Reprocessing instructions Tuning Fork C-512,without weights 125500
02-2021 © KARL STORZ It is not permitted to forward or reproduce this document or to use or disclose the contents thereof unless express consent is given. Publisher KARL STORZ SE & Co. KG Dr.-Karl-Storz-Straße 34 78532 Tuttlingen Postfach 230 78503 Tuttlingen Germany Phone: +49 7461 708-0 Fax: +49 7461 708-105 E-mail: [email protected] www.karlstorz.com
Table of contents
Table of contents 1
1 Target group ... 4 2 General information... 5 2.1 Read the reprocessing instructions ... 5 2.2 Read the reprocessing instructions for use ... 5 2.3 Read the instructions for use for the reprocessing unit... 5 2.4 National laws and regulations... 5 2.5 Additional information on the product ... 5 2.6 Description of warning messages... 5 3 Safety ... 7 3.1 Unsterile product ... 7 3.2 Contaminated products ... 7 3.3 Working with process chemicals ... 7 3.4 Creutzfeldt-Jakob disease ... 7 4 Overview of processes... 8 4.1 Reprocessing cycle for manual wipe disinfection... 8 4.2 Reprocessing cycle for standard products... 9 5 Manual wipe disinfection ... 10 6 Visual inspection ... 11 7 Life span... 12 7.1 Functional check... 12
NOT APPLICABLE FOR THE US Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
3
Target group
1 Target group These reprocessing instructions are intended for personnel with technical knowledge and expertise in the reprocessing of medical devices.
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
4
General information
2 General information 2.1 Read the reprocessing instructions If the reprocessing instructions are not followed, patients, users, or third parties may be injured or the product may be damaged. Read the reprocessing instructions for the product and its components carefully and follow all the safety notes and warnings.
2.2 Read the reprocessing instructions for use If the reprocessing instructions for use are not followed, patients, users, or third parties may be injured or the product may be damaged. Read and follow the “Cleaning, disinfection, care, and sterilization of KARL STORZ instruments” instructions for use (item no. 96216003). The cleaning, disinfection, and sterilization procedures are explained in detail in the reprocessing instructions for use. The reprocessing instructions for use can be downloaded from www.karlstorz.com.
2.3 Read the instructions for use for the reprocessing unit If the instructions for use are not followed, patients, users, or third parties may be injured or the product may be damaged. Read the instructions for use for the reprocessing unit carefully and follow all the safety notes and warnings. Carry out reprocessing in accordance with the instructions for use for the reprocessing unit.
2.4 National laws and regulations National laws and regulations must be observed in addition to the accompanying documentation.
2.5 Additional information on the product Additional general information on the product can be requested and downloaded from www.karlstorz.com.
2.6 Description of warning messages To prevent any injury to persons or damage to property, the warnings and safety notes in the instructions for use must be observed. The warning messages describe the following levels of danger. WARNING WARNING Designates a possible imminent risk. If this is not avoided, it could lead to death or serious injuries. CAUTION CAUTION Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries.
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
5
General information
ATTENTION ATTENTION Designates a possibly harmful situation. If this is not avoided, the products could be damaged.
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
6
Safety
3 Safety 3.1 Unsterile product The product is not sterile when delivered. The use of unsterile products poses a risk of infection for patients, users, or third parties. Before use, reprocess the product in line with the reprocessing instructions.
3.2 Contaminated products During work on contaminated products, the guidelines for personal safety must be observed.
3.3 Working with process chemicals Incorrect exposure time, concentration, life span, and range of action of chemicals can lead to a risk of infection for the patient, user, and third parties, as well as damage to the product. Note the information provided by the manufacturer of the chemicals and the microbiological range of action of the chemicals used.
3.4 Creutzfeldt-Jakob disease Products that come into contact with the central nervous system can become contaminated by organic residue containing prions. Prions lead to infection with Creutzfeldt-Jakob disease. If Creutzfeld-Jakob disease has been diagnosed or is suspected: Dispose of the product properly and do not continue to use it.
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
7
Overview of processes
4 Overview of processes The following reprocessing procedures have been approved for the product: –
Manual wipe disinfection
A detailed description of the validated processes is provided in the respective chapters in these instructions.
4.1 Reprocessing cycle for manual wipe disinfection
Purchasing
Use
Decontamination
Maintenance, functional test
Repair or disposal
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
8
Overview of processes
4.2 Reprocessing cycle for standard products
Use Storage Transport
Purchasing
Transport Pre-cleaning Sterilization, release
Packaging
Decontamination Maintenance, functional test
Repair or disposal
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
9
Manual wipe disinfection
5 Manual wipe disinfection Manual wipe disinfection is not a validated reprocessing procedure. Required materials: –
Disposable cloth and disinfectant Alternatively: ready-to-use disinfectant cloth
–
Dry, low-lint cloth
1.
Use a disposable cloth moistened with disinfectant or a ready-to-use disinfectant cloth to wipe the external surfaces of the product.
2.
Take up any excess moisture with a dry low-lint cloth.
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
10
Visual inspection
6 Visual inspection 1.
Check products for the following points: ‒ Visible contamination ‒ Damage and corrosion ‒ Completeness ‒ Dryness
2.
Subject any products displaying visible soiling to another complete cleaning and disinfection process.
3.
Discard damaged and corroded medical devices.
4.
Discard incomplete medical devices or replace missing parts.
5.
Dry the product by hand if necessary.
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
11
Life span
7 Life span The end of the product life is largely determined by wear, reprocessing processes, the chemicals used and any damage resulting from use.
7.1 Functional check If the product does not fulfill one of the points listed below or if damage can be identified, see chapter “Maintenance, repair, and disposal” in the instructions for use. The following tests must be carried out to detect functional limitations: 1.
Check the surface of the product for mechanical integrity and changes.
2.
Check the labeling for legibility.
3.
Check the product for mechanical integrity.
Reprocessing instructions • Tuning Fork C-512,without weights • 125500_EN_V1.1_02-2021_RI_CE
12