NeoPIP Quick Guide Oct 2009

Quick Guide

1 Page

NF-100 NeoPIP™  Pre-Use Check  Quick Guide  Adjust the gas supply to 6 L/min Use test lung if available or occlude T-piece opening  To check or set maximum pressure:  a) Occlude PEEP valve  b) Turn PIP control several full turns clockwise  c) Adjust Maximum Pop-Off Pressure  Set the PIP and PEEP:  a) Occlude PEEP valve  b) Adjust PIP control to the desired PIP on the manometer  c) Turn PEEP control to the desired value on the manometer  Turn off the gas supply and remove the test lung from the patient circuit T-Piece. If the device fails to operate as intended, remove from use and contact NeoForce.  Resuscitation Instructions Before use verify PIP and PEEP settings on manometer display. Fit T-piece to neonatal resuscitation mask or endo-tracheal tube. Resuscitate by placing and removing the thumb or index finger over the PEEP valve port at the top of the T-Piece to allow inspiration and expiration at the desired breath rate.  Always refer to Operation Manual for complete use instructions NeoForce Group Inc 35 Commerce Drive Ivyland, Pa 18974 www.neoforcegroup.com T (215) 672-6800 F (215) 672-1123  Authorized Representative: Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands T (31) (0) 70 345-8570 F (31) (0) 70 346-7299  No: NeoPIP™ Quick Guide 20091015  0843