SERVICE MANUAL VAVD CONTROLLER
Copyright All rights reserved. No part of this publication may be duplicated, adapted or translated without prior written permission, except within the framework of the copyright laws. © Copyright MAQUET Cardiopulmonary AG Subject to technical changes Owing to our policy of continuous product development, the illustrations and technical data contained in this Service Manual may differ slightly from the current version of the device. Manufacturer MAQUET Cardiopulmonary AG Hechinger Straße 38 D-72145 Hirrlingen Germany Phone: +49 (0) 7478 921-0 Fax: +49 (0) 7478 921-100 [email protected] www.maquet.com
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | Table of contents | 3 |
TABLE OF CONTENTS
1
General
|4
2
Service equipment
|5
3
Drawings + BOM
|6
4
Membrane change
| 14
5
Function test after service
| 16
6
Function test report
| 21
7
Spare part list
| 22
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| 4 | English | VAVD CONTROLLER |
1
GENERAL This service manual describes the service/maintenance of the VAVD CONTROLLER.
CAUTION! Assembling of the device have to be done according to the drawings and BOM including in this manual Only original spare parts have to be used After disassembling the described supplies have to be used according to the drawings All parts inside of the device have to be cleaned and disinfect after evacuated from blood After disassembling and assembling the function test have to be performed After service the device needs to be cleaned from the outside
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 5 |
2
SERVICE EQUIPMENT Following special equipment is needed to maintain the VAVD CONTROLLER: Diaphragm vacuum pump Model: Type/S [m³/h): ME 2S / 1,9 Manufacturer: VACCUUBRAND GmbH + Co. 2 x Pressure gauge (calibrated): Model: Pneumatic transducer tester, Model: DPM-BP Manufacturer: Bio-Tek Instruments-Inc. or FLUKE Pressure Reducer connectors for pressure gauge 1/4” x 1/16” tubes tube connectors Bolt with d=20±0,2mm Screw locking: Loctite 221 Lubricant: Berulub OX 40 EP
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| 6 | English | VAVD CONTROLLER |
3
DRAWINGS + BOM
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 7 |
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| 8 | English | VAVD CONTROLLER |
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 9 |
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| 10 | English | VAVD CONTROLLER |
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 11 |
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 13 |
Pos.
Quan. Description
28
1
Circlip
29
1
Zugfeder Standard
30
1
Gewindestift
32
1
Gewindestift
33
1
O-Ring NBR 70 / 9,1x1,6
34
1
Shoulder Bold
35
10
Flachkopfschrauben mit Innensechskant
36
3
Flachkopfschrauben mit Innensechskant
37
1
Cover
38
1
Kasse Bund
39
1
Bagplate
40
2
Vinkel Beslag
41
1
Membran
42
1
CYL. Pakning
43
1
Cylinder
44
1
Stempel
45
1
Vac. Kammer
46
1
Membran støtte
47
1
Vac. Rel. Stempel
48
1
Fjeder Fod
49
1
Fjeder Skrue
50
1
Fjeder aksel
51
1
Bøsning
52
1
ManoStag
53
1
Spender
54
1
Mem. Spend A VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| 14 | English | VAVD CONTROLLER |
Pos.
4
Quan. Description
55
1
Mem. Spend B
56
1
Told
57
1
Told Aksel
58
1
Schalthebel
59
1
Knap
60
2
Klik Kulisse
61
1
Typenschild
MEMBRANE CHANGE The membrane change needs to be performed in regular intervals, regarding to the following instruction.
CAUTION! Only the original Maintenance kit shall be used Assembling of the membrane only according to the pictures below Torques only according to drawings of chapter 2 If no torques described, tightened hand solid Only described supplies shall be used (see service equipment)
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 15 |
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| 16 | English | VAVD CONTROLLER |
5
FUNCTION TEST AFTER SERVICE CAUTION! Function test is required after each single service!
5.1
TEST 1: NEGATIVE PRESSURE METHOD: 1 The vacuum pump is connected to the vacuum connector and the pressure gauge is connected to the reservoir connector of the VAVD controller. 2 The ON/OFF button is set to the ON position. 3 The regulator knob of the VAVD is moved slowly with defined steps (-10, -20, -40, -60, -80 mmHg) to the maximum of –100 mmHg until the defined value is reached on the manometer of the VAVD controller. 4 For each defined step and the opening value of the negative pressure relieve valve, the values of the pressure gauge are noted.
SPECIFICATION: 1 The set vacuum value on the pressure gauge has to be reached within a tolerance of ±2 mmHg. 2 The negative pressure valve must open at –100 ± 10 mmHg, a humming sound and a oscillating needle in the VAVD’s gauge have to appear. The needle mustn’t move over a value of 110 mmHg, when trying to raise the vacuum. VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 17 |
5.2
TEST 2: POSITIVE PRESSURE METHOD: 1 The pressure source, the pressure reducer and the first pressure gauge are connected to the vacuum connector of the VAVD controller. 2 The second pressure gauge is connected to the reservoir connector of the VAVD controller. 3 The ON/OFF button is set to the ON position. 4 The regulator knob of the VAVD is set up to the maximum. 5 Start raising the incoming pressure with the pressure reducer slowly. 6 Note the pressure of the pressure gauge at vacuum connector, if the pressure gauge at the reservoir connector shows the first value (except 0). A flow needs to be present at the outlet on the backside (holes) of the device.
SPECIFICATION: 1 The positive pressure valve has to open at 3 ± 2 mmHg (value of pressure gauge in the outlet).
5.3
TEST 3: MANOMETER METHOD: 1 The vacuum pump is connected to the vacuum connector and the pressure gauge is connected to the reservoir connector of the VAVD controller. 2 The ON/OFF button is set to the ON position.
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| 18 | English | VAVD CONTROLLER |
3 The regulator knob of the VAVD is moved slowly to –40 mmHg and –80 mmHg on the manometer. 4 Note the value of pressure gauge on the reservoir connector. 5 Turn the regulator knob back to the beginning (0-Position). 6 Note the value of the manometer and the pressure gauge at the outlet.
SPECIFICATION: 1 The value on the pressure gauge has to be within the tolerance specified on the manometer; class I, 1.6 (1.6% of maximum reading): = ± 1,92 mmHg ≈ 2 mmHg. 2 After turning the regulator knob to the beginning (0-Position), the value on the manometer has to be 0 ± 5 mmHg and on the pressure gauge at the reservoir connector has to be 0 ± 3 mmHg.
5.4
TEST 4: ON/OFF SWITCH METHOD: 1 The vacuum pump is connected to the vacuum connector and the pressure gauge is connected to the reservoir connector of the VAVD controller. 2 The ON/OFF button is set to the ON position. 3 The regulator knob of the VAVD is moved slowly to –50 mmHg on the manometer. 4 The ON/OFF switch is set to the OFF position. 5 Note the value at the pressure gauge at the outlet.
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 19 |
6 The ON/OFF switch is set to the ON position. 7 Note the value at the pressure gauge reservoir connector and the manometer.
SPECIFICATION: 1 After switch off, the value at the pressure gauge on the reservoir connector needs to be 0 mmHg. 2 After switch on again, the value at the outlet and the manometer needs to be 50 mmHg ± 5 mmHg, the values of the manometer and the pressure gauge need to identical.
5.5
TEST 5: VISUAL TESTS METHOD: The device is visual checked for generally good condition. 1 cleanness 2 scratches 3 completeness 4 fittings + fixing bolt 5 gaps between enclosure parts 6 consistent finishing and structure 7 right label (type and position) 8 final production test report (for each serial-no.)
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| 20 | English | VAVD CONTROLLER |
SPECIFICATION: 1 device is clean 2 no scratches 3 device is complete 4 fittings + fixing bolt are present 5 no gaps between enclosure parts 6 consistent finishing and structure on all outside surfaces 7 right label on specified position mounted 8 final production test report (for each serial-no.) exists
5.6
TEST 6: HOLDER MECHANISM METHOD: It needs to be checked, if the holder mechanism works with a bolt of 20±0,2mm diameter. 1 Loose the threaded bolt on the backside of the device. 2 Insert the bolt (20±0,2mm diameter ) or the bolt of the original VAVD CONTROLLER holder into the hole on the bottom. 3 Tighten the threaded bolt on the backside of the device. 4 Perform function test. 5 Remove the bolt and tighten the threaded bolt. SPECIFICATION: 1 Bolt (20±0,2mm diameter ) needs be inserted easily. 2 After tighten the threaded bolt, it does not be possible to twist/turn the bolt with diameter 20mm.
VAVD CONTROLLER | Service Manual | 1.0 Copyright MAQUET Cardiopulmonary AG, Hirrlingen
| VAVD CONTROLLER | English | 21 |
6
FUNCTION TEST REPORT