eps320_cardiac_stimulator_service_manual_ver_1-4.pdf
Page 1
0120
EPS320 Cardiac Stimulator
SERVICE MANUAL
Version 1.4
CAUTION
Federal (U.S.A) law restricts this device to sale,
distribution and use by or on order of a physician.
Applies to Firmware V4.68
1
Page 2
EPS320 Service Manual
CONTENTS
1
INTRODUCTION
1.1
1.2
1.3
1.4
2
PURPOSE
SCOPE
RELATED DOCUMENTS
DEFINITIONS, ACRONYMS, AND ABBREVIATIONS
DOCUMENT OVERVIEW
2.1 DOCUMENT CONTENTS
2.2 SOFTWARE LICENSE AND W ARRANTY
2.3 SERVICE POLICY
3
SAFETY, WARNINGS AND PRECAUTIONS
BRIEF DEVICE DESCRIPTION
Intended Use
Description of system
Accompanying Documentation
W ARNINGS AND PRECAUTIONS
General Warnings
Warnings Specific to the EPS320 Stimulator
Warnings Related to the use of EPS320 Stimulator with RF Ablation Equipment
General Precautions in Handling the EPS320 Stimulator
ADVERSE EVENTS
Observed Adverse Events
POTENTIAL ADVERSE EVENTS
4
OVERVIEW OF STIMULATOR SYSTEM
4.1 GENERAL
4.2 HARDWARE OVERVIEW
5
PRINCIPLES OF OPERATION
5.1 BACKGROUND INFORMATION
5.2 SYSTEM FUNCTIONAL DESCRIPTION
6
DEVICE SPECIFICATION
ERROR! BOOKMARK NOT DEFINED.
7
DESCRIPTION OF SYSTEM COMPONENTS
7.1 SYSTEM CONFIGURATION OPTIONS
7.2 POWER REQUIREMENTS
7.3 SYSTEM INTERCONNECTION
7.4 ECG SENSING OPTIONS
7.5 THE EPS320 STIMULUS GENERATION UNIT (SGU) OVERVIEW
7.6 SGU HARDWARE-PATIENT INTERFACE SPECIFICATION
7.7 PATIENT INTERFACE OUTPUTS
7.8 CONNECTOR PIN OUTS
7.9 GU CIRCUIT DESCRIPTIONS
7.10 SGU MAIN AND FRONT PCB FUNCTIONAL DESCRIPTION
7.11 STIMULATOR PHYSICAL LAYOUT
4
4
4
4
4
6
6
6
7
8
8
8
8
8
10
10
11
11
12
13
13
14
17
17
17
19
19
20
23
23
25
25
27
29
29
31
32
33
33
35
8
ESP320 SYSTEM TROUBLESHOOTING
36
9
PACING AND SENSING TROUBLESHOOTING
38
9.1 PACING
9.2 ECG SENSING AND QRS SYNCHRONIZATION
10 SGU TROUBLESHOOTING
10.1EQUIPMENT REQUIRED
10.2GENERAL
10.3TROUBLESHOOTING OBJECTIVES
2
38
39
40
40
40
40
Page 3
10.4SGU PROBLEMS
11 COMPUTER SETUP & TROUBLESHOOTING
11.1COMPUTER HARDWARE SETUP
11.2CMOS SETUP
11.3PC OPERATING SYSTEM
11.4SOFTWARE INSTALLATION
12 SOFTWARE TROUBLESHOOTING
12.1SOFTWARE ERROR MESSAGES
12.2ERRORS RECORDED IN STIM.LOG
12.3HARDWARE-RELATED W ARNING MESSAGES ON THE COMPUTER DISPLAY
40
44
44
44
44
45
47
47
48
49
13 SUB-ASSEMBLY REPLACEMENT
50
14 TESTING AND RE-INSTALLATION OF REPAIRED EPS320 SYSTEM
52
15 EPS320 SPARE PARTS LIST
ERROR! BOOKMARK NOT DEFINED.
16 SERVICE RECORD
55
FIGURES
Figure 1 EPS Stimulator System........................................................................................................ 18
Figure 2 EPS320 Stimulator main components.................................................................................. 20
Figure 3 Typical EPS320 installation, with the Bard Duo EP Recording System. ................................ 26
Figure 4 External ECG sensing schematic......................................................................................... 28
Figure 5 Simpler set up for Catheter-tip ECG sensing........................................................................ 28
Figure 6 Auxillary Interface ................................................................................................................ 32
Figure 7 Patient Connector................................................................................................................ 32
Figure 8 Emergency Pace Connector ................................................................................................ 32
Figure 9 EPS320 System Schematic Layout...................................................................................... 33
Figure 10 SGU off PCB component connections................................................................................ 34
Figure 11 SGU Subsystem Physical Components Layout .................................................................. 35
TABLES
Table 1 Suggested connections for external ECG sensing. ................................................................ 27
Table 2 Accessing the catheter-tip ECG sensing ............................................................................... 28
Table 3 Intracardiac ECG Input Specifications ................................................................................... 29
Table 4 High Level ECG Input Specifications..................................................................................... 30
Table 5 Sync In Specifications........................................................................................................... 30
Table 6 Intracardiac ECG Input Specifications ................................................................................... 30
Table 7 Intracardiac Pace Output Specifications................................................................................ 31
Table 8 Sync Out Specifications ........................................................................................................ 31
Table 9 Emergency Pace Output Specifications................................................................................. 31
Table 10 Tabulated SGU Error messages ......................................................................................... 41
Table 11 SGU Error Code Number interpretation............................................................................... 42
Table 12 Table of File-related errors, recorded in the 'stim.log'........................................................... 48
Table 13 Table of warning and advisory messages............................................................................ 49
Table 14 EPS320 Spare parts list ...................................................................................................... 53
Table 15 EPS320 Test jig Spare parts list.......................................................................................... 54
3
Page 4
EPS320 Service Manual
1 INTRODUCTION
1.1 Purpose
This document is intended to enable the servicing of the EPS320 Cardiac Stimulator system by
Micropace service staff and by Distributor Service Department technicians suitable trained by
Micropace.
1.2 Scope
This manual covers the covers the servicing by sub-assembly replacement only of all EPS320
components including the SGU, PC and software and connecting cables. Issues of device
interaction between the EPS320 and connected equipment are also covered. Outside of the scope
of this document are the repair of the accessory computer, computer display, power supplies and
component level repair of the EPS320 printed circuit boards.
1.3 Related documents
Inputs:
TestJigManual.doc
EPS320-SGUAcceptanceProtocol.doc
q
q
Outputs:
q Nil
1.4 Definitions, Acronyms, and Abbreviations
The following abbreviations are applicable to this document.
SYMBOL
SGU:
ECG:
IECG:
RF:
JP:
TP:
EMC
EMI
PC
POST
Moniputer
LED
µP
4
EXPLANATION
Stimulus Generator Unit
Electrocardiogram
Intracardiac Electrocardiogram
Radio Frequency
Header or a connector
Test point
Electromagnetic compatibility. This will refer to emissions of
electromagnetic waves and the effect of these on the environment. It also
refers to the effects of electromagnetic waves on the equipment
Electromagnetic Interference
Personal computer, IBM compatible
Power On Self Test
Refers to a combined computer monitor/CPU
Light emitting diodes and seven segment displays when referring to display
information
Microprocessor, in this manual refers specifically to the PIC16C77
microcontroller, by Microchip
Page 5
Stimulus Generator Box controls and terms:
SGU or Stim Box
Backup mode
PC mode
Standby mode
'Pace Output' socket
'Emergency Output' socket
Switches:
'Power' switch
'Backup Enable' switch
'Pace On/Off' switch
= Stimulus Generator Unit
= 'Backup Manual Pace Control' mode, whereby SGU operates
independently of the PC as a simple pacemaker
= 'PC Pace Control' mode, whereby SGU opertation is
controlled by the Custom PC Software
= Low power SGU operational mode; Mains Power flashes
Green colour
= 'Pace Output' green coloured socket on SGU front membrane
= 'Emergency Fixed Pace Output' red coloured socket on SGU
front membrane providing backup pacing at 100bpm
independent of SGU or PC Software.
= SGU Power On/Off switch located at rear of SGU
= Blue membrane push button switch on front of SGU; invokes
alternatively Backup mode and Standby mode
= Red membrane push button switch on front of SGU; starts &
stops pacing in the Backup mode
Numeric Indicators on SGU front panel
'Pace Interval' display
= Backup Manual Pace Mode Pacing interval display,
adjustable by adjacent up and down push button keys
'Current' display
= Backup Manual Pace Mode Pacing current display, adjustable
by adjacent up and down push button keys
LED Indicators on SGU front panel
'Mains Power'
= Constant Green: Mains power connected and SGU Power
switch On and unit in Backup or PC Control mode; flashing
Green: Mains power connected and SGU Power switched On
and in unit in Standby state; Orange: Mains power connected
and SGU Power switch Off. Backup battery trickle charges
with any flashing or steady Green
'Battery Power'
= SGU power switch On but unit disconnected from mains
power and using battery power
'Battery Low'
= SGU internal backup 12V battery near depletion
'Atrium Pace' LED
= Stimulus issued on Atrial channel
'Ventricle Pace' LED
= Stimulus issued on Ventricular channel
'Atrium Check Lead' LED
= High impedance (>4Kohm) detected on channel
'Ventricle Check Lead' LED = High impedance (>4Kohm) detected on channel
'QRS' LED
= QRS complex (peak) detected on programmed ECG input
channel. Inactive when Software QRS detection selected in
software.
'Emergency Pace V' LED
= Stimulus issued on Emergency Pace channel;
'Emergency Battery OK' LED = Emergency Pace is engaged and 9V Lithium Battery has
adequate charge
5
Page 6
EPS320 Service Manual
2 DOCUMENT OVERVIEW
2.1 Document Contents
This document aims to provide the service technician with the resources to troubleshoot, repair,
obtain further help and and report on problems associated with the EPS320 Cardiac Stimulator
installation.
For technicians not familiar with electrophysiology, the section Background Information within the
Principles of Operation chapter gives an overview of the topic.
This manual contains the following:
(i)
(ii)
(iii)
(iv)
(v)
(vi)
(vii)
(viii)
(ix)
(x)
Safety, Warning and Precautions
Overview of the Stimulator System
Description of the principle of operations
Device Specification
Device Installation Requirements
Descriptions of the system components and interconnections
Troubleshooting guide
Testing and Re-installing Repair
Available spare parts list
Reporting Forms
2.2 Software License And Warranty
Refer to user instruction manual for the terms of software license and warranty.
6
Page 7
2.3 Service Policy
Micropace recommends servicing of the EPS320 by the various stakeholders according to the
following schema:
Site
User / Hospital
biomedical
Engineering
Distributor
Service Dept.
Micropace
Service Dept.
Recommended scope of repair / Service
1. Replacement of apparently faulty accessories,
such as power supplies, signal cables, isolation
transformer, LCD Display Screen.
2. Replacement of fuses in Isolation transformer
with Micropace supplied fused only.
3. Replacement of batteries internal to the SGU,
namely the 12V lead acid backup battery
(MP3109) and the 9V Lithium PP3 format battery
(MP3110), when SGU POST indicates faulty
battery. Use Micropace supplied batteries only.
4. Upgrade of product firmware by replacement
of EEPROM/ Microcontroller on SGU main PCB.
5. Upgrade of product software by installation of
software from FDD or by replacement of
Compact Flash Memory Card in PC.
1. As above, plus:
2. Repair of faulty PC by a Micropace-approved
and identified Computer service center and
subsequent reloading of software.
1. Repair of SGU internal subassemblies
including PCB’s, PCB header cables and
connectors or chassis components by
replacement.
2. Calibration of SGU current delivery system.
As above, plus
Repair of cables and connector subassemblies or
their replacement.
Repair of PCB’s through through-hole or SMT
component replacement or PCB repair.
Verification of repair
1. System functions OK.
2. System functions;
only supplied fuses
used.
3. POST passes OK;
only supplied batteries
used.
4. POST passes OK
5. POST passes OK and
software launches OK.
As above
Software tests own
checksum and essential
PC resources.
Verification with te.exe
test jig/test software.
As above.
Acceptance testing at
Micropace.
Performance of the full
EPS320-SGUAcceptanceProtocol.doc
There are no user serviceable parts inside the Stimulus Generation Box.
It is recommended that hospital service personnel contact the distributor with details of any faults
prior to attempted repairs to obtain recommendations for repair procedures. It is also
recommended that on-site repair be limited to repair of any off-PCB faults and visual inspection of
PCB. Apparently faulty PCB's should be tested by replacement of either the main PC board or
display PC board; defective PC boards should be returned to the manufacturer for repair.
7
Page 8
EPS320 Service Manual
3 SAFETY, WARNINGS AND PRECAUTIONS
Brief Device Description
Intended Use
The Micropace EPS320 Cardiac Stimulator is used to stimulate the heart via intracardiac
electrodes during electrophysiological cardiac studies. The stimulator system is a diagnostic tool
intended for provocative electrical stimulation of the heart for the purpose of measurement of
cardiac conduction properties and induction and diagnosis of arrhythmias. Note that the system is
not a life support device and is not intended to qualify as a life support device in respect to its
functionality, construction or reliability.
Description of system
The EPS320 Cardiac Stimulator by Micropace Pty Ltd. is a diagnostic external programmable
cardiac stimulator. The Stimulator consists of a self-contained two channel microcontroller-based
Stimulus Generator Unit independently capable of generating simple regular pacing pulses, but
under normal use externally interactively programmable by the user for complex pacing patterns
via a serial data link, using a custom software program implemented on a standard personal
computer. The two stimulation channels are independent isolated constant current pulse
generators capable of generating 0.5 to 10ms pulses at 0.1 to 25mA with a maximum output
voltage of 27V. The stimulus output may be used to stimulate the human heart during
electrophysiological studies via any third party legally marketed transvenous intracardiac pacing
catheters directly or via any third party legally marketed EP recording equipment intended to switch
pacing pulses of above description without distortion to selected specific catheters and electrodes.
The Stimulus Generator Unit is mains-powered via a DC power supply and has an internal tricklecharged backup battery in case of power failure. Its outputs are defibrillator and RF energy
protected and it features external ECG input ports as well as catheter-tip ECG sensing for
triggering of pacing events and three trigger output channels for synchronization with other
equipment. The system is powered from the mains via a medical grade isolation transformer.
The EPS320 software allows interactive programming of all aspects of the pacing stimulus: the
current amplitude, pulse width, delivery to channel 1 or 2 or both, and all stimulus parameters
including drive train number, timing and up to 6 extra-stimuli. The stimulator has an intuitive user
interface, with all commonly used stimulus and pacing protocol parameters located on fixed menus
on the one screen instantly adjustable via hotkeys including during actual pacing. All standard EP
stimulation protocols are available pre-programmed but may be reconfigured and automated
according to user requirements.
Accompanying Documentation
A reference package, comprising of a manual and a leaflet, is provided with the EPS 320 Cardiac
Stimulator. Included in the package are:
q User Instruction Manual
q Accessories Unit Contents and Instructions for Use Leaflet
It is strongly recommended that the Operator reads the User Instruction Manual document
in its entirety and is familiar with its contents before using the Stimulator on patients.
One further document is available on request, contact your distributor for copies of this document:
Software Reference Manual
This Service Manual applies to PC Software ver. 3.19.59 and SGU Firmware ver. 4.68.
q
8
Page 9
Compatible Equipment
The EPS320 Stimulator’s primary function is the generation of constant current predominantly
rectangular stimulation pulses with amplitudes of 0.1 to 25mA, duration of 0.5 to 10ms and with a
maximum voltage of +/- 27V. Third party switching equipment with the following special
characteristics must be used to carry the stimulus pulses to the intracardiac electrodes without
significant distortion:
q
q
q
q
Series resistance: < 100 Ohms at up to +/- 25mA
Shunt resistance: > 100,000 Ohms at up to +/- 27V
Frequency bandwidth: DC to 300 Hz
Interference RF energy sources: < 350Vpp at 400 kHz to 600 kHz, or 150W into a 300
Ohm load
Higher series resistance reduces maximal attainable pulse current amplitude; lower shunt
resistance reduces delivered current in all ranges; reduced frequency bandwidth may alter efficacy
of stimulation at any current level and higher RF energy exposure may activate over-voltage safety
elements, reducing RF energy delivery and overheating within the EPS320 system’s Stimulus
Connection Box.
Subject to these requirements, the EPS320 Stimulator is intended for use with the following
equipment; the user should contact Micropace Pty Ltd for compatibility information prior to use of
other equipment:
Diagnostic pacing electrode catheters
q Currently available legally marketed electrophysiological diagnostic electrode catheters,
including those manufactured by CR Bard, Cordis Biosense Webster, Daig, Medtronic and
EPT.
Ablation electrode catheters
q The EPS320 is tested for use with a number of legally marketed RF ablation catheters. Contact
Micropace Pty Ltd for further information (Refer also to “Warnings and Precautions” section
below).
EP Recording equipment
Computerised EP Recording systems manufactured by Bard Electrophysiology (LabSystemTM
TM
Duo ) and GE/Prucka (CardioLab 4000) have been tested for use with the EPS320
Stimulator.
q
RF Ablation Equipment
q RF ablation equipment manufactured by EPT (EPT1000XP) and Medtronic (Atakar RF
Generator) have been tested for use with the EPS320 Stimulator.
9
Page 10
EPS320 Service Manual
Warnings and Precautions
General Warnings
Warning: Stimulator must be used only under supervision by a cardiologist.
q
The Micropace EPS320 Cardiac Stimulator is used in procedures, during which
intentional or unintentional induction of potentially dangerous cardiac arrhythmias is
likely to occur. The Stimulator must only be used by appropriately qualified and
experienced personnel, in appropriate facilities and in conjunction with adequate safety
equipment.
Specifically, the EPS320 Stimulator System may be used on humans only by or under
the direct supervision at all times by a physician familiar with the operation of this
Stimulator.
Warning: Installation only by qualified personnel.
q
Only qualified personnel, such as representatives of Micropace, its authorized
distributor or hospital-appointed biomedical engineers with guidance from Micropace,
may carry out installation or modifications of the EPS320 Stimulator and its connection
to other equipment. Incorrect installation may create electrocution hazards or impair
performance of the Stimulator.
Warning: The Stimulator must only use isolated mains power supply.
q
q
To avoid electrocution hazards, the EPS320 Stimulus Generator Unit power supply,
computer power supply, and monitor power lead must all be connected to a mains
isolation transformer and never directly to a mains power outlet, (i.e. wall outlet).
To avoid electrocution hazards, do not plug cables or leads directly into mains power
outlet.
Warning: Connect Stimulator system only to legally marketed, mains-isolated electrical
equipment
q
To avoid electrocution hazards and system malfunction, the Stimulator outputs, RS232
communication port, Auxiliary port and ECG and Sync ports may be connected to other
equipment provided the other equipment is also isolated from the mains power supply,
and complies with IEC601-1, IEC601-2 and IEC601-1-1 standards and is legally
marketed in the country of use and CE marked for installations in the EU countries.
Warning: Use Stimulator only in ventilated areas and away from flammable gasses.
q
q
10
To avoid risk of explosion, the Stimulator should only be used in a ventilated area, as
gasses may be released during charging of backup battery, and should not be used in
rooms with flammable anesthesia.
Page 11
Warnings Specific to the EPS320 Stimulator
Warning: Stimulator is not a life support device and may fail to pace – To avoid injury to
patient from bradycardia, use only with immediately available backup temporary external
pacemaker.
q
q
The EPS320 Stimulator System is a diagnostic tool for provocative electrophysiological
testing of the human heart. The Stimulator System is not intended, designed or fit for the
purpose of life support. Two levels of backup pacing are provided in case of failure of
normal functioning of the Stimulator and are for use in non-life-threatening bradycardia or
until temporary external pacing is established in case of life-threatening bradycardia or
asystole.
A backup temporary external pacemaker must be immediately available for use in case of
occurrence of life-threatening bradycardia and should preferably be connected directly to
an intracardiac electrical catheter located in a ventricle, bypassing any switching apparatus
in case of failure or inappropriate settings of such switching apparatus.
Warning: Stimulator may not always function as expected – To avoid injury to patient from
arrhythmias, monitor function of Stimulator and patient’s vital signs continuously while
Stimulator is connected.
q
The EPS320 Stimulator software is warranted to operate substantially according to
specification however it is not guaranteed to do so at all times, in all possible
circumstances and in an uninterrupted manner, nor to be error-free. The Stimulator could
potentially fail to stimulate or unintentionally stimulate the patient; to avoid injury to the
patient, the Stimulator and the patient must be observed for abnormal function of
Stimulator at all times while the Stimulator is connected to the patient.
q
In case of recurrent or persistent unexplained life-threatening ventricular tachycardia or
ventricular fibrillation occurring despite cardioversion/defibrillation during the use of the
Stimulator, disconnect the patient from the Stimulator and any attached third party
switching equipment in case of occult malfunction in any of the equipment is causing microelectrocution or DC current stimulation and repeat cardioversion/defibrillation. It is
recommended to disconnect patient at the intra-cardiac electrode catheter connectors; the
Stimulator itself may be disconnected instantly by unplugging the green Pace Output plug
on the front panel.
Warning: Measurements by stimulator are for information only.
q
Measurements displayed by stimulator, including the Impedance measurement and SNRT
measurement are for facilitation of use of stimulator. The user should use third party legally
marketed measurement devices independent of the Stimulator to measure these
parameters for the purpose of clinical diagnoses.
Warnings Related to the use of EPS320 Stimulator with RF Ablation
Equipment
Warning: Use Stimulator only with Stimulus Connection Box (Micropace Parts MP3014 or
MP3065)
q
Use only supplied Stimulus Connection Box (Micropace Parts MP3014) or alternative
optional In-line RF Filter (MP3065) components to connect Stimulator output to patient
circuits. These components contain RF suppression filters to prevent large RF energies
from RF Ablation equipment not equipped with RF filters from reaching the Stimulator
output circuits. Use of other, including custom made connectors may bypass RF filtering
and potentially lead to repeated alarms and shutdowns of the Stimulator and possible
induction of unintended life-threatening arrhythmias during delivery of RF ablation energy.
Warning: Do not stimulate via ablation electrode during delivery of RF Ablation energy.
q
To avoid possibility of unintended arrhythmia induction, do not stimulate myocardium via
the ablation electrode during application of RF energy. Efficacy and potential for adverse
effects of stimulation of heated myocardium in the process of ablation have not been
established.
11
Page 12
EPS320 Service Manual
General Precautions in Handling the EPS320 Stimulator
Caution: Installation, Transport and Storage.
q
To ensure reliable operation of the Stimulator, install away from dust, excessive heat or
humidity, direct sunlight and splashing liquids and in a well-ventilated place.
q
To ensure that operator may see important error messages displayed during operation,
install the Stimulus Generator Unit with the front panel visible to the operator.
q
To avoid damage to the Stimulator, avoid exposure to chemical gases, excessive vibration,
impact, temperatures above 60 ºC or ambient air pressures less than 500 hPa during
transport and handling.
q
To ensure that backup battery remains fully charged, store Stimulator between uses with
the Stimulus Generator Unit connected to isolated mains power supply, switched on at the
rear panel switch (Mains Power green Led should blink) and in the Standby Mode .
Caution: Precautions prior to use.
q
Ensure that all cables are free from defects, are properly installed and secured.
q
Ensure isolation of the mains power supply to the Stimulator and that no newly attached
equipment is compromising the electrical safety of the installation.
q
If the stimulator has been unused or may have been disconnected from mains power
supply for more than 1 month, charge backup battery by leaving connected to the mains
supply in standby mode overnight and check that Emergency Backup Fixed Pacing and the
stimulator functions correctly.
q
Do not use the Stimulator if any component appears damaged, computer appears to start
up abnormally, or error messages appear on the computer screen or Stimulator front panel.
If in doubt, contact the Distributor or Micropace directly.
q
Ensure that the Operator is trained thoroughly on how to switch the Stimulator to Backup
Manual mode and Emergency Fixed Rate Pacing modes, and knows how to bypass any
third party switching apparatus to maintain pacing capability in case of failure of that
apparatus.
q
To prevent EPS320 software malfunction, do not attempt to install drivers, utilities or
software into the EPS320 PC other than that supplied by Micropace. The EPS320 software
will not execute if Terminate and Stay Resident (TSR) programs are detected by the
software.
Caution: Precautions during use.
q
Observe the Stimulator and patient at all times for abnormal function and rectify any
problem promptly or disconnect the patient from the Stimulator (by unplugging the green
Pace Output plug on the front panel).
q
Do not use the Stimulator and disconnect it from the patient if it repeatedly switches to
Backup Manual mode or repeatedly displays error messages on the front panel. Contact
your Micropace Distributor.
q
Use of excessive stimulation currents increases potential for induction of fibrillation, which
may be harmful to the patient and may be prognostically non-specific in ventricular
stimulation studies.
q
To ensure reliable pacing capture, the user must establish pacing current thresholds for
electrode catheters, including RF ablation catheters, prior to diagnostic stimulation studies
and set pacing current accordingly. Pacing thresholds, impedance and sensing
characteristics may differ amongst various diagnostic and RF Ablation electrode catheters
due to differing electrode geometries and associated electrical pathways.
q
To reduce chance of accidentally inducing ventricular fibrillation, ensure reliable ECG
sensing and use Synch-to-QRS function to avoid stimulating in the vulnerable diastolic
period where appropriate.
Caution: The Stimulus Generator Unit should be charging its backup battery while not in
use.
12
Page 13
q
The EPS320 Stimulator should be connected to external power; its Power switched on at
the rear panel switch and in the Standby Mode while not in use to ensure that backup
battery remains fully charged.
q
The Stimulator should be maintained according to “Maintenance” section of this manual.
Adverse Events
Observed Adverse Events
The EPS320 produces standard cardiac stimulation outputs similar to other existing programmable
cardiac stimulators in use for the past 30 years. The standard pulse output characteristics are well
defined in standard Electrophysiology texts and the application, safety and clinical efficacy of this
group of devices is well established in the medical field (2,3). and in one recent review by McLaughlin
et al (4).
Report of adverse events related to EPS320 arises from the following clinical experience with the
product subject to product vigilance (device exposure estimated by adopting an estimate of 14
patients per stimulator per month of use, each patient exposed for an estimated mean time of 2.5
hours)
Clinical use of the device in its evolving form since January 1998, evaluated in October 1999 by a
survey of 6 clinicians using 6 Stimulators with an accrued clinical experience with the product on
more than 1600 patients
q
A prospective field trial of safety and efficacy of the EPS320 performed in May to July 1999
concurrently at 4 hospitals in Sydney, involving 6 Stimulators / cardiologist users and a
total of 23 patients.
q
Clinical use of the EPS320 in Australia and Southeast Asia from July 1999 to October
2001, involving 22 Stimulators and an estimated 7400 patients.
q
Clinical use of the EPS320 in EC countries from January 2001 to October 2001, consisting
of 17 Stimulators accumulating clinical exposure to an estimated 2400 patients.
The EPS320 Stimulator has thus been used on an estimated total of 10,000 patients,
approximately half using the current firmware version, leading to an estimated device exposure of
25,000 hours.
No deaths or injuries related to the use of the EPS320 Stimulator have been reported during the
device exposure.
Two adverse events were reported, both due to device interaction with an RF Ablation Generator.
They were non-sustained Ventricular arrhythmias arising when operators simultaneously paced
and delivered RF ablation energy through the same ablation electrode; the EPS320 Stimulator
detected the interaction and shut down in both cases. Device interaction has been eliminated in the
field by the issue of a Safety Alert warning against this application and the addition of RF filters to
Stimulator’s Stimulus Connection Box to isolate Stimulator from RF signals.
13
Page 14
EPS320 Service Manual
Potential Adverse Events
Following is a list of potential adverse events which could be directly associated with diagnostic
cardiac stimulation using the EPS320 during electrophysiological studies (in alphabetical order):
Bradycardia or asystole
q
Should a patient develop bradycardia or asystole during the use of the Stimulator,
failure to promptly backup pace the patient may lead to hypotensive injury within
seconds to minutes. Such arrhythmia may typically occur (i) in patients with atrioventricular conduction defects, (ii) during placement of intra-cardiac electrode
catheters, (iii) after cardioversion/defibrillation or (iv) during RF ablation.
q
The Stimulator may fail to normally deliver stimuli due to (i) power failure or power
interruption, (ii) failure of PC, (iii) loss of communication between PC and Stimulus
Generator Unit (SGU), e.g. due to lead disconnection or (iv) failure of SGU itself due to
spontaneous fault or damage from external events such as defibrillation of RF ablation
energies.
q
To avoid patient injury, the user may use the battery powered Backup Manual Pace
mode in case of power failure, failure of PC or the communication link, and the
Emergency Fixed Pacing mode in case of failure of normal operation of the SGU. The
EPS320 Stimulator is not a life-support device; these functions are for use in non-lifethreatening bradycardia or until external temporary pacing is established in case of
asystole. Refer to “The EPS320 Stimulus Generator Unit” section below for instructions
on the use of these features.
DC voltages greater than 5000V or RF voltages greater than 350V pp applied to
Stimulator outputs may cause circuit failure; the Stimulator issues a range of alarms in
case of failure of output circuits – see Section “Hardware Error Messages on the
Stimulus Generator Unit.”
Explosion or fire
q
q
Explosion could arise from accumulation and ignition of explosive gasses vented from
charging of the backup battery within the Stimulator. Use Stimulator only in well
ventilated areas.
q
Explosion could also arise from electrical sparks within the Stimulator igniting explosive
anaesthetic gasses in the operating room. Do not use Stimulator in the presence of
volatile anaesthetic gasses.
Continuous RF voltages greater than 350V pp applied to Stimulator outputs in
contravention of labeling may cause overheating of the Stimulus Connection Box and
possibly cause a fire hazard. Similar hazards would probably be present in other
equipment with over-voltage protection. Special care will be required should RF
generators capable of delivering such voltages (corresponding to in excess of 150W
into 300 Ohm load) become available.
Myocardial injury
q
q
Excessive current flows through intra-cardiac electrodes, such as due to failure of
Stimulator output circuit causing excessive stimulation currents and inadvertent
shunting of defibrillation or RF ablation energies through a malfunctioning Stimulator
or equipment attached to its outputs could theoretically cause localized myocardial
damage at the pacing electrode.
In order to minimize risk of myocardial injury, connect EPS320 Stimulator stimulation
outputs only to legally marketed medical equipment compliant with IEC60601-1 and
IEC60601-1-1, and if error messages appear on the SGU, consult the section
“Hardware Error Messages on the Stimulus Generator Unit.” prior to using the
Stimulator again.
Operator electrocution
q
q
14
The EPS320 Stimulator auxiliary signal input/output ports and RS232 communications
port are optically isolated from patient circuits, but connect directly to Stimulator signal
ground plane, in accordance with IEC60601-1. Connection of these ports to equipment
Page 15
without appropriate mains isolation or protective grounding may cause electrocution of
the operator by mains-derived AC current in case of malfunction in the connected
equipment.
q
In order to minimize risk of electrocution to the operator, connect EPS320 Stimulator
input/output ports only to legally marketed medical equipment, isolated from mains
power and compliant with IEC60601-1 and IEC60601-1-1.
Ventricular tachycardia or fibrillation
q
The EPS320 Stimulator may be used to intentionally induce ventricular or atrial
arrhythmias including ventricular tachycardia and fibrillation. Causes of inadvertent
induction of such arrhythmias includes operator error, Stimulator malfunction and
device interactions:
(i) Myocardial stimulation with DC current.
Myocardial stimulation with DC current is likely to cause arrhythmias, including
intractable ventricular fibrillation (if operator is unaware of the fault and does not
remove the source from the patient, ventricular fibrillation may repeatedly recur
after defibrillation, possibly leading to patient death). DC current flow in the
stimulation circuit loop, may potentially be caused by Stimulator output circuit
failure, failure in other third party equipment inserted in the stimulation circuit loop
or device interaction between interconnected equipment. Should the EPS320
detect such a condition a DC-ERROR alarm is issued, (_,d,c,_,E,r,r )displayed on
SGU front panel and DC-ERROR alarm message on the computer), followed by a
safe state shut down of the Stimulator. Refer to section “Hardware Error Messages
on the Stimulus Generator Unit” below for instructions in this case. Regardless of
Stimulator alarm state, recurrent unexplained ventricular fibrillation should prompt
user to immediately disconnect the patient from the stimulation circuit, preferably at
the intracardiac electrode catheter connectors.
(ii) Intended cardiac stimulation in the vulnerable diastolic cardiac period
User should ensure that ECG sensing by the Stimulator is reliable and always use
the Synch-to-QRS function to avoid stimulating in the vulnerable diastolic period,
where appropriate. External amplified ECG signal is the recommended source of
ECG, however, where external ECG source is not available or unreliable, the
EPS320 may be configured to sense intra-cardiac ECG from either of the
stimulation output channels.
(iii) Inappropriately rapid cardiac stimulation
Cardiac stimulation at rapid intervals, typically less than 300 ms, may cause
undesirable arrhythmias including ventricular fibrillation. Inadvertent rapid
stimulation, may occur either through user error, inappropriate Stimulator software
safety parameter configuration or Stimulator hardware or software malfunction.
The Stimulator’s SGU monitors for unprogrammed rapid stimulation and if
detected, issues a Rate Error alarm, followed by a shut down safety state. Monitor
the Stimulator function, patient’s ECG and observe the patient continuously for
unexpected behaviour whenever patient is connected to the Stimulator; disconnect
the patient from the Stimulator and do not use if abnormal stimulation is observed.
(iv) Unprogrammed isolated stimulation pulses
Unintended cardiac stimulation with isolated stimulation pulses may cause
undesirable arrhythmias including ventricular fibrillation. Isolated stimulation pulses
may potentially occur due to Stimulator hardware or software malfunction. The
Stimulator’s SGU monitor detects unprogrammed stimulus pulses occurring within
300 ms of another stimulus pulse and issues a Rate Error alarm, followed by a
shut down safety state. Monitor the Stimulator function, patient’s ECG and observe
the patient continuously for unexpected behaviour whenever patient is connected
to the Stimulator; disconnect the patient from the Stimulator and do not use if
abnormal stimulation is observed.
(v) Mains derived AC current microshock
15
Page 16
EPS320 Service Manual
Mains derived AC leakage currents to ground may cause patient microshock
leading to arrhythmias, including intractable ventricular fibrillation. AC leakage
currents may potentially arise from a malfunction in the EPS320 Stimulator or
connected third party equipment if the mains power supply is not isolated in all
interconnected equipment. Ensure that the EPS320 Stimulator system, including
the PC and all connected third party equipment is connected only to isolated mains
power supply.
Adverse events which may be directly associated with diagnostic cardiac stimulation during
electrophysiological studies (in alphabetical order):
q
q
q
Arrhythmias
Death
Temporary chest pain
Adverse events may occur arising from use of third party equipment attached to the EPS320
Stimulator during EP studies. Refer to User Instructions for such third party equipment for
warnings, precautions and lists of adverse events.
16
Page 17
4 OVERVIEW OF STIMULATOR SYSTEM
4.1 General
The Micropace EPS320 is a programmable external cardiac pacemaker used in
electrophysiological studies. The device features two independent fully programmable and isolated
constant current stimulator channels used to pace the heart temporarily via transvenous
intracardiac electrodes.
The EPS320 system consists of a manufactured instrument case containing the Stimulus
Generator Unit (SGU) connected to and controlled by a Personal Computer (PC) via a serial
RS232 interface. A Stimulus Box connected to the SGU output is used to interface the stimulus
output to third party transvenous pacing electrodes and to filter out any unwanted electrical
interference. The transvenous electrodes are not part of the EPS320; they are a disposable
medical product.
Normal operation of the EPS320 SGU is under control of software running on the attached PC
which sends specific real-time commands to generate pacing pulses and control the associated
functions (see the software description for more details). In case of software or communication
failure, the SGU can operate independently in the Manual Backup mode, capable of providing
constant pacing at a rate and pulse current displayed on and programmable from the SGU's front
panel. A secondary backup facility is available, the Emergency Fixed Backup mode, which is a
completely separate analogue circuit pacemaker, powered by a ultra-long life battery, capable of
providing a fixed pacing train at 500 ms, 5mA and 2ms pulse width.
The two isolated outputs will deliver pulse widths from 0.5ms to 10ms, adjusted in 1ms steps. The
current amplitude can be adjusted from 0.1mA to 25mA in steps of 0.1mA. The maximum available
output voltage is 26V, so the maximum output pulse current will be less than 25mA for load
impedance greater than 1kΩ.
4.2 Hardware Overview
The EPS320 stimulator system:
•
•
•
•
•
•
•
A computer with power supply (CPU and separate Display as shown or all-in-one PC)
Stimulus Generator Unit (SGU)
Mains Isolation Transformer
15VDC power supply unit
Patient Connection Box
Auxiliary cables for connection to an EPS recording system
Various power cables as shown
17
Page 18
EPS320 Service Manual
Figure 1 EPS Stimulator System
18
Page 19
5 PRINCIPLES OF OPERATION
5.1 Background Information
The EPS 320 cardiac stimulator is a 2 channel constant current stimulator designed for delivery of
complex sequence of cardiac pacing stimuli for the purpose of clinical electrophysiological testing
of the human heart. Pacing stimuli are delivered to the heart by a third party transvenous pacing
electrode. The stimulator is a class 2B medical device according to MDD classification and class III
medical device according to FDA classification.
The stimulus parameters and timing sequences are programmed by the operator on the computer
which issues commands to the microprocessor controlled Stimulus Generation Unit at the
appropriate time. Stimulus protocols include simple constant rate stimulation, decremental
stimulation with ever decreasing inter-stimulus interval, single extra stimulus consisting of a drive
train of eight (8) constant interval stimuli followed by one (1) extra stimulus with progressively
decreasing coupling interval. More than one (1) extra stimuli may be called upon to be delivered
with or without synchronization with the cardiac intrinsic rhythm. In the VT induction protocol up to
six (6) independently timed extra stimuli are given at close coupling intervals to determine if the
heart can be provoked into abnormal ventricular arrhythmias. Potentially dangerous rhythms are
routinely induced in patients during these studies and external DC cardioversion or defibrillation is
frequently performed. Despite the frequent such rhythms and the frequent severe ill health of
patients undergoing the test, electrophysiological studies performed by trained personnel are
remarkably safe with expected death rates of approximately one (1) in 5000 studies.
What is an Electrophysiological Study?
Electrophysiological studies are diagnostic tests performed to investigate patients with complaints
of palpitations, fainting or resuscitated from sudden death. The information obtained from these
studies aids in clinical decisions regarding therapy, which may include drug, radio-frequency
ablation or implantation of an implantable defibrillator.
What is a Pacing Stimulus?
The heart may be electrically stimulated to contract by an electrical pulse if the current-duration
product exceeds the pacing threshold for that part. Pacing threshold may vary due to presence of
heart disease, drug therapy and depend heavily on the site of the stimulation within the heart.
Stimuli less than 0.03ms and less than 0.1mA generally do not produce cardiac depolarization
(myocardial capture). The usual pacing pulse parameters in the EPS laboratory are 2ms in
duration and applied from a constant current source at 0.1-5.0mA. Normal satisfactory pacing
threshold is less than 1.0mA for the ventricle and less than 1.5mA for the atrium. Current
amplitudes of up to 20mA may be required for patients with severe heart disease or in emergency
situations. Pacing impedance varies from 300-100 Ω and thus the constant current source must be
able to supply up to 20 volts to deliver 20mA into 1kΩ resistance. The EPS 320 delivers up to
25mA. Certain specialized applications such as transoesophageal stimulation require wider pulse
width.
Cardiac capture depends on pulse strength-duration product and thus is not sensitive to the shape
of the waveform, which should be approximately square with little effect from ringing, damping or
noise. Whereas chronic pacing arguably benefits from a biphasic pacing pulse, the application of
a small reverse polarity between pacing spikes prevent myocardial polarization, this is not an issue
in short term pacing involved in EPS studies and thus the pulse may be entirely positive. The
pacing electrode tip should be negative with respect to the reference electrode. Pacing during EPS
studies is usually bipolar. The current amplitude has modest accuracy requirements - stimulation is
usually performed in steps of 0.1mA up to 1mA, steps of 0.2mA up to 5mA and in 1mA steps up to
25mA.
19
Page 20
EPS320 Service Manual
Stimulation rates are typically 230ms to 700ms but may on occasion be as rapid as 100ms. Pacing
duration is typically from 1 second up to 1 minute. Extra stimulus coupling intervals vary typically
from 300 to 100ms but may be as short as 50ms.
5.2 System Functional Description
The EPS320 Stimulator is made up of two main components described below and illustrated in
Figure 2:
1. Desktop IBM compatible Computer (PC):
This computer executes a DOS
program with a windows-like graphical
user interface using the keyboard as
the only user input. The software
generates each stimulus by sending a
serial command to the Stimulus
Generation Unit to initiate a stimulus
pulse. Stimulus parameters may be
altered any time between stimuli by
other specific serial commands. Since
the PC software programs each
stimulus one at a time, its parameters
and its timing may be altered by the
user at any time prior to the actual
occurrence of the stimulus pulse,
affording the user the desired control
over stimulation at all times, as one
would have with most hardware
stimulators.
Monitor
Stimulus
Output:
Atrium &
Ventricle
Channels
COM1
EPS320
STIMULUS
GENERATOR
UNIT (SGU)
Computer
Emergency
Pace
Output
Auxiliary
Port
ECG1 In
ECG1 In
15 VDC
Sync1 Out
Power Supply
110-240VAC-15V
Front Panel Display/ Keys
Figure 2 EPS320 Stimulator main components
2. Stimulus Generation Unit (SGU):
The Stimulus Generation Unit contains a PIC16C77 µP and two optically isolated digitally
programmable constant current generators. Channel 0 is labeled and intended for pacing the
atrium and Channel 1 the ventricle, although they are electrically identically. The µP firmware
normally operates in one of two user modes:
(i)
(ii)
(iii)
20
In the Manual Backup mode, the Stimulus Generation Unit operates independently of the
PC, and is capable of generating regular pacing stimuli (always 2ms wide pulses into both
channels) at amplitudes and intervals adjustable by buttons
And displayed by 7-segment LED displays on the front panel. This mode is intended for
backup use only in the rare instance of the failure of the PC or its software at a time when
the patient urgently requires pacing.
In the PC Control mode, the Stimulus Generation Unit communicates with the PC via a
19,200-baud asynchronous serial link. Transmission takes place in three 8 bit word
st
nd
rd
'command strings', the 1 word the ‘frame’, 2 word ‘command’ and 3 word ‘data’. The
PC transmits to the Stimulus Generation Unit a watchdog command string every 47ms and
at other times any command strings produced by software logic or operator actions. The
Stimulus Generation Unit transmits to the PC a command string every 2ms with every
second transmission containing either an embedded 8 bit ECG data or status report word.
The Stimulus Generation Unit returns four words containing measured stimulus data after
executing each stimulus command from the PC. In case of communications failure, the PC
computer is able to reset the Stimulus Generation Unit microprocessor by raising the DTR
line in the RS232 port, which is wired to the microprocessor master reset.
Page 21
The isolated current generators consists of an 8 bit D to A converter with 0.1% voltage references
programming an operational amplifier capable of a 27V output swing and configured as a constant
current generator. Output current is DC coupled to the patient with multiple safeguards against
delivery of constant DC current. An 8 channel 8 bit A to D converter monitors output current and
voltage in addition to amplified ECG from catheter tip with 4 programmable gains. An analogue
peak detector also generates QRS-detected pulses. Digital converters communicate with the
microprocessor via optocoupled serial links using the SPI (serial peripheral interface) protocol. A
relay in the output circuit can shunt the output with a 1kΩ resistor for testing and self-calibration
which occurs on every power-up. The isolated channels are powered by a 2 watt 5kV isolated DC
to DC converter (Newport) which can be turned off by either a microprocessor command in the
case of a detected fault in the isolated channel or by a watchdog timer in the case of a
microprocessor failure.
In the case of communication failure with the PC, the microprocessor automatically changes to
Backup Manual mode and engages the front panel 7 segment LED displays and push buttons
enabling operator to program pacing current and interval. A second emergency backup channel is
provided in case of failure of microprocessor within the Stimulus Generation Unit. This is an
entirely independent circuit consisting of a 9V lithium manganese battery, an analogue multivibrator circuit and a fixed current source circuit for production of pacing stimuli. An extremely low
power monitoring circuit detects the connection of a load resistance to the output of this channel
and turns power on to the pacing circuit providing immediate fixed rate and current pacing at a
fixed 100 pulses per minute, 5mA amplitude and pulse a width of 2ms. Expected battery life is
equal to its shelf life of 10 years. The computer monitors approximate condition of this battery via
optocouplers.
The Stimulus Generation Unit hardware consists of one (1) main board and one (1) display board
located behind the front panel. The PCB’s are two layer, surface-mount and through-hole
component populated boards containing approximately 650 components. The Stimulus Generation
Unit is powered by an external 14.5VDC to 15.0VDC (at 750mA) power supply unit (PSU). This is
used to trickle charge the backup battery and supply three 5V regulated power networks (the latter
two may be switched off during stand-by mode):
(i)
(ii)
(iii)
Microcontroller power,
General circuit power
DC to DC converter power.
Electromagnetic emissions and susceptibility (EMC) compliance is achieved by use of surface
mount components (SMD), good PCB layout using separate grounds for digital, analogue and I/O
circuits with star grounding points, slow microprocessor speed of 8MHz, inductors and capacitors
on all interfaces and shielding of external cables. Note that the enclosure is not
electromagnetically sealed as this was found not to be required.
21
Page 22
EPS320 Service Manual
6 EPS320 Specifications:
vii. Protocol Automation
i. Computer
TM
CPU:
VIA-Eden ESP 566 MHz or similar
Weight: 4.2 kg, with monitor and keyboard
Dimension:
234x48x175 mm, 15” Monitor, 101
Keyboard
Hard Disk:
64Mb CompactFlash Card
Power Source:
100-240VAC to 12V DC 5.0A PSU
(via isolation transformer)
O/S Software:
Non-windows custom RTOS
ii. Stimulus Generator Unit
Main power source: 100-240VAC to 15 VDC 2 A PSU
Backup power:
12V 2.1 Ah sealed lead acid
Emergency power: 9V LiMnO2i PP3, 10y life
Weight: 5.3 kg including batteries, PSU
Dimensions:
31.5cm x 34.0 cm x 9.5cm
Power Consumption: 800 mA peak during operation
40 mA in standby mode
Operating Time:
Indefinite with mains connected
Two(2) hours on backup battery
iii. Pacing Channels
Isolated Channels(3): (i) Atrial and (ii) Ventricular via green
Redel 4 pin socket
(iii) Emergency Fixed Pace Output to Ventricle,
via red Redel 4 pin socket
Power Source:
Internal DC-DC converters
Circuit Isolation:
Compliant with AS3200 /IEC601,
5kV, common, differential modes
iv. Computer Controlled Stimulus Pulses
Current: 0.1 to 25 mA
Current Steps:
0.1 mA
Accuracy: ± 2% or ± 0.2 mA, which ever is greater
Pulse Duration:
0.5 ms, 1-10 ms in steps of 1ms
Accuracy: ± 0.15 ms
Load Impedance:
700 Ω max, for max current
Max Output Voltage: 27V
v. Inter-stimulus Intervals
S1 Range:
180 - 9990 ms(Pace)
30 - 9990 ms (Burst Pace)
Stability: Quartz computer clock,
o
± 30 parts per million @ 25 C
Extra-Stimuli:
6 max, S2-S7, independent
Coupling interval:
30 - 9990 ms,
Accuracy ± 1 ms or 0.1% whichever is greater
vi. Pre-programmed Protocols
♦
♦
♦
♦
♦
♦
♦
Stimulation Threshold ♦ Wenckebach Periods
Anterograde/Retrograde ♦ Programmed V Stimulation
SN Recovery Times
♦ Burst Pacing (to 30ms)
SN Conduction Times ♦ Auto-decremental Pacing
His-coincident extras
♦ AV sequential Pacing
Overdrive Pacing
♦ Combined AV drive-train
AICD-type ATP
♦ AF/VF induction
EPS320 Device Specifications Ver 2.4, 4.8.04
22
♦
♦
♦
♦
♦
♦
♦
Auto decrement / increment: S1, S2-S7,stim current
SNRT S1 intervals and RT calculation
Auto pace and sense - site selection in protocols
His-coincident extra-stimulus timing calculation
ATP S1 calculation from % of TCL
Trigger output on sensed ectopic beats
All automation subject to instant operator adjustment
viii. Backup Manually Controlled Stimulation
Power Source:
12V 2.1Ah sealed lead acid
battery
Pulse Current:
0.1 to 25mA
Accuracy: ± 2%, or ± 0.2mA which ever is greater
Pulse Duration:
2ms, fixed
Pulse Interval:
100 - 1400ms, accuracy:+1%
ix. ECG Sense/Trigger
High level external Two inputs:
Connector:
Input Ranges:
Maximum Input:
Frequency Range:
Pacing catheter tip Two channels:
Input ranges:
Frequency Range:
Either Source Threshold:
Lockout Period:
Pace Sync Delay:
Sampling- display:
(i) Ext_ECG1, (ii) Ext_ECG2,
6.5mm Phone socket
1V and 2V
2V AC pk - pk
1 to 250 Hz, typical
(i) Atrial, (ii) Ventricular
2mV to 36mV, 4 gains
30 to 250 Hz, typical
Adaptive software peak detect
50 - 1000ms in software
50 - 5000ms in software
500Hz, 8 bit
x. Auxiliary Sync Inputs/Outputs
Sync1 Output:
6.5 mm Phone socket
Timing: On any drive train pulse, on halt of pacing,
on extra in R-synced S2
Pulse: +5V CMOS level, 50 ms duration
Sync2 & 3 Output:
Used for internal functions
Sync1 & 2 Input:
Used for internal functions
xi Emergency Backup Pacing
Power Source:
9V Lithium Manganese PP3
battery, 10 years life
Pacing Pulse:
5mA +1.0 mA, 2ms + 0.5ms, 100
bpm
Pace output:
Separate red Redel 4 pin socket
Control: Pacing activated by connection of a load
<1MΩ, i.e. insertion of plug
connected to patient
xii Certification
MDD Device Class:
IEC60601-1
EMC compliance
CE Marking
US Regulations
Class IIb
certified
certified
certified
510(k) Accepted
Notes: 1. Specifications may change without notice
2. Computer specifications will vary according to
availability
Page 23
7 DESCRIPTION OF SYSTEM COMPONENTS
The EPS320 is always powered through an isolation transformer normally connects to a patient
through an external third party EP Recording System, such as the Bard Duo TM via stimulus (and
possibly trigger) output connectors and External ECG input cables.
7.1 System Configuration Options
There are four EPS320 Configurations, according to the computer used and mains power supply:
Configuration
Description
EPS320
First international configuration. Contains a Leo Aries III Model LS523
Moniputer 15" Pentium 75MHz All-in-one PC, for 220-240 VAC mains
supply.
EPS320N
Contains NEC Powermate 2000, Intel Celeron 433 MHz all-in-one PC,
with 15 inch Active TFT LCD Screen, for 110-120VAC mains supply.
EPS320H-US
Contains HP eVectra Intel Pentium III/733MHz Mini PC and NEC MultiSync LCD1530V (or LCD1540V) active TFT LCD display screen for 110120VAC mains supply
EPS320H-EU
As for EPS320H-US, but for 220-240VAC mains supply.
EPS320B-EU
Bona Light PC, NEC Multi-Sync LCD 1560V, 220-240VAC mains supply
EPS320B-US
Bona Light PC, NEC Multi-Sync LCD 1560V, 110-120VAC mains supply
23
Page 24
EPS320 Service Manual
EPS320B-US STIMULATOR & PC CARTON
Item
Torema Transformer Isolation, 110-120 VAC, 300VA
Keyboard with key labels Ver2.2 for Software 3.19
Accessories Box:
Isolated Mains Power Lead, IEC 3 to 2 Pin- US
Style
Stimulus Connection Box with RF Filter
Serial Boost Cable, Serial RS232
Isolation Transformer Power Lead, according to customer
country
Isolated Mains Power Lead IEC 3 to 3 Pin – US Style
Signal Cable, with Phone Plugs
Hipower Switched Mode 15.0 VDC 1 A Power Supply for
SGU
Accessories Booklet
Installation Kit:
Phone Plug – RCA socket adapter
RCA Plug – BNC socket adapter
Test LED, Yellow/Red
Fuses, 4A Semi-delay HBC M205
Cable tie
Cable tie adhesive anchor
Stimulus Generator Unit
Bona Light PC Computer
Power Supply for Bona Light PC
Documentation:
User Instruction Manual
Installation Procedure Booklet
Manufacturer’s Test Certificate
MS DOS6.22 Operating System Software License
Summary of Hotkeys
Part No.
MP3001US
MP3016
MP3055
MP3066US
Qty
1
1
1
1
MP3014
MP3033A
MP3059_
1
1
1
MP3030US
MP3034
MP3074
1
2
1
MP3064US
MP3063US
MP3056
MP3057
MP3058
MP3068US
MP3039
MP3061
MP3008
MP3093
MP3099
MP3036US
MP3069
MP3067
-
1
1
2
2
1
4
4
1
1
1
1
Part No.
LCD51V
MP3030US
Qty
1
1
1
1
1
1
1
1
1
NEC LCD Sceen Carton
Item
LCD Display (Micropace Part No. MP3076)
Video Signal Cable (Part of Micropace No. MP3076)
NEC LCD Screen Instruction Manual
Isolated Mains Power Lead, (IEC 3 to 3 pin) US Style
24
Page 25
7.2 Power Requirements
The PC must be powered from mains power via a mains-isolation transformer (Part No. MP3001 or
MP3001US) supplied with the system.
The Stimulus Generation Box is powered from a low voltage power supply, 14.5VDC for EU
(MP3002) or 15.0VDC (MP3074) for the US markets, rated at 750mA or higher. These power
supplies are double insulated and are to be powered only from the isolation transformer.
The Stimulus Generation Unit contains a backup sealed lead acid gel maintenance free 12 volt
2.1Ah battery in case of failure of mains power supply. The battery is float charged to 14.1V and
has an expected life span of five years. The battery is charged at all times when the MAINS
POWER light is lit green (unit on) or is flashing green (unit on Standby). It is NOT being charged
when the MAINS POWER light is glowing yellow (External Power available, SGU power
switched off) or not lit (no external power).
The EPS320 Cardiac stimulator system Power ratings are as follows:
Item
EU Low Voltage Transformer
Part no. MP3002
Mains Isolation Transformer
Part no. MP3001
US Low Voltage Transformer
Part no MP3074
US Mains Isolation
Transformer
Part no. MP3001US
Input Volts
220-240VAC
Output Volts
14.5@ 0.75A
Output VA
11 VA
220-240VAC
220-240VAC
400VA
90-264 VAC
15V @ 2.4A
36 VA
100-120VAC
100-120VAC
400VA
Item
Stimulus Generation Unit
Part no. MP3008
Computer + Monitor
Input Volts
14.5-18VDC, 750 mA
110-240VAC
The Emergency Fixed Rate Pacing channel is powered from a 9V Lithium/Manganese Dioxide
1200mAh PP3 style battery. This battery has a 10 year shelf life and will power the emergency
circuit for a minimum of two hours of continuous operation if pacing was commenced with green
battery OK LED lit.
Note that if the SGU has external power disconnected, while switched on or in the backup mode
(green Power LED blinking), a short beep is emitted by the SGU every few seconds to alert staff to
switch unit off completely. This is helpful particularly in laboratories where equipment is on a trolley
and gets disconnected and stored between EPS cases.
7.3 System Interconnection
The EPS320 is always powered through an isolation transformer normally normally connects to a
patient through an external third party EP Recording System, such as the Bard Duo TM via stimulus
(and possibly trigger) output connectors and Ext ECG input cables. Figure 3 shows a typical
installation of the stimulator.
Please refer to the full Installation Procedure supplied in the Distributor Handbook,
InstallProcedure.doc for details of installation.
25