Leica Microsystems
M525 F20 and M525 CT20 User Manual
User Manual
56 Pages

Preview
Page 1
Leica M525 F20 Leica M525 CT20 User Manual 10 715 156 – Version -
Thank you for purchasing a Leica Surgical Microscope System. In developing our systems, we have placed great emphasis on simple, self-explanatory operation. Nevertheless, we suggest studying this user manual in detail so that you know and are able to utilize all the benefits of your Leica surgical microscope in an optimum way. For valuable information about Leica Microsystems products and services and the address of your nearest Leica representative, please visit our website, www.leica-microsystems.com. Thank you for choosing our products. We hope that you will enjoy the quality and performance of your new Leica surgical microscope. Leica Microsystems (Schweiz) AG Surgical Division CH-9435 Heerbrugg
Leica M525 / Ref. 10 715 156 / Version -
Chapter overview
Introduction
3
Controls
8
Preparation Before the Operation
12
Operation
15
Accessories
30
Safety Notes
33
Care and Maintenance
41
What To Do If...?
46
Technical Data
48
Leica M525 / Ref. 10 715 156 / Version -
1
Contents Page Introduction User Manual Product identification Symbols in this user manual Design Function Controls Leica M525 surgical microscope with microscope carrier Control/illumination unit Swing arm Footswitch Handswitch CT20 ceiling mount Preparation Before the Operation Transport (F20 only) Preparations before the operation Performing function checks Positioning at the operating table (F20 only) Operation Adjusting the tilt of the microscope carrier (45°/70°) Balancing the Leica M525 surgical microscope Balancing the swing arm Adjusting the articulation brakes Setting position for "posterior fossa sitting" (sitting patient) Switching on the illumination Toggling the illumination Adjusting the field diameter Working with LASER and locking/releasing the multi-focal focus Determining diopter settings for users Setting up eyepieces Attaching the protective glass to the objective Attaching sterile controls Control unit with touch panel Controlling the zoom and focus Accessories 180° assistant’s attachment Binocular tube; can be tilted; with variable viewing angles 30°–150° Stereo attachment for second observer Video-zoom tube Adapter piece for accessories of the Leica M600 series Beam splitter Installing the accessories
2
3 3 3 4 6
8 9 10 10 10 11 12 13 13 14 15 15 17 17 17 18 19 20 20 21 21 21 22 22 28
Page Safety Notes Intended use Directions for the person responsible for the instrument Directions for the operator of the instrument Dangers of use Labels and symbols on the Leica M525 F20 surgical microscope (Part 1) Labels and symbols on the Leica M525 F20 surgical microscope (Part 2) Care and Maintenance Maintenance instructions Changing lamps Changing fuses Function check Re-preparation of re-sterilizable products What To Do If...? Malfunctions on the Leica M525 F20/CT20 surgical microscope Malfunctions on the stand Malfunctions on the TV, photography Technical Data Electrical data Auxiliary power socket Leica M525 F20/CT20 Accessories Floor stand Standards fulfilled Environmental conditions Limitations of use Dimensional drawing of F20 floor stand Dimensional drawing of CT20 ceiling mount
30 30 30 31 31 31 32
Leica M525 / Ref. 10 715 156 / Version -
33 33 33 38 39 40 41 41 43 43 44
46 47 47 48 48 48 49 49 49 49 49 50 51
Introduction
User Manual
Symbols used in this manual
This user manual contains important safety notes as well as information on using the instrument (see the chapter "Safety notes"). Before attempting to set up the product, carefully read through the user manual.
The symbols used in this User Manual have the following meaning:
Warning
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Caution
This indicates a potentially hazardous situation which could result in minor or moderate injury but possibly considerable material, financial or environmental damage.
Product identification The model code and serial number of your product are located on the identification label on the illumination unit. Enter this data in your User Manual and always refer to it when you contact us or the service workshop regarding any questions you may have. Type:
Serial No.:
Danger of burns from intense heat
Information about use that helps the user to employ the product in a technically correct and efficient way.
➩
Leica M525 / Ref. 10 715 156 / Version -
Call to action; here you are asked to do something.
3
Introduction
Design Leica M525 F20 surgical microscope 1
10 2
5
9
8 7 4
3
6
1 2 3 4 5
4
Swing arm Binocular tube Optics carrier Handle Microscope carriers
6 Base 7 Handle 8 Column 9 Control/illumination unit 10 Horizontal arm
Leica M525 / Ref. 10 715 156 / Version -
Introduction
Leica M525 CT20 ceiling mount
8
1
7 5 6
2
4
1 2 3 4 5 6 7 8
3
Swing arm Binocular tube Handle Optics carrier Microscope carrier Control/illumination unit Horizontal arm CT20 ceiling mount
Leica M525 / Ref. 10 715 156 / Version -
5
Introduction
Function
Brakes The Leica M525 surgical microscope has five mechanical brakes that stop the movements of the stand and surgical microscope: • In the rotary joint (1) • In the microscope carrier (2) • At the A/B sledges of the surgical microscope (3) • Horizontal arm (4) • Horizontal arm near column (5)
Balancing system The balancing system is used to Always balance the Leica M525 surgical microscope. All movements which must be carried out during the operation require only minimal force. Two directions of movement, A and B, are balanced on the Leica M525 surgical microscope.
1 2 A
5
B
4
3
Movement C is balanced out on the microscope carrier.
C
6
Leica M525 / Ref. 10 715 156 / Version -
Introduction Tilt of the microscope carrier On the Leica M525 surgical microscope, the microscope carrier can be tilted within a range between 45° and 70°. The 70° position offers the second surgeon more headroom on the 180° dual stereo attachment. The tilt is secured in the 45° or 70° position using screws under the cover (1) (refer also to "Adjusting the tilt of the microscope carrier (45°/70°)", page 15).
Illumination The illumination of the Leica M525 surgical microscope is located in the control/illumination unit (3). It consists of two xenon lamps. If one of the xenon lamps fails, you can toggle to the other xenon lamp using the toggle switch (2).
45 1
2
3
70
Leica M525 / Ref. 10 715 156 / Version -
7
Controls
Leica M525 surgical microscope with microscope carrier 1 2 10 9 3
21
8 15 16 4
7
3
6
5
4
1 Rotary knob for manual balancing of the C sledge 2 C sledge 3 Handle clamping lever 4 Handle 5 Optics carrier 6 B sledge 7 A sledge 8 Push button in – direction (optics carrier moves forward) 9 Push button in + direction (optics carrier moves backward) 10 Microscope carrier
14 11
13
20
19
17
Optics carrier (item 5) 15 Push button for reactivating the automatic adjustment of the field diameter 16 Rotary knob for manually changing the field diameter 17 LED for indicating the multifocal focus locking mechanism 18 Button for locking the multifocal focus 19 Rotary knob for manual adjustment of the working distance 20 Display of the working distance 21 Rotary knob for manual setting of magnification (for emergency operation only)
12
Handles (item 4): 11 Zoom 12 Focus 13 ALL FREE button (no braking function, but can be assigned other functions) 14 Selection button (no braking function, but can be assigned other functions) 8
18
Leica M525 / Ref. 10 715 156 / Version -
Controls
Control/illumination unit
1 1 Display 2 Power switch 3 On/off switch for illumination
Terminals under the control/illumination unit
2
5
6
13
12
7
8
7
3
5 6 7 8
11
10
9
Potential equalization socket Power supply Fuse holder Auxiliary power socket (max. output 100 VA); see Technical Data, page 48, for requirements for use Caution 1 Connecting unauthorized secondary devices to the auxiliary power socket can lead to damage to the Leica M525 surgical microscope and to the secondary device! ➩Never connect secondary devices to the auxiliary power socket unless they conform to the specifications. For requirements of use, see the Technical Data, page 48.
4 4 Toggle switch illuminator 1/2
9 12 V terminal (DC power supply for camera) 10 RS-232 interface (MDRS and service) 11 USB interface (service) 12 CAN-Bus interface 13 Footswitch/handswitch
Leica M525 / Ref. 10 715 156 / Version -
9
Controls
Swing arm
Handswitch
1
6
Z
2
F
5
4
A 1 Stop lever 2 Rotary knob for balancing
4 Automatic focus basic setting 5 Focus (freely assignable; see page 24) 6 Zoom (freely assignable; see page 24)
Footswitch 3
3
4
4
Y-
6
5
6 3 4 5 6
10
5
Adjustable supports Automatic basic focus position Focus (freely assignable; see page 24) Zoom (freely assignable; see page 24)
Leica M525 / Ref. 10 715 156 / Version -
Controls
CT20 ceiling mount
1
3
1 2 3
2
Remote control Ceiling mount down Ceiling mount up
Leica M525 / Ref. 10 715 156 / Version -
11
Preparation Before the Operation
Transport (F20 only)
1
Warning 2 Risk of injury from: • Uncontrolled swinging out of swing arm. • Tilting of the stand. • Feet in lightweight shoes becoming trapped beneath the base. ➩For transportation, always move the Leica M525 F20 surgical microscope into the transport position. ➩Never move the stand while the unit is extended. ➩Never roll over cables lying on the floor. ➩Always push the Leica M525 F20 surgical microscope never pull it. Caution 3 Surgical microscope can move without warning. ➩ Always lock the footbrake when you are not moving the system. Returning the Leica M525 F20 to the transport position ➩Pull the stop lever (1) and rotate it into a vertical position. ➩Grasp the handles. ➩Move the swing arm up and down until the transport lock engages. ➩Engage all brakes. ➩Bring swing arm into transport position. ➩Step on the footbrake release lever (3) to release the footbrakes (2). ➩Move the Leica M525 F20 using the handle.
12
Leica M525 / Ref. 10 715 156 / Version -
3
2
Preparation Before the Operation
Preparations before the operation
Performing function checks
Cleaning the optical accessories ➩Inspect the cleanliness of the eyepieces and objective as well as any photo or TV adapters as appropriate. ➩Remove dust and dirt.
Illumination
Installing the accessories
➩Switch on the illumination. ➩Using the toggle switch for illuminator 1/2 (4, page 9), check both xenon lamps for proper function.
Warning 4 Danger of fatal electrical shock! ➩The Leica M525 surgical microscope may be connected to a grounded safety socket only. Warning 5 Danger of tilting if load exceeds permitted load! ➩Load placed on the microscope interface must not exceed max. 6.5 kg. ➩Connect power cable. ➩Switch on the power switch. ➩Engage the footbrakes (see page 12). ➩Equip Leica M525 surgical microscope with accessories for use. ➩Position handles as needed for the forthcoming operation. ➩Connect the footswitch provided and check the settings. Balancing the Leica M525 surgical microscope ➩Always balance the Leica M525 (see page 15). Setting up eyepieces ➩Adjust the eyebase and interpupillary distance (see page 21). ➩Set diopter values for users (see page 21).
Always allow the illuminator to remain lit for at least five minutes, as otherwise the light output will be reduced quickly.
Footswitch/handswitch ➩Test all functions with footswitch/handswitch. Handle ➩Check zoom and focus functions. TV camera/monitor and still-photo camera (if present) ➩Check the image on the TV monitor. ➩Align TV and SLR camera to the microscopic image. ➩For photographs: Insert a daylight film. Sterility ➩Attach sterile components and sterile drapes. ➩Release brakes and check balancing. Balancing the microscope ➩Release the brakes in the AB sledge and C sledge. The microscope must be balanced.
Configuring control unit Refer to the section entitled "Control unit with touch panel", page 22. ➩Check the settings of the handles (refer to page 25). ➩Inspect all terminals and make certain that all accessories are seated firmly.
Leica M525 / Ref. 10 715 156 / Version -
13
Preparation Before the Operation Raising and lowering the CT20 ceiling mount The CT20 ceiling mount can be raised and lowered electrically. These functions can be controlled via buttons (1) on the remote control unit. Move telescopic arm to the desired height: ➩Press the button (1). 쒄 Raise the telescope arm 쒃 Lower the telescope arm
Positioning at the operating table (F20 only) Warning 6 Danger of injury due to the Leica M525 surgical microscope moving down. ➩Never change the accessories or attempt to rebalance the microscope while it is above the field of operation. ➩Always balance the Leica M525 after re-equipping it. ➩Release the brakes only when the instrument is in a balanced state. ➩Move the Leica M525 F20 surgical microscope to the transport position and release footbrakes (see page 12). ➩Move the Leica M525 F20 surgical microscope carefully over to the operating table by the handle and into the required position for the operation. ➩Set footbrake. ➩Position and connect the footswitch. ➩Connect the potential equalization to the control/illumination unit. ➩Inspect all terminals and make certain that all accessories are seated firmly. ➩Pack footswitch in a plastic bag if necessary. ➩Position stand so that pivot point on swing arm carrier is located over the operation area.
1
Under permanent-load conditions, the telescope may not be operated for more than 1 minute in a 10 minute period. After 2 minutes of uninterrupted operation, the built-in temperature switch deactivates the motor of the Leica ceiling mount.
14
Leica M525 / Ref. 10 715 156 / Version -
Operation
Adjusting the tilt of the microscope carrier (45°/70°)
Balancing the Leica M525 surgical microscope Warning 4 Danger of fatal electrical shock! ➩The Leica M525 surgical microscope may be connected to a grounded safety socket only. ➩Connect device to power supply.
1
Installing the accessories Warning 6 Danger of injury due to the Leica M525 surgical microscope moving down. ➩Never change the accessories or attempt to rebalance the microscope while it is above the field of operation. ➩Always balance the Leica M525 after re-equipping it. ➩Do not release the brakes when the instrument is in an unbalanced state.
➩Unscrew the cover (1). ➩Unscrew the 2 screws (2) and set the desired tilt.
Warning 5 Danger of tilting if load exceeds permitted load! ➩Load placed on the microscope interface must not exceed max. 6.5 kg. ➩Install the necessary accessories on the Leica M525 surgical microscope (see page 30). ➩Switch on the main switch (2, page 9) on the control unit. LED in power switch lights up. 2
➩Retighten the 2 screws (2). ➩Screw on the cover (1).
Leica M525 / Ref. 10 715 156 / Version -
15
Operation Balancing the Leica M525 surgical microscope ➩Move the tubes (binocular tube, attachment for second observer, videozoom tube) to the operating position. ➩Position the Leica M525 surgical microscope horizontally.
1 2
The Leica M525 surgical microscope tilts forward: ➩Push the push button in the + direction (1) until the Leica M525 surgical microscope no longer moves with the brake released. The Leica M525 surgical microscope tilts backward: ➩Push the push button in the – direction (2) until the Leica M525 surgical microscope no longer moves with the brake released. ➩Tilt the Leica M525 surgical microscope by 90°. ➩Tighten the brake knob (3). ➩Position the Leica M525 surgical microscope horizontally back to the working position. ➩Loosen brake knob (4). The surgical microscope falls toward the right: ➩Fold out the rotary knob (5) and turn it clockwise until the microscope carrier is balanced. ➩Fold the rotary knob (5) back in. The surgical microscope falls toward the left: ➩Fold out the rotary knob (5) and turn it counterclockwise until the microscope carrier is balanced. ➩Fold the rotary knob (5) back in.
3
➩Loosen the brake knob (3). The Leica M525 surgical microscope tilts forward: ➩Push the push button in the + direction (1) until the Leica M525 surgical microscope no longer moves with the brake released. The Leica M525 surgical microscope tilts backward: ➩Push the push button in the – direction (2) until the Leica5M525 surgical microscope no longer moves with the brake released. ➩Tilt the Leica M525 surgical microscope by 90°.
1 2
3
16
Leica M525 / Ref. 10 715 156 / Version -
4
5
Operation
Balancing the swing arm ➩Hold the microscope tightly. ➩Pull the stop lever (1) and rotate it into a horizontal position. The transport lock is released.
Setting position for "posterior fossa sitting" (sitting patient) ➩Align the surgical microscope in accordance with the figure below.
1
2
➩Test whether the microscope moves by itself (drifts). Microscope drifts downwards: ➩Adjust the rotary knob (2) clockwise until the microscope no longer moves downwards on its own. Microscope drifts upwards: ➩Adjust the rotary knob (2) counterclockwise until the microscope no longer moves upwards on its own.
Adjusting the articulation brakes 3 3
3 3
3
All the joints on the stand are equipped with articulation brakes. The resistance adjusts to make the joint easier or more difficult to move. To make the joint easier to move: ➩Loosen black brake knob (3). To make the joint more difficult to move: ➩Tighten the black brake knob (3). Leica M525 / Ref. 10 715 156 / Version -
17
Operation
Switching on illumination Warning 7 At a short focal length, the light source of the illumination unit may possibly be too bright for the operating physician and the patient! ➩Begin with the lower-intensity light source and slowly increase it until the operating physician has an optimally illuminated image. Always allow the illuminator to remain lit for at least five minutes, as otherwise the light output will be reduced quickly.
1
2
3
On the handswitch/footswitch/CAN handle: Depending on assignment (see page 24/25), the brightness of the main illumination can also be increased or decreased using two correspondingly assigned keys on the handswitch/footswitch/handle. BrightCare™ On the touch panel monitor, you can enable or disable the BrightCare™ function using the button (2). BrightCare™ is a safety function that automatically limits the maximum brightness depending on the working distance. This is because light that is too bright, in conjunction with a short working distance, can lead to burns on the patient. Explanation of light quantity: The optics of the Leica M525 F20/CT20 surgical microscope have a variable working distance of between 207 and 470 mm. The system is designed in such a way that it delivers sufficient light to produce a bright image even at a long working distance of 470 mm. According to the formula Ev=Iv/d2, the light quantity continually increases by 510 % when the working distance is changed from 470 to 207 mm. (Ev = light intensity, Iv = brightness, d = distance from the light source) This means that less light is required to work with the microscope at a shorter distance than at a greater distance. As a result, it is practical to begin with the light source at a lower intensity and then increase it until you have achieved an optimally illuminated image. When the microscope is shipped from the factory, the "BrightCare™" safety function is activated for all users.
➩Switch on main illuminator at the switch (3). Adjusting the brightness Brightness can be increased or decreased using the touch panel monitor, a handswitch/footswitch or a CAN handle. On the touch panel monitor: ➩Press the "+" or "–" button on the bar for adjusting the (1) brightness of the illumination. or ➩Press the brightness adjustment bar (1) directly. The brightness of the active main illuminator changes.
Explanation of heat development: Heat from invisible light (above 700 nm) is filtered out of the light of the xenon light source used. Nevertheless, white light also always develops heat. An excessive amount of white light can lead to overheating of tissue and metal objects. As a result, it is practical to begin with the light source at a lower intensity and then increase it until you have achieved an optimally illuminated image.
Clicking the "+" or "–" key changes the brightness value in increments of one. Holding down the mouse button with your finger changes the value in increments of five. The start value can be saved individually for each user. Even when switched off, the brightness setting can still be read. However, the brightness display bar appears darker.
18
Leica M525 / Ref. 10 715 156 / Version -