Blood Bank Refrigerators

Model RDBG2320MD Calibration Infomation

Calibration Infomation

1 Page

CALIBRATION INFORMATION MODEL :  RDBG2320MD  TYPE:  BLOOD BANK REFRIGERATOR  TEST PERIOD:  18 hours, readings taken every 1 min.  AMBIENT TEMPERATURE: 22.0 ± 1.0 °C  AMBIENT HUMIDITY: 32%RH ±5%  DATE OF TEST: 5/12/2007 TEST EQUIPMENT: Two Comark 6800 scanning thermometers each channel calibrated against an Ametek temperature calibrator UKAS Cert No: N1158632T unbuffered sensors  15 TEST POSITIONS  1  2 5 4  3 6  7  10 9  8  11 15 13  COTEST15.ECW  12  14  01/05/2007  READINGS Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15  Maximum (ºC) 5.4 5.2 5.5 6.2 5.3 5.1 5.1 5.3 5.7 5.2 5.0 5.2 5.4 5.5 5.2  Minimum (ºC) 3.0 3.2 3.3 4.5 2.5 2.3 2.3 2.8 3.6 2.5 3.0 2.8 2.0 2.0 2.9  Average (ºC) 4.4 4.3 4.5 5.2 4.1 3.9 4.0 4.2 4.8 4.1 4.1 4.1 4.1 4.1 4.2  Uncertainty of measurement ± 2.1 °C NOTE: the uncertainty of measurement incorporates the maximum variation in temperature recorder during the test. System performance complies with BS4376:Part1:1991 . Section 5.1