NF-100 NeoPIP™
Pre-Use Check
Quick Guide
Adjust the gas supply to 6 L/min Use test lung if available or occlude T-piece opening
To check or set maximum pressure:
a) Occlude PEEP valve
b) Turn PIP control several full turns clockwise
c) Adjust Maximum Pop-Off Pressure
Set the PIP and PEEP:
a) Occlude PEEP valve
b) Adjust PIP control to the desired PIP on the manometer
c) Turn PEEP control to the desired value on the manometer
Turn off the gas supply and remove the test lung from the patient circuit T-Piece. If the device fails to operate as intended, remove from use and contact NeoForce.
Resuscitation Instructions Before use verify PIP and PEEP settings on manometer display. Fit T-piece to neonatal resuscitation mask or endo-tracheal tube. Resuscitate by placing and removing the thumb or index finger over the PEEP valve port at the top of the T-Piece to allow inspiration and expiration at the desired breath rate.