Lantheus Medical Imaging
VIALMIX for Definity Vial User Guide Aug 2010
User Guide
46 Pages
Preview
Page 1
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 1
V IALMIX
®
ACTIVATION DEVICE FOR DEFINITY ® VIAL FOR (PERFLUTREN LIPID MICROSPHERE) INJECTABLE SUSPENSION
USER’S GUIDE ■■■■■■■■■■■■■■
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 2
Technical Service Reference: If you need technical assistance or have questions about Vialmix®, call Lantheus Medical Imaging Technical Services at 1-800-362-2668. Customer Service Reference: For customer orders, please call Lantheus Medical Imaging Customer Services at 1-800-299-3431.
Caution: The only intended use of Vialmix® is for the activation of Definity Vial for (Perflutren Lipid Microsphere) Injectable Suspension. CAUTION: US Federal law restricts this device to sale by or on the order of a licensed medical practitioner. ®
Distributed in USA by:
Distributed in Korea by:
Lantheus Medical Imaging 331 Treble Cove Road North Billerica, Massachusetts 01862 USA
Boo Kyung S.M. Co., Ltd. Seoul, Korea
Distributed in Singapore by: Distributed in Canada by: QT Instruments Pte. Ltd. Lantheus MI Canada, Inc Montréal, Canada Distributed in Australia by: Distributed in México by: Ralca S. A. de C.V. Edo. de México
Lantheus Medical Imaging Australia Pty. Ltd.
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 3
C O N T E N T S ■■■■■■■■■■■■■■■■
1
Q U I C K - S TA RT I N S T R U C T I O N S
2
INTRODUCTION
3
4
1.1
Safety Precautions
2.1
System Features Overview
2.2
Set-up
2.3
O P E R AT I O N
Powering On/Off
3.1
Loading the Vial
3.2
Activating the Vial
3.3
Removing the Vial
3.4
S A F E T Y F E AT U R E S
Cover Shut-off Switch
4.1
CANCEL Button
4.2
Thermal Sensor Shut-off Switch
4.3
5
L E D A L E RT M E S S A G E S
5.1
6
TROUBLESHOOTING
6.1
7
CLEANING
7.1
8
MAINTENANCE
8.1
9
S P E C I F I C AT I O N S
9.1
10
T E C H N I C A L S E RV I C E
10.1
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 4
Q U I C K S TA RT INSTRUCTIONS ■■■■■■■■■■■■■■■■
NOTE:
S E T- U P
Vialmix will not operate with cover open. ®
Do not open cover while Vialmix® is operating. Wait 7 minutes between activations.
Remove packaging, including red transport collar from base of vial carrier. Place Vialmix® on a flat surface and verify voltage compatibility. ■ Connect Vialmix® to the main power source using enclosed power cord. Non-grounded plug adaptors should never be used. ■ Power on Vialmix® by moving switch to “1.” ■ LED display briefly shows “88,” then “45.” ■ ■
O P E R AT I O N I N S E RT I N G V I A L A N D A C T I VAT I N G D E F I N I T Y ® V I A L F O R
( P E R F LU T R E N L I P I D M I C R O S P H E R E ) I N J E C TA B L E S U S P E N S I O N :
1 Press lever downward and to the left to open the carrier. 2 Hold Definity vial between thumb and fingers and load vial by first placing the top of the vial in the left cup of carrier; position vial between the two cups. ®
3 Release the lever; then close the Vialmix® cover. 4 Press START and countdown from “45” to “00” will begin. R E M OV I N G V I A L :
1 When the full 45-second cycle is complete, open Vialmix® cover. If the cycle i s interr upted fo r any reason, the Defin i t y® v ial is unusable. Retain vial and contact Lantheus Medical Imaging Technical Services. (See pg. 10.1) Keep cover closed when not in use.
1.1
515370-0810
2 Press lever downward and to the left to open the carrier. 3 Remove vial. 4 Close the cover. P O W E R I N G O F F V I A L M I X®
Vialmix® should be powered off at the end of the day by moving the switch on the back of the unit to “0.”
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 5
INTRODUCTION
■■■■■■■■■■■■■■■■
SAFETY PRECAUTIONS
IMPORTA N T: Read this section
■ Before the unit is put into operation, make sure that the operating
carefully before
voltage indicated on the type plate is the same as the supply voltage available. Incorrect voltage may damage the unit.
plugging in and
■ This unit must only be operated with the lid closed and intact. ■ During operation, the Definity® vial must be the only object placed
under the cover.
operating the unit. The unit must be used in strict accordance with these instructions.
■ The unit must be placed on a level surface.
We have no liability
■ A dangerous condition can result from condensation build-up. Before
for any damage
using the unit, allow the unit to reach room temperature, particularly when moving it from a colder to a warmer environment. ■ To safely operate the unit, insert the main plug only into an appropriate
arising from any other use of this unit.
outlet with a grounded conductor. If an extension cord is used, make sure that the grounded conductor is not broken. ■ Turn off and unplug the unit from the power outlet befo re changing fuses. ■ Use only correct fuses. Do not use repaired fuses or bypass the fuse holder. ■ In order to avoid electrical shock, DO NOT INSERT ANY OBJECTS INTO THE UNIT. ■ If, for any reason, it is possible that the safety of a unit has been
compromised, that unit must be re m oved from operation and identified in such a way that it is not inadvertently operated again by a third p a rt y. For example, do not use unit if it does not work as specified or is visibly damaged. ■ Keep solvents, inflammable liquids and powerful heat sources away from unit. ■ Do not allow cleansers to get into unit during cleaning; an electrical
s h o rt or a dangerous malfunction may occur. Do not immerse unit in any liquid. ■ Except for fuse replacement, all repairs should only be done by
Lantheus Medical Imaging approved personnel. Do not open unit housing. 2.1
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 6
INTRODUCTION
■■■■■■■■■■■■■■■■
IMPORTA N T: Shaking the Definity ® vial by hand will not activate the agent adequately. Using Vialmix® to shake Definity ® will ensure
S Y S T E M F E AT U R E S O V E RV I E W
Vialmix® is the activation device designed specifically for Definity® Vial for (Perflutren Lipid Microsphere) Injectable Suspension, an intravenous ultrasound contrast agent. Definity® is supplied as a liquid-filled glass vial and requires shaking in order to create the lipid-encapsulated microbubbles.The shaking rate and duration are controlled by Vialmix® to ensure reproducible activation of Definity®.
proper activation of the agent and consistent product performance.
Cycle Duration: 45 seconds
F E AT U R E S O F V I A L M I X ® I N C L U D E :
■ Proper Activation of Definity®
A number of LED messages inform the user of successful activation. In the event that the activation cycle is stopped short of the 45 seconds of required shaking time, or the shaking rate deviates from the acceptable range, the user is alerted via LED messages and an audible alarm. ■ Simple Operation
Vialmix® has only two operational buttons, one for starting the cycle, and another for canceling the cycle, if necessary.Vialmix® allows for easy loading of the Definity® vial into the vial carrier. Vialmix® has a fixed shaking rate and cycle time. Only one possible activation cycle is available.The cycle duration (45 seconds) is displayed as the default value after turning the power on. ■ Safe Operation
Vialmix® contains features that ensure safe operation.These include: ■ An automatic shut-off switch which is activated if the cover is
opened during operation. ■ A CANCEL button to allow the user to terminate shaking and a
thermal sensor shut-off to prevent the motor from overheating.
2.2
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 7
INTRODUCTION
■■■■■■■■■■■■■■■■
IMPORTA N T:
S E T- U P
Vialmix® is packaged with a power cord and this User’s Guide. Upon receipt of Vialmix®, and prior to use with Definity , the user must perform the following steps. ®
Vialmix® must be at room temperature before use.
1 Remove Vialmix® from the package, and remove the styrofoam inserts. 2 Remove the red transport collar from the base of the vial carrier. 3 Before plugging Vialmix® into an electrical outlet, verify that the voltage on the type plate on the back of Vialmix® corresponds to the local power line voltage. 4 Place Vialmix® on a flat surface. 5 Connect Vialmix® to the main power source using the enclosed power cord.
2.3
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 8
O P E R AT I O N
■■■■■■■■■■■■■■■■
P O W E R I N G O N V I A L M I X®
With Vialmix® properly connected to the main power, the unit can be powe red on by moving the switch on the back of the unit to the “1” position. During power-on, the LED will display “88” briefly.This will change to “45,” representing the cycle time in seconds.The display will remain at “45” until an activation cycle is initiated.
P O W E R I N G O F F V I A L M I X®
Vialmix® should be powered off at the end of the day.This is done by moving the switch on the back of the unit to the “0” position.
3.1
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 9
O P E R AT I O N
■■■■■■■■■■■■■■■■
LOADING THE VIAL
IMPORTA N T: To ensure cleanliness
With the cover open, the vial carrier is easily accessed for vial loading. Complete the following steps to load the Definity® vial:*
of the Definity ® vial septa, load the vial with the flip-top
1 The vial carrier is spring loaded and opens to accommodate the Definity® vial by means of the release lever. Press the lever downward to unlatch the vial carrier; then press the lever to the left to open the carrier for vial loading.
seal in place.
2 Holding the Definity® vial between your thumb and fingers, load the vial by placing the top of the vial in the left cup of the carrier.The cups of the vial carrier are molded to best accommodate the vial when the top of the vial is placed in the left cup. Either position (i.e., the top of the vial in either cup) is possible and will allow for acceptable activation of Definity®. 3 Release the lever and the spring action of the carrier will hold the vial securely. Make sure that the vial is evenly supported by both cups of the carrier. 4 Close the Vialmix® cover.
IMPORTA N T: Do not operate Vialmix® without a vial in the carrier.
*NOTE: Refer to the illustration on the Vialmix® unit, to the left of
the vial carrier for the loading procedure. 3.2
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 10
O P E R AT I O N
■■■■■■■■■■■■■■■■
IMPORTA N T:
A C T I VAT I N G T H E V I A L
Vialmix® will not start the activation cycle with the cover in the open position.
With Vialmix® powe red on, the vial in place, and the cover closed, the user is ready to activate Definity®.The cycle time of 45 seconds appears on the Vialmix® LED display. If it does not appear, see Section 6,“Troubleshooting.” Press the START button to begin the activation cycle.
IMPORTA N T: Unless there is a safety
Definity® is ready for use if:
concern, do not open the Vialmix® cover during the shaking cycle. Doing so will render the Definity ® vial unusable since acceptable activation is
■ Vialmix® shakes the vial for the full 45 second activation cycle. ■ Vialmix® stops shaking when the LED display shows “00” seconds
remaining. If the vial is not shaken for the full duration or at the acceptable rate, activation is not considered successful; an LED message will be displayed and an audible alarm will sound to alert the user. Refer to Section 5 for a list of LED messages and the conditions they indicate.
achieved only with an uninterrupted shaking cycle.
3.3
515370-0810
NOTE: After completion of an activation cycle, allow a minimum of 7 minutes before beginning another cycle.This allows the Vialmix® motor to cool.
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 11
O P E R AT I O N
■■■■■■■■■■■■■■■■
REMOVING THE VIAL
IMPORTA N T: For safety and mainte-
After successful activation of the Definity® vial (as indicated by the LED display of “00”), the vial can be removed for use.To remove the vial, open the cover and follow these steps:*
nance reasons, the cover should be closed when Vialmix® is not in
1 Press the lever down to unlatch the vial carrier, then press the lever to the left to fully open the carrier.
use.
2 Remove the vial from the carrier. 3 Release the lever and the spring action will close the carrier. 4 Close the cover. NOTE: The LED display changes from ”00” to “45” when the cover is opened.
*NOTE: Refer to the illustration to the left of the vial carrier for the
removal procedure. 3.4
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 12
S A F E T Y F E AT U R E S
■■■■■■■■■■■■■■■■
C O V E R S H U T- O F F S W I T C H
Vialmix® is equipped with an automatic shut-off switch that prevents the user from coming into contact with the rapidly moving shaking arm.The shut-off switch is triggered in the following situations: ■ When the cover is open, the shut-off switch will prevent the activation
cycle from beginning. Should the START button be pressed while the cover is open, the motor will not start; consequently, the activation cycle will not begin. ■ If the cover is lifted during operation, the shut-off switch will cause the
activation cycle to stop. The following table shows the LED message for each of the above situations, the associated Vialmix® response, likely cause of the message, and the action required of the user:
4.1
515370-0810
Likely Cause of Message
Action Required by User
Emits a brief audible alarm
Cycle initiation attempted while cover is open
Close the cover and re-start the cycle.
Stops shaking, emits an audible alarm
Opening the cover during the activation cycle
Power off the unit. Do not use the Definity® vial since it was not acceptably activated. Retain the vial and call Technical Serv i c e s. (See pg. 10.1)
LED Message
Associated Vialmix® Response
“CO” for 5 seconds
“EE” alternating with “CO”
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 13
S A F E T Y F E AT U R E S
■■■■■■■■■■■■■■■■
IMPORTA N T:
CANCEL BUTTON
Pressing the
Vialmix® is equipped with a button that allows the user to terminate the activation cycle. By pressing the CANCEL button after the shaking cycle is initiated, the motor shuts down and the shaking cycle stops.
CANCEL button will terminate the activation cycle and render Definity ®
The CANCEL button is available in the unlikely event that the activation cycle must be stopped for safety reasons (for example, should the vial become dislodged or a mechanical failure of Vialmix® occur).
unacceptable for use. Retain the vial and call Technical Services
The following table shows the LED message when the CANCEL button is pressed, the associated Vialmix® response, the likely cause of the message and the action required of the user:
LED Message “EE” alternating with “CL”
Associated Vialmix® Response Stops shaking, emits an audible alarm
Likely Cause of Message CANCEL button was pressed during operation.
for assistance. (See pg 10.1)
Action Required by User Power off the unit. Do not use the Definity® vial since it was not acceptably activated. Retain the vial and call Technical Services. (See pg. 10.1)
4.2
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 14
S A F E T Y F E AT U R E S
■■■■■■■■■■■■■■■■
T H E R M A L S E N S O R S H U T- O F F S W I T C H
Vialmix® is equipped with a sensor that monitors the motor temperature. In the unlikely event that the motor overheats, the sensor triggers the motor to shut down.The LED message (“EE” alternating with “LO”) and an audible alarm alert the user that the motor has overheated.
Under this condition,Vialmix® will not allow a cycle to begin until the motor has cooled. Should the START button be pressed prior to sufficient cooling, the user is alerted by the LED message,“CO,” and an audible alarm.
4.3
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 15
L E D A L E RT MESSAGES ■■■■■■■■■■■■■■■■
In order to ensure acceptable activation of Definity®, Vialmix® is programmed to display LED messages that alert the user in the event that the shaking cycle is outside the required duration or rate.
See table on following page.
The table on the following page shows the LED message, the associated Vialmix® response, the likely cause(s) of the message, and the action required of the user.
5.1
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 16
L E D A L E RT MESSAGES ■■■■■■■■■■■■■■■■
LED Message “LO” alternating with seconds remaining in cycle.
Associated Vialmix® Response
Likely Cause of Message
Continues shaking and counting down to “00”
■ Low line voltage
Stops shaking, emits an audible alarm
■ Mechanical
■ The shaking rate
is approaching the lower limit.
Shaking is acceptable. “EE” alternating with “LO.” The shaking rate is unacceptable since it is below the lower limit.
“EE” alternating with “HI.” The shaking rate is unacceptable since it is above the upper limit.
5.2
515370-0810
failure of Vialmix® affecting the shaking rate
■ Low line voltage ■ The motor has
overheated.
Stops shaking, emits an audible alarm
Mechanical problem with Vialmix®
Action Required by User Definity® vial is ready for use, since activation is acceptable. Contact Technical Services for assistance. (See pg 10.1) Power off the unit. Do not use the Definity® vial, since it was not acceptably activated. Retain the vial and call Technical Services for assistance. (See pg 10.1)
Power off the unit. Do not use the Definity® vial, since it was not acceptably activated. Retain the vial and call Technical Services for assistance. (See pg 10.1)
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 17
TROUBLESHOOTING
■■■■■■■■■■■■■■■■
In the event that Vialmix® does not respond as expected, the guide on the following page is given to aid the user in troubleshooting the problem. If the guide does not cover a particular problem, please call Technical Services for assistance. (See pg 10.1)
IMPORTA N T: Do not attempt to repair Vialmix®. All repairs to Vialmix® must be performed
Do not use the unit if it does not work as specified or is visibly damaged. Contact Technical Services for a Vialmix® replacement. (See pg 10.1)
by Lantheus Medical Imaging approved technical personnel. Contact Technical Services if there are any problems with the unit. (See pg 10.1) Do not open the Vialmix® housing. To avoid electrical shock, do not insert any objects into the main body of Vialmix.®.
6.1
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 18
TROUBLESHOOTING
■■■■■■■■■■■■■■■■
IMPORTA N T: If you need assistance, or have technical questions about Vialmix®, call Technical Services. (See pg 10.1)
6.2
515370-0810
LED Message
Likely Cause
Corrective Action Required
No main power
Check power supply and power cord connection.
Fuse is blown.
Contact Technical Services for assistance,or refer to page 8.1 fo r instructions to change the fuse.
Motor does not operate after START button is pressed.
Cover is open.
Close the cover.
Motor is overheated.
Allow the motor to cool down.
Vialmix® stops shaking during a cycle.
Cover has been opened.
Do not attempt to use Definity® vial since it was not acceptably activated. Retain the vial and call Technical Serv i c e s . (See pg 10.1) Do not open the cover during activation.
Unit deviated from acceptable shaking rate.
Do not use the Definity® vial since it was not properly activated. Retain the vial and call Technical Services. (See pg 10.1)
Fuse is blown.
Do not use the Definity® vial since it was not properly activated. Retain vial and call Technical Services for assistance.(See pg 10.1) See page 8.1 for instructions to change fuse.
Motor is overheated.
A l l ow motor to cool down. Do not use the Definity® vial since it was not pro p e r ly activated. Retain vial and call Technical Services for assistance.(See pg 10.1)
LED display does not operate when power switch is on.
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 19
CLEANING
■■■■■■■■■■■■■■■■
All plastic parts of Vialmix® should be cleaned with a soft cloth and, if necessary, with a mild cleanser. Under no circumstances should any solvents or abrasive detergents be used since these can damage the plastic.
IMPORTA N T: Do not allow cleansers to get into
The use of isopropanol to clean Vialmix is not recommended since over time it may cause the plastic housing to become brittle. ®
Vialmix®
during cleaning; an electrical short or a dangerous malfunction may occur. Do not immerse Vialmix® in any liquid.
7.1
515370-0810
Lantheus_VIALMIX_Final_manual_v2_20Aug10
8/20/10
2:45 PM
Page 20
MAINTENANCE
■■■■■■■■■■■■■■■■
SAFETY PRECAUTIONS:
All bearings in Vialmix® are maintenance-free.Therefore, maintenance of any internal components of the unit is not required.
Prior to changing the fuse, power
CHANGING THE FUSE
off Vialmix® and disconnect the power cord from the back of the unit.
■ Use only the correctly-rated fuse to replace the blown fuse. Refer
to the type plate for correct replacement fuse (as listed on back of Vialmix® or equivalent). ■ Do not use repaired fuses or bypass the fuse holder. ■ To avoid electrical shock, do not insert objects into the main body
of the unit. If the fuse blows, take these steps to change the fuse: 1 Turn the power off. Disconnect the power cord. 2 The fuse is located on the back of the unit next to the power switch. Place a screwdriver behind the small flap of the fuse box and gently pry it open. 3 Slide the fuse box out of the unit. 4 Lift the ratchet upward and pull the fuse clamp out of the fuse box. Remove the fuse and replace it with a new one. 5 Press the fuse clamp back into the fuse box until you hear the ratchet snap into place. Put the fuse box back into the unit.
IMPORTANT: The fuse is the only part of Vialmix® which may be replaced by the user. All other repairs must be performed by Lantheus Medical Imaging approved technical personnel. Contact Technical Services if there are any problems with Vialmix™®. (See pg 10.1)
8.1
515370-0810