TruClear System Elite Hysteroscope Cleaning and Sterilization Guide

3 Pages

CLEANING AND STERILIZATION GUIDE  TruClear™ System Elite Hysteroscopes  This guide is in reference to the following products only: Elite Hysteroscope Mini, product code 72204878, Elite Hysteroscope Plus, product code 72204879 Smith & Nephew/Wolf™* light adaptor, product code 4347, Storz™* light adaptor, product code 7207900  PREPARATION BEFORE CLEANING 1. The working channel insert (B) must be disconnected and removed from the hysteroscope (A).  5. The retaining cap (H) must be removed.Remove excess soil on the operative hysteroscope, seal cap, sheath, and obturator with disposable wipes. NOTE: The hysteroscope seal (G) is single-use and must be disposed of as biohazard waste following use. Do not attempt to reprocess or reuse the hysteroscope seal.  2. Disconnect by pressing the release button (C).  NOTE: The blue hysteroscope seal should remain in the working channel of the scope at all times.  2  6. Stopcocks must be disassembled. Disassemble by loosening thumb nuts (J) and removing stopcock plugs (I).  I  H 3  5  6  3. Slowly withdraw the working channel insert. 4. Adapters, Smith & Nephew/Wolf™* (E) and Storz™* (F) must be removed from light post (D).  G J  4 D  E F