Kendall

Pulse Oximetry

OxiMax forehead SpO2 sensor Application Guide

Application Guide

2 Pages

Nellcor™ OxiMax™ forehead SpO2 sensor Application Guide INTERNAL CAROTID ARTERY –THE KEY TO SENSOR PLACEMENT The lower forehead just above the eyebrow is ideal for sensor placement because its circulation stems from the internal carotid artery. Advantages are:  •	Little vasoconstriction compared with peripheral  or ear sensor sites.  •	Strong pulsatile signals without interference from  large vessels.  PROPER PLACEMENT IS ESSENTIAL •  P  lace the sensor just above the left or right eyebrow centring it with the iris.  •	Press the sensor firmly on the site for 10 seconds to  ensure adherence.  	For optimal results this position is key, as pulse perfusion is stronger in the lower forehead area and weaker higher up. Do not place the sensor on the temple.  BENEFITS OF USING THE HEADBAND 	The use of the headband is necessary  to assure optimum performance:  •	Pull the Velcro™* tab until the green arrows  on the elastic band are within the green tension indicator.  •	To secure the headband, fasten with the Velcro™*  tab.  •	Periodically check the headband.  Read the sensor’s “Directions for Use” insert for complete instructions for use, including all warnings, cautions and specifications, prior to use of this product.
File Type: PDF
File Size: 228 KB
File Name: oximax__forehead_spo2_sensor_application_guide.pdf

By clicking Download you are confirming this is the correct document for your purposes