catalog_2120_air_oxygen_blender_instruction_manual_.pdf
Page 1
BIO-MED DEVICES
AIR/OXYGEN
BLENDER
INSTRUCTION MANUAL
CATALOG # 2120 REV 100605
Page 2
TABLE OF CONTENTS
ADDENDUM 1- Low Flow Selector Knob ........................................................................ 1
ADDENDUM 2- NEO2 BLEND Blender ...........................................................................
1
1. WARNINGS, CAUTIONS AND NOTES .....................................................................
2
2. EXPLANATION OF SYMBOLS ...................................................................................
3
3. SPECIFICATIONS ......................................................................................................
3
4. INTRODUCTION AND OPERATION .........................................................................
4
5. SETTING UP THE BLENDER ................................................................................... 5
6. TESTING THE BLENDER ......................................................................................... 6
Page 3
7. USING THE BLENDER .............................................................................................. 6
8. TROUBLE SHOOTING GUIDE .................................................................................. 7
9. CLEANING INSTRUCTIONS ......................................................................................
8
10. BLENDER WARRANTY ...........................................................................................
8
11. SERVICE RECOMMENDATIONS ............................................................................
8
APPENDIX ......................................................................................................................
9
ADDENDUM 1- Low Flow Selector Knob
A low flow selector knob has been installed on some of the Bio-Med Devices Blenders
(standard on units mounted on the Crossvent ventilators) in lieu of the auxiliary side port.
This enables the user to maintain accurate concentrations using the single primary output
for both high and low flows with a simple turn of the knob.
A label attached to the side of the blender indicates how to position the knob for accurate
concentrations at settings less than or greater than the flow rate indicated. The knob
must be pushed in prior to turning.
ADDENDUM 2- NEO2 BLEND Blender
The NEO2 BLEND is a low flow blender with two flowmeters mounted to it, one on each
side. Any time this blender is used below 3 lpm, an internal bleed is required to maintain
Page 4
the accuracy of the O2 concentration being delivered. Due to the unique design of the
NEO2 BLEND, this bleed is automatically activated when the flowmeter on the right is set
to the “ON” position as described below. To conserve gas, this bleed may be turned off
when this flowmeter is not in use. The flowmeter on the right side is for delivering flows
below 3 lpm. The left flowmeter is generally used for flows above 3 lpm, but may be used
below 3 lpm as explained below.
The flow rate for these flowmeters should be set using the center of the ball.
RIGHT SIDE FLOWMETER
The right side flowmeter and corresponding bleed is inactive when it is angled towards the
front of the blender. To activate it and initiate the required bleed, push the flowmeter in
towards the side of the blender and then rotate it clockwise (towards the back) to its
vertical position. The internal bleed will now be active and the flow rate may be set using
the knob on the flowmeter. To return the flowmeter and bleed to its off (inactive) state,
return it to its angled position by pushing it in and rotating it counterclockwise (towards the
front).
LEFT SIDE FLOWMETER
CAUTION: The flowmeter on the left side is stationary. Do not try to turn it.
This flowmeter is meant for flows above 3 lpm. It may, however, be used below 3 lpm with
the following consideration; when set below 3 lpm, turn on the required bleed by setting
the right side flowmeter to its vertical position following the procedure as described above.
Even though the flowmeter on the right is not being used, having it in the vertical position
activates the internal bleed thereby guaranteeing O2 accuracy while using the left side
flowmeter at flows below 3 lpm.
1. WARNINGS, CAUTIONS AND NOTES
WARNINGS
If the pressure of the oxygen or air gas source increases or decreases resulting in a 20
psi* (138 kPa) difference (*30 psi [207 kPa] in the case of overseas devices and those
manufactured for Draeger / Hill Rom / Air-Shields), the alarm will sound. This will affect
the blender’s output flow and oxygen concentration.
The blender alarm will sound if the air or oxygen gas source fails. This indicates to the
user that the oxygen concentration or flow may not be accurate. A physician must
determine the correct FIO2 setting.
The blender must not be exposed to extremely high temperatures, as in the case of steam
autoclaving (which could reach 145 degrees F).
The alarm should not be obstructed, removed or tampered with in any way.
The blender is designed to operate from a 50 psig (345 kPa) source of air and oxygen.
Page 5
An oxygen analyzer should be used to verify the initial patient gas concentration and
when making any adjustments to the concentration (Bio-Med Devices’ M-2 Oxygen /
Temperature Monitor or M-10 ten-function Ventilation Monitor may be used in this
capacity).
The bleed port on the bottom of the blender must not be covered at any time.
Never leave a ventilator patient unattended, or without remote monitoring.
Some special order blenders may not have a bleed when using the auxiliary outlet. When
this is the case, the flow specifications for the primary outlet apply to the auxiliary outlet.
CAUTIONS
Moisture or dirt can affect the operation of the blender; a clean dry gas source must be
used at all times. The air must meet USP compressed air and ANSI Z86.1-1973 grade F
and water vapor content cannot exceed a dew point of 5 degrees F. below the lowest
ambient temperature to which the blender and accessories are exposed.
A water trap assembly and filter must be used to avoid malfunction should water
accidentally get into the gas supply sources.
Do not use in a MRI room unless the blender has been built by Bio-Med Devices to be
used for such an environment. This will be indicated by “MRI” on the blender.
The flowmeter on the left side of the NEO2 BLEND is stationary. Do not try to turn it.
If the blender does not pass the performance test, do not place the unit into service; call
your dealer or service representative.
NOTES
The NEO2 BLEND with two flowmeters conforms to the model #2003 configuration with
flow limited by the flowmeters. Refer to Addendum 2 in the beginning of this manual.
2. EXPLANATION OF SYMBOLS
Attention, See instructions for Use
Date of Manufacture
Page 6
Serial Number
Catalog Number
The CE mark displayed on this product signifies that this device is in compliance
with the European Medical Devices Directive (Council Directive 93/42/EEC). As a
prerequisite for the CE mark, Bio-Med Devices operates under an ISO 13485 compliant
quality system (covering the design and manufacture of medical devices). The four-digit
code underlying the CE mark (0086) pertains to Bio-Med's Notified Body, the British
Standards Institute, whose function is to investigate and attest to the validity of CE-mark
claims.
3. SPECIFICATIONS
Bio-Med Devices' line of blenders delivers accurate FIO2 mixtures from either one or two
outlet ports and has an overall standard flow range of 2 - 120 LPM. They can be used
with ventilators, nasal cannulas, mask CPAP and resuscitation bags. By adding an
additional side port to the standard model, three outlet ports are able to provide combined
flows from 2 -100 LPM, allowing it to power three items at once. The 0-50 LPM Blender is
a perfect compromise between the High Flow and Low Flow blenders as it requires less of
a bleed for accuracies below 6 LPM than the High Flow while allowing greater maximum
flow than the Low Flow Blender. The Low Flow version of the blender provides flows from
3 to 30 LPM with no gas bleed. Bio-Med Devices also offers MRI compatible versions that
are made entirely of non-magnetic materials.
CAUTION: Do not use in a MRI room unless the blender has been built by Bio-Med
Devices to be used for such an environment. This will be indicated by “MRI” on blender.
Oxygen % Range: 21 to 100%
Oxygen % Accuracy: ±3% of full scale
Supply Pressure: 30-75 psi (207-517 kPa) Air & oxygen must be within 10 psi (69 kPa)
of
each other.
Page 7
Maximum Flow: ≥120 LPM (≥50 LPM, 0-50 Flow blender; ≥30 LPM, Low Flow blender)
@
60% setting & 50 psi (345 kPa) inlet pressures
Standard Flow Ranges: refer to table in Section 2
Custom Configuration Flow Ranges: refer to addendums and table in Section 2
Pressure Drop: <6 psi (42 kPa) at 50 psi (345 kPa) inlet pressure and 40 LPM flow (10
LPM, Low Flow blender)
Alarm/Bypass Reset: when inlet gas pressure differential is ≥6 psi (42 kPa).
Alarm Intensity: 80 dB at 1 foot
Dimensions:
Weight:
Height 3 1/2" (8.9 cm)
Width 2 1/4"
(5.7 cm)
Depth 2 7/8" (7.3 cm)
2
3/4
lbs
(1.25
kg).
4. INTRODUCTION AND OPERATION
The Bio-Med Air/Oxygen Blender is a precision proportioning device for mixing medical
grade air and oxygen to any concentration from 21% to 100% oxygen and delivering it to
a variety of respiratory care devices. The blender uses source air and oxygen at a
pressure of 50 psi (345 kPa) connected to two D.I.S.S. fittings on the bottom of the
blender. Each fitting has a built-in 30 micron particulate filter. The gas source then
passes through a duckbill check valve which prevents reverse gas flows from either
source.
The blender uses a double stage balancing system with the gas entering into the first
stage to equalize the operating pressure of the gas sources before entering the
proportioning stage.
The gases then flow into the proportioning stage where they are mixed to the percentage
dialed in on the front panel knob. This stage has a double-ended valve with valve seats
on either end. Each one of these valve seats controls the passage of the air or oxygen to
the outlet of the blender.
Page 8
Depending on the model, there may be up to three gas outlets on the Bio-Med Devices
blender as outlined in the table below. The model number can be found on the back of the
1
blender. Use the first four digits to identify the configuration . Under certain custom
configurations, some ports may not be present. The flow limitations listed below apply,
regardless of what is attached to the indicated port.
Note 1. For 2002K2 with 0-50 LPM on the front of the blender, use 2000K for its configuration.
Blender Ports
B
AIR O
Model
2000K
2001
2001K
2002
2003
2004
Ports A,
Flow Range - Port “A” 6-50
Flow Range - Port “B” Flow Range - Port “C” N/A
Knob @C A,
C A, Knob
@C A B, C A,
B, C
lpm (No Bleed) 0-50 lpm (6 lpm
Bleed) 15-120 lpm (No Bleed)
15-120 lpm (No Bleed) 2-108 lpm
(10-12 lpm Bleed) 15-120 lpm (No
Bleed) N/A 15-120 lpm (No Bleed)
N/A Knob Set >6 Knob Set< 6 N/A 2-108 lpm (10-12 lpm Bleed)
N/A N/A Knob Set >15 Knob Set< 15 N/A N/A 3-30 lpm (No Bleed)
0-30 lpm (3 lpm Bleed) 15-120 lpm (No Bleed) 2-108 lpm (10-12
Bleed)
WARNING: Some special order blenders may not have a bleed when using the
auxiliary outlet. When this is the case, the flow specifications for the primary outlet
apply to the auxiliary outlet.
NOTE: The NEO2 BLEND with two flowmeters conforms to the model #2003 configuration
with flow limited by the flowmeters. Refer to Addendum 2 in the beginning of this manual.
The blender has an audible alarm built in to detect if either of the gas sources changes by
more than 20 psi* (138 kPa) from the other. This will warn the user that they are running
out of one of the gas sources or that there is a severe pressure drop in one source. If both
gas sources drop or increase together such that a 20 psi* (138 kPa) difference cannot be
detected, then no alarm will sound. If the blender is connected but not being used and a
20 psi* (138 kPa) difference in gas sources develops, the blender will not alarm.
*(30 psi [207 kPa] in the case of overseas devices and those manufactured for Draeger /
Hill Rom / Air-Shields)
The blender alarm/bypass function will provide > 90 LPM (the full 30 LPM, Low Flow
Page 9
Blender) upon the loss of air or oxygen, if the remaining gas is at 50 psi (345 kPa).
5. SETTING UP THE BLENDER
The Bio-Med Devices Blender can be either pole-, wall-, or rail-mounted for easy use for
any desired application. The inlet fittings are located on the bottom of the blender and
conform to Diameter Index Safety System (D.I.S.S.) so that air and oxygen connections
cannot be reversed. Connect an air high pressure hose to the air fitting and an oxygen
high pressure hose to the oxygen fitting on the bottom of the blender. Bio-Med Devices
recommends an air inlet water trap be used between the air hose and inlet fitting to
prevent moisture from entering the blender.
The primary outlet (see Addendum1) on the bottom of the standard high flow blender is
appropriate for high flow situations, as with most ventilators requiring flows up to 120 LPM.
Flows of less than 15 LPM (6 LPM, 0-50 LPM blender; 3 LPM, Low Flow blender) require
the auxiliary outlet (side port or knob). If both outlets are used simultaneously, neither one
will
deliver
its
maximum
flow.
6. TESTING THE BLENDER
The following checks should be performed before first placing the blender into service.
Note: If the blender does not pass these checks do not place the unit into
service; call Bio-Med Devices Service Department at (203) 458-0202.
First, connect the 50 psi (345 kPa) air and oxygen sources to the appropriate fittings and
set the blender to 60% (the alarm should not activate). Check to see that the oxygen
concentration is actually 60% by using a calibrated oxygen analyzer. Disconnect the
oxygen source from the blender and listen for the audible alarm. Once it alarms,
reconnect the oxygen to stop the alarm and verify the oxygen concentration again. Next
disconnect the air source from the blender and listen for the audible alarm. Once it alarms,
reconnect
the
air
and
verify
the
oxygen
concentration
again.
7. USING THE BLENDER
Connect the gas outlet of the blender either directly or via a high pressure hose to the
ventilator or other equipment with which it is being used. Set the control on the front panel
to the desired oxygen concentration. Turn on the 50 psi (345 kPa) air and oxygen
sources and set the controls on the ventilator or equipment being used. Use a calibrated
oxygen analyzer to check the accuracy of the patient gas. When changing oxygen
concentration, wait sixty seconds (equilibration time) before checking it against the
analyzer.
Page 10
To use the standard high flow blender for low flow applications, connect a flowmeter to the
secondary outlet (see Addendum 1 at the beginning of this manual), and set the
concentration with the knob on the front panel. Then turn on the source gases, set the
flowmeter and check the output with a calibrated oxygen analyzer.
8. TROUBLE SHOOTING GUIDE
PROBLEM
CAUSE OF PROBLEM
TO SOLVE PROBLEM
OXYGEN ANALYZER
DOESN'T AGREE WITH
SETTING OF BLENDER
ANALYZER OUT OF
CALIBRATION BLENDER
OUT OF CALIBRATION
CALIBRATE OXYGEN
ANALYZER CALL
BIO-MED SERVICE
DEPARTMENT
DIRTY GAS SUPPLY
BLEED ON BOTTOM OF
BLENDER IS
RESTRICTED
BLENDER ALARMING
THE ONLY TIME THE
BLENDER IS ACCURATE
IS WHEN THE SOURCE
PRESSURES ARE
CALL BIO-MED SERVICE
DEPARTMENT
CALL BIO-MED SERVICE
DEPARTMENT
AIR IS FLOWING INTO
PIECE OF EQUIPMENT
BEING USED AND
DILUTING
CONCENTRATION
CORRECT SITUATION
BY STOPPING THE
FLOW OF AIR
AIR AND OXYGEN
SOURCE PRESSURES
HAVE GREATER THAN
20 PSI* (138 kPa)
DIFFERENTIAL
BRING THE SOURCE
PRESSURES WITHIN
THE 20 PSI* (138 kPa)
RANGE
ALARM SYSTEM IS OUT
OF CALIBRATION
CALL BIO-MED SERVICE
DEPARTMENT
DIRTY GAS IS
CONTAMINATING
ALARM SYSTEM
CALL BIO-MED SERVICE
DEPARTMENT
PRESSURE BALANCE
CHAMBER NOT
WORKING PROPERLY
CALL BIO-MED SERVICE
DEPARTMENT
Page 11
EXACTLY THE SAME
*(30 psi [207 kPa] in the case of overseas devices and those manufactured for Draeger /
Hill Rom / Air-Shields)
9. CLEANING INSTRUCTIONS
Bio-Med Devices' line of blenders should only be cleaned by wiping the outside surfaces
with alcohol applied to a tissue or cloth. These blenders should never be sprayed with or
immersed in any other liquid.
10. BLENDER WARRANTY
The Bio-Med Devices, Inc. warranty lasts for one year from date of purchase. This
warranty covers parts and labor. Shipping costs are covered up to six months from date
of purchase. This warranty is limited to defects in parts and workmanship; Bio-Med
Devices will not be held responsible for misuse or abuse of the product.
All service must be done by Bio-Med Devices or an authorized service representative of
Bio-Med Devices. Bio-Med Devices will not be held responsible for unauthorized service
work on any blender.
11. SERVICE RECOMMENDATIONS
Periodic preventive maintenance should be performed to insure continued proper
operation of the blender. The frequency of preventative maintenance is determined
by many factors, some of which are:
Frequency & length of use
Quality of the compressed gas source(s)
Environmental conditions
Recommended Maintenance Schedule
Interval
Recommended Procedures
Prior to each use
Performance test
Every year between PM’s
Calibration certification
Every 2 years
Major overhaul, cleaning and calibration
Recommend return to factory for this service
Page 12
APPENDIX European Agent
Bio-Med Devices’ Official Agent in Europe is:
HORST HÖRNLA
H + H Intermed
Schwedenstraße 32
87463 Dietmannsried-Reicholzried
United Germany
Telefon:
Fax:
(08
(08
31)
31)
6
6
UK Agent is:
INSPIRATION HEALTHCARE LTD
Gildor House
West Street
Earl Shilton
Leicester
LE9 7EJ
United Kingdom
Telfon: 0044 (0) 1455 840555
Fax: 0044 (0) 1455 841464
31
09
86
54