medtronic

Medtronic Nerve Stimulators

Pulsatron II Nerve Stimulator PRODUCT INFORMATION AND INSTRUCTIONS

PRODUCT INFORMATION AND INSTRUCTIONS

28 Pages

PULSATRON® II NERVE STIMULATOR REF 8262015 PRODUCT INFORMATION AND INSTRUCTIONS DEVICE DESCRIPTION The Medtronic Pulsatron II Nerve Stimulator is a hand-held battery-operated device that supplies timed pulses of electrical current at a rate of approximately 80 pulses per second as the stainless steel probe tip touches tissue. The stimulator is a pulsed, constant-voltage device (3.75 V peak-to-peak maximum) which delivers a maximum current intensity of 0.7 mA depending on tissue impedance. An attached needle electrode (ground) is provided for subcutaneous insertion. The LED light is used to test the device. INTENDED USE / INDICATIONS FOR USE The Medtronic Pulsatron® II Nerve Stimulator is intended for the identification of exposed motor nerves during surgery. CONTRAINDICATIONS The device is not indicated for use when paralyzing anesthetic agents are being used as they will significantly reduce, if not completely eliminate, EMG responses to direct or passive neural stimulation. WARNINGS • • •  • • • • • • • •  Excessive pressure or prolonged stimulator contact with the nerve, as indicated by the LED, could result in neural injury. This device is not designed to determine the degree of neural viability. Due to variations in tissue impedance and, hence, current delivered, activation of the LED may not be always be achieved despite delivering current during surgery. Firmly depress the switch and confirm functionality by touching the probe tip and needle electrode. During surgery, a lit LED confirms delivery of current. If the LED is not activated intraoperatively, the device can be tested by touching the probe tip to the needle electrode. Do not bend the probe tip as the device may become damaged and malfunction. Do not use in the presence of any flammable material, there is a potential for ignition. Improperly placed or bent needle electrodes, and attempts to straighten bent electrodes, increase the risk of the needle breaking off in the patient. False negative responses (failure to locate the nerve) may result from neuromuscular fatigue from prolonged or repeated stimulation. Contaminated single use electrodes and probes must be disposed of in an appropriate sharps container in accordance with user facilities policy. The device should not be used while a NIM™ EMG monitor is in use, as it will cause false positive responses. Stimulating current is most effectively delivered through a single point of tissue contact. Avoid multiple points of tissue contact along the probe shaft, as this will shunt current away from the desired point of stimulation.  PRECAUTIONS • After depressing the tan rubber switch, the device has a useful life of 4-20 hours. Continuous stimulation will result in a useful life of about 4 hours. Typical use (i.e. intermittent stimulation) will result in a useful life of 8-20 hours. • Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this Guide. • Portable and mobile RF communications equipment can affect Medical Electrical Equipment. • The Pulsatron should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the Pulsatron should be observed to verify normal operation in the configuration in which it will be used. INSTRUCTIONS FOR USE . Unwind the ground wire. 2. Firmly depress the tan rubber switch at the rear of the handle until it locks down. Allow five seconds for the device output to stabilize. 3. Remove protective covering from grounding needle and probe tip. 4. Test unit for functionality by touching the probe tip to the needle electrode. This will cause the LED light located beside the rubber switch to illuminate. The LED will turn off when the probe tip and needle electrode are separated. Do not attempt to use a device that fails this test. 5. Ground the stimulator by inserting the needle electrode subcutaneously near the operation site. The device must be grounded to insure a complete circuit. Avoid fatty tissue, as this tends to be a poor conductor.  
File Type: PDF
File Size: 551 KB
File Name: pulsatron_ii_nerve_stimulator_product_information_a.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions