KARL STORZ ENDOSKOPE
Connecting Cable Reprocessing Instructions
17 Pages
Preview
Page 1
en
Reprocessing instructions Connecting Cable 28340770
02-2021 Copyright © All product illustrations, product descriptions, and texts are the intellectual property of KARL STORZ SE & Co. KG. Their use and reproduction by third parties require the express approval of KARL STORZ SE & Co. KG. All rights reserved.
Table of contents
Table of contents 1
1 Target group ... 4 2 General information... 5 2.1 Read the reprocessing instructions ... 5 2.2 Read the reprocessing instructions for use ... 5 2.3 Read the instructions for use for the reprocessing unit... 5 2.4 National laws and regulations... 5 2.5 Additional information on the product ... 5 2.6 Description of warning messages... 5 3 Safety ... 7 3.1 Unsterile product ... 7 3.2 Contaminated products ... 7 3.3 Working with process chemicals ... 7 3.4 Creutzfeldt-Jakob disease ... 7 4 Overview of processes... 8 4.1 Reprocessing cycle for manual wipe disinfection... 8 4.2 Reprocessing cycle for standard products... 9 5 Manual wipe disinfection ... 10 6 Requisite materials... 11 7 Initial treatment at the site of use... 12 7.1 Transport to the reprocessing site... 12 8 Cleaning and disinfection... 13 8.1 Reprocessing with automated decontamination ... 13 8.1.1 Automated cleaning / thermal disinfection ... 13 9 Visual inspection ... 14 10 Life span... 15 10.1 Functional check... 15 11 Packaging ... 16
NOT APPLICABLE FOR THE US Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
3
Target group
1 Target group These reprocessing instructions are intended for personnel with technical knowledge and expertise in the reprocessing of medical devices.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
4
General information
2 General information 2.1 Read the reprocessing instructions If the reprocessing instructions are not followed, patients, users, or third parties may be injured or the product may be damaged. Read the reprocessing instructions for the product and its components carefully and follow all the safety notes and warnings.
2.2 Read the reprocessing instructions for use If the reprocessing instructions for use are not followed, patients, users, or third parties may be injured or the product may be damaged. Read and follow the “Cleaning, disinfection, care, and sterilization of KARL STORZ instruments” instructions for use (item no. 96216003). The cleaning, disinfection, and sterilization procedures are explained in detail in the reprocessing instructions for use. The reprocessing instructions for use can be downloaded from www.karlstorz.com.
2.3 Read the instructions for use for the reprocessing unit If the instructions for use are not followed, patients, users, or third parties may be injured or the product may be damaged. Read the instructions for use for the reprocessing unit carefully and follow all the safety notes and warnings. Carry out reprocessing in accordance with the instructions for use for the reprocessing unit.
2.4 National laws and regulations National laws and regulations must be observed in addition to the accompanying documentation.
2.5 Additional information on the product Additional general information on the product can be requested and downloaded from www.karlstorz.com.
2.6 Description of warning messages To prevent any injury to persons or damage to property, the warnings and safety notes in the instructions for use must be observed. The warning messages describe the following levels of danger. WARNING WARNING Designates a possible imminent risk. If this is not avoided, it could lead to death or serious injuries. CAUTION CAUTION Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
5
General information
ATTENTION ATTENTION Designates a possibly harmful situation. If this is not avoided, the products could be damaged.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
6
Safety
3 Safety 3.1 Unsterile product The product is not sterile when delivered. The use of unsterile products poses a risk of infection for patients, users, or third parties. Before use, reprocess the product in line with the reprocessing instructions.
3.2 Contaminated products During work on contaminated products, the guidelines for personal safety must be observed.
3.3 Working with process chemicals Incorrect exposure time, concentration, life span, and range of action of chemicals can lead to a risk of infection for the patient, user, and third parties, as well as damage to the product. Note the information provided by the manufacturer of the chemicals and the microbiological range of action of the chemicals used.
3.4 Creutzfeldt-Jakob disease Products that come into contact with the central nervous system can become contaminated by organic residue containing prions. Prions lead to infection with Creutzfeldt-Jakob disease. If Creutzfeld-Jakob disease has been diagnosed or is suspected: Dispose of the product properly and do not continue to use it.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
7
Overview of processes
4 Overview of processes The following reprocessing procedures have been approved for the product: –
Manual wipe disinfection
–
Reprocessing with automated decontamination
A detailed description of the validated processes is provided in the respective chapters in these instructions.
4.1 Reprocessing cycle for manual wipe disinfection
Purchasing
Use
Decontamination
Maintenance, functional test
Repair or disposal
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
8
Overview of processes
4.2 Reprocessing cycle for standard products
Use
Purchasing
Storage Transport
Transport Pre-cleaning Sterilization, release
Packaging
Decontamination Maintenance, functional test
Repair or disposal
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
9
Manual wipe disinfection
5 Manual wipe disinfection Manual wipe disinfection is not a validated reprocessing procedure. Required materials: –
Disposable cloth and disinfectant Alternatively: ready-to-use disinfectant cloth
–
Dry, low-lint cloth
1.
Use a disposable cloth moistened with disinfectant or a ready-to-use disinfectant cloth to wipe the external surfaces of the product.
2.
Take up any excess moisture with a dry low-lint cloth.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
10
Requisite materials
6 Requisite materials The reprocessing accessories used must be clean and functional. The following reprocessing accessories are required: Application
Material
Initial treatment at the site of use
Moist compresses, possibly disposable cloth
Cleaning and disinfection Manual drying and/or after-drying
Medical compressed air from compressed air gun, item no. 27660 Alternatively: syringe 60 cc
Maintenance Packaging
Standardized and approved packaging
Suitable reprocessing accessories are listed in the following catalog: –
HYGIENE – Care, Sterilization, Storage Techniques (item no. 96211004)
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
11
Initial treatment at the site of use
7 Initial treatment at the site of use Reprocessing of the product should start within 2 hours of use to ensure the effectiveness of the reprocessing processes listed in the reprocessing instructions. 1.
Wipe the surfaces of the product with a compress or disposable cloth to remove gross soiling, corrosive solutions, and drugs.
2.
Irrigate surfaces with cold water.
7.1 Transport to the reprocessing site 1.
Right after using it, place the dry product in a suitable transport container.
2.
Transport the securely positioned product to the site of reprocessing.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
12
Cleaning and disinfection
8 Cleaning and disinfection The following procedures are validated and approved for cleaning and disinfection of the product: –
Automated cleaning: thermal disinfection
8.1 Reprocessing with automated decontamination 8.1.1 Automated cleaning / thermal disinfection A washer-disinfector for thermostable devices must be used for the device. The washerdisinfector must meet the requirements of standard ISO 15883. The A0 value of the disinfection process must be observed. Required materials: –
Suitable slide-in trolley and, if necessary, suitable instrument holder. The selection must be made in consultation with the manufacturer of the WD.
1.
Place the device in the slide-in trolley and, if necessary, in the instrument holder.
2.
Use the adaptor to connect the lumina of the device to the slide-in trolley.
3.
Connect the lumina of the device using the irrigation facilities of the slide-in trolley.
4.
The parameters of the automated cleaning and disinfection process validated by KARL STORZ are specified in the document “Cleaning, Disinfection, Care, and Sterilization of KARL STORZ Instruments” (item no. 96216003).
Steps 1. Pre-irrigation 2. Cleaning 3. Intermediate irrigation 4. Thermal disinfection 5. Drying WARNING Risk of infection due to residual liquid! If devices are not adequately dried following disinfection, the effectiveness of the validated reprocessing processes is not guaranteed. Use compressed air or a syringe filled with air to dry devices fully following disinfection. Check if the device is dry and dry it by hand if necessary, see chapter Checking [p. 14].
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
13
Visual inspection
9 Visual inspection 1.
Check products for the following points: ‒ Visible contamination ‒ Damage and corrosion ‒ Completeness ‒ Dryness
2.
Subject any products displaying visible soiling to another complete cleaning and disinfection process.
3.
Discard damaged and corroded medical devices.
4.
Discard incomplete medical devices or replace missing parts.
5.
Dry the product by hand if necessary.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
14
Life span
10 Life span The end of the product life is largely determined by wear, reprocessing processes, the chemicals used and any damage resulting from use.
10.1 Functional check If the product does not fulfill one of the points listed below or if damage can be identified, see chapter “Maintenance, repair, and disposal” in the instructions for use. The following tests must be carried out to detect functional limitations: 1.
Check the surface of the product for mechanical integrity and changes.
2.
Check the labeling for legibility.
3.
Check the product for mechanical integrity.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
15
Packaging
11 Packaging The packaging material must always be matched to the sterilization process being used. Required materials: –
Standardized packaging materials and packaging systems that are approved for the product (EN 868 Parts 2–10, EN ISO 11607 Parts 1 + 2, DIN 58953)
Within the scope of validation, the following packaging material was used: Package the product according to the instructions of the packaging manufacturer.
Reprocessing instructions • Connecting Cable • 28340770_EN_V1.1_02-2021_RI_CE
16