CALIBRATION INFORMATION MODEL:
RSBG1030MD
TYPE:
LABCOLD GLASS DOOR BLOODBANK REFRIGERATOR
TEST PERIOD:
24 hours, readings taken every 15 sec.
AMBIENT TEMPERATURE: 20.0 ± 1.0 °C
AMBIENT HUMIDITY: 60%RH ±5%
DATE OF TEST: 10-05-2007 TEST EQUIPMENT: Two Comark 6800 scanning thermometers each channel calibrated against an Ametek temperature calibrator UKAS Cert No: N1158632T
3
8
2
1 5 4
6 10
7 9
TEST POSITIONS 10 POINTS COTEST10.ECW
01/05/2007
READINGS Position 1 2 3 4 5 6 7 8 9 10
Maximum (ºC) 4.3 4.8 4.6 4.5 4.3 4.6 4.2 4.4 4.3 4.4
Minimum (ºC) 3.3 3.6 3.3 3.0 3.2 3.0 2.2 3.2 2.9 3.1
Average (ºC) 3.8 4.2 3.9 3.7 3.7 3.8 3.1 3.8 3.5 3.7
Uncertainty of measurement ± 2.7 °C NOTE: the uncertainty of measurement incorporates the maximum variation in temperature recorder during the test. System performance complies with BS4376:Part1:1991 . Section 5.1