Test Data Form
1 Page
Preview
Page 1
Covidien Pty Ltd A.C.N. 003 143 502 A.B.N. 51 003 143 502 52A Huntingwood Drive, Huntingwood NSW 2148
TEST DATA SHEET – SCD 700 Series (product code: 295255) Date : __________ Job No : __________ Customer : ___________________
Serial No : _____________ Software Version :_______ Pass Fail
N/A
General Function Test (Test Mode 02)
[ ]
[ ]
[ ]
Pressure Transducer Calibration Verification (Test Mode 04)
[ ]
[ ]
[ ]
Self Test (Test Mode 05)
[ ]
[ ]
[ ]
Performance Test (Test Mode 06)
[ ]
[ ]
[ ]
Check E 12 displayed on LCD screen and Alarm sound with no cuff connected
[ ]
[ ]
[ ]
Check corresponding port Sleeve display on LCD Screen with one cuff connected
[ ]
[ ]
[ ]
FUNCTIONAL ALARMS TESTING
ELECTRICAL SAFETY TEST – Class 1 Type B (Refer AS3551:2004) Test
Condition
Limits
Pass Fail
N/A
[<0.3 ohm]**
[ ]
[ ]
[ ]
from shorted active/neutral earth pin on plug [> 1 Mohm]* exposed conductive parts [> 1 Mohm]* * no more than 20% variance from acceptance test
[ ] [ ]
[ ] [ ]
[ ] [ ]
[ ] [ ]
[ ] [ ]
[ ] [ ]
Protective earthing: earth on plug to 5.3.2 equipment chassis ** if >0.2 a) Cord must be <0.2and b) Recepticle pin to chassis <0.1 Insulation Resistance: 5.3.3 to:
Earth Leakage Current: 5.3.4
Control #_______________
normal neutral open
[< 500 uA] [< 1000 uA]
Tested by: ___________________________ FM-SC-181 SCD 700 Series Test Data Form v1. 2 February 2012