Operation Manual
98 Pages
Preview
Page 1
Operation Manual
1-1
S e r i a l n u m b e r s a n d
d a t e o f i n s t a l l a t i o n
Operation Manual
i n f o r m a t i o n
m u s t b e s u p p l i e
S e n u m b e r s
Operation Manual
i
1 Introduction ...1 1.1 Certifications... 1 1.2 Intended Use... 1 1.3 Training Information... 1 1.4 Contraindications... 1 1.5 Restricted Sales ... 1 1.6 Required Training ... 1 1.7 Disclaimers... 1
2 Symbols...3 2.1 Manufacturing Symbols ... 3 2.2 Shipper Symbols... 3 2.3 Notified Body ... 4 2.4 Regulatory Classifications... 4 2.5 MR Icons and Classifications... 5 2.6 Warning Labels and Symbols ... 5
3 Warnings, Cautions, and Notices ...7 3.1 Warnings... 7 3.2 Cautions... 8 3.3 Notices... 8
4 System Overview ...9 4.1 System Diagram... 9 4.2 Moving the System...10 4.3 Fluid Delivery Basics ...11 4.3.1 Protocol...11 4.3.2 I Checked for Air ...12 4.3.3 Test Injection...12 4.3.4 KVO (Keep Vein Open)...12 4.3.5 Protocol Manager...12 4.3.6 Fluid Delivery System Design ...12 4.3.7 Fluid Pressure and Pressure Limiting...13 4.3.8 Response to Occlusions ...13 4.3.9 Volume and Flow Rate Protection ...13 4.4 Syringe Installation Features ...13 4.5 Basic Informatics ...14 4.6 Programming Mode...14 4.7 Using the Pedestal with Integrated IV Pole...14
5 Understanding the Display and Pod ...15 5.1 Home Screen ...15 5.2 Pod Description ...18
6 Understanding the Injector Head...19 6.1 Injector Head Controls ...19 6.2 Injector Head Components ...21 6.2.1 Manual Knobs...21
ii
MEDRAD® MRXperion
7 System Lights and Indicators...23 7.1 Injector Head Lights and Indicators ...23 7.2 Injector Status Lights...24 7.3 Pod Power Indicator...24 7.4 Hand Switch Light ...25
8 Powering Up and Shutting Down the System ...27 8.1 Powering Up the System...27 8.2 Powering Down the System ...28 8.2.1 Hard Shutdown...28 8.3 Restore from Injector Shutdown...28 8.4 Restore from System Shutdown...28
9 Calculators...29 9.1 Enabling Calculators ...29 9.2 Calculator Setup ...29 9.2.1 Setting Up the eGFR Calculator...30 9.2.2 Setting Up the Weight-Based Dosing Calculator...34 9.3 Using the Calculators...34 9.3.1 Using the eGFR Calculator ...34 9.3.2 Using the Weight Based Dosing Calculator ...35
10 Protocol Management ...37 10.1 Create or Edit a Protocol ...37 10.2 Save a Protocol ...40 10.3 Recall a Saved Protocol ...41
11 Preparing for Injection ...43 11.1 Control Room Preparation...44 11.1.1 Prepare Injection Protocol ...44 11.2 Scan Room Preparation ...44 11.2.1 Installing a Syringe ...44 11.2.2 Filling a Syringe: Automatically or Manually ...45 11.2.3 Attach and Prime the Tubing ...46 11.2.4 Connect the Tubing to Patient ...47
12 Arming and Injecting ...49 12.1 Add Volume Indicator ...49 12.2 Arming the Injector...49 12.2.1 Arming from the Control Room ...49 12.2.2 Arming from the Scan Room ...51 12.3 Starting KVO (Optional) ...51 12.4 Initiating an Injection ...52 12.5 Operator Initiated Hold ...52 12.6 Aborting an Injection ...52 12.7 Viewing Injection Progress...53 12.8 Reminders ...54
Operation Manual
iii
13 Completing an Injection ...55 13.1 Injection Complete ...55 13.2 Injection Aborted...56 13.3 Exiting Injection Complete ...57 13.3.1 Conducting Another Injection ...57
14 Removing Disposable Syringes and Connector Tube ...59 14.1 Removing Disposable Syringes and Connector Tube...59
15 Advanced Configurations ...61 15.1 System Setup ...61 15.1.1 System Setup Configurable Items ...62 15.2 Calculator Setup...62 15.3 Protocol Manager Setup...63 15.3.1 Delete a Protocol...63 15.3.2 Rearrange Protocol List ...64 15.3.3 Hide/Show a Region ...64 15.3.4 Rename a Region...65 15.4 Fluid Delivery Setup...65 15.4.1 Fluid Delivery Setup Configurable Items ...67 15.4.2 Contrast Type Setup ...67 15.5 Help ...70 15.5.1 Accessing the Help System...70 15.6 Status ...71
16 System Messages...73 16.1 Type 1 Messages...73 16.2 Type 2 Messages...73 16.3 Type 3 Messages...74
17 Maintenance and Checkout...75 17.1 Daily...75 17.1.1 Cleaning the Injector Head, Syringe Piston, Syringe Interface, and Scan Room Unit Power Supply ...75 17.1.2 Cleaning the Pedestal...76 17.1.3 Inspecting the Injector Head and Mounting ...76 17.1.4 Cleaning the Display...76 17.1.5 Inspecting the Display...76 17.2 Monthly ...77 17.2.1 Operational Checkout...77 17.3 Annually ...78 17.3.1 Injection System Calibration ...78 17.3.2 Checking Electrical Leakage ...78
18 Specifications...79 18.1 Display Specifications ...79 18.1.1 Display Dimensions...79 18.1.2 Display Connections ...79
iv
MEDRAD® MRXperion
18.1.3 Display Input Power Requirements...80 18.1.4 Pod Input Power Requirements ...80 18.2 Injector (Scan Room Unit) Specifications...80 18.2.1 Injector (Scan Room Unit) Dimensions ...80 18.2.2 Scan Room Unit Power Supply Dimensions...81 18.2.3 Input Power Requirements...81 18.3 Environmental Specifications ...81 18.3.1 Non-Operating (Transportation and Storage)...81 18.3.2 Operating ...81 18.3.3 Protection Against Electrical Shock ...82 18.3.4 EMI/RFI ...82 18.3.5 Protection Against the Ingress of Fluids...82 18.3.6 Mode of Operation ...82 18.4 System Capabilities ...83 18.5 Approximate Heat Generation...83 18.6 Over and Under Infusion Protection ...83 18.7 System Fluid Performance ...84 18.8 System Performance...84 18.9 Forward and Reverse Controls...84 18.10 Ground Continuity...84 18.11 Classifications ...84
19 Options and Accessories ...85 19.1 MEDRAD® MRXperion MR Injection System...85 19.2 MEDRAD® MRXperion Injection System Syringe Kit...85 19.3 Hand Switch ...85 19.4 IV Poles...85 19.5 Manuals...85 19.6 Penetration Panel Filter Kit...85
20 Compliance to EN/IEC 60601-1-2: 2007...87
Operation Manual
1
1-1
Introduction
This manual applies to the MEDRAD® MRXperion MR Injection System, referred to as the system throughout this document. Read all of the information contained in this manual. Understanding this information will assist users in operating the system in a safe manner. NOTE: Operating specifications and feature availability may vary by country. Check with local product representatives and refer to country-specific operating instructions.
1.1 Certifications This device is equipped to operate at 100-240 VAC, 50/60 Hz and is designed to comply with IEC60601-1:2012 (Edition 3.1). Special precautions regarding ElectroMagnetic Compatibility (EMC) are required for installation and use of this injector system. Detailed EMC information can be found in "Chapter 20 - Compliance to EN/IEC 60601-1-2: 2007".
1.2 Intended Use The MEDRAD® MRXperion Injection System is a syringe-based fluid delivery system indicated for delivery of contrast media and saline during MR procedures. It is intended to be used for the specific purpose of injecting intravenous MR contrast media and saline into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) applications with MRI scanners that have a magnetic field strength between 0.7 Tesla and 3.0 Tesla. Only trained healthcare professionals are intended to operate this device.
1.3 Training Information This manual is intended as an extension of the user interface of the MEDRAD® MRXperion Injection System to provide procedural and technical information. Additional MEDRAD® MRXperion training information will be available in the following formats: Q
On-site initial installation and additional training as requested
Q
MEDRAD® MRXperion Quick Guides
Q
Syringe instructions for use (IFU)
Please contact Bayer HealthCare Services or your local Bayer representative if any of these resources are needed.
1.4 Contraindications None known.
1.5 Restricted Sales Federal (USA) Law restricts this device to sale by or on the order of a physician.
1.6 Required Training This device is intended to be used by individuals with adequate training and experience in magnetic resonance (MR) imaging diagnostic studies.
1.7 Disclaimers External wiring and modifications disclaimers: Bayer disclaims liability for any modifications or interfaces with other equipment that are not in conformity with the specifications and information contained within this manual. Accessory equipment connected to the device must be certified according to IEC60601-1:2012 (Edition 3.1). Anyone who connects additional equipment to the signal input or output part configures a medical system and is therefore responsible that the system complies with the requirements of the standard IEC60601-1:2012 (Edition 3.1). To obtain on-site consulting or consulting references, contact Bayer HealthCare Services.
1-2
MEDRAD® MRXperion
The system is not intended to deliver contrast agents and/or saline to more than one patient from the same saline container and Bayer disclaims any and all liability or consequences resulting from use of the same container on more than one patient. Use of syringes and other disposables from a source that is not approved by Bayer may cause injury as set forth herein and such use may also void the warranty associated with this product.
Operation Manual
2
2-3
Symbols
2.1 Manufacturing Symbols Manufacturer
Authorized representative in the European community
2.2 Shipper Symbols
Temperature range
Humidity range
Atmospheric pressure range
Serial number
Catalog number
This side up
Keep dry
Fragile
2-4
MEDRAD® MRXperion
2.3 Notified Body Indicates that this device conforms to the requirements of the European Medical Device Directive 93/42/EEC
2.4 Regulatory Classifications Identifies a type BF applied part complying with EN 60601-1 standards Medical - General Medical Equipment As To Electrical Shock, Fire, and Mechanical Hazards Only In accordance with ANSI/AAMI ES60601-1 (2005, 3rd Ed.) CAN/CSA-C22.2 No. 60601 (2008), IEC 60601-1: 3rd Ed. See accompanying documentation. This symbol indicates the user shall refer to the instructions-for-use to ensure safe operation.
CLASS 1
Indicates the injection system is Class I medical equipment as defined by IEC606011:2012 (Edition 3.1) standards for protection from electric shock
Class II Equipment, Double Insulated.
IPX0 Code that indicates non-protection against ingress of water (IEC 60529)
IPX1 Code that specifies the degree of protection against vertically falling water drops (IEC 60529)
Indicates separate collection for Electrical and Electronic Equipment per Directive 2002/96/EC. Refer to the following website for additional information: http://www.medrad.com/en-us/resources/Pages/WEEE.aspx Indicates that this product contains certain toxic or hazardous substances or elements and can be used safely during its environmental protection use period, indicated by the number in the center of the logo. This product should be recycled immediately after its environmental protection use period has expired. Identifies the Equipotential connection. The Equipotential Connector (EPC) is an electrically bonded terminal on the injector that is used as a connection point between other medical electrical equipment. The EPC's function is to minimize any voltage potentials differences between all connected equipment. The EPC is not designed to be an electrical safety ground.
Does not contain serviceable parts.
Operation Manual
2-5
Indicates that a component contains an electrical fuse that can be replaced.
2.5 MR Icons and Classifications
MR
MR Conditional
Has been demonstrated to pose no known hazards in a specified MR environment with specified conditions of use as defined by the ASTM International Standards for MRI Device Marking.
MR Unsafe
Known threat or poses a hazard in all MR environments as defined by the ASTM International Standards for MRI Device Marking.
2.6 Warning Labels and Symbols
Pushing prohibited in this area. Do not push the injector above this label.
Do not coil the cable to the Power Supply on or near the injector or magnet. Excess cable should be tie wrapped in a figure-eight configuration and positioned as far away from the magnet as possible. Looped or coiled cable could create image artifacts.
Caster brakes must be used to prevent the injection system from drifting during use.
The injector (scan room unit) power supply must be more than 6ft (1.8m) away from the magnet.
The optional Penetration Panel Filter must be more than 6ft (1.8m) away from the magnet.
2-6
MEDRAD® MRXperion
Caution: Transportable Weight- 43.4kg Safe Working Load- 0.75kg
Warning: Refer to warnings on Instructions for Use packaged in each carton.
Caution: Refer to cautions on Instructions for Use packaged in each carton.
Indicates hazardous voltage.
Consult operating instructions
WARNING
Indicates that the information is a warning. Warnings advise you of circumstances that could result in injury or death to the patient or operator. Read and understand the warnings before operating the injection system.
CAUTION
Indicates that the information is a caution. Cautions advise you of circumstances that could result in minor or moderate injury to the patient or operator. Read and understand the cautions before operating the injection system.
NOTICE
Indicates that the information is a notice. Notices advise you of circumstances that could result in damage to the device. Read and understand the notices before operating the injection system.
Note
Indicates that the information that follows is important information or a tip related to the proper functionality of the system that will help the operator to recover from an error or point to related information within the manual. Read and understand the notes before operating the system.
Operation Manual
3
Warnings, Cautions, and Notices
3.1 Warnings WARNINGS Air Embolism Hazard - Serious patient injury or death may result. Q
Use only Bayer approved syringes, connector tubing, and spikes.
Q
Use only catheters and connectors with pressure ratings that are compatible with this system.
Q
Use only accessories and options approved by Bayer.
Q
The operator needs to check for air to confirm that the syringes and tubing have been inspected for the presence of air.
Environmental Contamination Hazard - Serious patient or worker injury or death may result. Q
Visually inspect all components before use.
Q
Do not use damaged components.
Electric Shock Hazard - Serious patient and/or worker injury or death may result. Q
Only use the power cord supplied with the system or by the Bayer HealthCare Services representative.
Q
Equipment must only be connected to a supply mains with protective earth.
Q
Do not remove or open any enclosure, as hazardous voltages exist within system components.
Q
Do not immerse any system components in water or soap solutions.
Q
Disconnect the system from supply mains before cleaning.
Q
Do not use the system if any worn or damaged cabling is detected. Contact Bayer HealthCare Services for replacement.
Fire Hazard - Serious patient and/or worker injury or death may result. Q
Use the correct fuse type.
Q
Fuse should only be replaced by Bayer or Bayer-trained personnel.
Explosion Hazard - Serious patient and/or worker injury or death may result. Q
The system contains a lithium ion battery pack. Replacement and disposal shall be performed only by a qualified service engineer. Please contact Bayer HealthCare Services for assistance.
Compromised Efficacy Hazard - Serious patient injury or death may result. Q
Do not use this system to deliver any fluid other than intravenous MR contrast media and saline.
Q
Do not retract pistons with connector tube installed.
Procedure Delay Hazard - Serious patient and/or worker injury or death may result. Q
Turn off any other equipment that could generate high electromagnetic fields or high-level electrostatic discharge.
Personal injury or equipment damage may result when standing on the Power Supply Enclosure. Q
Do not stand on the power supply enclosure.
3-7
3-8
MEDRAD® MRXperion
WARNINGS Patient and/or worker injury or equipment damage may result. Q
Display is MR unsafe. The display is a known threat or poses a hazard in all MR environments as defined by the ASTM International Standards for MRI Device Marketing.
Q
The Injector (scan room unit) and power supply are MR conditional. The scan room unit and power supply have been demonstrated to pose no known hazards in a specific MR environment with specified conditions of use as defined by the ASTM International Standards for MRI Device Marketing.
Q
Do not make any unauthorized modifications to the injector or any of its parts.
3.2 Cautions
CAUTIONS Mechanical Hazard - Minor or moderate patient and/or worker injury may result. Q
Only use the system as defined in this manual. Follow the instructions given in system communication or fault messages.
Q
Do not use the system in the presence of flammable (such as anesthetics) or combustible gases or other agents.
Q
Use only non-magnetic tools to install any scanner/magnet room components.
Q
Dispose of system components and accessories according to local regulations or contact Bayer HealthCare Services.
Q
When moving the system, hold the pedestal no higher than 45inches (114.3 cm) above the floor.
Electric Shock Hazard - Minor or moderate patient and/or worker injury may result. Q
Only plug the system into a direct mains access point.
Q
Make sure the direct mains access point is accessible during operation of the system.
Q
Do not plug the system power cords into an extension cord or multi-outlet power strip.
Q
Verify that the voltage and frequency as labeled matches the voltage and frequency of the electrical outlet.
Q
Use only Bayer-approved accessories.
Do not adjust or transport the system with fluid containers attached. Minor or moderate patient and/or worker injury may result. Q
The fluid containers and/or holder may fall, causing patient injury and/or property damage.
Q
Attach containers only after the system is in a stationary position.
3.3 Notices NOTICE Electro-Mechanical Hazard - Equipment Damage may result. Q
Do not use the system immediately after it has been brought indoors from extreme outside temperatures.
Q
Allow the system to stabilize at room temperature before use.
Q
Condensation may cause electrical damage to the injection system.
Mechanical Hazard - Equipment Damage may result. Q
Do not touch the display screen with a sharp object.
Q
Component damage may occur if not installed properly. Ensure all connections are secure; do not over-tighten. This will help minimize leaks, disconnection, and component damage.
Operation Manual
4
4-9
System Overview
4.1 System Diagram The system is comprised of a scan room unit (injector) and a control room unit (touch screen display and Pod). These components are connected by a fiber optic communications link. NOTE: When using the system with an open-bore scanner, it is recommended that the injector be placed a minimum of 60 inches (152.4 cm) from the facade of the scanner. NOTE: When using the system with a Siemens MAGNETOM 1.5T or 3.0T scanner, it is recommended that the injector be placed a minimum of 18 inches (45.72 cm) from the facade of the scanner. NOTE: Consider limits on the length of the connector tubing and patient position when placing the injector. NOTE: Do not place the scan room unit in a magnetic field greater than 1000G without the casters being locked.
Figure 4 - 1: System Diagram - Power Supply in Scanner Room Scan Room
Control Room
1 - Scan Room Unit (Injector)
6 - Control Room Unit (Display and Pod)
2 - Power Cable
7 - Display and Pod Power Supplies
3 - Scan Room Unit Power Supply
8 - Fiber Optic Communication Link
4 - Fiber Optic Communication Link 5- Fiber Optic Quick Disconnect
4 - 10
MEDRAD® MRXperion
Figure 4 - 2: System Diagram - Power Supply Outside Scanner Room Scan Room
Control Room and MRI Equipment Room
1 - Scan Room Unit (Injector)
6 - Penetration Panel Filter
2 - Power Cable
7 - Scan Room Unit Power Supply
3 - Fiber Optic Communication Link
8 - Display and Pod Power Supplies
4 - Fiber Optic Quick Disconnect
9 - Fiber Optic Communication link
5- Penetration Panel
10- Control Room Unit (Display and Pod)
4.2 Moving the System WARNING Patient injury could result from movement of the scan room unit (injector) after the patient is connected to the fluid path. Q
Do not move the injector with the patient connected.
Q
Lock the casters at the base of the unit to prevent unintended movement.
Prior to moving the system, ensure that: Q
the patient is disconnected
Q
the injector head is sitting above the base of the system
Q
there are no fluid containers attached
Operation Manual
4 - 11
Once the injector head column is in place, move the system by holding the handle or by holding the pedestal below the point indicated by the label on the pedestal arm. 1.
To rotate the injector head column, turn the knob at the base of the injector head column counterclockwise to unlock it.
2.
Rotate the injector head column so the injector head sits above the base as shown in Figure 4 - 3.
3.
Once the injector head has been moved into place, turn the knob clockwise to lock the column to prevent the injector head from rotating about the column.
4.
Lock the casters at the base of the unit to prevent unintended movement.
Lock Symbol: Indicates which direction to turn the knob to lock or unlock the column. Locking the column prevents rotating the injector head about the column.
Figure 4 - 3: Moving the System
4.3 Fluid Delivery Basics The system is a programmable power injector that enables an operator to generate a defined fluid delivery protocol and execute a fluid injection.
4.3.1 Protocol A protocol defines how the fluid injection will proceed and is comprised of three elements: phases, a pressure limit, and optional reminders.
4.3.1.1 Phase Phases are steps that define the actual movement of fluid, pauses, and/or holds. There are three types of phases: Q
Fluid Delivery Phase: Defines the flow rate, volume, and duration of a fluid to be injected.
4 - 12
MEDRAD® MRXperion
Q
Pause Phase: Defines a set amount of time that fluid injection will be paused. The next phase will execute once the set time has elapsed.
Q
Hold Phase: Places the fluid delivery in Hold. The operator must press the “Start” button on the Pod, injector head or hand switch.
4.3.1.2 Pressure Limit The pressure limit defines the maximum fluid pressure the system is permitted to generate during the execution of the protocol. The default maximum pressure limit is 325psi (2,240 kPa) when using the recommended disposable syringes from Bayer. The operator may choose to reduce this pressure limit based on the conditions for use of the protocol.
4.3.1.3 Reminders Reminders are optional, operator-defined timing notifications that appear during the injection sequence or after the sequence completes. Reminders provide a notification to the operator after a defined time has elapsed. There are two types of reminders: Q
Standard reminders: Occur during the protocol execution and are triggered by the initiation of the injection.
Q
Post-injection reminders: Occur after the protocol is complete and are triggered by the completion of the injection prior to disconnecting the patient.
4.3.2 I Checked for Air Before the system can be armed, the operator must visually check that all air has been expelled from the syringes and tubing. After the operator has determined that all air has been expelled from the syringes and tubing, the operator can either press the I Checked for Air confirmation button, or acknowledge on the notification screen on the display.
4.3.3 Test Injection An optional, operator-defined saline test injection can be delivered before the protocol is initiated.
4.3.4 KVO (Keep Vein Open) An operator-defined function that delivers small boluses of saline from Syringe B at configurable intervals during programming, pre-and post-injection, between multiple injections, and/or during Pause and Hold phases.
4.3.5 Protocol Manager Using Protocol Manager, the operator can name, save, and recall protocols.
4.3.6 Fluid Delivery System Design The system is flow-controlled by three parameters: flow rate, duration, and volume. The system permits the operator to define any two of these three parameters and will automatically calculate the third: Q
The flow rate specified in a phase is the value the system targets to deliver during the injection until the defined phase volume is delivered.
Q
The duration is the defined time the phase will take to perform the volume injection at the stated flow rate.
Q
Volume is the amount of fluid that will be delivered in a phase.
Operation Manual
4 - 13
4.3.7 Fluid Pressure and Pressure Limiting The fluid pressure is measured by the system during the execution of a phase and ensures the protocol Pressure Limit is not exceeded. Fluid pressure is dependent on the following: Q
Flow rate
Q
Fluid viscosity
Q
Fluid temperatures
Q
Attached syringes and connector tube
Q
Catheter type/size
Pressure Limiting occurs when the measured fluid pressure exceeds the protocol Pressure Limit setting. If this occurs, the system informs the operator of the condition and initiates Flow Rate Reduction for the Fluid Delivery phase to reduce the fluid pressure.
4.3.7.1 Flow Rate Reduction A slower flow rate is automatically initiated by the system to prevent the fluid pressure from exceeding the Pressure Limit. The system will deliver the phase volume at this reduced flow rate. The fluid delivery phase will exceed the programmed duration when Flow Rate Reduction occurs.
4.3.8 Response to Occlusions When injecting into an occlusion, a stall condition or high pressure disarm results. A stall condition occurs when the actual flow rate is less than 10% of the defined rate. If a stall or high pressure disarm occurs, check the fluid path for blockage and inspect the syringes and connector tube for damage or kinks. If no issue is found, consider increasing the catheter size, removing administration sets, or decreasing the flow rate. If an occlusion occurs during KVO, the system will detect the condition after 4 or less KVO boluses fail to be delivered. (The occlusion will be detected after 1 minute or less if the KVO interval is set to 15 seconds or after 5 minutes or less if the KVO interval is set to 75 seconds.) Refer to "Section 15.4 - Fluid Delivery Setup" for information about setting KVO intervals.
4.3.9 Volume and Flow Rate Protection The system provides the following protection against over and under volume or flow rate conditions: Q
Warnings on the Safety Screen to remind the operator to check the protocol prior to the system being armed.
Q
On-screen indications of insufficient volume when the total programmed volume is greater than the amount of fluid in the syringe.
Q
Injection monitoring to detect over flow rate or over volume conditions due to system faults. If either of these conditions is detected, the injection is stopped.
4.4 Syringe Installation Features
CAUTION Mechanical Hazard - Minor or moderate patient and/or worker injury may result. Q
Ensure the syringe is properly snapped into the front of the injector head before injecting.
Q
Improper engagement or rotating syringe may cause the syringe to leak, become damaged, or to disengage from the injector head during the injection.
The system is designed with four features that decrease the time and steps to install and remove syringes to the injector head. (For more information about configuration of these features, see "Section 15.4 - Fluid Delivery Setup".) 1.
Non-rotational orientation: When installing a syringe onto the injector head, alignment is unnecessary. Push the 65mL syringe into opening A and the 115mL syringe into opening B.
4 - 14
MEDRAD® MRXperion
2.
Auto Docking: When Auto Advance is configured to be ON and a new empty syringe is installed, the injector piston automatically advances and docks with the syringe plunger.
3.
Auto Advance: When Auto Advance is configured to be ON and a new empty syringe is installed on the injector head, the piston automatically docks with the syringe plunger and advances it to the full forward position.
4.
Auto Retract: When Auto Retract is configured to be ON and the syringe is removed, the piston rod will automatically retract into the injector head.
4.5 Basic Informatics If the optional Informatics platform is included with the system, please refer to the Certegra Workstation Informatics Panel and Modality Worklist Operations Manual for overview and functionality.
4.6 Programming Mode The system allows the operator to enter values for two of the three parameters in the Fluid Delivery Phase. Based upon these operator-defined values, the system calculates the value of the third parameter. The system default configuration is for the operator to enter values for flow rate and volume and the system calculates the duration. Using Programming Mode (see "Section 15.4 - Fluid Delivery Setup"), the operator can select which of the three parameters (flow rate, volume, or duration) the system will automatically calculate based on their entries for the remaining two parameters.
4.7 Using the Pedestal with Integrated IV Pole The pedestal with integrated IV pole is on casters and can be located anywhere within the MR scan room to facilitate the enhanced MRI procedure.
WARNINGS Electric Shock Hazard - Serious patient and/or worker injury or death may result. Q
Turn off any equipment that could generate a high level electrostatic charge.
CAUTIONS Mechanical Hazard - Minor or moderate patient and/or worker injury may result. Q
Tighten all screws, clamps, and knobs during assembly and as needed during use. Loose components may cause the pedestal to collapse.
Q
Do not move or manipulate the injector by pulling or pushing the integrated IV pole. Using the IV pole to maneuver the injector could bend the IV pole or cause the injector to fall over. Maneuver the injector as instructed in "Section 4.2 - Moving the System".
Q
Do not move the injector when anything is hanging on the Integrated IV pole.
Q
Use care and diligence in folding and unfolding the IV pole hooks. Keep hands and fingers clear of all pinch point areas.
Q
The IV pole folding hooks are designed to hold a maximum weight of 0.33 lbs (0.15 kg) or 150mL of fluid. Do not exceed the weight limit.
Q
Do not coil or hang cables from the IV pole.
Q
The pedestal is intended to support the injector head in an MR scanner room. Do not attempt to use the pedestal for any other purposes.
Electric Shock Hazard - Minor or moderate patient and/or worker injury may result. Q
Only plug the system into a direct AC outlet.
Q
Do not plug the system power cords into an extension cord or multi-outlet power strip.
Q
Use only approved adapters and accessories from Bayer.
Operation Manual
5
5 - 15
Understanding the Display and Pod
5.1 Home Screen
Figure 5 - 1: Home Screen Name
Icon (if applicable)
Description Shows the volume in the syringes. An outline of the syringe displays if no syringe is present.
1
Syringe Volume Information
2
Add Volume Indicator (if applicable)
If there is not enough volume in the syringe to complete the current protocol, the “Add” box displays the volume that needs to be added to complete the current protocol.
3
Protocol
Displays the name of the protocol. When modifications have been made to a protocol, an asterisk (*) appears to the right of the protocol name.
4
Phase Button (Edit Phase)
Displays the programmed phase type. Select to edit a phase type.
5
Flow Rate
Displays the programmed flow rate. Select to modify (if enabled).
6
Volume
Displays the programmed volume. Select to modify (if enabled).
7
Duration
Displays the duration of the programmed injection. Select to modify (if enabled).
8
Phase Button (New Phase)
Select to enter a new phase type.
9
Date and Time
Shows the current date and time.
If applicable, a dotted line displays on the syringe graphic to indicate that there is not enough volume in the syringe to complete the current protocol.