Medartis AG
APTUS Plates, Screws and Instruments Instructions for Use
4 Pages
Preview
Page 1
Instructions for Use for Medartis APTUS Plates, Screws and Instruments Introduction These instructions for use are for a product line of Medartis AG, Hochbergerstrasse 60E, 4057 Basel/Switzerland Phone +41 61 633 34 34, Fax +41 61 633 34 00, www.medartis.com. All instructions provided in this document must be followed.
Notes Regarding the Delivered Goods The individual parts of the system may only be accepted when the manufacturer’s label and packaging are undamaged and unopened at the time of delivery. If this is not the case, the rejected goods must be returned to Medartis AG, Basel/Switzerland or to the relevant Medartis territory consultant or distribution partner within ten working days. Implants are intended for single use only and are not designed to be reused. All components that are delivered NON-STERILE must be appropriately prepared before first use. All packaging must be removed before preparation.
Product Materials APTUS implants, plates and screws, are made of pure titanium (ASTM F67, ISO 5832-2) or titanium alloy (ASTM F136, ISO 5832-3). All of the titanium materials used are biocompatible, corrosion-resistant and non-toxic in a biological environment. K-wires and staples are made of stainless steel (ASTM F138, ASTM F139); instruments are made of stainless steel, PEEK, aluminum, Nitinol or titanium.
Color Coding Concept System Size
Color Code
APTUS 1.2
red
APTUS 1.5
green
APTUS 1.7
turquoise
APTUS 2.0
blue
APTUS 2.2
purple
APTUS 2.3
brown
APTUS 2.5
purple
APTUS 2.8
orange
APTUS 3.0
yellow
APTUS 3.5
green
APTUS 4.0
brown
APTUS 5.0
dark blue
APTUS 7.0
turquoise
Plates, Screws and Blades Special implant plates, screws and blades have their own color: Implant plates gold
Fixation plates
Implant plates blue
TriLock plates (locking)
Implant screws gold
Cortical screws (fixation) and cannulated compression screws
Implant screws blue
TriLock screws (locking) Screws for blade fixation
Implant screws pink
Cancellous screws (fixation)
Implant screws silver
TriLock Express screws (locking) and transfixation screws
Implant screws green
SpeedTip screws (self-drilling)
Implant spiral blades blue
Spiral Blades Proximal Humerus
APTUS Forearm • Treatment of fractures and osteotomies of the radius and ulna shaft APTUS Radial Head • Management of proximal radius fractures and osteotomies APTUS Foot • Fractures, osteotomies and arthrodesis of small bones, in particular of the tarsals, metatarsals and phalanges APTUS Calcaneus • Fractures and osteotomies of the calcaneus APTUS Ankle • Fixation of fractures, osteotomies, malunions and non-unions of the distal tibia and fibula APTUS Cannulated Compression Screws APTUS headed Cannulated Compression Screws • Treatment of fractures, osteotomies and arthrodesis of bones with the appropriate screw size APTUS K-Wire System • The APTUS K-Wire System is intended for use in fixation of bone fractures, for bone reconstruction, and as guide pins for insertion of other implants APTUS Distal Humerus • The APTUS Distal Humerus System is indicated for fractures, osteotomies and non-unions of the distal humerus APTUS Proximal Humerus • The APTUS Proximal Humerus System is indicated for fractures, osteotomies and nonunions of the proximal humerus APTUS Proximal Humerus XL Plates • The APTUS Proximal Humerus XL Plates are indicated for fractures, osteotomies and nonunions of the proximal humerus and fractures extending to the humeral shaft APTUS Clavicle • Treatment of fractures, osteotomies, malunions and non-unions of the clavicle
Contraindications • • • • •
Possible Complications In most cases, potential complications have a clinical source as opposed to arising from the implants/instruments. These include among other things: • Loosening of the implant from insufficient fixation • Hypersensitivity to metal or allergic reactions • Bone necrosis, osteoporosis, insufficient revascularization, bone resorption and poor bone formation that can cause premature loss of fixation • Soft tissue irritation and/or nerve damage through surgical trauma • Early or late infection, both superficial and deep • Elevated fibrotic tissue reaction around the surgical area • Complications in implant removal from improper explantation of the implant In consideration of patient’s clinical condition and medical history, the treating physician shall ensure that the use of APTUS implants can be justified based on a patient-specific benefit/risk assessment.
Warnings and Precautionary Measures • • •
TriLock plates (locking) are marked with the following symbol:
Intended Use
• •
The APTUS fixation systems are used for fractures, osteotomies and arthrodesis of the hand, forearm, shoulder and foot. The APTUS cannulated compression screws are used for bone fractures, osteotomies and arthrodesis.
Indications APTUS Hand • Fractures of the distal, middle and proximal phalanges as well as of the metacarpals • All transverse fractures, spiral fractures, fractures near joints with or without joint involvement, shaft fractures, comminuted fractures, dislocated fractures and ligament/bone avulsions • Arthrodeses in the hand APTUS Radius • Intra- and extra-articular fractures • Correction osteotomies • Radiocarpal fusions (arthrodesis) APTUS 2.0/2.3 Four Corner Fusion Plate • The APTUS 2.0/2.3 Four Corner Fusion Plate, an addition to the APTUS Titanium Fixation System, is designed specifically for fusion of carpal bones including: hamate, capitate, lunate, triquetrum
Pre-existing or suspected infection at or near the implantation site Known allergies and/or hypersensitivity to implant materials Inferior or insufficient bone quality to securely anchor the implant Patients who are incapacitated and/or uncooperative during the treatment phase Growth plates are not to be blocked with plates and screws
•
• •
•
APTUS Wrist Arthrodesis • The APTUS Wrist Arthrodesis Plates are indicated for wrist arthrodesis
•
APTUS Ulna • Management of fractures and osteotomies of the ulna
•
The products may only be used by medical personnel who hold relevant qualifications Medartis, as manufacturer, recommends that the user reads all available documents before first use and contacts other users who have practical experience with this type of treatment Never use products that have been damaged by transport, improper handling in the hospital, or in any other way! All of the implant components are intended for single use and may not be reused under any circumstances Necessary care must be observed for storage and use of the products: Damages (e.g. from improper cutting or bending) to and/or scratches on the instruments/implants can substantially impair the strength of the product and lead to premature breakage Repeatedly bending the plate in opposite directions may cause the plate to break during postoperative treatment All of the system components have been developed and manufactured for a specific purpose and are therefore precisely adapted to each other. The user may not alter any of the components or replace them with an instrument or product from another manufacturer even if the size or shape is similar or exactly corresponds to that of the original product. The use of materials from other manufacturers, structural changes resulting from the use of thirdparty products and/or material impurities, as well as minor deviations or imprecise fit between the implants and instruments, or similar, can represent a risk for the user, patient or third parties The sterilizing cases, instrument trays and implant containers shall not be vigorously shaken or tipped over since the individual components may become damaged or fall out Twist drills and reamers: It is recommended not to exceed a maximum drilling speed of 1’000 revolutions per minute to avoid overheating the bone. With reamers, it is advisable to use a speed of less than 1’000 revolutions per minute, or to use a handle for controlled, manual reaming. Reusable, non-sterile packed twist drills and reamers may only be used for a maximum of ten times. Sterile packed twist drills and reamers are for single use only, and may not be reused under any circumstances. Use the indicated screwdriver for the respective system size. Make sure that the screwdriver/screw head connection is precisely aligned in axial direction. If not, there is a greater risk of damage to the implant and screwdriver blade. When inserting the screw, ensure that a sufficient axial force is used between blade and screw. At the same time, the axial force should be in certain limits in order not to damage the bone structure The APTUS products have not been evaluated for safety and compatibility in the MR environment. The APTUS products have not been tested for heating or migration in the MR environment. Therefore, MR-assisted imaging techniques cannot be recommended Implants can cause artifacts in various imaging procedures such as CT, MR
APTUS-00000101_ Rev. U 2020/05/13 Instructions for Use for Medartis APTUS Plates, Screws and Instruments Medartis, APTUS, MODUS, TriLock, HexaDrive and SpeedTip are registered trademarks of Medartis AG/Medartis Holding AG, 4057 Basel/Switzerland
1 of 4
Multidirectional, Angular Stable TriLock Locking System Correct locking (±15°) of the TriLock screws in the plate: Visual inspection of the screw head projection provides an indicator of correct locking. Correct locking has occurred only when the screw head has locked flush with the plate surface (Fig. 1 and 3). However, if there is still a noticeable protrusion (Fig. 2 and 4), the screw head has not completely entered the plate and reached the locking position. In this case the screw has to be retightened to obtain full penetration and proper locking. In case of poor bone quality a slight axial pressure might be necessary to achieve proper locking. Due to the system characteristics, a screw head protrusion of around 0.2 mm exists when using plates with 1.0 mm thickness. Do not overtighten the screw, otherwise the locking function cannot be guaranteed anymore. Correct: LOCKED
Incorrect: UNLOCKED
Fig. 1
Fig. 2
Correct: LOCKED
Incorrect: UNLOCKED
Fig. 3
Fig. 4
Instructions for Selecting the Appropriate APTUS Products Medartis, as manufacturer, does not recommend a specific surgical procedure for a specific patient. The operating surgeon is solely responsible for choosing the appropriate implant for the specific case. The follow-up treatment as well as the decision of whether to retain or explant the implant is the responsibility of the user. The treating physician should beforehand become thoroughly familiarized with the procedure, for example by: • Carefully studying all the product documentation • Carefully reviewing the current professional literature • Consulting with colleagues experienced in this field and with the use of this system • Practice in handling the system, practice of the surgical procedure and postoperative treatment Generally, implants are designed to remain in the body temporarily and be removed after sufficient (osseous) healing has taken place. They are not designed for long term bone replacement. Where they are mechanically supporting the osteosynthesis, the regular operating period of the implants is expected to be between 30 days and 6 months. In consideration of the individual fracture situation as well as the compliance of the patient, the surgeon shall ensure an adequate postoperative relief of the osteosynthesis in terms of adaption- or mobilization stability (e.g. splinting and/or immobilization). Postoperatively, the fixation achieved by the implants must be treated carefully until osseous healing is completed. The doctor’s aftercare instructions have to be strictly observed by the patient in order to avoid adverse loads of the implants. Early load bearing can increase the risk of loosening, migration or breakage of the devices. In the case of complications, it might be necessary to remove the implants. For removal use the indicated screwdriver. Make sure that the screwdriver/screw head connection is precisely aligned in axial direction.
Additional Information Additional information on the products (e.g. the surgical technique, handling instructions for sterile plates, screws and instruments, care, cleaning, disinfection and sterilization of non-sterile products) can be requested from your local Medartis Territory Consultant or distribution partner. In addition, all relevant information can be found on the internet at www.medartis.com.
Single-Use Device
The product is intended for one single application in a single patient. Application of an already used device may compromise the structural integrity of the device and/or lead to device failure, which may result in patient injury.
Sterile Packaged Products
The product has been subjected to a validated irradiation sterilization process and is supplied in sterile packaging. Prior to use, check the product expiration date and verify the integrity of the sterile packaging. Do not use any product where the sterile packaging has been opened or damaged and do not remove them from the packaging until immediately before use. Once the sterile packaging has been opened, the product cannot be resterilized. Sterility of the device must be ensured at all times. The device is for single-use only and may not be reused under any circumstances. Reuse or reprocessing (e.g. cleaning and resterilization) may compromise the structural integrity of the device and/or lead to device failure, which may result in patient injury. Sterile packaged twist drills and reamers: Single-use, sterile packaged cutting tools such as twist drills and reamers are intended for single-use and must not be reused. Reuse or reprocessing (e.g. cleaning and resterilization) may compromise the structural integrity of the device and/or lead to device failure, which may result in patient injury. All single-use cutting tools have to be discarded after the operation following the local requirements.
Non-Sterile Products
Instructions Regarding Cleaning, Disinfection and Sterilization of Non-Sterile Products All implants, instruments and containers in the APTUS systems that are delivered NONSTERILE must be cleaned, disinfected and sterilized before each use. This also applies to the first use after delivery (after removal of the protective transport packaging). Thorough cleaning and disinfection are essential for effective sterilization. All implant components are intended for one single application in a single patient. Implants that were used in a patient and removed, have to be discarded following the local requirements. Application of an already used device may compromise the structural integrity of the implants and/or lead to device failure which may result in patient injury. Furthermore, application of an implant that has already been used may create a risk of contamination e.g. due to the transmission of infectious material from one patient to another. This could result in injury of the patient or user. Implants that have not come into direct contact with a patient may be reprocessed. Implants that have come into direct contact with blood or other bodily fluids or show visual contamination must be cleaned and disinfected separately before they can be placed back into the implant tray. It is your responsibility to ensure that the implants and instruments are completely sterile when used, to use device- and product-specific procedures for cleaning/disinfection and sterilization that are sufficiently validated, to regularly service and inspect the employed devices (disinfector, sterilizer), and to ensure that the validated and/or manufacturer's recommended parameters are maintained for each cycle. The statutory regulations applicable in your country and the hospital's hygiene requirements must also be observed. This applies in particular to the various instructions for effectively deactivating prions.
Basic Instructions If possible, use an automated procedure (disinfector) for cleaning and disinfecting. Do not use a manual procedure even with an ultrasonic bath due to the significantly reduced efficiency and potential damage. Pretreatment is required in both cases. Choosing Detergents, Disinfectants and Equipment Observe the following aspects when choosing detergents, disinfectants and equipment for all steps: • They must be suitable for their intended use (e.g. cleaning, disinfection or ultrasonic cleaning) • The detergents and disinfectants must be aldehyde-free (otherwise blood residues may dry and attach firmly to surfaces) • The disinfectant used must have a proven effectiveness (such as approval by VAH/DGHM or a CE mark) • The detergents and disinfectants must be suitable and compatible for use with the products • The manufacturers’ instructions, such as those regarding concentration, exposure time and temperature, must be followed For cleaning materials and accessories, both for precleaning and manual cleaning, observe the following: • Use only clean, lint-free cloths (e.g. Peform classic from Schülke & Mayr) and/or soft brushes (e.g. Justman Brush from VWR International). Never use metal brushes or steel wool. • When necessary, use materials and accessories such as cleaning stylets, syringes, cannulas and bottle brushes for cannulated products or products with a lumen For drying accessories, Medartis recommends lint-free disposable wipes (e.g. Perform classic from Schülke & Mayr) or medical compressed air. For water quality, Medartis recommends that demineralized and purified water (e.g. Aqua purificata) is used for cleaning, disinfection and subsequent rinsing steps. Medartis instrument trays (steel or plastic) and implant trays made from aluminum or plastic are intended for the sterilization, transportation and storage of products. They are not intended for cleaning and disinfection when loaded. The products must be removed from the trays and then cleaned and disinfected separately. Implant trays made of steel can undergo automated cleaning and disinfection when loaded. For manual cleaning, the implants must be removed from the system and then cleaned and disinfected separately. Remove major contaminants in the operating room before segregating dirty instruments. Preferably use dry preparation for the transportation to the cleaning/sterilization department. If a wet preparation method is used, place the instruments in a prepared solution directly after usage. The instruments must be disassembled and opened as much as possible. All products (including grooves, holes, lumens, etc.) must be sufficiently covered with solution. To avoid damage to the materials, do not leave them in the solution for longer than directed. Pretreatment Prior to Cleaning, Disinfection and Sterilization Pretreatment Process • Disassemble and open the instruments as far as possible. When doing so, follow the assembly and disassembly instructions, which can be found at www.medartis.com • Empty the instrument trays completely and remove the lid if necessary • Empty the aluminum or plastic implant trays completely and remove the lid if necessary; • For steel implant trays, the implants can be left in the tray but the lid must be removed during the rinsing process and rinsed separately • Clean products and individual parts under running water using soft brushes (shift moveable parts back and forth, use cleaning wire, syringes and cannulas for cannulated products; for larger lumina, use a bottle brush if necessary) • Visually inspect the products and repeat pretreatment as required until visible contamination is no longer evident The disassembled instruments and trays should remain dismantled for the following cleaning and disinfection process.
APTUS-00000101_ Rev. U 2020/05/13 Instructions for Use for Medartis APTUS Plates, Screws and Instruments Medartis, APTUS, MODUS, TriLock, HexaDrive and SpeedTip are registered trademarks of Medartis AG/Medartis Holding AG, 4057 Basel/Switzerland
2 of 4
Manual Cleaning and Disinfection For manual cleaning and disinfection, the trays have to be empty. Instruments and trays must be opened and disassembled as far as possible. Implants must be removed from the system and must be cleaned and disinfected separately. Manual Cleaning Process • Place the (disassembled) products in the cleaning bath with enzymatic cleaning solution (e.g. CIDEZYME® Enzymatic Detergent Solution, 1.6 % v/v) for 5 minutes. The products must be adequately covered and the individual components should not be in a position to damage each other. Follow the enzymatic cleaner manufacturer’s instructions for use for correct exposure time, temperature and concentration. • Clean with a soft plastic brush (e.g. Justman Brush from VWR International) • Shift moveable parts back and forth several times • Clean large lumina with a bottle brush • Cannulated products (with cavities whose diameter is less than or equal to 1/6 of the device‘s length), e.g. cannulated drills, must be cleaned by inserting the dedicated cleaning stylet and rinsed using a suitable cannula and disposable syringe (rinse volume: 30 ml) • Clean the products in the ultrasonic bath for 15 minutes using a suitable detergent (e.g. CIDEZYME® Enzymatic Detergent Solution, 1.6 % v/v). Follow the enzymatic detergent manufacturer’s instructions for use for correct exposure time, temperature and concentration • Rinse with cold (T < 40°C) or warm (T > 40°C) water for at least one minute (lumina and cannulated products must also be rinsed inside using syringes and suitable cannulas); hand-held water jets can also be used • Visually inspect the products and repeat the cleaning process as required until visible contamination is no longer evident • Inspect the products (see the section «Inspection») Manual Disinfection Process • Place the (disassembled), cleaned and inspected products in the disinfection bath for 15 minutes (e.g. CIDEX® OPA Solution). The products must be adequately covered and the individual components should not be in a position to damage each other. Follow the enzymatic disinfection solution manufacturer’s instructions for use for correct exposure time, temperature and concentration. • Shift moveable parts back and forth several times • Large lumina must also be filled on the inside • Cannulated products (with cavities whose diameter is less than or equal to 1/6 of the device‘s length), e.g. cannulated drills, must be filled with disinfectant and rinsed using a syringe and suitable cannula (rinse volume: 30 ml) • Rinse with cold (T < 40°C) or warm (T > 40°C) water for at least one minute (lumina and cannulated products must also be rinsed inside using syringes and suitable cannulas); hand-held water jets can also be used • Visually inspect the products and repeat the cleaning and disinfection process as required until visible contamination is no longer evident • The products must be completely dried directly afterwards (it is recommendable to dry them using medical compressed air) • Inspect the products (see the section «Inspection») and service them (see the section «Product Care») • Pack the products preferably immediately or if necessary after giving them additional time to dry Automated Cleaning and Disinfection For automated cleaning and disinfection instruments have to be removed from the trays. Instruments have to be opened and disassembled! Implant trays made of aluminum or plastic are not intended for cleaning and disinfection when loaded. Implants must be removed from the trays and must be cleaned/disinfected separately. Implant trays made of steel can undergo automated cleaning and disinfection when loaded. Make sure the implant trays have been properly sealed with their lid prior to automated cleaning/disinfection. The above recommendations must also be followed when choosing detergents and disinfectants for this process. For automated cleaning, ensure that the products have been rinsed thoroughly and that there is no remaining foam. When selecting the disinfector, make sure: • That the cleaning process includes the following phases in accordance with EN ISO 15883: Phase
Temperature
Duration
Action
Cleaning
55°C (± 2°C) (131°F; ± 3.6°F)*
10 min.*
Adding detergent*
Neutralization
Cold (T < 40°C/104°F)
2 min.
Neutralize with cold water
Rinsing
Cold (T < 40°C/104°F)
1 min.
Rinse with cold water
Thermal disinfection (Ao value > 3’000)
≥ 90°C (194°F)
5 min.
With demineralized and purified water; do not add additional detergent
Rinsing
Device-specific / cold (T < 40°C/104°F)
1 min.
Rinse with demineralized and purified water
Dry
Device-specific**
Device-specific**
Drying process
Inspection (Implants and Instruments) Before assigning the implants to the implant containers/trays, check them after cleaning and disinfection for damage and contaminants, and remove damaged and contaminated implants. After the instruments are cleaned and disinfected, check them all for damage (e.g. corrosion, damage to surfaces, chipping, etc.), contaminants and function. Remove damaged instruments. In addition, instruments with lumina (e.g. cannulated drills) have to be checked for free passage without obstructions, cutting instruments must be checked for sharpness and rotating instruments must be checked for bending. Instruments that are still soiled must be cleaned and disinfected again. You can find further details at www.medartis.com in «Instructions for Cleaning, Disinfection and Sterilization». Product Care Carefully apply maintenance products (paraffin-based/white oil-based, biocompatible, steamsterilizable and steam-permeable) to the articulations, closures or threads and sliding surfaces. Do not use maintenance products containing silicone. The disassembled instruments and trays should be reassembled for the following sterilization process. Sterilization Medartis recommends sterilizing the products in the specially designed APTUS sterilization containers, implant containers and instrument trays . If the total weight of the loaded module is over 10 kg, the module must not be sterilized in a sterilization container; rather, wrap it in sterilization paper and sterilize it according to state of the art techniques and using approved methods. Steam Sterilization All NON-STERILE products can be sterilized in an autoclave (EN 13060 and EN 285). For both initial and subsequent sterilization, the following parameters were validated by Medartis in accordance with the requirements of the current sterilization standards, EN ISO 17665 and ANSI/AAMI ST79: Procedure
Fractionated and Dynamic Prevacuum Process
Flow and Gravitation Processes
Exposure time
≥ 4 min.
≥ 15 min.
Temperature
132°C/134°C
132°C/134°C
Drying time
> 20 ─ 30 min.
> 20 ─- 30 min.
Medartis recommends that sterilization is performed in accordance with the above validated processes. If the user utilizes other processes (e.g. flash sterilization), these must be validated by the user. The ultimate responsibility for validation of sterilization techniques and equipment lies with the user. Outside the USA: the sterilization time can be extended to 18 minutes to meet the recommendations of the WHO and the Robert Koch Institut (RKI). Medartis products are designed for these sterilization cycles. Do not use hot-air sterilization, radiation sterilization, formaldehyde sterilization, ethylene oxide sterilization or substitute procedures for sterilizing thermolabile products such as plasma or peroxide sterilization for Medartis products. After sterilization, the products must be stored in a dry and dust-free environment. Reusability (Implants and Instruments) All implant components are intended for one single application in a single patient. Implants that were used in a patient and removed, have to be discarded following the local requirements. Application of an already used device may compromise the structural integrity of the implants and/or lead to device failure which may result in patient injury. Furthermore, application of an implant that has already been used may create a risk of contamination e.g. due to the transmission of infectious material from one patient to another. This could result in injury of the patient or user. Implants that have not come into direct contact with a patient may be reprocessed. Implants that have come into direct contact with blood or other bodily fluids or show visual contamination must be cleaned and disinfected separately before they can be placed back into the implant tray. The instruments can be reused if corresponding precautions are observed and if they are undamaged and uncontaminated. No liability is assumed by the manufacturer in case of non-observance. Medartis recommends: if products come in contact with pathogens that are difficult to identify such as variations of Creutzfeldt-Jakob’s disease (confirmed or suspected pathogen), they must be discarded. Manufacturer Medartis AG Hochbergerstrasse 60E 4057 Basel/Switzerland
Consult instructions for use
Article number / Order number
* The information provided is based on the use of «Neodisher MediClean forte» by Dr. Weigert; validation has been performed with a concentration of 0.2 % at 50°C; however, exposure times, temperatures and concentrations may vary if a different detergent is used; follow the applicable information provided by the manufacturer.
Lot number
Non-sterile
** Drying temperature must be < 141°C When loading the disinfector, use the loading layouts provided by the manufacturer; also follow the detailed information provided in «Instructions for Cleaning, Disinfection and Sterilization» at www.medartis.com.
Do not reuse
APTUS-00000101_ Rev. U 2020/05/13 Instructions for Use for Medartis APTUS Plates, Screws and Instruments Medartis, APTUS, MODUS, TriLock, HexaDrive and SpeedTip are registered trademarks of Medartis AG/Medartis Holding AG, 4057 Basel/Switzerland
3 of 4
Do not resterilize
Do not use if package is damaged
Sterilized using irradiation
Use by date
Manufacturer
Date of manufacture
Marking for Risk Class I medical devices, sterile, I with measuring function, IIA and IIB
Marking for Risk Class I medical devices, non-sterile and without measuring function
This document is subject to continuous revision. Please verify that the current printed version is identical to the one at www.medartis.com.
APTUS-00000101_ Rev. U 2020/05/13 Instructions for Use for Medartis APTUS Plates, Screws and Instruments Medartis, APTUS, MODUS, TriLock, HexaDrive and SpeedTip are registered trademarks of Medartis AG/Medartis Holding AG, 4057 Basel/Switzerland
4 of 4