BiliBed
®
SERVICE MANUAL
BiliBed
®
SERVICEANLEITUNG
Valid as of production date: July 2006 Medela AG 200.0076 /12.06/A
Service Manual Contents 1.
page
Technical data ... 4 1.1 What the symbols mean ...5 1.2 Technical standards...5
2.
Safety tests... 6 2.1 Visual check...6 a Cleaning and disinfecting...6 2.2 Functional tests...6 a Changing the fluorescent tube ...6
3.
Disposal... 7
4.
Electrical wiring diagram... 8
5.
General overview and spare parts list ... 9 5.1. Baby support...9 5.2. Light unit ...10
Service Manual / Serviceanleitung
BiliBed®
Seite 3
1.
Technical Data Light unit ¾ Power supply ¾ Power consumption ¾ Irradiation (according to IEC 60601-2-50) ¾ Tube life ¾ Air pressure ¾ Relative operating humidity ¾ Relative storage and transport ¾ Operating temperature ¾ Storage and transport temperature ¾ Noise level ¾ Standards
Power adapter (TR30RAM120): Input: 100 - 240V AC, ~47-63Hz , 0.8-0.4A Output: 12V DC, 2.5A 30 W with compact fluorecent lamp (Osram Dulux L18W/71 or Ralutec long 18W/71) 2 Ebi max = 49 W/m approx. 1'500 hours 70 – 106 kPa 20 – 80% 20 - 95% humidity +5 to +30°C -20 to +50°C <20 dB(A) CE (93/42/EWG) EN 60601-1:1990 (IEC 60601-1:1988 + A1:1991 + A2:1995) EN 60601-2-50:2002
Baby support ¾ Outer dimensions ¾ Cover material
630 x 326 x 42 mm Skin-friendly polyurethane (PUR)
Complete device ¾ Total dimensions ¾ Working conditions ¾ Storage conditions ¾ Weight
630 x 326 x 134 mm + 5°C to + 30°C - 20°C to + 50°C 4.0 kg
Bilicombi ¾ Material: ¾ Light permeable fabric material:
75% cotton, 25% polyester 100% cotton
Service Manual / Serviceanleitung
BiliBed®
Seite 4
1.1. What the symbols mean BiliBed Ensure that the patient is positioned correctly. The BiliBed should not be used in an incubator On/Off switch for lamp Refer to instructions Model B unit
IPX4 Protection against splashing water Power adapter Direct current Protection class II
Must not be disposed of together with household refuse
For indoor use only
1.2
Technical Standards ¾
CE (93/42/EWG), EN 60601-1:1990, (IEC 60601-1:1988 + A1:1991 + A2:1995), IEC 60601-2-50:2000
¾
CE marking according to 93/42/EWG
¾
Quality Management System EN ISO 13485:2003
¾
National test marks available on request
Service Manual / Serviceanleitung
BiliBed®
Seite 5
2.
Safety tests We recommend performing the following safety checks yearly:
2.1
Visual check The BiliBed should be examined for damage and dirt on the inside and outside of the unit, and repaired or cleaned if necessary. a
Cleaning and disinfecting ¾
Unplug the unit from the power supply.
Light unit and baby support: The outside surfaces can be cleaned with a damp cloth. For disinfecting, alcohol may also be used. Cleaning or disinfecting agents should not be used in concentrated form! Important Do not use cleaning solutions such as appliance disinfectants which use formaldehyde. Do not autoclave the baby support or disinfect by using heat. Bilicombi therapy blanket: Please refer to the BiliBed instructions for use.
2.2 Functional tests Every test function of the BiliBed shall be performed according to the Instructions for Use and Service Manual. Intensity of irradiation: Check the "Light Tube Change Log" (in the inside of the light unit) for the date and/or "actual hours" column of the last tube change. If necessary (every 1,500 hours), change the tube (see 2.2.a). a
Changing the fluorescent tube After a 1,500 hour-interval, the light tube requires replacement. By using the "total h" hour meter, this time interval can be monitored accurately. ¾
Unplug the light unit from the power supply.
¾
Remove the baby support from the light unit.
¾
Release the 10 screws with a screwdriver and remove the Perspex cover.
¾
Lift the metal clamp and carefully remove the light tube from its socket and from its clip.
¾
Put the tube aside immediately; in order to avoid confusion.
¾
Insert the new tube into the clip and push into the socket until the tube snaps in.
Service Manual / Serviceanleitung
BiliBed®
Seite 6
¾
Readjust the metal clamp to securely hold the lamp.
Warning Use only Medela-approved replacement tubes. Otherwise, the success of the therapy cannot be guaranteed! ¾
Check that the reflector is clean and shiny. If required, clean reflector.
¾
Re-mount the Perspex cover with the 10 screws.
Fan: The BiliBed has an electronic safety circuit which switches the lamp off if the fan does not run at the proper speed or is blocked. This protects the system from overheating. Stop the fan with a finger and ensure that the lamp switches off automatically. If necessary, contact a Medela service technician.
3.
Disposal ®
The BiliBed is of modular design. It can thus be disassembled into its individual components. The BiliBed is mainly made of technical plastics which can be disposed of in accordance with local regulations. The fan, lamp and electric parts should be disposed of separately.
Service Manual / Serviceanleitung
BiliBed®
Seite 7
Service Manual / Serviceanleitung
920.0049
Mains Adapter
900.0339
On / Off Switch
900.0688
J1/3 J1/4
3 4
J1/2
J3/2
2
braun
braun
J3/1
J1/1
3
3
weiss
weiss
1
1
1
12V In
On / Off
900.0087
Lüfter
Lampe
J2/1
schwarz
gelb
nc
J2/3 J2/2
rot
J2/4
Dulux DL 18/71
Fluorescent Lamp
Blower
936.0015
Pelko C6025L12
910.0076
schwarz
gelb
rot
gelb
J4/5
900.0418
3
gelb
grün
J4/4
4
2
grün
1
weiss
braun
nc
weiss
braun
J4/3
J4/2
J4/1
4. Electrical wiring diagram
BiliBed®
Seite 8
5.
General overview and spare parts list 5.1
Baby support
Item Description baby support: Frame, complete Plastic cover 1 2 3 4 5 6 7 8
Hinged joint, LH, complete Profile tube Hinged joint, RH, complete Rectangular tube Corner piece, RH Corner piece, LH Drive screw, 3.5 mm, x 6.35 Leafspring
Service Manual / Serviceanleitung
L 63,0 x W 32,6 cm L 63,0 cm
L 63,0 cm L 27,9 cm
BiliBed®
Order no. 600.0425 808.0018 600.0381 537.0030 600.0380 537.0031 810.0298 810.0297 433.0003 522.1467
Seite 9
5.2
Light unit
1 2 21
3
22 23 4
15 16 17
5 6
24
18 7 19 20
8 9 10 11
25 26
12 13 14
Pos. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27
27
Beschreibung Strahlereinheit Best.-Nr. Slotted screw M4x10 ... 412.0017 Plexiglass cover BB4 ... 815.0060 Foam strip 4x10x1480 ... 850.0012 4-core Cable BB4 ... 900.0418 PT countersunk screw KA 40x8 ... 436.0028 Cable tie ... 899.0062 Quicklock Benzing ... 442.0006 Metal clamp lamp holder ... 524.0064 PT countersunk screw KA 30x16 ... 436.0008 Compact lamp support ... 524.0056 Reflector ... 522.1476 Compact fluorescent tube, blue... 936.0015 PT countersunk screw 40x10 ... 436.0011 Deflector ... 522.1519 PT countersunk screw KA 40x10 ... 436.0011 PCB BB4 complete... 900.0087 Fan 12V BB4 ... 910.0076 Foam 8x20x180... 850.0013 PT countersunk screw KA 40x10 ... 436.0011 Lamp holder... 899.0081 Cover electronics BB4 ... 810.0767 Foam ... 850.0027 Power adapter BB4 ... 920.0049 Push button (start/stop) ... 900.0339 2-core Cable, plug PCB... 900.0688 Adjustment pegs... 810.0376 Base unit BB4... 810.0766
Service Manual / Serviceanleitung
BiliBed®
Seite 10
OOO Medela Fersmana Street 5A 117312 Moskau Phone +7 (095) 980 6194 Fax +7 (095) 718 1700 e-mail [email protected] www.medela-russia.ru
Canada Medela, Inc. 4090B Sladeview Crescent, Unit 2 Mississauga, Ontario Canada, L5L 5Y5 Phone +1 905 608 7272 Fax +1 905 608 8720 e-mail [email protected] www.medela.ca
Spain Medela España c/ Manuel Fernández Márquez, 49 08918 Badalona Phone +34 93 320 59 69 Fax +34 93 320 55 31 e-mail [email protected] www.medela.es
France Medela France Sarl 29, avenue des Grenots 91150 Etampes Phone +33 (0)1 69 16 10 30 Fax +33 (0)1 69 16 10 32 e-mail [email protected] www.medela.fr
Sweden Medela Medical AB Box 7266 187 14 Täby Phone +46 (0)8 630 09 40 Fax +46 (0)8 630 09 48 e-mail [email protected] www.medela.se
Germany Medela Medizintechnik GmbH & Co. Handels KG Postfach 1148 85378 Eching Phone +49 (0)89 31 97 59-0 Fax +49 (0)89 31 97 59 99 e-mail [email protected] www.medela.de
United Kingdom Medela UK Ltd. Huntsman Drive Northbank Industrial Park Irlam, Manchester M44 5EGw Phone +44 870 950 5994 Fax +44 870 389 2233 e-mail [email protected] www.medela.co.uk
Italy Medela Italia Srl Via Stelloni 39/O 40010 Sala Bolognese Phone +39 051 681 52 76 Fax +39 051 681 52 77 e-mail [email protected] www.medela.it
USA Medela, Inc. P.O. Box 660 McHenry, IL. 60051-0660 Phone +1 815 363 1166 Fax +1 815 363 1246 e-mail [email protected] www.medela.com
Japan Medela K.K. Fuji Chuo Building 5F 2-1-7, Shintomi Chuo-ku, Tokyo 104-0041 Phone +81 3 3553 3571 Fax +81 3 3553 3573 e-mail [email protected] www.medela.co.jp Netherlands/Belgium Medela Benelux BV De Steenbok 12 5215 ME 's-Hertogenbosch Phone +31 73 690 40 40 Fax +31 73 690 40 44 e-mail [email protected] e-mail [email protected] www.medela.nl www.medela.be
Switzerland & International Sales Medela AG, Medical Technology Lättichstrasse 4b 6341 Baar/Switzerland Phone +41 (0)41 769 51 51 Fax +41 (0)41 769 51 00 e-mail [email protected] www.medela.ch Russia