Stellant Service Manual Rev B
127 Pages

Preview
Page 1
201681 Rev. B
Table of Contents 1 - Introduction ............................................................................................................ 1-3 Copyright Notice ........................................................................................................... 1-3 Trademarks ................................................................................................................... 1-3 Restricted Sale .............................................................................................................. 1-3 Disclaimer ..................................................................................................................... 1-3 Address / Telephone ..................................................................................................... 1-4 Applicability ................................................................................................................... 1-4 Purpose ......................................................................................................................... 1-4 Important Safety Notice ................................................................................................ 1-4 Intended Use ................................................................................................................. 1-4 Certifications ................................................................................................................. 1-4 Symbols and Descriptions ............................................................................................ 1-5 WARNINGS ................................................................................................................ 1-12 CAUTIONS ................................................................................................................. 1-13 2 - Maintenance and Checkout ................................................................................ 2-15 Recommended Maintenance Schedule ...................................................................... 2-15 MEDRAD Service ....................................................................................................... 2-15 Inspection Procedures ................................................................................................ 2-16 Single and Dual Head .......................................................................................... 2-16 Syringe Heat Maintainer ...................................................................................... 2-16 Display and Base ................................................................................................. 2-16 Wall Bracket ......................................................................................................... 2-16 Adjustable Height Pedestal .................................................................................. 2-17 CS Mounting Systems ......................................................................................... 2-17 ISI ........................................................................................................................ 2-17 Electrical Leakage Check ........................................................................................... 2-18 Ground Continuity Check ............................................................................................ 2-18 Cleaning and Operational Checkout ........................................................................... 2-18 System Labels ............................................................................................................ 2-19 Power Up .................................................................................................................... 2-19 Programming .............................................................................................................. 2-19 3 - Troubleshooting .................................................................................................. 3-21 System Malfunction Codes ......................................................................................... 3-21 General Troubleshooting Guidelines .......................................................................... 3-21 System Messages ....................................................................................................... 3-22 Type 1 Messages ................................................................................................ 3-22 Type 2 Messages ................................................................................................ 3-22 Type 3 Messages ................................................................................................ 3-23 White Screen/System Lock Up Condition ................................................................... 3-24 Listing of Type 3 Errors ............................................................................................... 3-24 4 - Theory of Operation ............................................................................................ 4-33 Single Head Interconnect Drawing ............................................................................. 4-34 Dual Head Interconnect Drawing ................................................................................ 4-36 1
MEDRAD Stellant CT Injection System Display Unit ................................................................................................................. 4-38 Display - Block Diagram Description ........................................................................... 4-38 Display Top Block Diagram ......................................................................................... 4-40 PCMCIA Block Diagram .............................................................................................. 4-41 Base Unit - Block Diagram Description ....................................................................... 4-42 Three Phase Motor Function Operation ...................................................................... 4-42 Base Unit Block Diagram ............................................................................................ 4-44 Single Head Block Diagram Description ..................................................................... 4-45 Power Drive Card - Block Diagram Description .......................................................... 4-47 Single Head Block Diagram ........................................................................................ 4-48 Dual Head Block Diagram ........................................................................................... 4-49 Cable Wiring Pin Out Diagrams .................................................................................. 4-50 5 - Disassembly / Assembly & Replacement Parts ................................................ 5-51 Parts List Components ................................................................................................ 5-51 Symbols Guide ............................................................................................................ 5-52 Drawings ..................................................................................................................... 5-52 Display ................................................................................................................. 5-53 Base ..................................................................................................................... 5-65 Single Head ......................................................................................................... 5-74 Dual Head ............................................................................................................ 5-96 ISI 800 ............................................................................................................... 5-119 ISI 700 ............................................................................................................... 5-121 Pedestal ............................................................................................................. 5-123 Miscellaneous Replacement Parts ............................................................................ 5-124 2
Copyright Notice Trademarks Restricted Sale Disclaimer
1 - Introduction 1 - Introduction Copyright 2006 by MEDRAD, INC.. All rights reserved. No part of this manual may be reproduced in any form without prior written permission of MEDRAD. Printed and assembled in the U.S.A. MEDRAD Stellant CT®, Qwik-Fit Syringe®, FluiDot®, Quality for Life®, and MEDRAD® are registered trademarks of MEDRAD, INC. Federal (U.S.A.) law restricts the sale of this device on or by the order of a physician. Disclaimers MEDRAD makes no warranties on the contents of this manual, and specifically disclaims any implied warranties of merchantability or fitness for any purpose. MEDRAD reserves the right to change specifications and the contents of this manual without obligation. External wiring modification disclaimer: MEDRAD disclaims liability for any modifications or interfaces with other equipment which are not in conformity with the specifications and information contained within this manual. Such unauthorized action could jeopardize injector operation, safety, or reliability. Accessory equipment connected to the MEDRAD Stellant CT Injector must be certified according to IEC 60601-1 standards. Furthermore, all configurations shall comply with the system standard IEC 60601-1-1. Anyone who connects additional equipment to the signal input or output, configures a medical system, and is therefore responsible that the system complies with the requirements of the system standard IEC 60601-1-1. To obtain on-site consulting or consulting references, contact MEDRAD Service. All drawings in this manual are for reference purposes only, and may not reflect the construction of units produced prior to the publication of this manual. Reproduction quality of these drawings may have been effected by the level of reduction required. Call MEDRAD Service if assistance in drawing interpretation is required. The Stellant CT Injector is not for portable use.
1 - 3
MEDRAD Stellant CT Injection System
Address / Telephone
If you experience problems with the MEDRAD Stellant CT System, contact your MEDRAD authorized dealer or:
MEDRAD, INC. One Medrad Drive Indianola, Pa 15051-0780 U.S.A Phone: 412 767-2400 Fax: 412-767-4128
MEDRAD Europe B.V. Postbus 205 6190 AE Beek The Netherlands Phone: +31(0)43-3585601 Fax: +31(0)43-3656598
Nihon MEDRAD KK 9F Central Shin-Osaka Building 4-5-36, Miyahara, Yodogawa-ku Osaka, 532-0003 Japan Phone: +81-6-6350-0680 Fax: +81-6-6398-0670
Applicability Purpose Important Safety Notice Intended Use Certifications
This manual applies to the MEDRAD Stellant CT Injection System. The purpose of this manual is intended to provide instructions for servicing the MEDRAD Stellant CT Injection System safely and accurately. It is intended for those qualified to service the injection system, whether they be MEDRAD Service Personnel, Certified Laboratory Service Technicians or MEDRAD authorized international dealers. The information in this manual is intended for people with adequate backgrounds and experience in electronics and electromechanical devices. Any attempt to repair a sophisticated medical device such as the injector may result in personal injury, property damage, or patient injury. Refer to the Stellant CT Injection System operation manual. MEDRAD Stellant CT Injection Systems are equipped to operate at 100 - 240 VAC, 50/60 Hz, and designed to be in compliance with EN 60601 -1 (Safety), and EN 60601-1-2 (EMC/ Emissions).
1 - 4
Symbols and Descriptions
1 - Introduction The following international symbols are used on the MEDRAD Stellant CT Injector and throughout this manual. Attention, consult accompanying instructions. Indicates that this device conforms to the requirements of the European Medical Device Directive 93/42/EEC. Indicates on/off switch for the Control Room Unit. Indicates hazardous voltages. Indicates alternating current. Identifies a type BF applied part complying with EN 60601-1 standards. Indicates the injection system is Class 1 medical equipment as defined by EN 60601-1 standards. Identifies the degree of protection against fluid as drip proof. Identifies connection of the handswitch. Forward and Reverse Piston Controls. Identifies the direction of manual knob rotation relative to plunger movement. Identifies the ENABLE key. Indicates the AIR EXPELLED button on the injector head. When illuminated yellow on the touch screen, also indicates that the operator has acknowledged inspecting the fluid path for air. Identifies the Equipotential connection. Identifies the Earth Ground point. Identifies the Service Connection Port. Indicates design for indoor use only. 1 - 5
Indicates the presence of no serviceable parts.
MEDRAD Stellant CT Injection System
Identifies the Control Room Unit brightness controls. Indicates the Protocol Lock is engaged. Start / Hold. Stop. Reset. Help. Setup. Labels on the system or statements in this manual preceeded by any of the following words and/or symbols are of special significance, intended to help you to operate the system in a safe and successful manner: Protocol is unlocked or unengaged. Phase View. Profile Preview. Pressure Graph View.
1 - 6
1 - Introduction
ISI Symbols Indicates scanner hookup. Indicates DC power. Indicates injector hookup.
Identifies the CAN Interface TRANSMIT LED. Identifies the CAN Interface RECEIVE LED. Identifies the Ethernet LINK LED. Identifies the Ethernet ACTIVITY LED.
DO NOT dispose of in municipal waste. Wheeled bin symbol indicates separate collection for electrical and electronic equipment. (WEEE Directive 2002/96/EEC).
Symbols on the Power Switch:
I - On O - Off
The following icons may appear on the injector screen when using ISI 700: The user is not able to arm the injector. The injector is configured for ISI, and it is communicating to the ISI module. The injector and ISI module are functional, but the scanner signals are in a state that will NOT allow arming the injector. The injector is configured for ISI and it is communicating to the ISI module. The scanner signals are in a state that will allow arming the injector.
Communication is lost between the injector and the ISI module. OR The ISI module is in need of repair. Refer to the ISI System Messages section of this manual.
The following icons may appear on the injector screen when using ISI 800:
(No Highlight) Indicates that communication between the injector and the MEDRAD ISI 800 Module has been established.
(Highlighted in Yellow) Indicates that communication between the Injector and the scanner has been established.
(No Highlight) Indicates that communication has been lost between the injector and the MEDRAD ISI 800 Module. The MEDRAD ISI 800 Module may be in need of repair. Refer to the ISI 800 System Messages section of this manual.
(Highlighted in Yellow) Indicates that the injector is working in conjunction with the scanner ("coupled"). This symbol will only appear on the screen when the system is in Control or Tracking modes. (No Highlight) Indicates whether or not the injection protocol has been locked. When this icon is highlighted in yellow, the protocol has been locked.
1 - 7
MEDRAD Stellant CT Injection System NOTE: The configurations below are determined by the scanner manufacturer. Refer to the scanner manufacturer instructions for further information. Indicates that the injector will request the initiation of a procedure. When either or both system symbol is gray, that system is not ready and the injection will not proceed. When both symbols are highlighted in yellow, both systems are ready and the injection can proceed. Indicates that the scanner will request the initiation of a procedure. The Start Button has been locked out on the Injector. When either or both system symbol is gray, that system is not ready and the injection will not proceed. When both symbols are highlighted in yellow, both systems are ready and the injection can proceed. Indicates that the procedure request can be initiated from EITHER the scanner or the injector. When either or both system symbol is grayed out, that system is not ready and the injection will not proceed. When both symbols are highlighted in yellow, both systems are ready and the injection can proceed. 1 - 8
1 - Introduction
Additional Information Regarding Compliance to IEC 60601-1-2 / 2001 2nd Edition This section is intended to reflect conformance to IEC-60601-1-2 / 2001 2nd edition Standards. CAUTIONS The following statements are cautions. Cautions advise of circumstances that could result in damage to the device. Read and understand these cautions before operating the injector system. Injector may disarm or fail to operate when exposed to high magnetic fields. Portable and mobile RF communications equipment can affect the injector. FOR PROPER OPERATION, use only accessories and options provided by MEDRAD that are designed specifically for the injector system. Other non-MEDRAD approved accessories or options may cause equipment damage or may result in increased emissions or decreased immunity of the injector system. Injector system accessories listed in it’s operation manual comply with the requirements of electromagnetic emissions and immunity standards IEC-60601-1-2 / 2001 2nd edition. DO NOT USE INJECTOR ADJACENT TO OR STACKED WITH OTHER EQUIPMENT. If adjacent or stacked use is necessary, the injector should be observed to verify normal operation in the configuration in which it will be used.
Recommended separation distances between portable and mobile RF communications equipment and the injector
The injector is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the injector can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the injector as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter W
Separation distance according to frequency of transmitter m
150 KHz to 80 MHz d = [3.5/V1] p
80 MHz to 800 MHz d = [3.5/E1] p
800 MHz to 2.5 GHz d = [7/E1] p
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
1 - 9
MEDRAD Stellant CT Injection System
INJECTOR REQUIRES SPECIAL PRECAUTIONS REGARDING EMC. Install and put into service according to the EMC information provided below:
Guidance and manufacturer's declaration - electromagnetic emissions
The injector is intended for use in the electromagnetic environment specified below. The customer or user of the injector should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The injector uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2
Class B Stellant Class B Stellant
The injector is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Voltage fluctuations/flicker emissions IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The injector is intended for use in the electromagnetic environment specified below. The customer or user of the injector should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD) IEC 61000-4-2
+6 kV contact +8 kV air
+6 kV contact +8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with a synthetic material, the relative humidity should be at least 30%.
Electrical/fast transient/ burst IEC 61000-4-4
+2 kV for power supply lines +2 kV for power supply lines +1 kV for input/output lines +1 kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
+1 kV differential mode +2 kV common mode
+1 kV differential mode +2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles
<5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles
Mains power quality should be that of a typical commercial or hospital environment. If the user of the injector requires continuous operation during power mains interruptions, it is recommended the injector be powered from an uninterruptible power supply or battery.
70% UT (30% dip in UT) for 25 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 sec
<5% UT (>95% dip in UT) for 5 sec
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
3 A/m
3 A/m
NOTE UT is the a.c. mains voltage prior to application of the test level.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
1 - 10
1 - Introduction
Guidance and manufacturer's declaration - electromagnetic immunity
The injector is intended for use in the electromagnetic environment specified below. The customer or user of the injector should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the injector, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF IEC-61000-4-6
3 V rms 150 kHz to 80 MHz
3 V rms
d = 1.17 p
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
3 V/m
d = 1.17 p 80 MHz to 800 MHz
d = 2.33 p 800 MHz to 2.5 GHz
Where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the injector is used exceeds the applicable RF compliance level above, the injector should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the injector. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
1 - 11
WARNINGS
MEDRAD Stellant CT Injection System Introduction to Warnings and Cautions This manual contains important information about safe servicing of the MEDRAD Stellant CT Injection System. MEDRAD urges the service technician to read this manual carefully, become familiar with the procedures and system functions that it describes, and follow its recommendations to assure proper servicing of the system. Warning labels on the MEDRAD Stellant CT system or Warning statements in this manual preceeded by any of the following words and/or symbols are of special significance: WARNING: Indicates a potentially hazardous situation. If not avoided, this could result in death or serious injury. WARNING: Indicates electrical hazards which could result in death or serious injury. CAUTION: Indicates potential hazards or unsafe practices which could cause product, system, or property damage. NOTE: Indicates helpful information is being offered. Injury may result from exposure to hazardous voltages existing within the system. The system should be opened and serviced by qualified service personnel only. Disconnect the system from line power before cleaning or attempting to perform any maintenance. Explosion hazard in the presence of flammables. Do not use the system in the presence of anesthetic gases and equipment. Patient or operator injury can occur from use of worn power cords or control cables. Examine power cords and cables for cuts, frays, or any other visible damage. Do not use the system if any of the cords or cables show signs of damage. Any damaged or worn connection cables or power cords should be replaced. Unsafe operation may result from using improper accessories and replacement parts. Use only accessories, options, and parts designed for this system, and provided by MEDRAD. Fire hazard: to avoid an electrical fire, assure the correct type of fuse is used for replacement. The fuse must be replaced with type T, 250V, 3.15 A fuse by qualified personnel only.
1 - 12
CAUTIONS
1 - Introduction Damage may occur as a result of failure to follow electrostatic discharge (ESD) protection practices. ESD protection practices must be followed when servicing any component of this system. Damage can occur as a result of incorrect voltages. Check the voltage and frequency marked on the back of the Power Supply. Ensure that the outlet providing power to the injector supplies a voltage, frequency, and volt-ampere rating within the range specified on the unit. Environmental damage may result from improper disposal of system components or accessories. Electronic assemblies contain potentially hazardous materials. Follow all local regulations for the recycling or disposal of electronic assemblies, or contact MEDRAD Service for assistance. Damage can occur as a result of abrupt interruption or application of supplies. To avoid damage to sensitive circuits on the boards, disconnect the power cord before removing or replacing PC boards. Allow system temperature to stabilize before use. To avoid damage to sensitive electronic circuits, allow the system to stabilize to room temperature before servicing when exposed to extreme temperature changes. Perform regular preventive maintenance. To ensure that your MEDRAD Stellant CT System remains properly calibrated, and that all primary and backup circuits are functioning properly, regular preventive maintenance is recommended. An annual preventive maintenance package is not included in the new machine warranty. Contact your local MEDRAD Service Representative for details. Damage may result from improper or careless cleaning methods. While cleaning any outside portion of the system, avoid allowing any water to seep inside system components. NOTE: All relevant guidelines for institutional, local, or national safety recommendations related to cable routing and installation should be followed.
1 - 13
MEDRAD Stellant CT Injection System 1 - 14
2 - Maintenance and Checkout
2 - Maintenance and Checkout
This section contains recommended procedures for maintenance, and an operational checkout of the MEDRAD Stellant CT Injection System. Routine maintenance and inspection will:
• Ensure continued performance of the injection system • Reduce the possibility of equipment malfunction
Recommended Maintenance Schedule
Your MEDRAD Stellant CT Injection System must be properly maintained to ensure that it is in peak operating condition. Your individual maintenance system and schedule depends upon how your injection system is used, the type of procedures performed, and frequency of use. The following maintenance schedule is recommended for the system: Daily: The piston rod or rods should be thoroughly cleaned after each use. Before use each day, the system should be cleaned and inspected, using the procedures outlined in this section. Ensure that all system safety and warning labels are in place and are legible. Monthly: Once a month, the entire system should be thoroughly inspected and cleaned, and an Operational Checkout should be performed. Annually: As part of an annual maintenance program performed by a qualified MEDRAD Service Representative or authorized dealer, both Electrical Leakage and Ground Continuity checks should be performed. NOTE: Local regulations or hospital protocol may require electrical leakage checks at more frequent intervals. If this applies, local regulations for leakage must be followed. MEDRAD also recommends that a complete system calibration and performance checkout be performed annually. Contact MEDRAD Factory Service, or your local MEDRAD office for complete details. In the United States, Canada, Japan, and Europe, the MEDRAD Service Department offers Preventive Maintenance Programs. These annual programs greatly assist in maintaining accuracy and reliability, and can also extend the life of the system. Contact MEDRAD for details. In Europe, contact your local MEDRAD office or your local authorized dealer for further information. Refer to the back of the title page of this manual for address, telephone and FAX information. NOTE: Failures which occur due to lack of proper maintenance will not be covered under warranty.
MEDRAD Service
MEDRAD Service will make available upon request:
• Circuit diagrams, component parts lists, or other information that will assist qualified technicians to repair components classified as repairable. • On site consulting upon request
2 - 15
MEDRAD Stellant CT Injection System
Inspection Procedures
The following procedures are recommended for daily inspection of all components in the MEDRAD Stellant CT Injection System. If any defects are detected, either repair the system, or call MEDRAD for service. Do not use the system until the problem is corrected.
Single and Dual Head
1. Inspect the housing for any damage or cracks that could allow fluid to leak inside, or weaken the structural integrity of the unit. 2. Inspect all cables connected to the unit: Look for cuts, cracks, worn spots or other obvious damage to the cables. Ensure that all connectors are properly seated. 3. Inspect for contrast media build-up in the syringe interface area. Follow the cleaning guidelines outlined in this section. 4. Ensure that all mounting bolts and screws are secure. 5. Inspect the pivot points. The head and support arm must pivot freely. 6. Ensure that the keypad is clean and free of cracks.
NOTE: All relevant guidelines for institutional, local, or national safety recommendations related to cable routing and installation should be followed.
Syringe Heat Maintainer
1. Ensure that the device is warm to the touch while attached to the head. 2. Ensure that the LED indicator is not illuminated or flashing. 3. Inspect the cable and conncector for cracks, worn areas, or other obvious damage. 4. Ensure the heater has extension cable attached P/N 3013737.
Display and Base
1. Inspect all cables connected to the unit: Look for cuts, cracks, or worn spots, or other obvious damage. Ensure that all connectors are properly seated. 2. Inspect the housing for any damage or cracks that could allow fluid to leak inside, or weaken the structural integrity of the unit.
Wall Bracket
1. Inspect all parts of the bracket for cracks and other defects that would weaken the assembly. 2. Ensure that the bracket is securely attached to the wall. 3. Ensure that all cables are secured to the display control unit and do not interfere with the movement of the mounting bracket.
2 - 16
Adjustable Height Pedestal
2 - Maintenance and Checkout 1. Inspect the stand, base and support arm for cracks and other defects that could weaken the structure. 2. Ensure all mounting bolts and screws are secure. 3. Ensure that the casters roll smoothly with no binding or scraping. 4. Ensure all locking mechanisms on the casters are functional. 5. Verify that the vertical height adjustment of the column shaft moves freely without binding or scraping.
CS Mounting Systems
1. Inspect all parts of the arm and mounting system for cracks and other defects that would weaken the system. 2. Ensure that the mounting system is securely assembled, with no loose parts. The area should be stable with the head installed. 3. Ensure that the arm moves smoothly in all directions, with no binding or scraping. 4. Verify that all cabling is tied back and does not interfere with the movement of the supporting parts or the injector head.
NOTE: All relevant guidelines for institutional, local, or national safety recommendations related to cable routing and installation should be followed.
ISI
1. Inspect all parts of the ISI, ensure all cables are secured to the ISI.
2. Ensure ISI cables are free of cuts, cracks or other obvious damage.
2 - 17
MEDRAD Stellant CT Injection System
Electrical Leakage Check Ground Continuity Check Cleaning and Operational Checkout
To ensure safe operation of the MEDRAD Stellant CT Injection System, an electrical leakage check must be part of regular maintenance. Use a commercial leakage tester such as one of the following:
MANUFACTURER Bio-Tek Instruments, Inc. Electrical Safety Analyzer Bender Bapco
MODEL Model 601 PRO Unimet 1000 ST IEC601L
1. With the AC ground open, power applied, and the line at normal, leakage should be less than 100 micro amps at 110V or 300 micro amps at 220 V. 2. With the AC ground open, power applied and the line reversed, leakage should be less than 100 micro amps at 110V or 300 micro amps at 220 V. 3. Disconnect the leakage test device.
A ground continuity check must also be part of regular maintenance of the MEDRAD Stellant CT system. 1. Disconnect the system from the power source. 2. Using an ohm meter, measure the resistance between the ground terminal, the power cord and the equipotential connector on the Base and the front casting of the head. The resistance measured must be less than 0.2 ohms. Cleaning Guidelines Deposits of contrast media can interfere with proper operation of the MEDRAD Stellant CT Injection System. The following guidelines should be followed when removing deposits, or cleaning any portion of the system. WARNING: Serious injury or death may result from exposure to hazardous voltages existing within the system. Disconnect the system from line power before cleaning or attempting to perform any maintenance. Ensure that the system is completely dry before connecting to the power source and applying power. CAUTION: Improper or careless cleaning methods may result in equipment damage. Do not soak or immerse any part of the injection system in water. While cleaning any outside portion of the system, avoid allowing any water to leak inside system components. • If contrast medium has leaked inside any component of the system, the affected subassembly should be disassembled and cleaned. This cleaning procedure can be done in the field by trained MEDRAD Service personnel, or returned to MEDRAD Service. • Care must be taken not to get water or cleaning solutions inside any system components. Do not use strong industrial cleaning agents or solvents such as acetone. Warm water and a mild disinfectant such as antibacterial hand soap are all that is required. • To clean the syringe interface area of the injector head, fully retract the piston. Using a paper towel moistened with warm water or a mild disinfectant, gently wipe the inner syringe installation area. Do not insert any sharp instruments into this area during the cleaning process.
2 - 18
System Labels Power Up Programming
2 - Maintenance and Checkout
Operational Checkout A basic functional checkout of the MEDRAD Stellant CT Injection System should be included as part of regular maintenance. Verifying proper operation of the injection system will help in detection of any problems that may not be noticed in day to day operation. The following procedure represents a suggested series of activities which encompass typical operation of the system. Read the following procedure carefully before beginning the checkout. If problems are detected, refer to the General Troubleshooting Procedure. CAUTION: Any problems detected during this or any other procedure should be corrected before using the injection system in patient procedures.
Ensure that all system safety and warning labels are in place and legible. 1. Apply power to the system. 2. Verify that the Safety screen is displayed after system diagnostics occur. 3. Press Continue to acknowledge the messages on the Safety screen. 4. Verify that audible tones are functioning on the Display Control Unit and injector head. 5. Verify that the following controls are functioning properly. • Forward and Reverse • Start • Stop • Fill 6. Verify that the arm lights on the injector head are functioning. 7. After the Main screen is displayed, verify that the following controls are functioning properly. 8. At the rear of the Display, press the Display Brightness Up key until the screen is lightened to its fullest extent. Press the Display Brightness Down Key until the screen is darkened to its fullest extent. Adjust the screen appearance to return to a desirable contrast level. 9. Fully advance and reverse the pistons by using the ENABLE key and the forward/reverse controls. 10. Ensure the Auto Dock, Auto Advance, and Auto Retract are enabled. 11. Verify that both pistons respond to the forward and reverse controls. Enter the following protocol: Dual Flow Rate
Phase Phase 1: Phase 2: Phase 3: Phase 4:
Syringe Syringe A: Syringe B: Syringe A: Syringe B:
Flow Rate 10 ml/s 2.5 ml/s 5.0 ml/s 0.1 ml/s
Volume 70 ml 29 ml 100 ml 1 ml
2 - 19
MEDRAD Stellant CT Injection System
Single Flow Rate
Phase Phase 1: Phase 2: Phase 3: Phase 4:
Syringe Syringe A: Syringe A: Syringe A: Syringe A:
Flow Rate 10 ml/s 2.5 ml/s 5.0 ml/s 0.1 ml/s
Volume 70 ml 29 ml 100 ml 1 ml
12. Install empty syringes(s) and ensure the piston automatically docks with syringe plunger. The plunger(s) should advance to full forward position. 13. Fully reverse the plunger(s). 14. Arm in single mode and inject. In one of the phases, activate the HOLD feature for at least 10 seconds. 15. Press Start/Hold again. Verify the injection completes normally, and the Injection Complete screen values are similar to the protocol values. 16. Advance plunger(s) to the full forward position, remove syringes(s) and ensure the piston(s) automatically retracts. 17. Inspect the heat maintainer. Make sure it is warm and the fault indicator is not illuminated while attached to the head.
2 - 20
3 - Troubleshooting
System Malfunction Codes General Troubleshooting Guidelines
3 - Troubleshooting Conditions can occur which will prevent the injection system from being armed, or even interrupt an injection that is in progress. These conditions may be operator induced or caused by a system malfunction. Error Codes which inform the user of these conditions are displayed on the control panel. Consider the following guidelines before troubleshooting any condition. These guidelines may help in resolving the condition quickly: Remember, try the simple things first. CAUTION: Damage may occur as a result of failure to follow electrostatic discharge (ESD) protection practices. ESD protection practices must be followed when servicing any component of this system. CAUTION: Damage could result from improperly handled components. Before touching any of the circuit cards in the system, discharge yourself to grounded metal. If memory components are to be shipped, place the components in conductive carriers (as supplied through MEDRAD). CAUTION: Disconnect the power cord before removing or replacing PC boards. Sensitive circuits on the boards can be damaged by abrupt interruption or application of supplies. • Try removing the power cord from the power source for one minute. Allow the system to reset completely, then reapply power and retry. The condition could be intermittent, or caused by a voltage transient. If the condition persists, continue troubleshooting. • To verify the existence of a condition, attempt to recreate the problem. Follow the Checkout Procedure outlined in Section 2 of this manual to check for proper (or improper) operation of the system. • Some faults can be caused by a noisy electrical environment. If these conditions persist, contact MEDRAD Factory Service* for further assistance. * Indicates contact MEDRAD Factory Service or an Authorized Dealer.
3 - 21
System Messages Type 1 Messages
MEDRAD Stellant CT Injection System The system will display messages on the screen as conditions or events occur. There are three basic types of messages: Type 1 messages provide information regarding the current system status, and will clear automatically. These messages typically display in the lower right corner of the screen.
Type 2 Messages
Type 2 messages convey information that must be explicitly acknowledged before proceeding. Press the button(s) within the yellow dialog box to dismiss the message.
3 - 22
Type 3 Messages
3 - Troubleshooting Type 3 messages are system malfunction messages which require power to be removed from the system. Some Type 3 messages provide suggestions to prevent the condition from recurring. If the condition cannot be corrected, record the code and number from the lower left corner of the dialog box, then call MEDRAD Service for assistance. WARNING: Patient injury may result from a system malfunction. If a system malfunction occurs, immediately remove and disconnect the system from the patient. If a fault message is displayed that cannot be corrected, and/or the system is not operating correctly, do not use the injection system. Call MEDRAD for assistance.
Type 3 errors are divided into categories of function level. Each category is also divided into specific errors. Below is a list of the categories and suggested repair sequences. These are to be tried in order, not to be performed all at once. Before replacing any parts, cycle power to injector. This will initiate a system self-test. If this does not correct the problem, replacement of the Display, Base, or Head as needed is suggested. For further assitance, contact MEDRAD Service or an authorized dealer. NOTE: Replacement of the Servo/CPU Card requires calibration/verification.
3 - 23
MEDRAD Stellant CT Injection System
White Screen/System Lock Up Condition Listing of Type 3 Errors
A white screen that displays, “MEDRAD” indicates a communication or software problem. • Software mismatched between Display and Head. Software update needed. • Faulty Head extension cable. Replace Head extension cable. • Replace the Servo/CPU card CAL - Calibration Errors with a prefix of CAL mean that the system has found an error with the system calibration values missing or corrupt. • Calibrate system potentiometer, flex ring and pressure. • Replace the Servo/CPU Card. DOCK SENSA - Docking Senor A Errors with a prefix of DOCK SENSA mean that the system has detected that Piston A docking sensor is closed with no syringe attached on side A. • Verify syringe is installed correctly. • Clean syringe interface area. • Replace size Sensor A. DOCK SENSB - Docking Senor B Errors with a prefix of DOCK SENSB mean that the system has detected that Piston B docking sensor is closed with no syringe attached on side B. • Verify syringe is installed correctly. • Clean syringe interface area. • Replace size Sensor B. ENCA - Encoder A Errors with a prefix of ENCA indicate a problem with the Encoder on motor A or the associated circuitry. • Ensure P4 is connected. • Replace the Power Drive card. • Replace the motor. ENCB - Encoder B (Dual only) Errors with a prefix of ENCB indicate that the system has found a problem with the Encoder on motor B or the associated circuitry. • Ensure that P4 is connected correctly. • Replace the Power Drive card. • Replace the motor.
3 - 24
3 - Troubleshooting FRA - Flex RingA Errors with a prefix of FR indicate a problem with the Flex Ring on side A. • Clean housing. • Replace housing FRB - Flex RingB Errors with a prefix of FR indicate a problem with the Flex Ring on side B. • Clean housing. • Replace housing HOVER - Switch Card Errors with a prefix of HOVER indicate that the system has found a problem with the head, DCU overlay switch card or the associated circuitry. • Ensure there where no switches pressed during power up diagnostic test. • Ensure that the head cable is seated properly. • Replace the switch card of the module posint the error. • Replace the Servo/CPU card. HSW - Handswitch Errors with a prefix of HSW indicate a problem with the Handswitch or associated circuitry. • Ensure handswitch is is properly seated. • Replace handswitch. • Replace CPU card in moduel where handswitch is installed. HUB - Ethernet Hub Errors with the prefix HUB, have detected a problem in the Ethernet HUB or the associated circuitry. • Ensure all cables are fully connected into the HUB (power and communication). • Replace the HUB. ICD - Interface/Switch Card In Display Errors with a prefix of ICD indicate a problem with the Display Interface Card in the Display top section or the associated circuitry. • Ensure the card has all the connectors fully seated. • Replace the Display Interface Card. • Replace the SBC Card. ICE - Interconnect Card Base Errors with a prefix of ICE indicate a problem with the Expansion Base Interface Card or the associated circuitry. • Ensure that the card connectors are fully seated. • Replace the Base Interface Card. 3 - 25
MEDRAD Stellant CT Injection System ICS - Interface Card In Head Unit Errors with a prefix of ISC indicate a problem with the Head Interface Card(s) or the associated circuitry. • Ensure that the Head connectors are fully seated. • Replace the Head Interface Card. • Replace the Servo/CPU card. MOTA - Motor A Errors with a prefix of MOTA indicate a problem with motor A or the associated circuitry during the H Bridge test. • Ensure that P3 is connected correctly onto the Power Drive card. • Replace the Power Drive card. • Replace the motor. MOTB - Motor B (Dual only) Errors with a prefix of MOTB indicate a problem with motor B or the associated circuitry during the H Bridge test. • Ensure that P3 is connected correctly onto the Power Drive card. • Replace the Power Drive card. • Replace the motor. PCMCIA - PCMCIA Card Errors with a prefix of PCMCIA indicate a problem with the PCMCIA card or the associated circuitry. • Remove and re-apply power to the unit. This will trigger a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • Ensure that the card is fully seated into the SBC card. • Replace the PCMCIA card. • Replace the SBC card. POTA - Pot A Errors with a prefix of POTA indicate a problem with the Potentiometer A or the associated circuitry. • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • Ensure the connector is fully seated on the interface card. • Realign potentiometer A. • Replace potentiometer A. • Replace Head Interface Card. • Replace the Servo/CPU card. 3 - 26
3 - Troubleshooting POTB - Pot B (Dual only) Errors with a prefix of POTB indicate a problem with Potentiometer B or the associated circuitry. • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • Ensure the connector is full seated on the Interface card. • Realign potentiometer B. Align with gear train and recalibrate. • Replace Potentiometer B. • Replace Head Interface Card. • Replace the Servo/CPU card. POWR - Power Drive Card Errors with a prefix of POWR indicate a problem with the power card or the associated circuitry. • Verify power supply voltages are at the Power Drive card. • Ensure that connectors are fully seated. • Replace the Power Drive card. • Replace the Servo/CPU Card. PRSSRA - Pressure Transducer A Errors with a prefix of PRSSRA indicate a problem with Pressure Transducer A or the associated circuitry. • Verify Power Supply voltages are a the Power Drive Card. • Ensure the Transducer connector is fully seated on the Interface Card. • Replace the Interface card. • Replace the Servo/CPU card. PRSSRB - Pressure Transducer B (Dual only) Errors with a prefix of PRSSRB indicate a problem with the Pressure transducers B-side or the associated circuitry. • Ensure the Transducer connector is fully seated on the Interface Card. • Replace the Interface card. • Replace the Servo/CPU card. RATE - Over Rate Injection Errors with a prefix of RATE indicate that the system has detected an over rate condition. • Verify Calibration • Ensure potentiometers are fully seated into gears. • Replace potentiometer of the side with the problem. • Replace the motor of the side with the problem. • Replace the Servo/CPU card. 3 - 27
MEDRAD Stellant CT Injection System SBC - Single Board Computer (SBC) Errors with a prefix of SBC indicate a problem with the Single Board Computer or the associated circuitry. • Ensure the card has all the connectors fully seated. • Replace the SBC card. • Replace the Display Interface Card. SER - Servo/CPU Card Errors with SER typically indicate the Head Card. This may be a condition that must be corrected, or simply an electrical transient that disrupted the system. Try the following: • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • If the problem continues, replace the Servo/CPU card. • If the condition persists, call MEDRAD Factory Service. SSD - Start Switch (Display) Errors with a prefix of SSD typically indicate a problem with the Start Switch. If a Start Switch is not present, contact MEDRAD Service. If one is present, try the following to return the injector to service: • Check the Start Switch connection at the Display to assure the connector is fully seated. If the connection is loose, remove and reattach the connector. • Cycle power to the unit to reset the system. • Disconnect the patient and perform a trial injection using the Start Switch. If the system performs correctly, return the system to use. • If the error reappears, remove the Start Switch. • Cycle power. If the error is resolved, order a replacement Start Switch. • If the error persists, contact MEDRAD Service. SSH - Start Switch (Head) Errors with a prefix of SSH typically indicate a problem with the Start Switch. If a Start Switch is not present, contact MEDRAD Service. If one is present, try the following to return the injector to service: • Check the start switch connection at the Head to assure the connector is fully seated. If the connection is loose, remove and reattach the connector. • Cycle power to the unit to reset the system. • Disconnect the patient and perform a trial injection using the Start Switch. If the system performs correctly, return the system to use. • If the error reappears, remove the Start Switch. • Cycle power. If the error is resolved, order a replacement 3 - 28
3 - Troubleshooting SW - Software Problem Errors with SW typically indicate a problem with internal software. This may be a condition that must be corrected, or simply an electrical transient that disrupted the system. Try the following: • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • If the problem continues, contact MEDRAD to reload software. • If the problem continues, replace the SBC card. • If the problem continues, replace the Servo/CPU card. SWMTCH - Software Mismatch Errors with SWMTCH typically indicate a problem with internal software not matching between the Head and the Display. This may be a condition that must be corrected, or simply an electrical transient that disrupted the system. Try the following: • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • If the problem continues, contact MEDRAD to reload software. SWSBC - CPU software Errors with a prefix SWSBC typically indicate a problem with internal software on the display CPU card. Try the following: • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • If the problem continues, contact MEDRAD to reload software. • If the problem continues, replace the SBC card. SWSER - Servo software Errors with a prefix SWSER typically indicate a problem with internal software on the Head CPU card. Try the following: • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • If the problem continues, contact MEDRAD to reload software. • If the problem continues, replace the Servo/CPU card. 3 - 29
MEDRAD Stellant CT Injection System SYR SENA - Size Sensor A Errors with a prefix of SSSA indicate a problem with the size sensor on A side or the associated circuitry. • Ensure no syringes are on during self test. • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • Ensure the connector between the Syringe Size Sensing Module and the Head Interface Card is fully seated. • Replace the Size Sensing Assembly. • Replace the Interface Card. SYR SENB - Size Sensor B (Dual only) Errors with a prefix of SSSB indicate a problem with the size sensor on B side or the associated circuitry. • Ensure no syringes are on during self test. • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • Ensure the connector between the Syringe Size Sensing Module and the Head Interface Card is fully seated. • Replace the Size Sensing Assembly. • Replace the Interface Card. TCHSCR - Touch Screen Errors with a prefix of TCHSCR indicate a problem with the touch screen or the associated circuitry. • Ensure that nothing is pressing the touch screen during self test. • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • Recalibrate the Touch Screen • Insure ribbon cable for the Touch Screen to the SBC is fully connected. • Replace the Touch screen. • Replace the SBC card. UOM - Unintended opposite motion Errors with the prefix UOM indicate that a software routing running the DSP did not report in the allowed time. • Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify proper unit operation. If the problem is corrected, there was probably a voltage transient in the unit, now cleared, enabling system use. • Replace the Servo/CPU card. 3 - 30
3 - Troubleshooting VOL - Over Volume Injection Errors with a prefix of VOL indicate that the system has detected an over volume condition. • Ensure potentiometers are fully seated into gears. • Replace potentiometer of the side with the problem. • Replace the motor of the side with the problem. • Replace the Servo/CPU card. 3 - 31
MEDRAD Stellant CT Injection System 3 - 32
4 - Theory of Operation
4 - Theory of Operation The Stellant system consists of 3 basic parts: • Display Unit • Base Unit • Head Unit (Single or Dual) Each component in the system is interconnected with an external cable from the Base Unit, the system center. In this section, basic circuit block descriptions of each of the sub-systems are explained. A block diagram of each primary component is followed by a functional block description for each of the sub-systems.
Description
Page
Single Head Interconnect Drawing
4-32
Dual Head Interconnect Drawing
4-34
Display Block Diagram Description
4-36
Display Top Block Diagram
4-38
PCMCIA Block Diagram
4-39
Base Unit Description
4-40
Three Phase Motor Function
4-40
Base Unit Block Diagram
4-42
Single Head Block Diagram Description
4-43
Power Drive Card Block Diagram Description
4-45
Single Head Block Diagram
4-46
Dual Head Block Diagram
4-47
Cable Wiring Pin Out
4-48
4 - 33
MEDRAD Stellant CT Injection System Single Head Interconnect Drawing 4 - 34
Single Head Interconnect Drawing
4 - Theory of Operation
4 - 35
MEDRAD Stellant CT Injection System Dual Head Interconnect Drawing Dual Head Interconnect Drawing 4 - 36
4 - Theory of Operation 4 - 37
MEDRAD Stellant CT Injection System
Display Unit Display - Block Diagram Description
The Display interface board is the primary PCB in the Display enclosure. Support circuits are placed on this PCB including the physical support for the SBC. All connectors and switches, such as the start switch and interface cable from the Base are terminated at the Display interface card. Brightness Control Two buttons. One button sends a voltage to the up count of an up-down counter, increasing the count to increase the amount of voltage sent to the inverter card; the other button causes the counter to count down, decreasing the voltage sent to the inverter card. This inverter card output is also sent to the SBC card for monitoring. CCFT Inverter The brightness on the LCD is developed by the use of two CCFT (Cold Cathode Filament Tube). The CCFT inverter supplies a regulated, adjustable AC potential across the CCFT to start and maintain fluorescence. The brightness control input of 0-3 VDC is inversely proportional to the intensity of light emitted by the CCFT. +12VDC Regulator The +15 VDC power supply feeds power to a regulator providing a +12 VDC power level to the CCFT inverter board required to light the CCFT on the LCD panel and supplies power to the audio amplifier. +5VDC Regulator The +15 VDC power supply feeds power to a regulator providing a +5 VDC power level to most logic circuits. +3.3VDC Regulator The +15 VDC power supply feeds power to a regulator providing a +3.3 VDC power level to most logic circuits. +12VDC Audio Regulator The +15 VDC power supply feeds power to a regulator providing a +12 VDC power level to the CCFT inverter board required to light the CCFT on the LCD panel and supplies power to the speaker amplifier.
4 - 38
4 - Theory of Operation Single Board Computer (SBC) The SBC is the master controller for system functions and system communications. • Ethernet interface: 10Base T • PCMCIA Interface* • Power Management • LCD Panel Interface Speaker Amplifier The amplifier is driven by an audio signal provided by the SBC card the signal is amplified and sent to the speaker located on the rear cover. Volume and pitch are controlled by the SBC card. PCMCIA Card This card contains the Real Time Clock for the system, 32 Kbytes of Non-Volatile (NV) ram and 256 Kbytes of Boot Flash. Start Switch The start switch signals are buffered by the Display interface card and sent to the SBC card. The signals are fed through a set of normally open and normally closed contacts along with Start Switch Present line. Arm light drive for Start Switch The start switch contains an LED driven from a drive circuit on the Display Interface card from the SBC. Touch Screen The touch screen is an analog output device that is run directly from the SBC card. LCD The primary visual interface to the operator is a 640 x 480 Color Liquid Crystal Display (LCD) panel with a physical dimension of 10.4” diagonally. This is driven directly from the SBC card. On/Off Switch The on/off switch provides the means to power down the system. The +15 VDC main power is switched through a single pole, single throw lighted switch. The light is run off +5VDC DCU Copilot This signal enables and disables the +5VDC regulator on the Base Interface Card. Power Down Indicator This circuit provides an early warning to the processor when the +15VDC supply rail decays to +12VDC. *Depending on system configuration, this may not be present. 4 - 39
Display Top Block Diagram
MEDRAD Stellant CT Injection System
4 - 40
PCMCIA Block Diagram
4 - Theory of Operation
4 - 41
MEDRAD Stellant CT Injection System
Base Unit - Block Diagram Description Three Phase Motor Function Operation
This unit supplies and controls DC power for the system. It is a communicaton interface between the Head and the Display and any external communication inputs.
Ethernet Hub The Ethernet hub is the connection point for all Ethernet communications. The network is arranged in a star topology with each segment (SBC, FO communications, Service Port, and an external interface) connecting to a port on the hub. Packets received on one port are copied to all other ports, allowing all ports on the network to see all packets. The hub is powered by the 5 volts from the Base interface cards.
Power Sense The Power sense circuit watches a dedicated digital line (DCU Co pilot) between the Display and Base for an indication that the Display unit has been turned on. The circuit then energizes the 5 VDC power to the Base circuitry and HUB.
48 V Switch Circuitry From the power supply, 48 volts is fed onto the base interface card. On the base interface card it is filtered and monitored so that it does not fluctuate from 48 volts. If the 48V drops to 42V for more than 2 seconds, the monitoring circuits disable the switching circuit to shut off the 48 volts feeding the head. Normal operation of the 48V switch cirucuit is controlled by the power sense line via the 5 volt VCC. Without 5 volts VCC, the 48 volts will not be switched to the head.
To understand how the Drive Card works, one must have a good understanding of how the entire servo system works, which includes the Servo/CPU Card, the Base Interface card and the BLDC (Stepper) motor. One must also understand BLDC motor principles and commutation techniques. This is a basic explanation on how motor stator excitation is performed, so as to produce rotary motion.
COMMUTATION TABLE CW ROTATION
POWER AMPLIFIER OUTPUT
∅A
∅B
∅C
L
H
X
L
X
H
X
L
H
H
L
X
H
X
L
X
H
L
SYRINGE A BLDC MOTOR
A
C
B
PHASE C PHASE B PHASE A
LEGEND:
L - Low Side H - High Side X - OFF
ACH ABH
The commutation table shown above shows the AAH
switching order that the power amplifier FETs
must follow in order to cause clockwise motor AAL
shaft movement. The first 60× of rotation is
ABL ACL
accomplished by setting ABH and AAL high
while holding all other signal low. Current is
sourced from Phase B of the bridge, and sunk
through Phase A as shown in Figure 1.
Figure 1
4 - 42
4 - Theory of Operation
The next 60° of rotation is caused when ABH is shut off, AAL is left on, and ACH is switched on. Current sinking will still occur through the Phase A motor winding, however, current will now be sourced through Phase C as shown in Figure 2. SYRINGE A BLDC MOTOR
A
C
B
PHASE C PHASE B PHASE A
ACH ABH AAH
AAL ABL ACL Figure 2 Once complete, then AAL is switched to the off state, and ABL is switched on, while ACH is kept on. The current sinking path is now transferred to Phase B. By following the commutation table, one can see that for a 360× cycle, each switching device is held on for 120× thus allowing current to flow through that particular phase for 120×. Figure 3 is a graphical display of the current flow through the motor windings for the given example.
MOTOR CURRENT
Phase A Phase B Phase C
ELECTRICAL DEGREES 0-59 60-119 120-179 180-239 240-299 300-359
Figure 3
4 - 43
Base Unit Block Diagram
MEDRAD Stellant CT Injection System
4 - 44
4 - Theory of Operation
Single Head Block Diagram Description
The head consists of 6 separate cards and the mechanical drive. The cards are: • Power Regular card • Servo/CPU card • Head interface card • Power Drive card (2 needed for a dual head) • LED Card • Syringe size sensing assembly (Two needed for Dual) Power card The power card is to regulate power from the base unit and distribute it ot the other cards. Block diagram Descriptions Heat Maintainers The heat maintainer runs on the 48 V supplied from the base units. 5V DC Regulator This cirucuit uses a 5V DC switching regulator to produce a +5V supply for the head circuitry. This voltage is derived from the +15V DC from the base. Servo/CPU Card The CPU/Servo card contains a microprocessor that performs the following functions in the head: • Card power management • Data management I/O - two ports • Footswitch control • Network communications in conjunction with the Tower • Motor Drive control • JTAG - programming port - not used. Head Interface Card The Head card is the electrical interconnect point between the internal electronics of the head to the Servo CPU card. Block Diagram Descriptions POT_REF Regulator This circuit uses a +4.096 VDC reference to produce the +4.096V at the 0.05 Amps reference voltage for the potentiometers. This voltage is derived from +5 VDC. Potentiometer Buffers The Pot signals are routed to the Servo/CPU. This signal is buffered and filtered on the Head interface card before being routed to the Servo/CPU. Pressure V+ Loopback This signal monitors a reference voltage from the top of the pressure bridge. The reference is an output from the pressure sensor and is buffered before being sent to the Servo/CPU card where it is monitored as a critical voltage.
4 - 45
MEDRAD Stellant CT Injection System Pressure Buffered The signal from the pressure sensor is sent to the Servo/CPU card. It is buffered and filtered on the Head Interface card. Pressure A Syringe This circuit uses a full Wheatstone bridge arrangement to produce a differntial signal which corresponds to the force applied to the rear plate due to the pressure in the syringe. This signal is sent to the Servo/CPU card. Start Switch Interface Start switch signals are conditioned and routed to the Servo/CPU card through the Head Interface card. The signals pass through a set of normally open and normally closed contacts along with the Hand Switch Present line. Hand Switch Arm Light Drive The Hand Switch Arm Light drive signal comes from the Servo/CPU card to a driver on the Head Interface card. LED Board LEDs on the board are driven by the A driver on the Head Interface card. Potentiometer A The Pot provides an analog signal from 0 to 4.096 VDC indicating the absolute position of the piston during power up. It also serves as a backup to the syringe encoder during a move. This signal is buffered and filtered in the Head Interface card and passed to the Servo/PCPU card. Flex Ring Sensor The Flex Ring system is an electro/mechanical design this is used by the Stellant Dual and Single Syringe injector to determine if a syringe or adapter has been inserted or removed from the injector assembly. Autodocking Sensor The Autodocking sensor consits of a spring energized mechinical pin whie, when depressed, interrupts an optical path, thereby closing a transistor switch. The sensor is used to determine when the piston has encountered the rear of the syringe, known as the plunger. Speaker Drive - Head The speaker on the head is a piezo device (buzzer) that is driven by an external oscillator residing on the Head Interface board. A resistor controls the oscillator frequency. Syringe Size Sensing (SSS) The SSS Module consists of 2 PCBs containing an LED and an array of photodiodes, as well as various alignment/focus mechanical components. The LED illuminates the Syringe whie, through a series of grooves, disperses the light in a characteristic pattern unique for a particular syringe, or Pre-filled Adapter. 4 - 46
Power Drive Card Block Diagram Description
4 - Theory of Operation SYRINGE A PWM ISOLATION This block provides electrical isolation of the Pulse Width Modulation (PWM) switching signals for Syringe A. These signals are generated on the Servo/CPU Card via DSP A and are used to command and control the power amplifier current delivery to the motor. SYRINGE A DRIVER The Syringe Driver stage performs voltage translation and current capacity enhancement to the PWM signals for the Syringe motor. Once the PWM signals have passed through the isolation barriers, they are routed to the Driver circuits. It is the responsibility of the Driver to translate the 0-5V PWM signals to 0-12V on the low channels, and 0-12V plus boost voltage on the high side. The Driver has the capacity of sinking and sourcing high switching currents normally associated with power FETs. The Drive Inhibit signal disables the Driver output signals. When Drive Inhibit is active high, the outputs are disabled. When low, the outputs are enabled. The Driver has builit-in shoot-through protection to inhibit high side and low side outputs from being turned on simultaneously. SYRINGE A POWER AMPLIFIER The Syringe Power Amplifier is configured in a 3-phase bridge arrangement comprising 6 NChannel Enhancement mode power FETs. Each bridge leg is configured in a high-low arrangement with a motor phase (A,B, or C) tapped at the junction of the upper and lower power FET. The Syringe Power Amplifier converts the low level PWM signals from DSP on the Servo/CPU to high current motor signals that are used for the excitation of the stator windings of the Syringe BLDC motor. The switching rate for the Power Amplifier is approximately 16 KHz. DRIVE INHIBIT CONTROL This block provides shutdown control of the Drivers, which in turn disable the Syringe Power Amplifiers. This signal is generated on the Servo/CPU card and is passed through optical isolation circuitry where it is applied to the Shutdown pins of all drivers. When Drive Inhibit is active high, the outputs of all drivers are disabled. DRIVE INHIBIT STATUS This block is a feedback to the Servo/CPU card indicating the status of the Drive Inhibit signal. This signal is optically isolated from the drive section of the board. SYRINGE BRIDGE CURRENT TEST FEEDBACK This block is responsible for measuring, isolating and filtering each Syringe Power Amplifier bridge node voltage. The rationale is: During bridge testing, these nodes will provide vital voltage information back to the Servo/CPU Card on the status of the bridge and motor windings. By exploiting the resonant frequency of the motor windings series inductance and distributed capacitance, each phase of the bridge can be excited with a single, short duration pulse. These pulses would be long enough to induce current into the motor windings but not long enough to produce motion. By observing the voltage at each node during excitation, the status of both the bridge and motor windings can be ascertained. HIGH SIDE BOOST The HIGH SIDE BOOST block is to generate a floating voltage that is applied to the high-side FET gates at the Driver stage. This voltage (VBOOST) allows the gate voltage of the high side FETs to remain elevated above the source voltage when that particular leg of the bridge is sourcing motor current. +5VDC REGULATOR This block provides 5VDC power to the motor side of the following subsystems for both Syringe A and B axes: Bridge Test Current Feedback, PWM Isolation, Drivers.
4 - 47
Single Head Block Diagram
MEDRAD Stellant CT Injection System
4 - 48
Dual Head Block Diagram
4 - Theory of Operation
4 - 49