Continuum Infusion System Max Flow Rate Programming Instructions April 2014

Rate Programming Instructions

2 Pages

Instructions for programming maximum flow rate with Continuum To access the adjustable settings:  1... ress the INFO button twice... croll to “Adjust Setting”, then press START/OK... MEDRAD, INC... 3... croll to “More…” then press START/OK... nter Code 901 then press START/OK... croll to “Maximum Flow Rate”, then press START/OK... 6... Enter the rate and press START/OK... ate should be set at 500 ml/hr... hen done, choose Restart Pump from the menu... onfirm that the correct Maximum flow rate has been entered: The Show Setting feature is designed to allow you to view the programmable options and other important information about the pump... n Show Setting, the current settings can be viewed, but cannot be changed... 10... ress INFO twice... he screen below is displayed  100 Global View Drive Warrendale, PA 15086 U... A... 412) 767-2400 www... yer... m   Page 2 of 2  11... ress the DOWN ARROW button to select Show Setting... ress START/OK and the current settings are displayed, but cannot be changed... o view the setting for a particular option, select the desired option, then press START/OK... 2... aximum flow rate should be set at 500 ml/hr... RAD-MR-14-06441  April 2014