Overview
9 Pages
Preview
Page 1
The iPAD SP1 Automatic External Defibrillator
Overview
SP1 Layout
Patient Mode Switch: - Easily switch from Adult to Child mode without changing pads.
Pad Connector
Power Button
- Safety cover prevents accidental switching.
Shock Button
Speaker: Ambient noise detector measures level of background noise and adjusts the volume of the voice prompts accordingly. Ideal for noisy environments such as public places, factories, warehouses, schools etc.
Disposable Battery Pack
- Electrode pads are stored,
- High Capacity - Type DC 12 Volt 4.2Ah, Lithium Manganese Dioxide
pre-connected, in a clear compartment on the underside of the unit.
- Capacity - Minimum 200 shocks (150J)
Smart Pads Storage:
- Lifespan - 5 years - assumes unit is kept within operable temperature range and after the initial check, the unit is left in standby mode.
SP1 Layout Cont. Visual Indicator: Visual Indicator shows at a glance - Battery Life - Unit Status - Pad Status
Information Button: 'i' button performs the following functions; Allows responder to select if CPR metronome is heard during CPR Checks for faults and errors Provides 'handover' information to Ambulance crew (usage time and shocks delivered) Allows the SP1 to be programmed with CPR protocols such as compression rate, number of compressions, breaths and cycles etc, ensuring that the SP1 is always up to date
Indicator Lights:
Downloads data for review
Indicator Lights operate in time with the voice prompts to prompt the responder with;
Checks the units software version
- Pad placement
Data Recording
- Staying clear of casualty
- IrDA Port - Wireless Transmission of data to PC
- CPR compressions
- SD Card Port - Data can be transferred easily - Internal Memory (Nand-Flash) ECG, Event - Storage Capacity - Multi recording 5 events / max 3 hours - Data review program - CU-EX1 Software (for PC)
Key Features The new iPAD SP1 from CU Medical is the latest in the range of intelligent public access defibrillators from one of the Worlds leading manufacturers of clinical defibrillator monitors.
Designed to be used by minimally trained individuals, the iPAD SP1 is suitable for use almost anywhere that people gather, whether it's within the workplace, a shopping centre or a hospital, the iPAD SP1 makes saving lives easier than ever.
Easily switch from Adult to Child mode without changing pads Safety cover prevents accidental switching Smaller child pads also available
Ambient noise detector measures level of background noise and adjusts the volume of the voice prompts accordingly. Ideal for noisy environments such as public places, factories, warehouses, schools etc.
Integrated pad storage - electrode pads are stored, pre-connected, in a clear compartment on the underside of the unit. Pads are easily removed by pulling on the exposed tab The pads are always ready for use and easily checked
CPR Detection - the iPAD SP1 detects if CPR is being performed when appropriate If CPR is not being performed, voice prompts encourage the responder to 'perform CPR' If CPR is being performed, voice prompts encourage the responder to 'continue CPR'
Key Features Cont. Visual Indicator shows at a glance - Battery Life - Unit Status - Pad Status
Integrated pad storage - electrode pads are stored, pre-connected, in a clear compartment on the underside of the unit. Pads are easily removed by pulling on the exposed tab The pads are always ready for use and easily checked
Internal memory stores the last 5 events/3 hours of data. Data can be transferred via the built in SD card and IrDA ports Data can be reviewed via the 'CU-EX1' software (for pc)
'i' button performs the following functions; - Allows responder to select if CPR metronome is heard during CPR - Checks for faults and errors - Provides 'handover' information to Ambulance crew (usage time and shocks delivered) - Allows the SP1 to be programmed with CPR protocols such as compression rate, number of compressions, breaths and cycles etc, ensuring that the SP1 is always up to date - Downloads data for review - Checks the units software version
Visual Indicator Statuses Indicator Status LCD Device Operation Status LCD Device Operation Status LCD Battery Level Indicator Status LCD Battery Level Indicator Status LCD Battery Level Indicator Status LCD Battery Level Indicator Status LCD Pad Status Status LCD Pad Status Status LCD Pad Status Do Not Touch Patient Indicator: Off Do Not Touch Patient Indicator: Light
Description The device is functioning normally. The device has an error. The battery is fully charged. Less than half battery power remains. Less than a quarter battery power remains. Battery is low. The expiration date of the pad is more than 3 months. The pad will expire within 3 months. The pad is used or expired.
You may touch the patient.
You may not touch the patient.
CPR Detection Indicator: Light
Indicates that CPR is being performed.
CPR Detection Indicator: Flashing
Indicates that CPR is not performed or not properly performed.
i-Button: Flashing in Red
Shock Button: Flashing in Orange
The device detected an error. Press the i-Button for more information. The device is ready to deliver a defibrillating shock. Press the Shock Button to deliver a shock.
Note
Specifications Physical Category
Nominal Specifications
Dimensions
260㎜ x 256㎜ x 69.5㎜ (Width x Length x Height)
Weight
2.4㎏ (Including the battery pack and pads)
Environmental Category
Nominal Specifications
Operational Status (The device is in emergency use) Temperature: 0°C ~ 43°C (32°F ~ 109°F) Humidity: 5% ~ 95% (non condensing) Altitude
0 to 15,000 feet (operational and storage)
Drop Vibration
Withstands 1.2-meter drop to any edge, corner, or surface Operating: Meets MIL-STD-810G Fig.514.6E-1, random Standby: Meets MIL-STD-810G Fig.514.6E-2, swept sine(helicopter)
Sealing
IEC 60529: IP55
ESD
Meets IEC 61000-4-2:2001
EMI (Radiated) EMI (Immunity)
Meets IEC 60601-1-2 limits, method EN 55011:2007 +A2:2007, Group 1, Class B Meets IEC 60601-1-2 limits, method EN 61000-4-3:2006 +A1:2008 Level 3 (10V/m 80MHz to 2500MHz)
Defibrillator Category
Nominal Specifications
Operating Mode
Semi-automated
Waveform
e-cube biphasic (Truncated exponential type)
Output Energy
150 J at 50 Ω load for adults 50 J at 50 Ω load for children
Charge Control
Controlled by an automated patient analysis system
Charging Time
Within 6 seconds of the voice prompt “Shock Advised”.
AED Comparison Table IPAD SP1
Cardiac Science G5
Cardiac Science G3
Philips/ Laerdal FR3
ZOLL AED Plus
Physio Control Lifepak CR+
Heartsine
DefibTech
Samaritan
Lifeline
Electrodes Tested For Presence Daily, Weekly, Monthly Self Tests
Weekly
Daily
All Tests Automated
Status Indicator Showing Clinical Readiness Battery Capacity Gauge Pad Life Indicator Used Pad Detection Pad Life Upto
36 months
2 Year
2 Years
2 Years
Uses Same Set of Pads of Adult And Child
5 Years
2 Years
3.5 Years
2 Years
2
Ambient Noise Detector
Integrated Pad Storage CPR Detection Illuminated Prompts Internal Memory
15 hours
90 Minutes
1 Hour
8 Hours via memory card
7 Hours
20 Minutes
45 Minutes
Dependant On Card
IR / SD Card
USB
Direct Connect
Direct Connect
Direct Connect
IR
Direct Connect
Data Card
Fluid Ingress Protection
IP 55
IP55
IP 24
IP 55
IP 55
IP 55
IP 56
IP54
Battery Capacity (Shocks/Shelf)
200/5 years
500/4 Years
290/5 years
/5 years
/5 years
/2 years
/3.5 years
125/5Years 300/7 Years
Capacity
Data Transfer Method
User Reprogrammable CPR Metronome Paramedic Handover Information
Carry Case Supplied As Standard Manual Self Test Option All Status Indicators Visible When Case Is Closed 1
Via key.
Data supplied by Community Heartbeat Trust
SP1 Accessories SP1 Disposable Adult/Child Smart Electrode Pads SP1 Adult/Child "Smart" Pads, capable of 'talking' to the device to share expiry notice and previous usage. These pads are useable with paediatric patients thanks to the intuitive adult/ paediatric switch on the device.
SP1 Disposable Battery Pack This Lithium-Manganese Oxide (LiMnO2) disposable battery. With up to five years life or in excess of 200 shocks it is high performing while cost effective.
SP1 Orange Carry Case This highly visible carry case is the perfect companion for your SP1 device. Currently available at no additional cost. The bag has space for an additional set of pads, a spare battery and an AED Start Kit. There is also a detachable shoulder strap. The case offers protection while still being able to use the device.
Additional Options Electrode Pads Adapters These pad adapters will allow the electrode pads from Zoll, Phillips and PhysioControl to be used with the iPAD SP1. This means the same pads can be unplugged from iPAD and plugged straight in to the defibrillator on the front line vehicle. This reduces risk, saves time, reduces trauma and is cost-effective. SP1 Wall Bracket This sturdy metal wall bracket is the perfect means for keeping the iPAD at hand, ready to use. The SP1 is securely cradled and held in place, with a Velcro strap for quick release.
SP1 Wall Cabinet This rigid plastic wall cabinet is ideal for holding your i-PAD within an indoor environment. The device is held firmly in place until required. There is also a battery operated alarm, which is triggered when the door is