PROVEO 8 User Manual Ver 02 Oct 2020
User Manual
84 Pages

Preview
Page 1
MEDICAL DIVISION
PROVEO 8 User Manual 10 733 910 – Version 02
Thank you for purchasing a Leica surgical microscope system. In developing our systems, we have placed great emphasis on simple, selfexplanatory operation. Nevertheless, we suggest studying this user manual in detail in order to utilize all the benefits of your new surgical microscope. For valuable information about Leica Microsystems products and services, and the address of your nearest Leica representative, please visit our website:
Contents 1
Introduction 1.1 About this user manual 1.2 Symbols in this user manual 1.3 Optional product features
2 2 2 2
2
Product identification
2
3
Safety notes 3.1 Intended use 3.2 Directions for the person responsible for the instrument 3.3 Directions for the operator of the instrument 3.4 Dangers of use 3.5 Signs and labels
3 3
4
Design 4.1 F42 floor stand 4.2 CT42 telescope mount 4.3 PROVEO 8 Optics carrier modules
12 12 13 14
5
Functions 5.1 Balancing system 5.2 Brakes 5.3 Illumination 5.4 Fusion Optics
15 15 15 17 17
6
Controls 18 6.1 PROVEO 8 optics carrier including focus, tilt and XY with IVC 18 6.2 PROVEO 8 optics carrier including focus, tilt and XY with IVA 19 6.3 Surgeon panel 20 6.4 Balancing 20 6.5 Control unit 20 6.6 F42 stand 20 6.7 F42 Terminals 21 6.8 CT42 Terminals on the rack controller 22 6.9 CT42 horizontal arm 23 6.10 Footswitch and handles 23
7
Preparation before surgery 7.1 Transportation (F42) 7.2 Installing a binocular tube 7.3 Setting the binocular tube 7.4 Adjusting the eyepiece 7.5 Change the 0° assistant to another position 7.6 Changing the camera 7.7 Balancing and locking the parallelogram 7.8 Positioning on the operating table
www.leica-microsystems.com Thank you for choosing our products. We hope that you will enjoy the quality and performance of your Leica Microsystems surgical microscope. Leica Microsystems (Schweiz) AG Medical Division Max-Schmidheiny-Strasse 201 CH-9435 Heerbrugg Phone: +41 71 726 3333
Legal disclaimer All specifications are subject to change without notice. The information provided by this manual is directly related to the operation of the equipment. Medical decision remains the responsibility of the clinician. Leica Microsystems has made every effort to provide a complete and clear user manual highlighting the key areas of product use. Should additional information regarding the use of the product be required, please contact your local Leica representative. You should never use a medical product of Leica Microsystems without the full understanding of the use and the performance of the product. Liability For our liability, please see our standard sales terms and conditions. Nothing in this disclaimer will limit any of our liabilities in any way that is not permitted under applicable law, or exclude any of our liabilities that may not be excluded under applicable law. II
PROVEO 8 / Ref. 10 733 910 / Version 02
3 4 5 8
26 26 27 28 29 29 30 32 33
7.9 7.10 8
Attaching sterile controls and drape Function check
35 36
Operation 8.1 Switching the microscope on 8.2 Surgeon panel 8.3 Positioning the optics carrier 8.4 Adjusting the optics carrier 8.5 Integrated slit illumination 8.6 Transport position (F42) 8.7 Shutting down the surgical microscope
37 37 38 38 39 41 42 42
9
Control unit with touch panel function 9.1 Menu structure 9.2 Selecting users 9.3 Menu – User Settings
43 43 43 45
10
Menu 10.1 Menu – "Maintenance" 10.2 Menu – "How to..." 10.3 Menu – "Service"
57 57 58 58
11
Accessories 59 11.1 Devices and accessories manufactured by Leica Microsystems (Schweiz) AG 59 11.2 Accessories list for advanced configurations 60 11.3 Drapes 60
12
Care and maintenance 61 12.1 Maintenance instructions 61 12.2 Cleaning the touch panel 61 12.3 Maintenance 61 12.4 Changing fuses 62 12.5 Notes on reprocessing of resterilizable products 62
13
Disposal
65
14
What to do if ...? 14.1 Malfunctions 14.2 Malfunctions documentation accessories 14.2 Error messages on the control unit
65 65 66 66
15
Specifications 15.1 PROVEO 8 15.2 F42 floor stand 15.3 CT42 telescope mount 15.4 CT42 rack controller 15.5 Ambient conditions 15.6 Standards fulfilled 15.7 Limitations of use (only F42) 15.8 Dimensional drawings
68 68 69 69 70 70 70 70 72
16
17
Manufacturer's declaration of electromagnetic compatibility (EMC) 16.1 Table 1 from EN 60601-1-2 16.2 Table 2 from EN 60601-1-2 16.3 Table 4 from EN 60601-1-2
75 75 76 77
Annex 17.1 Checklist before the operation
79 79
PROVEO 8 / Ref. 10 733 910 / Version 02
1
Introduction
1
Introduction
1.1
About this user manual
1.3
In this user manual the surgical microscope PROVEO 8 is described. The present user manual applies to the PROVEO 8 with the following stand variants: F42 and CT42. In addition to notes on the use of the instruments this user manual gives important safety information (see chapter "Safety notes").
X Read this user manual carefully before operating the
product.
The full name of this operation microscope is PROVEO 8.
1.2
Symbols in this user manual
Optional product features
Different product features and accessories are optionally available. The availability varies from country to country and is subject to local regulatory requirements. Please contact your local representative for availability.
2
Product identification
The model and serial numbers of your product are located on the identification label on the stand system closed to the power plug. X Enter this data in your user manual and always refer to it when you contact us or the service workshop regarding any questions you may have. Type
Serial no.
...
...
The symbols used in this user manual have the following meaning: Symbol
Warning word
Meaning
Warning
Indicates a potentially hazardous situation or improper use that could result in serious personal injuries or death.
Caution
Indicates a potentially hazardous situation or improper use which, if not avoided, may result in minor or moderate injury.
Note
Indicates a potentially hazardous situation or improper use which, if not avoided, may result in appreciable material, financial and environmental damage Information about use that helps the user to employ the product in a technically correct and efficient way.
X
Action required; this symbol indicates that you need to perform a specific action or series of actions.
MD
Medical Device
2
PROVEO 8 / Ref. 10 733 910 / Version 02
Safety notes
3
Safety notes
X When instructing new users, do so thoroughly and explain the
The PROVEO 8 surgical microscope is state-of-the-art technology. Nevertheless, hazards can arise during operation. X Always follow the instructions in this user manual, and in particular the safety notes.
3.1
maintenance. Monitor compliance with this. X The PROVEO 8 is intended for professional use only.
WARNING
Intended use
• The PROVEO 8 surgical microscope is an optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment. • The major field of use is Ophtalmology. • The microscope device is not for use in microsurgery (Neuro, Plastic/Reconstructive, Ear, Nose and Throat surgeries). • The PROVEO 8 surgical microscope may be used only in closed rooms and must be placed on a solid floor or mounted to the ceiling. • It is available on the floor stand F42 and on the telescope mount CT42. The F42 is a floor stand for three-dimensional positioning of the PROVEO 8 in the room. The CT42 is a ceiling mounted electrical linear actuator for lifting and lowering the PROVEO 8. It is operated by a hand switch. • The PROVEO 8 surgical microscope is subject to special precautionary measures for electromagnetic compatibility. It must be installed and commissioned in accordance with the guidelines and manufacturer's declarations and recommended safety distances (according to EMC tables based on EN60601‑1‑2). • Portable and mobile as well as stationary RF communications equipment can have a negative effect on the reliability of the PROVEO 8 surgical microscope's functionality. • Always release the brakes to move or relocate the PROVEO 8 surgical microscope. • The essential performance of the PROVEO 8 is to provide illumination and mechanical stability of the optics carrier in any position.
3.2
meanings of the warning signs and messages. X Allocate responsibilities for commissioning, operation and
Directions for the person responsible for the instrument
X Ensure that the PROVEO 8 surgical microscope is used only by
persons qualified to do so. X Ensure that this user manual is always available at the place where the PROVEO 8 surgical microscope is in use. X Carry out regular inspections to make certain that the authorized users are adhering to safety requirements.
Risk of electric shock! X Connect this equipment only to a supply mains with protective earth. X Only use the PROVEO 8 surgical microscope if it is free of
defects. X Inform your Leica representative or Leica Microsystems (Schweiz) AG, Medical Division, 9435 Heerbrugg, Switzerland, immediately about any product defect that could potentially cause injury or harm. X If you use accessories from other manufacturers with the PROVEO 8 surgical microscope, make sure that these manufacturers confirm that the combination is safe to use. Follow the instructions in the user manual for those accessories. • Modifications, installations to or service on the PROVEO 8 surgical microscope may be carried out only by technicians who are explicitly authorized by Leica to do so. For further details refer to the installation manual. • Only original Leica replacement parts may be used in servicing the product. • After service work or technical modifications, the device must be readjusted in accordance with our technical specifications. • If the instrument is modified or serviced by unauthorized persons, is improperly maintained (as long as maintenance was not carried out by a qualified trained service engineer), or is handled improperly, Leica Microsystems will not accept any liability. • The effect of the surgical microscope on other instruments has been tested as specified in EN 60601-1-2. The system passed the emission and immunity test. Comply with the usual pre cautionary and safety measures relating to electromagnetic and other forms of radiation. • The electric installation in the building must conform to the national standard, e.g., current-operated ground leakage protection (fault-current protection) is suggested. • Like any other instrument in the operating theater, this system may fail. Leica Microsystems (Schweiz) AG therefore recommends that a backup system be kept available during the operation.
PROVEO 8 / Ref. 10 733 910 / Version 02
3
Safety notes
3.3
Directions for the operator of the instrument
X Follow the instructions described here. X Follow the instructions given by your employer regarding the
organization of work and safety at work. Phototoxic damage to the retina during eye surgery WARNING Damage to the retina due to prolonged exposure! The light of the instrument may be harmful. Risk of retina damage increases with the duration of exposure. X During exposure to the light from this instrument, do not exceed the hazard reference values. If the exposure time exceeds the value in the table on page 4 with this instrument at maximum output power, the hazard reference value will be exceeded.
The following table is intended to serve as a guideline and make the surgeon aware of the potential hazard. The data have been calculated for the worst-case scenario: • Eye with aphakia • Completely unmoving eye (continuous irradiation of the same region) • Uninterrupted light exposure, e.g. no surgical instruments in the eye • Pupils dilated to 7 mm The calculations are based on the corresponding ISO standard1) and the exposure limit values recommended in that standard. . Main light Light setting
Recommended maximum exposure time according to 1) [min.]
25%
6.5
50 %
2.5
75%
1.5
100 %
1
Retina Protection Function Activated
16.5
Coaxial Red Reflex illumination
4
Light setting
Recommended maximum exposure time according to 1) [min.]
25%
10
50 %
4.5
75%
3
100 %
2
Retina Protection Function Activated
14
Sources: 1) DIN EN ISO 15004-2; Ophthalmic instruments - Fundamental requirements and test methods - Part 2: Light hazard protection. If both illuminations are used, the lower of the two values for the permitted exposure time must be used according to the configured light output. The two hazards do not have to be set off against each other, as their reflections on the retina are not superimposed. This also applies to the use of the Slit Illumination "Phototoxic damage to the retina during eye surgery" on page 4. Slit illumination
The following table shows the allowed surgery durations and their possible extension when reducing the slit width: Slit width [mm]
Time [min.]
>6
1
5
1.2
4
1.5
3
2
2
3
Protect the patient with the following safeguards: • Short exposure times • Low brightness settings • Switching off the illumination during breaks in the operation It is recommended to adjust the brightness to the minimum necessary for the surgery. Infants, patients with aphakia (whose eye lens has not been replaced by an artificial lens with a UV protection screen), small children and persons with diseases of the eye are at greater risk. The risk is also increased if the person being treated or operated on has, within the last 24 hours, already been exposed to illumination from the same or any other ophthalmological instrument that uses a bright visible light source. This applies especially to patients that have been examined via retinal photography. Decisions about brightness must be made case by case. In any event, the surgeon must evaluate the risks and benefits of the used light intensity. Despite all efforts to minimize the risk of retinal injury by surgical microscopes, damage may still occur. Photochemical retinal damage is a possible complication of the necessity to use bright light to make eye structures visible during difficult ophthalmological processes. In addition, the Retina Protection function can be activated during the surgery to reduce the main light intensity below 10% and RedReflex below 20%.
PROVEO 8 / Ref. 10 733 910 / Version 02
Safety notes
3.4
Dangers of use
WARNING
WARNING Risk of electric shock! X Connect this equipment only to a supply mains with protective earth. WARNING Damage to the retina due to prolonged exposure! The light of the instrument may be harmful. Risk of retina damage increases with the duration of exposure. X During exposure to the light from this instrument, do not exceed the hazard reference values. If the exposure time exceeds the value in the table on page 4 with this instrument at maximum output power, the hazard reference value will be exceeded. WARNING Risk of human errors due to scaring noises during surgery! X Don't plug/unplug external devices at the "Line In" input during surgery. X Make sure that the volume level on audio device is set to the minimum before connecting to the "Line In" input. WARNING Danger of injury due to: • uncontrolled lateral movement of the parallelogram • tilting of the stand • feet in lightweight shoes could become trapped beneath the casing of the base X For transportation, always move the PROVEO 8 surgical microscope into the transport position. X Never move the stand while the unit is extended. X Never roll over cables lying on the floor. X Always push the PROVEO 8 surgical microscope; never pull it.
Risk of injury due to downward movement of the surgical microscope! X Complete all preparations and adjustments to the stand before the operation. X Never change the accessories or attempt to rebalance the microscope while it is above the field of operation. X Before changing accessories, always lock the parallelogram (see chapter 7.7.2). X Balance the PROVEO 8 after re-equipping it. X Do not release the brakes when the instrument is in an unbalanced state. X Before re-equipping during the operation, first swing the microscope away from the operating field. WARNING Risk of injury due to downward movement of the surgical microscope! X Never change the accessories or attempt to rebalance the microscope while it is above the field of operation. X After re-equipping, always rebalance the microscope on the parallelogram. WARNING Risk of injury from downward movement of surgical microscope! X Complete all preparations and adjustments to the stand before the operation. X If settings need to be altered during the operation, first swing the microscope away from the operating field. X If the microscope needs to be re-equipped, do this before the operation. X Before re-equipping, always lock the parallelogram. X Do not use the handles or remote brake release when the instrument is in an unbalanced state. WARNING
WARNING Danger to the patient due to failure of the integrated slit illumination! If the integrated slit illumination motor fails, the slit illumination can be manually activated/deactivated using the rotary knob (2).
Risk of injury due to downward movement of the surgical microscope! X Always lock the parallelogram: • when transporting the microscope • when re-equipping
PROVEO 8 / Ref. 10 733 910 / Version 02
5
Safety notes WARNING
CAUTION
Risk of injury from parts falling down! X Before the operation, make sure that the optical components and accessories are sufficiently secured and cannot move.
Damage of the touch panel! X Operate the touch panel using your fingers only. Never use hard, sharp or pointed objects made out of wood, metal or plastic. X Never clean the touch panel using cleaners that contain abrasive substances. These substances can scratch the surface and cause it to be become dull.
WARNING Risk of infection! X Always use the PROVEO 8 surgical with sterile controls and knobs. WARNING Danger of fatal electrical shock! X Operate the system only with all equipment in its proper position (all covers fitted, doors closed). WARNING Danger of fatal electrical shock! X The PROVEO 8 surgical microscope may be connected to a grounded socket only (F42).
Surgical microscope can move without warning! X Always lock the footbrake when you are not moving the system. CAUTION Damage to the PROVEO 8 surgical microscope due to uncontrolled tilting! X Hold the handle when releasing the brake. CAUTION
Motors return to their park position! X Before switching on the microscope, ensure that the travel paths of XY-unit and focus are free of obstructions.
Damage to the PROVEO 8 surgical microscope during transportation! X Never move the stand in the extended condition. X Never roll over cables lying on the floor. X Do not transport or store the system in areas with an elevation angle bigger than 10°.
WARNING
CAUTION
Light that is too intense can damage the retina! X Observe the warning messages in the chapter on "Safety notes".
Damage to the housing of the stand or to the touch panel of the control unit! If the optics carrier is moved into the transport position or from the transport position to the operating position: X Ensure that the transport lock must be locked.
WARNING
WARNING Danger to the patient due to failure of the magnification motor! If the magnification motor fails, the magnification can be manually adjusted using the rotary knob. WARNING Danger to the patient due to downward movement of the CT42 telescope mount! X Do not use the up/down movement when the microscope is over the patient.
6
CAUTION
CAUTION There is a risk of damage to the surgical microscope from uncontrolled tilting! X Firmly hold the handles before triggering the "All Brakes" function. CAUTION Danger of damage! X Before lifting the optics carrier make sure that the area above the parallelogram is clear to avoid collisions with OR lamps, ceiling etc.
PROVEO 8 / Ref. 10 733 910 / Version 02
Safety notes CAUTION Risk of infection! X Leave sufficient space around the stand to ensure that the sterile drape does not come into contact with nonsterile components. CAUTION Risk of overheating of the system! Covering the air inlet can result in a controlled shutdown of the system due to overheating. X Make sure that there is always some space between the air inlet and the drape. CAUTION Damage of the magnification motor! X Only adjust the magnification manually if the magnification motor is defective. CAUTION Danger to the patient due to changes in the user settings! X Never change the configuration settings or edit the user list during an operation. CAUTION Risk of injury! X Pay special attention to the required safety distances if you use the Combination Mode function together with accessories from other manufacturers that can reduce the working distance to less than 140 mm (non-contact wide-angle observation systems), since focus together with Combination Mode is a semi-automated function. CAUTION Danger of collision! The surgical microscope can collide with surrounding components, the ceiling or lamps. X Check the danger area before moving the swing arm. X Carefully move the ceiling mount upwards, and observe ceiling and lamps.
PROVEO 8 / Ref. 10 733 910 / Version 02
7
Safety notes
3.5
Signs and labels
3.5.1
F42 floor stand
13
13
12
11
13
14
10
11 12 13
9
8
8
7
1
2
8
15
3
PROVEO 8 / Ref. 10 733 910 / Version 02
4
5
4
6
6
Safety notes 1
Leica Microsystems (Schweiz) AG Max Schmidheiny-Strasse 201 CH-9435 Heerbrugg MODEL STAND
MD
Type label
2
8
Trained personnel
Utilisation réservée à un personnel formé 9
Equipotential bonding
10
max. 8.0 kg (17.64 lb)
Max. load for Optics Carrier
11
Open
12
Closed
13
Crushing hazard label
Follow the User Manual
Grounding reliability can only be achieved when EQUIPMENT is connected to equivalent receptacle marked "Hospital only" or "Hospital Grade". For US and Canada only
Grounding label
La fiabilité de la mise à la terre n'est assurée que si l'équipement est connecté à une prise équivalente, marquée "Hôpital seulement" ou "Qualité hôpital".
5
Only to be operated by trained personnel
Prefix number Leica system article no. Serial number Incremental number starting at 1 for each batch JJ = year (2 ditis) MM = month (2 digits) TT = day (2 digits) Production starting date
3
4
System weight label
Medical device Fabrication label
e f g h
7
50 - 60 Hz
MD
a b c d
Light risk label
PROVEO 8 F42 (OCT Ready) 100 – 240 V 600 VA
6
14
15
MET label
max. 16 kg (35.27 lb)
Monitor weight label
UDI Label
Device Identifier (DI) GS1 Data Matrix Code Production Identifier (PI) Serial number Date of manufacturing
PROVEO 8 / Ref. 10 733 910 / Version 02
9
Safety notes
3.5.2
CT42 telescope mount
A
B
10
PROVEO 8 / Ref. 10 733 910 / Version 02
Safety notes 1
Leica Microsystems (Schweiz) AG Max Schmidheiny-Strasse 201 CH-9435 Heerbrugg
MODEL
PROVEO 8
STAND
C42/CT42
MODULE
Rack Controller
Type label
MD
6
INMETRO label
7
Equipotential bonding
100 –240 V 50 - 60 Hz 500 VA
PRIM
2x T6.3 AH 250 V
2
Fabrication label
a b c d e f g h
9
Open
10
Closed
11
Crushing hazard label
12
Follow the User Manual
max. 16 kg (35.27 lb)
13
Monitor weight label SKF type label
Grounding label
La fiabilité de la mise à la terre n'est assurée que si l'équipement est connecté à une prise équivalente, marquée "Hôpital seulement" ou "Qualité hôpital".
5
Max. load for optics carrier
Prefix number Leica system article no. Serial number Incremental number starting at 1 for each batch JJ = year (2 digits) MM = month (2 digits) TT = day (2 digits) Production starting date
Grounding reliability can only be achieved when EQUIPMENT is connected to equivalent receptacle marked "Hospital only" or "Hospital Grade". For US and Canada only
Max. 8.0 kg (17.64 lb)
3
4
8
14
MODEL PROVEO 8 STAND C42 / CT42 MODULE Swing Arm
15
UDI Label
Type label
MET label
Device Identifier (DI) GS1 Data Matrix Code Production Identifier (PI) Serial number Date of manufacturing
PROVEO 8 / Ref. 10 733 910 / Version 02
11
Design
4
Design
4.1
F42 floor stand
1 PROVEO 8 optics carrier 2 Parallelogram 3 Video monitor (optional) 4 Control unit with touch panel 5 Camera and video control units (optional)) 6 Footbrake 7 Base 8 Terminals (e.g. electrical, video, etc.) 9 Handrail 10 Monitor arm (optional) 11 Speaker With its open architecture the PROVEO 8 provides space for holding the camera and video control units.
12
PROVEO 8 / Ref. 10 733 910 / Version 02
Design
4.2
CT42 telescope mount
1 2 3 4 5 6 7 8 9
PROVEO 8 optics carrier Parallelogram Monitor arm (optional) Video monitor (optional) CT42 telescope mount Horizontal arm Remote control telescope Up Down
4.2.1
CT42 rack controller
7
1 2 3 4 5 6
8
Mains Holder for footswitch Video control units and recording systems (optional) Control unit with touch panel Wireless foot switch Wall plate
7 Terminals (electical, video etc.) 8 Rack Compartment Set (optional)
PROVEO 8 / Ref. 10 733 910 / Version 02
13
Design
4.3
PROVEO 8 Optics carrier modules
4.3.1
PROVEO 8 with IVC module
4.3.3
PROVEO 8 with IVA module
• Optics carrier with C-mount interface, without integrated inverters
• Optics carrier with integrated camera and integrated inverters
4.3.2
PROVEO 8 with IVC module without inverter
• Optics carrier with integrated camera, without integrated inverters
14
PROVEO 8 / Ref. 10 733 910 / Version 02
Functions
5
Functions
5.1
Balancing system
5.2
With a balanced surgical microscope PROVEO 8 you can move the optics carrier to any position without tilting or moving down. After balancing all movements during operation only need a minor force. The parallelogramm balances the up/down movement. Balancing the parallelogram see chapter 7.7.1.
Brakes
The PROVEO 8 surgical microscope has 4 electromagnetic brakes which stop the movements of the stand, the ceiling mount and the surgical microscope: The electromagnetic brakes can be released via handle or nurse switch.
5.2.1
Brakes at the F42 floor stand
Balancing the optics carrier at the F42 floor stand
1
Balancing the optics carrier at the CT42 telescope mount
1 2 3 4
Rotation of the tower Rotation of parallelogram Up/Down movement of parallelogram Rotation of microscope carrier
PROVEO 8 / Ref. 10 733 910 / Version 02
15
Functions
5.2.2
Selected brakes at the F42 floor stand
5.2.3
Brakes at the CT42 telescope mount
With the function "Selected brakes" the user can release any individual brake.
This function can be activated by a qualified person only.
1 2 3 4
Rotation of horizontal arm Rotation of parallelogram Up/Down movement of parallelogram Rotation of microscope carrier
5.2.4
Selected Brakes at the CT42 telescope mount
With the function "Selected brakes" the user can release any individual brake. These function can be activated by a qualified person only.
1 2 3 4
Rotation of tower Rotation of parallelogram Up/Down movement of parallelogram Rotation of microscope carrier
Don't move the system without released brakes.
1 2 3 4
16
Rotation of horizontal arm Rotation of parallelogram Up/Down movement of parallelogram Rotation of microscope carrier
PROVEO 8 / Ref. 10 733 910 / Version 02
Functions
5.3
Illumination
The illumination of the surgical microscope PROVEO 8 consists of two LED modules and are located in the optics carrier. There are two lamps, main lamp and Red Reflex lamp.
5.4
Fusion Optics
This feature provides an augmentation in resolution and depth of field for an ideal 3-D optical image. Fusion Optics operates with two separate beam paths with different information: the left beam path is optimized for high resolution, the right beam path for optimum depth of field. The human brain merges these two very different images to a single, optimal spatial image.
PROVEO 8 / Ref. 10 733 910 / Version 02
17
Controls
6
Controls
6.1
PROVEO 8 optics carrier including focus, tilt and XY with IVC
19
1 XY Coupling 2 Rotary knob for changing the 0° assistant position 3 Rotary knob "Inverter" (emergency operation only, with version see chapter 4.3.1) 4 Fine focus for integrated camera 5 Rotary knob "Magnification" (emergency operation only) 6 Rotary knob "Slit illumination" (emergency operation only) 7 Slot for filter slide 8 Handle 9 Rotary knob for Red Reflex illumination diameter 10 PROVEO 8 optics carrier
18
18
17
11 Fine focus for assistant 12 Rotary knob "Inverter" assistant (emergency operation only, with version see chapter 4.3.1) 13 0° assistant 14 Surgeon panel 15 2× CAN sockets - only for Leica accessories 16 Cooling slots 17 Objective 18 Mounting threads for accessories 19 Socket for BIOM
PROVEO 8 / Ref. 10 733 910 / Version 02