, then the pump will
start and run for the set time.
.
The flow resistance is also monitored for changes in
restrictions in the air stream that do not represent an
occlusion.
When the set point is exceeded, sensing that the ULPA
filter is no longer effectively capturing particles, the
19