Leica Microsystems
M620 F20 User Manual 2013
User Manual
56 Pages

Preview
Page 1
Medical Division
Leica M620 F20 User Manual 10 714 371 - Version 00
Thank you for purchasing a Leica surgical microscope system. In developing our systems, we have placed great emphasis on simple, self-explanatory operation. Nevertheless, we suggest studying this user manual in detail so that you know and are able to utilize all the benefits of your Leica surgical microscope in an optimum way. For valuable information about Leica Microsystems products and services and the address of your nearest Leica representative, please visit our website, www.leica-microsystems.com Thank you for choosing our products. We hope that you will enjoy the quality and performance of your new Leica Microsystems surgical microscope. Leica Microsystems (Schweiz) AG Medical Division Max Schmidheiny-Strasse 201 CH-9435 Heerbrugg Phone: +41 71 726 3333 Fax: +41 71 726 3334
II
Leica M620 F20 / Ref. 10 714 371 / Version 00
Chapter overview
Introduction
3
Controls
5
Preparation Before Surgery
8
Use
18
Accessories
23
Safety Notes
31
Care and Maintenance
40
Disposal
44
What To Do If...?
45
Technical Data
47
Leica M620 F20 / Ref. 10 714 371 / Version 00
1
Contents Page
Page
Introduction User manual...3 Product identification...3 Symbols in this user manual...3 Design and function...4
Safety Notes Intended use of instrument... 31 Information for the person responsible for the instrument... 31 User qualifications... 31 Information for the user... 31 Manufacturer's declaration of electromagnetic compatibility (EMC)... 33 Table 2 from EN 60601-1-2:2007... 34 Table 4 from EN 60601-1-2:2007... 35 Table 6 from EN 60601-1-2:2007... 36 Dangers of use... 37 Signage... 39
Controls Control unit...5 Lamp housing...5 Tilt head/focus unit...5 Footswitch (standard configuration)...6 Stand...6 Optics carrier...7 Binocular tube, eyepiece, second observer tubes...7 User interface of the control panel...7 Preparation Before Surgery Changing accessories of the surgical microscope and balancing the swing arm...8 Installing binocular tube, eyepiece and objective...8 Fitting adapters for accessories...10 Adjusting the second-observer tube...10 Fitting documentation accessories...11 Selecting documentation accessories...12 Adjusting the eyebase and eyepoint...13 Adjusting the parfocality...13 Show/switch footswitch type...13 Reversing + and – directions of the XY unit...14 Transporting the surgical microscope...14 Positioning the surgical microscope at the operating table...15 Installing sterile components...16 Checking the function of the lamp...16 Preparing the surgical microscope for use...17 Use
Positioning the microscope...18 Adjusting the focus...18 Adjusting the magnification...18 Adjusting the illumination...19 Switching from main to spare illumination...19 Setting the illumination type and working distance... 20 Operating the control panel... 20 Decomissioning... 22
Care and Maintenance Care instructions... 40 Cleaning the control panel... 40 Maintenance... 40 Care and Maintenance of the Leica Footswitch... 40 Changing the fuse...41 Changing the bulb...41 Function check...41 Notes on reprocessing of resterilizable products... 42 Instructions... 42 Maintenance... 43 Disposal Disposal... 44 What To Do If...? General malfunctions... 45 Microscope... 45 TV, photography... 46 Specifications Electrical data... 47 Microscope... 47 Optical data... 47 Control unit ... 47 Stand... 48 Ambient conditions... 48 Standards fulfilled... 48 Limitations of use... 48 Configurations and Weights... 48 Dimensions... 49
Accessories Components and accessories... 23 Video accessories for Leica M620 F20... 28 Load table... 29
2
Leica M620 F20 / Ref. 10 714 371 / Version 00
Introduction
User manual
Symbols in this user manual
This user manual contains important safety notes as well as information on using the instrument (see the chapter "Safety Notes").
The symbols used in this user manual have the following meanings: Warning Indicates an imminently hazardous situation or improper use that could result in serious personal injuries or death.
Read the user manual carefully and thoroughly before placing the product in operation.
Caution Indicates a potentially hazardous situation or improper use which, if not avoided, may result in minor or moderate injury and/or appreciable material, financial and environmental damage.
Product identification The model code and serial number of your product are provided on the nameplate found on the underside of the horizontal arm. Copy this information to the line below so that it is handy in case you have questions for our representatives or service locations. Type: Serial No.:
Information about use that helps the user to employ the product in a technically correct and efficient way. ¸¸
Action required; this symbol indicates that you need to perform a specific action or series of actions.
Leica M620 F20 / Ref. 10 714 371 / Version 00
3
Introduction
Design and function
1 2 3 4 5 6 7
9
8
10
1 Swing arm 2 XY-unit (optional) 3 Tilt head 4 Focus unit 5 Binocular tube 6 Eyepieces 7 Optics carrier 8 Lens 9 Control unit 10 Base
4
Leica M620 F20 / Ref. 10 714 371 / Version 00
Controls
Control unit
Lamp housing
1
1
2
1 Main switch 2 Control Panel
2
1 Lamp replacement cover 2 Slide button for quick-change lamp mount
3 Socket for potential equalization for connecting the Leica M620 F20 to a potential equalization device. This is part of the customer's building installation. Observe the requirements of EN 60601-1 (§ 8.6.7). 4 Power cable strain relief device 5 Power socket 6 Connection for footswitch Here, only foot and hand switches delivered by Leica Microsystems (Schweiz) AG, Medical Division may be connected.
Tilt head/focus unit
1
2
3
Tilt head 1 Tilt head fine adjustment Focus unit 2 Reset button, XY-unit 3 Reset button, focus Leica M620 F20 / Ref. 10 714 371 / Version 00
5
Controls
Footswitch (standard configuration)
Stand
1
3
+ 7
-
1 2
2
1
1
6
5 4
1 2 3 4 5 6 7
3
XY adjustment Microscope illumination darker No Function Microscope illumination on/off Zoom up/down Focus bottom and top Microscope illumination brighter See page 22 for alternative footswitches and free programming of footswitches.
4 5
1 2 3 4 5
6
Articulation brakes Rotary knob for balancing Retaining pin Footbrake release lever Footbrake
Leica M620 F20 / Ref. 10 714 371 / Version 00
Controls
Optics carrier
User interface of the control panel For information on operating the control panel, see page 20. "Main" menu 3
1
4
2
6
5
7
8
1 3
2
9
1 Insert for filter (UV protection filter GG475, protection filter 5×) 2 Jalousie blind rotary knob for fade out/fade in 6° illumination 3 Handles
Binocular tube, eyepiece, second observer tubes 2
1
1 2 3 4 5 6 7 8 9
Illumination lighter/darker Zoom up/down Defective lamp warning XY reversal active "Speed" menu Reset All Reset XY Reset focus User selection 1-4
"Speed" menu
+5
–5 +5
–5 4 1
2
3
5
2
1 Drive knob for interpupillary distance adjustment 2 Diopter adjustment 3 Knurled ring for image correction
1 2 3 4 5
Zoom speed setting Focus speed setting XY-speed setting Back to the "Main" menu Switch to the service area (password-protected)
Leica M620 F20 / Ref. 10 714 371 / Version 00
7
Preparation Before Surgery
Changing accessories of the surgical microscope and balancing the swing arm Warning Risk of injury from downward movement of surgical microscope! ¸¸Never change the accessories or attempt to rebalance the microscope while it is above the field of operation. ¸¸After every modification, balance the swing arm.
– + 3 Balancing the swing arm ¸¸Hold the microscope firmly. ¸¸Pull the retaining pin (2).
The swing arm is now released.
1
¸¸See whether or not the microscope drifts.
Microscope drifts downwards:
¸¸Turn the rotary knob (3) clockwise (+).
Microscope drifts upwards:
¸¸Turn rotary knob (3) counter-clockwise.
2
2
Installing binocular tube, eyepiece and objective A range of options enables the surgical microscope to be matched to the requirements of the task at hand. Installing the binocular tube
Locking the swing arm ¸¸Position the swing arm approximately horizontally. ¸¸Press retaining pin (2) and move the swing arm up and
Warning Risk of injury through falling binocular tube! ¸¸Tighten the clamping screw securely.
down until the retaining pin snaps into place. The swing arm is now locked. Cleaning the optical accessories
¸¸Inspect the cleanliness of the eyepieces and objectives as
well as any photo or TV adapters as appropriate. ¸¸Remove dust and dirt. Mounting accessories
¸¸Outfit the microscope and all accessories for use.
8
1
¸¸Tighten the clamping screw (1). ¸¸Push the binocular tube into the dovetail ring. ¸¸Tighten the clamping screw securely.
Leica M620 F20 / Ref. 10 714 371 / Version 00
Preparation Before Surgery Fitting the eyepiece
Inserting additional filters
¸¸Set the eyepiece in place. ¸¸Tighten the rotary ring (1).
The Leica M620 F20 features two slots for additional filters. Temperature-resistant 32 mm diameter filters may be used. For further information, please contact your Leica Microsystems representative. ¸¸Remove cover (1) of the filter slots. ¸¸Mount the filter in the filter holder with a tension ring.
Eyepieces Eyepiece 8.33×, adjustable Eyepiece 10×/21B, adjustable Eyepiece 12.5×/17B, adjustable Fitting the objective protection glass ¸¸Screw the holder for the objective protection glass (2) into
the objective and position it such that the mark (3) faces backwards. ¸¸Position the objective protection glass (5) and turn it slightly to the right. The objective protection glass will snap into place and the markings (3) and (4) will match.
1
¸¸Insert the filter holder (2).
The objective protection glass is not autoclavable.
2
2
3
4 5
Leica M620 F20 / Ref. 10 714 371 / Version 00
9
Preparation Before Surgery
Fitting adapters for accessories
Adjusting the second-observer tube
Installing the beam splitter/stereo adapter
Install stereo attachment/second observer tube
¸¸Loosen the clamping screw (1). ¸¸Push the beam splitter/stereo adapter into the dovetail ring. ¸¸Tighten the clamping screw.
¸¸Insert the stereo attachment/second observer tube into the
beam splitter.
¸¸Tighten the rotary ring (1).
1
Beam splitter with 50/50 % observation, alternative: beam splitter with 70/30 % observation Stereo attachment for second observer The dual stereo attachment can be attached to the left or right on the beam splitter and turned as required.
Stereo adapter for accessories For installing accessories with M600 interface (dovetail ring) under beam splitter with Leica M500 interface (e.g. "Wild" type laser filter).
Adjusting the second observer tube ¸¸Rotate the monocular second observer tube in the desired
direction.
¸¸Adjust the diopter setting at the eyepiece. ¸¸Correct the image with the knurled ring (2).
Fitting the adapter
Adjusting the stereo attachment for second observer
¸¸Insert the adapter in the beam splitter. ¸¸Tighten the rotary ring (1).
¸¸Turn the attachment for the second observer in the desired
1
direction.
¸¸Align the binocular tube horizontally. ¸¸Adjust the diopter setting at the eyepiece. ¸¸Correct the image with the knurled ring (2).
2
10
Leica M620 F20 / Ref. 10 714 371 / Version 00
Preparation Before Surgery Using the assistant's stereomicroscope Warning Risk of injury through falling surgical microscope! ¸¸Check the mounting of the assistant's stereomicroscope before every use. The assistant's stereomicroscope can only be used with the following objectives: • WD=175 mm • WD=200 mm
Fitting documentation accessories Warning Risk of injury from parts falling down! ¸¸Complete all preparations and adjustments to the optics carrier before the operation. ¸¸Never rebalance or re-equip the optical components and accessories with the instrument over the operation area. ¸¸Check that the optical components and accessories fit well and are fastened securely before the operation. ¸¸Before changing equipment during an operation, swivel the microscope away from the operation area. Adjusting the magnification
1
2
3
Focusing knob
¸¸Mount adapter with included screws (see manual of
assistant's stereomicroscope).
¸¸Focus assistant's image with sterile-operable focus drive (2). ¸¸Moving assistant's stereomicroscope to left/right: ¸¸Remove sterile handles (3) with hub. ¸¸Loosen clamping screw (1) and turn the assistant's
stereomicroscope.
¸¸Tighten the clamping screw (1). ¸¸Turn the hub and reattach sterile handles (3).
Leica M620 F20 / Ref. 10 714 371 / Version 00
11
Preparation Before Surgery
Selecting documentation accessories Zoom Video Adapter
TV adapter
35 mm
55 mm
Photo/TV dual attachment
60 mm
TV adapter
70 mm
Photo/TV dual attachment
85 mm
Zoom Video Adapter
100 mm
1/4”
1/3”
1/2”
2/3”
1”
Photo/TV dual attachment
250 mm
350 mm
35 mm
Field of view
Digital Photo camera
Monitor/image
12
Leica M620 F20 / Ref. 10 714 371 / Version 00
TV adapter
107 mm
Preparation Before Surgery
Adjusting the eyebase and eyepoint Interpupillary distance scale
2. Focus on the test object ¸¸Set the maximum magnification. ¸¸Focus the microscope. ¸¸Set the minimum magnification.
3. Adjusting the diopter settings ¸¸Adjust the diopter setting for each eye in turn, until the
image is seen in sharp focus.
75 65 55
Setting wheel The interpupillary distance and preferred degree of contact with the eyepieces vary from user to user. Prepare the microscope for surgery using a user table. Adjusting the eyebase ¸¸Set the eyelenses of the eyepieces to "0" or select the
required diopter value.
¸¸Set the magnification changer to step 10. ¸¸Look into the eyepieces and adjust the tubes with the
setting wheel or by hand (in the case of binocular tubes without a setting wheel) until a circular field of view is visible.
¸¸Set the maximum magnification. ¸¸Focus the microscope again. ¸¸Set the minimum magnification. ¸¸Inspect the diopter settings, readjusting them if necessary
so that both images are sharp.
4. Checking parfocality ¸¸Zoom through the whole range, observing the test
object. The image sharpness must remain constant at all magnifications. If it does not, then repeat points 2 to 4 of this procedure.
Show/switch footswitch type See page 22.
Adjusting the parfocality Rotary ring for adjusting the diopter settings
Adjusting the diopter settings Adjust the diopter settings accurately for each eye separately; only this method will ensure that the image will stay in focus throughout the entire zoom range (parfocal). 1. Prepare the microscope ¸¸Switch on the microscope at the control unit and place
a flat test object such as a piece of paper beneath the objective. ¸¸Set the maximum brightness. ¸¸Set the minimum magnification. ¸¸Shift the microscope until the test object is visible in the center of the field of view and reasonably sharp.
Leica M620 F20 / Ref. 10 714 371 / Version 00
13
Preparation Before Surgery
Transporting the surgical microscope
Reversing + and – directions of the XY unit The "XY-reverse" function must be assigned to the footswitch (see page 22). ¸¸Use the "XY-reverse" button (1) on the footswitch to reverse the + and – directions of the XY unit to – and +. The "XY-reversal enabled" symbol (2) on the control panel lights up.
-Y
+Y +X
-X
-X
+X
-Y
+Y
Warning Risk of injury from: • Uncontrolled lateral movement of the swing arm. • Tilting of the stand. • Feet in lightweight shoes could become trapped beneath the casing of the base. • Abrupt braking of the surgical microscope at a threshold that cannot be crossed. ¸¸For transportation, always move the Leica M620 F20 surgical microscope into the transport position. ¸¸Never move the stand while the unit is extended. ¸¸Never roll over cables lying on the floor. ¸¸Always push the Leica M620 F20 surgical microscope - never pull it. Transport position Locking the swing arm
1
1
2
¸¸Position the swing arm approximately horizontally. ¸¸Press in the retaining lever (1), rotate it horizontally and
move the swing arm up and down until the retaining lever snaps into place. The swing arm is now locked. ¸¸Release the articulation brakes (2), fold the swing arm. ¸¸Move the surgical microscope into transport position and tighten the articulation brakes.
1
14
Leica M620 F20 / Ref. 10 714 371 / Version 00
Preparation Before Surgery Transport the surgical microscope and secure it in its installation location ¸¸Pull the power plug from the socket and wrap the power
cable around the handle.
¸¸Hook in the footswitch to the clip. ¸¸Step on the footbrake release lever (4) to release the
footbrakes.
¸¸Push the surgical microscope to its new location by the
Positioning the surgical microscope at the operating table Caution Danger of fatal electrical shock! ¸¸The surgical microscope may be connected to a grounded socket only.
handle (3).
¸¸Lock the footbrakes at the installation location.
Caution Surgical microscope can move without warning! ¸¸Always secure the footbrakes when you are not moving the system.
2 2
¸¸Carefully move the surgical microscope to the operating
2
3
table using the handle and position it for the forthcoming operation. ¸¸Tighten the footbrakes. ¸¸Plug the power cable (3) into the socket and secure it with cable ties (2). ¸¸Connect the potential equalization to the socket (1) on the control unit. ¸¸Connect the footswitch to the socket (4) on the control unit. 1 2
4
4
3
4 4
Leica M620 F20 / Ref. 10 714 371 / Version 00
15
Preparation Before Surgery
2
1
1
Installing sterile components Standard components: • One handle, large, black (tilt head) • 2 handles, transparent • 2 drive knobs
1
The handles are steam and gas-sterilizable. ¸¸Sterilize the handles and drive knobs. ¸¸Attach the sterile drive knobs on the binocular tube and the drive knob to block ambient illumination. ¸¸Attach sterile handles to optics carrier and tilt head.
Checking the function of the lamp ¸¸Switch on the microscope using the power switch.
The main lamp will light up.
¸¸Move the slide button for the quick-change lamp mount (3)
to the other side. The auxiliary lamp will light up. ¸¸Move the slide button for the quick-change lamp back to the main lamp. The slide button for the quick-change lamp mount (3) must be slid fully against its stop, otherwise the lamp will not light. ¸¸Adjust the tilt of the footswitch. Rotate the footrest in or out. ¸¸Check that all accessories are correctly connected and
securely attached.
¸¸Release the articulation brakes and set them to a light
resistance.
Make the joint easier to move:
¸¸Release the articulation brake (1).
Make the joint more difficult to move:
¸¸Tighten the articulation brake (1). ¸¸Pull the retaining lever (2). ¸¸Extend the swing arm. ¸¸Check the weight setting on the swing arm by raising and
lowering the microscope and correct as necessary, see page 8.
Caution Risk of infection! ¸¸Leave sufficient space around the stand to prevent contact with non-sterile components.
16
Leica M620 F20 / Ref. 10 714 371 / Version 00
3
Preparation Before Surgery
Preparing the surgical microscope for use ¸¸Switch on the power switch (2).
2 The XY unit will move to its middle position. The LED (page 18, position 2) is lit when the mid position has been reached. The focus will travel to its home position (1/3 up, 2/3 down) at maximum speed. The LED (page 18, position 3) is lit when the focus reset position has been reached. Basic setting: When switching the Leica M620 F20 on with the power switch, the zoom will travel to the same position it was in when the instrument was turned off. The brightness of the lamps and drive speeds remain at the programmed values of the last user.
Leica M620 F20 / Ref. 10 714 371 / Version 00
17
Use Auto reset:
Positioning the microscope
The "auto reset" function can be deactivated on the control panel in the service area.
Adjusting the mid position: ¸¸Press the "Reset XY-unit" key (4).
The XY unit will move to its middle position. LED (2) is lit when the mid position has been reached.
2
3
¸¸Hold both handles and move the microscope to its
uppermost position. All drives will move to their reset positions and the illumination will go out. When the microscope is returned to its working position, the illumination will switch on at the configured starting value.
Adjusting the focus (in accordance with factory settings)
1
Basic setting: ¸¸Press "reset focus" button (5).
4
5
6
Adjusting the focus: ¸¸Press the programmed button (8) or (9) on the footswitch. ¸¸Adjust the speed on the control panel (see page 21).
+ –
9
The focus will travel to its home position (1/3 up, 2/3 down) at maximum speed. LED (3) is lit when the focus reset position has been reached.
7
8
Adjusting the magnification ¸¸Adjust the magnification on the control panel (see page 20).
Coarse positioning ¸¸Hold both handles (6) to position the microscope.
Fine positioning ¸¸Actuate the XY unit via the footswitch (7). ¸¸Adjust the travel speed of the XY unit on the control panel
(see page 21).
Tilt fine adjustment: ¸¸Use the tilt head fine adjustment (1) to set the tilt of the
microscope precisely.
18
Leica M620 F20 / Ref. 10 714 371 / Version 00