KARL STORZ ENDOSKOPE

Mini Straight Forward Telescope 0°,1.3mm Reprocessing Instructions

19 Pages

Table of contents  Table of contents 1  1 Target group ...  4  2 General information ... 2.1 Read the reprocessing instructions ... 2.2 Read the reprocessing instructions for use ... 2.3 Read the instructions for use for the reprocessing unit ... 2.4 National laws and regulations ... 2.5 Additional information on the product ... 2.6 Description of warning messages ...  5 5 5 5 5 5 5  3 Safety ... 3.1 Unsterile product ... 3.2 Contaminated products ... 3.3 Working with process chemicals ... 3.4 Creutzfeldt-Jakob disease ...  7 7 7 7 7  4 Overview of processes ... 4.1 Reprocessing cycle for standard products ...  8 8  5 Requisite materials ...  9  6 Bedside Pre-Cleaning ... 10 6.1 Transport to the reprocessing site ... 10 7 Disassembly ... 11 7.1 Removing the light inlet socket ... 11 8 Cleaning and disinfection ... 12 8.1 Reprocessing with automated decontamination ... 12 8.1.1 Pre-cleaning ... 12 8.1.2 Automated cleaning / thermal disinfection ... 12 9 Visual inspection ... 14 10 Assembly ... 15 11 Life span ... 16 11.1 Functional check ... 16 12 Packaging ... 17 13 Sterilization ... 18 13.1 Steam sterilization ... 18  NOT APPLICABLE FOR THE US Reprocessing instructions • Mini Straight Forward Telescope 0°,1.3mm • 11575A_EN_V3.1_11-2021_RI_CE  3
File Type: PDF
File Size: 262 KB
File Name: KARL STORZ ENDOSKOPE - 11575A_EN - Mini Straight Forward Telescope 0°,1.3mm Reprocessing Instructions - 2021-11.pdf