Inditherm Patient Warming Systems Operating Instructions Rev. 2.8 Dec 2007
Operating Instructions
16 Pages

Preview
Page 1
Patient Warming Systems Operating Instructions
Contents: Page Introduction
1
Description of Parts and Range of Products
2
Symbols
3
Safety
4
System Description
5
Setting Up
6
Connecting & Disconnecting
7
Storage
7
Use of Clamps with Mattress
7
Operation of Control Unit (MECU1) Before Use Control Unit Start Up Cycle Control Descriptions Language Selection Temperature Selection
8 8 8 9 9 9
Use and Care of Mattresses and Blankets
10
Infection Control & Cleaning Instructions
11
Problem Solving & Warning Messages
12
Technical Specification
13
© Inditherm plc, 2007 Copyright: All rights reserved. No part of this handbook may be reproduced, stored in a retrieval system or transmitted in any form or by any means electronic, mechanical, photocopying, recording or otherwise, without the prior permission of Inditherm plc.
Inditherm Patient Warming Systems
Rev. 2.8 : 17 -12-07
Introduction:
The Inditherm Patient Warming Systems are designed for use in the operating room, recovery room, anaesthetic room, intensive care and emergency department. They provide safe and controlled warming to assist patients to maintain normal body temperature. The warming medium is available as: • A mattress for use under a patient • A recovery blanket or operating room blanket to be placed over a patient. In addition to warming, the mattresses also include a pressure-relieving pad. This is achieved using an integral viscoelastic foam pad under the flexible heating surface, ensuring no degradation of heating performance. Different sizes and models of mattress and blanket are available for various procedures and situations. They are sealed to prevent ingress of fluids and allow cleaning. Integral straps are provided to ensure mattresses can be securely fastened to the operating table or trolley. The system is powered and controlled by an electronic control unit. The mattresses and blankets are powered at low voltage, ensuring safety for patients and operators. The temperature is controlled automatically to user-selected level. An over-temperature safety cutout is integrated into each mattress and blanket. The control unit can be placed on a flat surface or mounted on an IV pole, anaesthetic machine, etc. It has a standard mains supply input and a working voltage of 24Vac to the mattress or blanket. The system has 4 pre-set operating temperatures of 37, 38, 39 and 40°C and is designed to be operated continuously, maintaining a uniform heat under or over the patient.
Inditherm Patient Warming Systems
-3-
Rev. 2.8 : 17-12-07
Description of Parts and Range of Products Note: Any mattress / blanket can be used with any control unit.
Model
Size
Description
OTM1
1900 x 585 x 40 mm
OTM1 N
1900 x 535 x 40 mm
OTM2
1200 x 585 x 40 mm
OTM2 N
1200 x 535 x 40 mm
GTM1
1070 x 585 x 40 mm
GTM1 N
1070 x 535 x 40 mm
PTM1
560 x 500 x 40 mm
Operating Room Paediatric Mattress
OTB1
500 x 1070 x 40 mm
Operating Room Blanket, Long
OTB2
870 x 500 x 40 mm
Operating Room Blanket, Short
RB1
1660 x 800 x 40 mm
Recovery Room Blanket, Full length
RB2
1660 x 1200 x 40 mm
Recovery Room Blanket, Wide
MECU1
160 x 240 x 230 mm
Electronic Control Unit
Operating Room Mattress, Full Length
Operating Room Mattress, ¾ Length
Operating Room Gynaecology Mattress
Inditherm Patient Warming Systems
-4-
Rev. 2.8 : 17-12-07
Symbols: Warning Read fully these operating instructions as they may affect safety or prevent damage
Type BF Equipment
Class II Equipment
Mains Switch I (On) Power connected to the mains supply O (Off) Power disconnected from mains supply
Alternating Current
Product complies with the provisions of the Medical Device Directive (93/42/EEC) 0473
Within European Union product must be disposed of in accordance with WEEE Directive
Enable
Scroll
Alarm Cancel / Reset Inditherm Patient Warming Systems
-5-
Rev. 2.8 : 17-12-07
!
Safety: Indications for Use Designed for use in the operating room, recovery room (PACU), anaesthetic room, intensive care, emergency department, on medical and surgical wards, and for patient transport, the Inditherm patient warming system provides safe and controlled warming to assist patients to maintain normal body temperature. In addition to providing warming, the mattress also includes a pressure relief pad.
Contraindications It is the responsibility of the user to determine whether warming is appropriate for each individual patient. The patient warming system should not be used on patients where clinical considerations indicate that warming of the patient is not advisable.
Warnings Do not use the warming system if there is any sign of damage. Ensure any damaged item is checked and/or repaired by a qualified engineer or technician. Do not use the system in the presence of flammable anaesthetic gases or oxygen enriched atmospheres as these may pose an explosion hazard.
Precautions Refer to section on Storage to prevent damage to the mattress or blanket. Ensure the mattresses and blankets are not folded or creased, and do not store any other product or item on top of the mattress. Refer to section on Setting Up to ensure patient safety. The mattress must be fitted to the operating room with the printed side towards the table and the plain surface uppermost. Refer to section on Use and Care of Mattresses and Blankets to prevent damage, ensure patient safety and obtain best performance. Refer to section on Use of Clamps to prevent damage to the mattress and ensure patient safety. Refer to section on Operation of Control Unit to prevent damage to the product and ensure safety of the user and other staff. Refer to sections on Infection Control and Cleaning to ensure safety. Pooling of most surgical prep solutions under the patient is specifically contraindicated, due to risk of skin irritation. Warming may exacerbate this, so ensure instructions for use of prep solution are followed. The Inditherm patient warming system is not sterile. Where necessary the user should take appropriate precautions to protect the sterile field. Caution: United States Federal law restricts this device to sale by or on the order of a physician
Inditherm Patient Warming Systems
-6-
Rev. 2.8 : 17-12-07
System Description Control Unit The control unit is a precision temperature control unit to be used in conjunction with the electrically heated blanket or mattress. Mains voltage is supplied to the control unit via a captive mains lead. A recess is provided to allow the mains cable to be safely coiled around the unit when not in use. There is a power switch on the right side of the unit that illuminates when the power is on. At the top is a carrying handle and to the rear is a clamp to allow the unit to be mounted on a pole or anaesthetic trolley. Warning Indicator
On the front of the unit is a display and control panel from which all the functions can be accessed, there are three control buttons, and an LCD display. The cable for the mattress or blanket is at the front. The control unit input is protected by a 1600mA fuse, which is accessible from the base of the unit. The control unit provides a 24Vac supply to the mattress or blanket and receives temperature information back. The temperature setting is indicated on the display. The unit automatically adjusts to the type of mattress or blanket connected.
LCD Screen
Enable Button
Scroll Button
Alarm Cancel / Reset Button
Mattress The mattress is available in various sizes and is only designed to operate under the patient. The mattress provides an even temperature over the whole surface and is extremely flexible. The basic construction of the mattresses remains the same throughout the range. The mattresses are water and solvent resistant and have been tested for bio-compatibility to allow skin contact during use. All seams are fully sealed by RF Welding.
Inditherm Patient Warming Systems
-7-
Rev. 2.8 : 17-12-07
The internal patented Inditherm carbon polymer material provides the heat source, and an internal temperature sensor provides the output to the control unit for temperature control. The mattress has a thermal protector that will not allow the mattress to overheat. A pressure relief pad is integrated into the mattress, underneath the flexible warming surface. This helps reduce the incidence of pressure sores without any attenuation of the warming performance. Integral straps are used to retain the mattress in-situ.
Blanket The blanket is designed to operate over the patient. There are two types of blanket available, both are constructed in the same manner as the mattresses, but without the pressure relief pad, which is replaced by a soft insulating material.
!
Setting Up Ensure the surface has no folds or creases and check that the surface is not damaged prior to securing the product to the operating table. Do not use if there is any sign of damage or wear. The mattress must be fitted with the printed side towards the table and the plain surface uppermost. The mattress can be fixed to the table via the straps and press-studs as shown. Straps should be secured on both sides. A number of pressstuds are provided on each strap; use those which allow most secure fastening of the mattress to the table or trolley. It is important to ensure that the mattress is firmly secured and that the straps cannot slide off the operating table brackets. Note that strap positions vary depending on size of mattress. Mattresses can be manufactured with customised strap positions and lengths if required. See section on Operation of Control Unit (MECU1) for further information on setting up.
Inditherm Patient Warming Systems
-8-
Rev. 2.8 : 17-12-07
!
Connecting & Disconnecting A connection cable is provided to take power and control signals between the mattress or blanket and the control unit. Note: Do not force the connector into the socket. Ensure it is correctly orientated. The black connector plugs into the connector fitted to the mattress or blanket. Align the connectors and push together. It should be possible to feel a “click” as the connectors engage. To disconnect pull the connectors apart. Note:
!
Do not attempt to rotate or unscrew the connectors. It is important to pull them apart straight, do not bend or twist them
Storage & Care To ensure the mattress is not damaged during storage, either keep flat or loosely roll the mattress with the printing on the inside. Do not fold the mattress. Do not store any other product or item on top of the mattress! Check the mattress or blanket for signs of damage before each use. Avoid storing Control Unit, mattresses or blankets in direct sunlight.
!
Use of Clamps with the Mattress The use of operating table clamps and similar devices may cause damage to the product and result in loss of the heating function and / or localised heat build-up in the damaged area. When using clamps, it is essential to avoid clamping the mattress. If clamping to the mattress is likely to be unavoidable, please contact Inditherm to discuss your special needs.
Inditherm Patient Warming Systems
-9-
Rev. 2.8 : 17-12-07
!
Operation of Control Unit (MECU 1) Before Use Check that the control unit is configured for the local supply voltage, as shown on the rating label on the underside of the control unit. Clamp the control unit to an IV or accessories pole or place in a suitable position on a flat surface. Connect the unit to a suitable mains supply and position the lead so it does not cause a hazard. Excess length of mains lead can be wrapped around the cable stowage channel on the control unit. Connect the black socket on the control cable to the plug attached to the mattress or blanket.
Control Unit Start-Up Cycle Turn on the power switch (green button) on the side of the unit and ensure the switch illuminates to indicate power on. The system will follow the cycle below: Display Screen Readout
Action Required
Information Bleep will be heard on switch on First screen indicating initialisation started and software version If required language is set, take no action and system will automatically continue Bleep will be heard
Inditherm V2.03E
None
English UK > to change language
Option to change Language See Language Selection
Language English UK
None
Screen Indicates Language
System Check
None
Bleep will be heard Automatic system test Bleep will be heard
System OK
None
Alarm Check
None
Alarm OK
None
∧ 37 °C ∧
> 37 °C <
∨ 37 °C ∨
Inditherm Patient Warming Systems
User indication that system is OK
User can change temperature at this point See Temperature Selection User can change temperature at this point See Temperature Selection User can change temperature at this point See Temperature Selection
- 10 -
Bleep will be heard Automatic test of alarm sound Bleep will be heard User indication that alarm ok Bleep will be heard Display showing temperature. Unit automatically selects last temperature used. Arrows point upwards, indicating system is warming up, until desired temperature is achieved. Display showing temperature. Arrows point horizontally, indicating desired temperature is achieved. Display showing temperature. Arrows point downwards, indicating system is cooling down, until set temperature has been achieved.
Rev. 2.8 : 17-12-07
Control Descriptions Warning Indicator
LCD Screen
Enable Button
Scroll Button
Alarm Cancel / Reset Button
Language Selection The language can only be changed at the appropriate point in the start-up sequence, see section on Control Unit Start-Up Cycle. The language can be set to one of the following: English UK, English US, French, German, Spanish, Italian When Language is displayed in the start-up sequence, press and hold down the Enable Button. While holding down the Enable button, press and release the Scroll button to set the required language, the unit will cycle sequentially through the available languages each time the Scroll button is pressed and released. When the required language is shown release the Scroll and Enable buttons. English US shows temperatures in Fahrenheit and all other languages show temperatures in Celsius.
Temperature Selection The temperature can be set at any time after the self-test start up cycle has completed, the following temperatures can be set: 37°C 38°C 39°C 40°C To adjust the temperature, press and hold down the Enable Button and then press and release the Scroll button to set the required temperature. The unit will cycle sequentially through the temperatures each time the Scroll button is pressed and released. When the required temperature is shown release the Scroll and Enable buttons. When the system is first switched on it will default back to previous temperature setting that was used.
Inditherm Patient Warming Systems
- 11 -
Rev. 2.8 : 17-12-07
!
Use and Care of Mattresses and Blankets Do not allow sharp objects to penetrate the mattress / blanket. Inspect the surface before and after each use. Mattresses and blankets should not be used if there is any sign of damage. Do not use the blanket / mattress with any but the supplied control unit and connecting leads. Gel pads are not required. Use of gel pads may cause a loss of warming performance, as the mattress would have to heat the gel pad first. A clinical study on healthy volunteers suggests that the Inditherm system may provide better pressure relief than gel pads or standard operating table mattresses1. The operator should consider whether it is necessary to restrain the patient under or on the blanket / mattress.
General Performance On initial start up, if the ambient temperature is low, the mattress / blanket will take longer to achieve the desired temperature. If the desired temperature is not achieved within 40 minutes the alarm will sound. Should this happen select the Alarm Cancel / Reset button and the unit will restart and will allow a further 40 minutes for the desired temperature to be achieved. The mattress will not cause any environmental heating and contains a special conductive heating material. The heat given off by this material will only be felt by the user / patient when applying pressure / weight to the mattress. It is normal that the mattress or blanket do not feel particularly warm to the touch when left uncovered. The control unit will monitor temperature performance, and when the mattress/blanket reaches the desired temperature the control unit will stop heating the mattress/blanket. If the maximum allowed temperature is exceeded the over temperature alarm will sound. An internal safety cut-out temperature management system will operate if any fault condition causes the mattress/blanket to exceed a temperature of 43°C. For best performance it is recommended that the system is switched on and left to run at set temperature in advance of placing the patient.
1
Baker EA, Leaper DJ: Pressure-relieving properties of an intra-operative warming device. J Wound Care, 2003, 12:4, p. 156-140.
Inditherm Patient Warming Systems
- 12 -
Rev. 2.8 : 17-12-07
!
Infection Control and Cleaning Instructions Infection Control Infection control and routine cleaning should be carried out in accordance with the local Infection Control Policy. Products should be decontaminated prior to servicing or return to the supplier. If returning product to Inditherm or local distributor, then a decontamination certificate must accompany the product and a copy of the certificate should be affixed securely to the outside of the packaging.
Cleaning Guidelines Ensure the control unit is disconnected from the mains electricity supply before cleaning. Clean mattress and control unit with care using a damp soapy cloth and / or disinfectant wipes. After cleaning control unit dry thoroughly before use. Where higher level of decontamination is required, use Sodium Hypochlorite (bleach) solution, or similar, (up to 10,000 ppm available Chlorine). Note that the mattress should be rinsed and dried after cleaning with bleach solution, in accordance with normal instructions for use.
All cleaning should be carried out in accordance with the instructions for use of the manufacturer of the cleaning agent. Do not immerse the mattress, blanket or control unit in fluids. Do not use high temperature autoclave or Phenolic based products for cleaning.
Inditherm Patient Warming Systems
- 13 -
Rev. 2.8 : 17-12-07
!
Problem Solving & Warning Messages The Inditherm system has a number of features integrated that check for correct performance. These will generate warning messages if appropriate, which will be shown on the display and sound the audible alarm. These are described below. Connect Mattress / Blanket: Indicates that either the mattress / blanket is disconnected, or possibly there is a fault with the connection cable or connector. Check that the connector is fully engaged. If this fails to clear the fault then press the Alarm Cancel button. If the error persists have the system checked by a qualified engineer. This error may also indicate a fault on the mattress / blanket internal temperature sensor. Mattress Fault: Normally indicates there is a fault within the mattress / blanket or if the thermal safety cut-out operates due to a fault. Press the Alarm Cancel button. If problem persists then have the system checked by a qualified engineer. This message is also displayed if the internal fuse on the control unit output is blown. High Temperature Detected: Indicates that the temperature has exceeded the maximum setting. In some cases this will not be experienced as the thermal safety cut-out will operate first. Low Temperature Detected: Indicates either that mattress has failed to reach temperature in the prescribed period after switching on or that the temperature of the mattress / blanket has fallen below the set value. During the warm-up period if the set temperature is not reached after 40 minutes then this warning will be displayed. If Reset is pressed then a further 40 minute warming up period will be initiated. If temperature is still low after the second period then the message will be displayed again. In this case the system should be checked by a qualified engineer.
If any other problems are encountered, then a qualified engineer should be consulted.
Inditherm Patient Warming Systems
- 14 -
Rev. 2.8 : 17-12-07
Technical Specification Mattress Construction:
Inditherm® flexible polymer heating sheet, with 18mm viscoelastic foam pressure relief pad under and 205g.m-2 expanded polyester comfort lining over. Encapsulated in latex-free, nylon fabric cover, with non-microporous polyurethane coating, fully sealed with RF welded seams. In-built temperature sensor and over-temperature thermal cut-out. Connection cable, 20 cm long, with strain relief, fully sealed entry grommet and IP61 rated waterproof connector. Sensors and cables let into pressure relief pad for patient comfort.
Temperature Range:
37°C to 40°C (99°F to 104°F) in steps of 1°C (2°F) Over-temperature safety cut-out at 43°C (109°F)
Power:
Control Unit:
230 Vac or 110Vac or 100Vac (±6%), 50Hz/60Hz 75 W
Mattresses:
24 Vac (nom.), 50Hz 25 W to 65 W, depending on size
Dimensions:
Control Unit:
160 x 240 x 230 mm
Mattresses:
Part Standard Narrow OTM1 1900 x 585 mm 1900 x 535 mm OTM2 1200 x 585 mm 1200 x 535 mm GTM1 1070 x 585 mm 1070 x 535 mm PTM1 560 x 500 mm OTB1 1070 x 500 mm OTB2 870 x 500 mm RB1 1660 x 800 mm RB2 1660 x 1200 mm Other dimensions available on request
Cable Length:
4m
Compliance:
Type Mattress Mattress Mattress Mattress Blanket Blanket Blanket Blanket
EN60601-1, Class IIa, Type BF EN60601-1-2 EN60601-2-35 93/42/EEC, EEC Medical Devices Directive 73/23/EEC, EEC Low Voltage Devices Directive
Environmental: Ambient Temperature (Operating):
15°C to 40°C (59°F to 104°F)
Ambient Temperature (Storage):
-10°C to 55°C (14°F to 131°F)
Relative Humidity:
30% to 70%
Inditherm Patient Warming Systems
- 15 -
Rev. 2.8 : 17-12-07