Novametrix
Novametrix Non Invasive Cardiac Output Monitors
Model 7300 NICO2 Non Invasive Cardiac Output Monitor Users Manual
Users Manual
100 Pages
Preview
Page 1
™
User’s Manual Non-Invasive Cardiac Output Monitor Model 7300
February 1, 2001 Catalog No. 9226-23-05
Copyright © 2001. Novametrix Medical Systems Inc. 5 Technology Drive, Wallingford, Connecticut, U.S.A. 06492. All rights reserved.
Thank you … Thank you for purchasing the NICO® Non-Invasive Cardiac Output monitor from Novametrix. NICO® measures cardiac output through respiratory gas analysis based on the well accepted Fick Principle, providing continuous and accurate display of cardiac output. The monitor also operates in Respiratory Mechanics-only mode, providing the clinician with a respiratory profile of the patient through a combination of capnography, airway flow and pressure, and pulse oximetry. We expect that you will find the application and use of NICO® extremely simple, making it easy to adopt this exciting technology into your clinical practice. NICO® can provide accurate cardiac output values without the need for invasive procedures, benefitting the patient, the clinician, and the health care system in general. We appreciate your patronage and look forward to developing a long-term relationship with you and your institution. Sincerely,
USA TOLL FREE 1-800-243-3444 PHONE 205-265-7701 FAX 205-284-0753 WORLD WIDE WEB: http://www.novametrix.com E-MAIL: Customer Service [email protected] Technical Service [email protected]
Rev. 05
NICO User’s Manual
iii
Introduction About this manual
This manual is written for clinical personnel using the Novametrix NICO® Non-Invasive Cardiac Output Monitor, Model 7300, and the sensors and accessories intended for use with the monitor. This document contains information which is proprietary and the property of Novametrix Medical Systems Inc., and may not be reproduced, stored in a retrieval system, translated, transcribed, or transmitted, in any form, or by any means, without the prior explicit written permission of Novametrix Medical Systems Inc. Novametrix reserves the right to change specifications without notice.
NICO® Monitor Technical Description
Per requirements of IEC 601-1, the NICO® monitor is classified as class II equipment, internally powered, with type BF applied part, and an enclosure protection rating of IPX0. The NICO® monitor is Year 2000 compliant. Transport/Storage: -10 to +55° C (14-131° F), 10-95% R.H. non-condensing Operating Conditions: 10 to +40° C (50 to 104° F), 10-90% R.H. non-condensing The NICO® monitor, Model 7300, contains no user serviceable parts. Refer servicing to qualified service personnel. A technical Service Manual is available for use by technical personnel.
Manufacturing Quality & Safety
The Novametrix Medical Systems Inc. manufacturing facility is certified to both ISO 9001 and EN46001 (MDD93/42/EEC Annex II). Novametrix’ products bear the “CE 0086” mark. The product is certified by Underwriter’s Laboratories (UL) to bear the UL mark; and tested by TÜV Rheinland to IEC 601-1/EN60601-1.
Declaration of Conformity with European Union Directive
The Authorized Representative for Novametrix equipment is:
Trademarks and Patents
CAPNOSTAT CO2 Sensor and NICO are registered trademarks (®); NICO2 and the stylized NICO2 with CO 2 shadow, NICO Sensor, NICO Loop and CObar (cardiac output confidence bar), SuperBright and Y-Sensor are trademarks (™) of Novametrix Medical Systems Inc. Other trademarks and registered trademarks are the property of their respective owners.
D.R.M. Green European Compliance Services Limited, Oakdene House, Oak Road, Watchfield Swindon, Wilts SN6 8TD United Kingdom
NICO ® and its sensors and accessories are covered by the following USA patents: 4,859,858, 4,859,859, 4,914,720, 5,146,092, 5,153,436, 5,190,058, 5,206,511, 5,251,121, 5,347,843, 5,369,277, 5,379,650, 5,398,680, 5,535,633, 5,616,923, 5,693,944, 5,789,660, 5,793,044, 5,820,550, 5,891,026, 5,999,834, 6,098,622, 6,126,610, D424,193, 6,179,784. Other patents pending.
Manual Revision History
iv
NICO User’s Manual
11-Mar-99 Release at Rev. 00. 23-Mar-99 Revision 01. 05-Oct-99 Revision 02. R-N677 20-Mar-00 Revision 03. R-N741 19-Oct-00 Revision 04, R-N807 01-Feb-01 Revision 05, R-N850
Rev. 05
Table of Contents Welcome to NICO®... 1 General Description... 1 Front Panel... 1 Rear Panel... 2 Symbols... 3 AC/Battery Operation... 4 NICO® Parameter List... 5 Principles of Operation... 8
Navigating in Cardiac Output mode... 10 Areas of the Display... 10 Navigating the Display System... 10 KNOB selectable Monitoring Screens... 11 MENU key Screen Displays... 13 DATA ENTRY key Screen Displays... 14
Navigating in Respiratory Mechanics mode... 16 Areas of the Display... 16 Navigating the Display System... 16 KNOB selectable Monitoring Screens... 17 MENU key Screen Displays... 18 DATA ENTRY key Screen Displays... 19
Safety... 21 Monitoring Cardiac Output... 24 Preparing for Use... 24 Begin NICO® monitoring... 25 Entering Patient Data... 28 Entering ABG Data... 29 C.O. Averaging... 31
Respiratory Monitoring... 32 Preparing for Use... 32 Begin Respiratory Monitoring... 34 Entering Patient Data... 34 Entering ABG Data... 36
Monitoring and Setup Screens... 39 Cardiac Output mode... 39 Respiratory Screens (Optional)... 41 Respiratory Mechanics mode... 46
Rev. 05
NICO User’s Manual
v
Notes on Patient Monitoring... 53 Automatic Purging... 53 Manual Purging... 53 Intra-Aortic Balloon Pump... 54 Configuration Menu... 55 Reference Handbooks... 55
Alerts... 56 NICO® Alert Priorities... 56 Responding to Alert Audio... 56 Parameter Limit Alerts... 57 No Respiration Alert... 58
NICO Sensor™... 59 Disposable NICO Sensors™... 59
CO2/Flow Sensors... 62 Choosing a CO2/Flow Sensor... 63 Connecting a CO2/Flow Sensor... 63
CAPNOSTAT® CO2 Sensor... 65 The CAPNOSTAT® CO2 Sensor... 65 CAPNOSTAT® Adapter Zero... 66
Pulse Oximetry Sensors... 67 Oximetry Sensors... 67 Finger Sensor... 68 Y-Sensor™... 68 Single Patient Use Sensors... 71
Messages... 73 Message Areas - Cardiac Output mode... 73 Message Areas - Respiratory Mechanics mode... 78
External Devices... 82 GE Medical Systems Solar® Interface... 82 Agilent Technologies VueLink Interface... 83 ASCII Output... 85 Analog Output... 86 Alerts and Messages... 87
Maintenance... 88 Cleaning and Sterilization... 88 Monitor Maintenance Schedules... 89 Battery Maintenance... 89
Specifications... 90 NICO® Accessories... 92 Warranty... 94
vi
NICO User’s Manual
Rev. 05
General Description
Welcome to NICO® General Description NICO ®, a Non-Invasive Cardiac Output monitor from Novametrix Medical Systems Inc., noninvasively measures and displays cardiac output (C.O.). The NICO® monitor, Model 7300, also displays cardiac index and stroke volume, as well as various respiratory monitoring parameters including CO 2 elimination (VCO2) and alveolar minute ventilation. In Respiratory Mechanics mode, NICO® can be used as a respiratory profile monitor, without cardiac output displayed. In either mode, NICO® provides the clinician with important information to aid in precise and efficient patient management.
Indications
The NICO ® monitor is indicated for use by technically skilled clinical personnel. In Cardiac Output mode, the monitor is used for the monitoring of cardiac output and various respiratory parameters of adult patients receiving mechanical ventilation. In Respiratory Mechanics mode, the NICO® monitor is used for monitoring the respiratory parameters of adult, pediatric and neonatal patients. NICO® is not intended for any other purpose.
Contraindications
In Cardiac Output mode, use of the NICO® monitor is contraindicated in patients in whom a small rise (3-5 mmHg, 0.4-0.67 kPa) in their PaCO2 level cannot be tolerated.
Front Panel The NICO ® monitor’s front panel includes a display screen, sensor input connectors, a control knob, and operational push button keys and indicators that are explained below.
7 8 9
10 11 12
13 1
Rev. 05
2
3
4
5
6
NICO User’s Manual
1
Rear Panel
1
Display Screen. The screen displays NICO® data, respiratory mechanics, trends, waveforms and messages, along with setup and configuration data.
2
CAPNOSTAT® CO2 Sensor Input Connector. Connect only a Novametrix CAPNOSTAT® CO2 Sensor, Catalog Number 9567-00 here.
3
Connector Isolation Icon. Identifies the connector to either side of this icon as a type BF patient isolation connection.
4
Pulse Oximetry Sensor Input Connector. Connect only Novametrix pulse oximetry sensors and extension cables approved for use with the NICO® monitor.
5
NICO Sensor™ Input Connector. Connect only Novametrix NICO Sensors™, Catalog Number 8950-00, 8951-00 and 8952-00 or Novametrix CO2/Flow sensors, Catalog Number 9765-00, 9766-00 and 9767-00.
6
KNOB. The KNOB is used to select monitoring screens, scroll through menus and to change or enter values. The KNOB is generally turned to access different monitoring screens and to highlight menu options, and pressed to accept or change those selections.
7
DATA ENTRY key. Press to activate the DATA ENTRY screen and illuminate the key’s green icon. Press the key again to return to the previously displayed screen. From the DATA ENTRY screen, you can enter patient information including height, weight and respiratory gas mixture, and access the ABG DATA ENTRY screen.
8
MENU key. Press to activate the SELECT A SCREEN menu and illuminate the key’s green icon. Press the key again to return to the previously displayed screen. From the SELECT A SCREEN menu you can, by turning the knob, highlight the screen you wish to display. Press the MENU key or the KNOB to display that selected screen.
9
SILENCE key. The SILENCE key is used to mute/prevent audible alerts. It also visually indicates the presence of a “High Priority Alert”. The Silence feature operates in two modes; a temporary “2 Minute Silence” mode and an “Audio Disabled” mode. • 2 Minute Silence - Press and release to activate or deactivate the two minute silence. The key’s icon illuminates amber when active and audible alerts will be muted for two minutes, after which the icon turns off and any audible alert will sound. • Audio Disabled - Press and hold for one second to prevent or allow any audible alerts. The key’s icon illuminates and flashes amber to indicate that all audible alerts are being suppressed. • High Priority Alerts - (See “NICO® Alert Priorities” on page 56) The SILENCE key’s icon illuminates and flashes red to indicate High Priority Alert is active. The icon alternately flashes red and amber if the audio is disabled and a High Priority Alert is active.
10
STOP/CONTINUE REBREATHING key. Press to start NICO® monitoring and the automatic rebreathing process. Subsequent presses will stop (amber indicator illuminated) or continue (amber indicator off) the rebreathing process. Press and hold for two seconds to reset the NICO® algorithm; the C.O. value and averaging filter will be cleared. The STOP/CONTINUE REBREATHING key will be amber and inactive in Respiratory Mechanics mode.
11
AC Mains Power Indicator. This icon illuminates green to indicate AC Mains power is applied to the monitor. To illuminate the icon, the monitor must be plugged into the AC outlet and the monitor’s rear panel power switch must be on (“|”).
12
OPERATE/STANDBY key. Press this key to turn the monitor on. If connected to the AC outlet, the monitor uses AC power, otherwise it powers up using its internal battery (provided the battery is charged). Press the OPERATE/STANDBY key again to put the monitor into Standby mode (if using AC power) or to turn it off (if using battery power).
13
Kickstand (front and rear). The NICO® monitor can be positioned for better viewing from above or below by extending the kickstand at the front or rear of the monitor.
Rear Panel The NICO ® monitor’s rear panel includes an AC Mains power input module, three RS232 serial communications ports, an analog input/output port, equipotential connector, fan and ventilation slots, and the monitor’s serial number label. These are explained below.
2
NICO User’s Manual
Rev. 05
Symbols
14
15
19
16
20
17
21
18
22
14
AC Mains Power Cord Connection. Connect only approved hospital-grade line cords to this connector.
15
AC Mains Power Switch. This switch controls the flow of AC current into the NICO® monitor. Press the “|” portion of the switch to supply the monitor with AC power, or the “O” portion of the switch to interrupt the flow of AC power. If supplied with AC power, the monitor illuminates the front panel AC Mains Power Indicator, energizes the fan and recharges the internal battery.
16
RS232 Communications Ports. Three 9-pin serial communications ports provide for digital communications with the NICO® monitor. (See “RS232 Communications” ).
17
Fan. The fan draws air in through the monitor. Do not block the fan’s air intake slots.
18
Serial Number Label. The NICO® monitor’s serial number is shown here. Refer servicing to qualified personnel.
19
Power cord retaining clip. If desired, remove the screw, slip the cord through the clip and insert and tighten the screw. Use only the supplied screw to secure the clip.
20
Equipotentiality. Connection to the monitor’s chassis (earth ground system).
21
Analog Input/Output Port. This 15-pin connector provides analog signal output capability for the NICO® monitor (Input reserved for future use).
22
Ventilation Slots. Do not block the air ventilation slots.
Symbols Th es e sy mb ols ma y b e f ound on t he N I CO ® mon it or, i ts se ns ors, a cc es sori es a nd documentation.
!
2
Rev. 05
Attention Consult manual for detailed information.
Patient Isolation Identifies the patient isolation connection as type BF.
Single Patient Use Treat in accordance with protocol for “single patient use” items.
AC Mains Power Switch “|” ON-connection to mains; “O” OFF-disconnection from mains
NICO User’s Manual
3
AC/Battery Operation
Mains Fuse Rating Mains rating for replacement fuses
Equipotentiality Connection to monitor’s chassis.
Separate collection Take appropriate steps to ensure that spent batteries are collected separately when disposed of. This symbol is found on the internal battery and the monitor enclosure.
Heavy Metal Content Indicates heavy metal content, specifically lead. This symbol is found on the internal battery and the monitor enclosure.
250V
Pb
Recyclable item This symbol is found on the internal battery and the monitor enclosure.
AC/Battery Operation The NICO ® monitor is designed to be operated from AC Mains power. An internal battery provides uninterrupted monitoring and trending capability for short periods (no more than 45 minutes) if the AC power is removed.
AC Mains Operation
To operate NICO® from AC Mains power:
1
Plug the line cord into the rear panel connector and the AC Mains power outlet.
2
Set the rear panel power switch to the On “|” position. • The front panel AC Mains Power indicator illuminates. • The monitor’s fan turns on. • The internal battery starts to recharge. Operate /Standby
3
Battery Operation
Press the front panel Operate/Standby key to turn the monitor on and off.
The NICO® monitor automatically switches to battery power from AC Mains power if the AC source is removed. A fully charged battery will power the monitor for up to 45 minutes. While on battery power, NICO® displays a battery icon that “drains” as power is consumed.
Battery icon shows capacity
full
half
depleted
The battery icon starts to flash when approximately 5 minutes of battery power remain. An audible alert tone also sounds. Re c on n e ct to A C M ai ns pow e r o r the m on it or w i ll automatically shut off. A depleted battery may require 12-16 hours to fully recharge. To operate NICO® on battery power:
1
Unplug the line cord or set the rear panel power switch to the Off “O” position. • The front panel AC Mains Power indicator turns off.
2
4
NICO User’s Manual
Press the front panel Operate/Standby key to turn the monitor on and off.
Rev. 05
NICO® Parameter List
NICO® Parameter List Cardiac Output mode
The NICO® monitor displays the parameters described in this table.
Label
Parameter
Range/Units
Description
Screen Display
C.O.
Cardiac Output
0.5-19.9 L/min
Volume of blood pumped by the heart each minute
All
CO-a
Average Cardiac Output
0.5-19.9 L/min
C.O. averaged value, displayed when the fast-mode cardiac output mode is chosen for large display.
Last Completed Cycle rebreathing curve & Tabular Data
CO-f
Fast-mode Cardiac Output
0.5-19.9 L/min
C.O. unaveraged value, displayed when the average cardiac output mode is chosen for large display.
Last Completed Cycle rebreathing curve & Tabular Data
Cdyn
Dynamic Compliance
0-500 ml/cmH2O
Volume the lungs expand for a given pressure
Numerics & Tabular Data
Note that if the ventilator is set for an inspiratory pause that is detected by NICO, Cdyn becomes Cstat. CI
Cardiac Index
0-9.9 L/min/m2
C.O. divided by body surface area
All
ETCO2
End Tidal Carbon Dioxide
0-150 mmHg 0-20.0 % 0-20.0 kPa
Maximum CO2 plateau value at the end of the breath (reflects alveolar CO2)
Numerics, Respiratory Numerics, CO2/SpO2, SBCO2, Tabular Data & rebreathing curves
Insp CO 2
Inspired Carbon Dioxide
3-50 mmHg 0.4-6.7 % 0.4-6.7 kPa
Maximum CO2 value observed during the baseline portion of the Inspiratory phase of the breath (baseline shift above zero point)
General Message Area (if above 3 mmHg for 10 sec (0.4 % or kPa))
MAP
Mean Airway Pressure
0-100 cmH2O
Mean (average) pressure in the airway throughout the breath
Numerics & Tabular Data
MV
Minute Volume
2-40 L/min Adult
Volume (in liters) of gas delivered to the patient per minute
Numerics, Respiratory Numerics & Tabular Data
MV alv
Alveolar Minute Volume
0.05-16 L/min
MV less deadspace (wasted) ventilation
Numerics, Respiratory Numerics & Tabular Data
PCBF
Pulmonary Capillary Blood Flow
0.5-19.9 L/min
Portion of the cardiac output that is effective in gas exchange
3 min Cycle in Progress rebreathing curve & Tabular Data
PeCO2/ FeCO2
Mixed expired CO2
0-100 mmHg, 0-13.2 kPa or %
Volume weighted average CO2 in the breath
Respiratory Numerics & Tabular Data
PEEP
Positive End Expiratory Pressure
0-99 cmH2O
Pressure in the lungs at the end of expiration
Numerics, Flow/ Pressure & Tabular Data
PEF
Peak Expiratory Flow
2-180 L/min
Highest absolute flow rate during expiration
Tabular Data
PIF
Peak Inspiratory Flow
2-180 L/min
Highest absolute flow rate during inspiration
Tabular Data
PIP
Peak Inspiratory Pressure
0-120 cmH2O
Peak (highest) pressure in the airway during inspiration
Numerics, Flow/ Pressure & Tabular Data
Rev. 05
NICO User’s Manual
5
NICO® Parameter List
Label
6
Parameter
Range/Units
Description
Screen Display
Pulse Rate
30-250 bpm
Number of pulse beats per minute
Numerics, CO2/SpO2 & Tabular Data
Raw
Airway Resistance
0-100 cmH2O/L/sec
Pressure required to cause gas flow at a given rate
Numerics & Tabular Data
RR
Respiration Rate
2-120 br/min
Number of breaths per minute
Numerics, CO2/SpO2 & Tabular Data
RSBI
Rapid Shallow Breathing Index
0-1000 br/min/L
Respiratory rate divided by average spontaneous tidal volume (only calculated when RR < 57)
Respiratory Numerics & Tabular Data
SpO2
Oxygen Saturation
0-100 %
Oxyhemoglobin as a percentage of total hemoglobin less dysfunctional hemoglobin
Numerics, CO2/SpO2, Tabular Data & rebreathing curves
SV
Stroke Volume
0-250 ml
Volume of blood pumped by the heart each beat
All
SVI
Stroke Volume Index
0-125 ml
Stroke volume divided by body surface area
Tabular Data
SVR
Systemic Vascular Resistance
0-5000 dynes sec/ cm5
Resistance exerted by the blood vessels on blood flow and is an indicator of left ventricular afterload.
SVR Calculation & Tabular Data
SVRI
Systemic Vascular Resistance Index
0-5000 dynes sec/ cm5
SVR normalized to body surface area
Tabular Data
VCO2
Carbon Dioxide Elimination
0-3000 ml/min
Volume of CO2 eliminated through the breath each minute
Numerics, Respiratory Numerics, SBCO2, Tabular Data & rebreathing curves
Vd Aw
Airway deadspace
0-500 ml
Includes added mechanical deadspace proximal to the flow sensor
Respiratory Numerics, SBCO2, Tabular Data
Vd/Vt
Deadspace to tidal volume ratio
0-1.00 ml
(PaCO2-PeCO2)/PaCO2
Respiratory Numerics & ABG Data Entry
Vd alv
Alveolar deadspace
0-500 ml
Difference between physiologic and airway deadspace
Respiratory Numerics & ABG Data Entry
Vt alv
Alveolar tidal volume
0-2400 ml
Tidal volume less airway deadspace
Respiratory Numerics, SBCO2 & Tabular Data
Vte
Expired Tidal Volume
200-3000 ml
Volume of gas exhaled per breath
Respiratory Numerics, Flow/Pressure & SBCO2
Vte-m
Expired Tidal Volume mechanical
200-3000 ml
Volume of mechanically exhaled gas, per breath
Tabular Data
Vte-s
Expired Tidal Volume spontaneous
200-3000 ml
Volume of spontaneously exhaled gas, per breath
Tabular Data
Vti
Inspired Tidal Volume
200-3000 ml
Volume of gas inhaled per breath
Respiratory Numerics, Flow/Pressure & SBCO2
Vti-m
Inspired Tidal Volume mechanical
200-3000 ml
Volume of mechanically inhaled gas, per breath
Tabular Data
Vti-s
Inspired Tidal Volume spontaneous
200-3000 ml
Volume of spontaneously inhaled gas, per breath
Tabular Data
NICO User’s Manual
Rev. 05
NICO® Parameter List
Respiratory Mechanics mode
The NICO® monitor displays the parameters described in this table.
Label
Parameter
Range/Units
Description
Screen Display
Cdyn
Dynamic Compliance
0-500 ml/cmH2O
Volume the lungs expand for a given pressure
Numerics & Tabular Data
ETCO2
End Tidal Carbon Dioxide
0-150 mmHg 0-20.0 % 0-20.0 kPa
Maximum CO2 plateau value at the end of the breath (reflects alveolar CO2)
All
I:E
I:E Ratio
1:9.9 or 4:1
Ratio of inspiratory time (ti) to expiratory time (te)
Flow/Pressure
MAP
Mean Airway Pressure
0-100 cmH2O
Mean (average) pressure in the airway throughout the breath
Numerics & Tabular Data
MV
Minute Volume
0.4-40 L/min adult 0.06-30 L/min pedi. 0.01-5 L/min neonatal
Volume (in liters) of gas delivered to the patient per minute
All except Flow/ Pressure & Loops
MV alv
Alveolar Minute Volume
0-16 L/min adult 0-8 L/min pediatric 0-4 L/min neonatal
MV less deadspace (wasted) ventilation
Numerics & Tabular Data
NIP
Negative Inspiratory Pressure
0 to -120 cmH2O
Maximum negative pressure during inspiratory cycle
Loops
PeCO2/ FeCO2
Mixed expired CO2
0-100 mmHg, 0-13.2 kPa or %
Volume weighted average CO2 in the breath
Numerics, SBCO2 & Tabular Data
PEEP
Positive End Expiratory Pressure
0-99 cmH2O
Pressure in the lungs at the end of expiration
Numerics, Flow/ Pressure & Tabular Data
PEF
Peak Expiratory Flow
2-180 L/min adult 0.5-100 L/min pedi. 0.25-25 L/min neo.
Highest absolute flow rate during expiration
Loops & Tabular Data
PIF
Peak Inspiratory Flow
2-180 L/min adult 0.5-100 L/min pedi. 0.25-25 L/min neo.
Highest absolute flow rate during inspiration
Loops & Tabular Data
PIP
Peak Inspiratory Pressure
0-120 cmH2O
Peak (highest) pressure in the airway during inspiration
Numerics, Flow/ Pressure & Tabular Data
Pulse Rate
30-250 bpm
Number of pulse beats per minute
All
Airway Resistance
0-100 cmH2O/L/sec adult/pediatric
Pressure required to cause gas flow at a given rate
Numerics & Tabular Data
Raw
0-500 cmH2O/L/sec neonatal RR
Respiration Rate
2-120 br/min adult 2-150 br/min pedi. 10-150 br/min neo.
Number of breaths per minute
All
RSBI
Rapid Shallow Breathing Index
0-1000 br/min/L (adult only)
Respiratory rate divided by average spontaneous tidal volume (only calculated when RR < 57)
Loops
SpO2
Oxygen Saturation
0-100 %
Oxyhemoglobin as a percentage of total hemoglobin less dysfunctional hemoglobin
All
Rev. 05
NICO User’s Manual
7
Principles of Operation
Label
Parameter
Range/Units
Description
Screen Display
VCO2
Carbon Dioxide Elimination
1-3000 ml/min adult/pediatric
Volume of CO2 eliminated through the breath each minute
All except Flow/ Pressure & Loops
0-300 ml/min neonatal Vd Aw
Airway deadspace
0-500 ml
Includes added mechanical deadspace proximal to the flow sensor
Numerics, SBCO2, Tabular Data
Vd/Vt
Deadspace to tidal volume ratio
0-1.00 ml
(PaCO2-PeCO2) / PaCO2
Numerics, SBCO2 & ABG Data Entry
Vd alv
Alveolar deadspace
0-500 ml
Difference between physiologic and airway deadspace
Numerics, SBCO2 & ABG Data Entry
Vt alv
Alveolar tidal volume
0-2400 ml adult 0-1200 ml pediatric 0-160 ml neonatal
Tidal volume less airway deadspace
Numerics, SBCO2 & Tabular Data
Vte
Expired Tidal Volume
200-3000 ml adult 30-400 ml pediatric 1-100 ml neonatal
Volume of gas exhaled per breath
All except Flow/ Pressure & Loops
Vte-m
Expired Tidal Volume mechanical
200-3000 ml adult 30-400 ml pediatric 1-100 ml neonatal
Volume of mechanically exhaled gas, per breath
Tabular Data
Vte-s
Expired Tidal Volume spontaneous
200-3000 ml adult 30-400 ml pediatric 1-100 ml neonatal
Volume of spontaneously exhaled gas, per breath
Tabular Data
Vti
Inspired Tidal Volume
200-3000 ml adult 30-400 ml pediatric 1-100 ml neonatal
Volume of gas inhaled per breath
All except Flow/ Pressure & Loops
Vti-m
Inspired Tidal Volume mechanical
200-3000 ml adult 30-400 ml pediatric 1-100 ml neonatal
Volume of mechanically inhaled gas, per breath
Tabular Data
Vti-s
Inspired Tidal Volume spontaneous
200-3000 ml adult 30-400 ml pediatric 1-100 ml neonatal
Volume of spontaneously inhaled gas, per breath
Tabular Data
Principles of Operation Non-Invasive Cardiac Output (NICO)
NICO® calculates cardiac output (C.O.) non-invasively based on respiratory gas analysis, using a technique known as “differential Fick partial rebreathing.” The key to this technique is a NICO Sensor™, consisting of a rebreathing valve and a combined CO 2/Flow sensor placed in the breathing circuit. The NICO Sensor™ is placed into the ventilator circuit between the patient elbow and ventilator wye. The rebreathing valve is automatically controlled by the monitor. When the valve is activated, the flow of the inspired and expired gas is diverted through a rebreathing NICO Loop.™ When the valve is deactivated, this additional rebreathing volume is bypassed and normal ventilation resumes. Every three minutes, a baseline, rebreathing and stabilization phase occurs. (See “The NICO® Cycle” on page 27.) A non-invasive cardiac output calculation is made following the end of each three minute cycle. The calculation is based on the changes induced in CO2 elimination and end tidal CO2 in response to the rebreathing volume. The increase in end tidal CO2, which reflects the increase in PaCO2, is usually 3-5 mmHg (0.40.67 kPa) and returns to baseline in less than 30 seconds. The Fick equation using CO 2 as an indicator states that cardiac output is equal to CO2 elimination divided by the venous-arterial difference in the CO2 content: VCO2/(CvCO2-CaCO2). The partial rebreathing method yields a differential form of the Fick equation, eliminating the need to measure mixed venous CO 2 (assumed constant during the rebreathing period and therefore cancels out of the equation). This indirect Fick method is then corrected for shunt,
8
NICO User’s Manual
Rev. 05
Principles of Operation
based on the Nunn’s iso-shunt curves using SpO2 (or entered PaO2) and a user-entered value for FiO2 (INSP O2).
Carbon Dioxide (CO2)
NICO ® uses the CAPNOSTAT® CO 2 Sensor to measure CO2 by using the infrared absorption technique. The principle is based on the fact that CO 2 molecules absorb infrared (IR) light energy of specific wavelengths, with the amount of energy absorbed being directly related to the CO 2 concentration. When an IR beam is passed through a gas sample containing CO2 , the electronic signal from the photodetector (which measures the remaining light energy) can be obtained. This signal is then compared to the energy of the IR source and calibrated to accurately reflect CO2 concentration in the sample. To calibrate, the photodetector’s response to a known concentration of CO2 is stored at the factory in the monitor’s memory. A reference channel accounts for optical changes in the sensor, allowing the system to remain in calibration without user intervention.
Flow and Pressure
Flow and pressure measurements in the NICO® monitor are made by a fixed orifice differential pressure pneumotachometer. Respired gas flowing through the flow sensor causes a small pressure drop across the two tubes connected to the sensor. This pressure drop is transmitted through the tubing to a differential pressure transducer located inside the monitor, and is correlated to flow according to the factory stored calibration. User calibration is not required due to the ability of the plastic injection mold to repeatedly produce precision flow sensors. The pressure transducer is automatically “zeroed” to correct for changes in ambient temperature and electronics. The NICO ® monitor system software compensations allow accurate flow and volume measurements in the presence of high oxygen concentrations, anesthetic gases and helium-oxygen mixtures. When compensated, gas density and viscosity effects do not cause significant errors in flow measurement.
Carbon Dioxide Elimination (VCO2)
Carbon dioxide elimination (VCO2) is a key measurement for NICO® calculations. It is calculated based on a mathematical integration of the measured flow and CO2 signals. These signals are obtained from practically the same point at the patient’s airway, thereby insuring optimal accuracy. Both the flow and CO2 sensors are integral components of the NICO Sensor™.
Oxygen Saturation (SpO2) & Pulse Rate
Oxygen saturation (SpO2) is used by the NICO® monitor to calculate the shunt correction of the NICO® calculation, and the pulse rate is used to calculate stroke volume. SpO2 is determined using sensors containing red and infrared light emitting diodes (LEDs). The light from each LED is beamed through a pulsating vascular bed such as the patient’s finger or toe. The remaining light not absorbed by the tissue reaches a photodiode light receptor in the sensor. Oxygen saturated blood absorbs different amounts of light at each wavelength as compared to unsaturated blood. Therefore, the amount of light absorbed by the blood in each pulse can be used to calculate oxygen saturation. NICO ® is calibrated to display “functional” saturation. This differs from the “fractional” saturation value displayed by most co-oximeters. Functional saturation represents the amount of oxyhemoglobin as a percentage of the hemoglobin that can be oxygenated. Dysfunctional hemoglobins, (COHb and METHb) are not included in the measurement of functional saturation. • Functional Saturation = HbO2/100-(COHb+METHb); HbO2 is fractional hemoglobin, COHb is carboxyhemoglobin, and METHb is methemoglobin. Pulse Rate, derived from the pulse oximetery sensor, is calculated by measuring the time interval between the peaks of the infrared light waveform. The inverse of this measurement is displayed as pulse rate.
Rev. 05
NICO User’s Manual
9
Areas of the Display
Navigating in Cardiac Output mode Areas of the Display The major sections of the Cardiac Output mode screen are identified below. Stroke Volume (SV) and Cardiac Index (CI) are displayed in all views. A battery icon also appears if on battery power.
Cardiac Output (C.O.) value is displayed in all views. The CObar™ confidence indicator is replaced by FAST MODE when in fast mode.
A General Message area for status, alert and error information. Displayed in all views (here shown blank). A Cardiac Output Message area for C.O. information is displayed in all views. The Rebreathing Bar is displayed during the rebreathing portion of the cycle.
The lower half of the display presents trend, waveform, respiratory and numeric data to the user. Various data entry, setup and alert menus are also presented here. Use the KNOB and the MENU and DATA ENTRY keys to select the various displays.
Navigating the Display System Use the KNOB, MENU, and DATA ENTRY keys to navigate the NICO® display system (as outlined in the following sections).
10
NICO User’s Manual
Rev. 05
KNOB selectable Monitoring Screens
KNOB selectable Monitoring Screens The KNOB is used to page through monitoring screens, scroll through menus and make selections, and to change or enter values. The KNOB is generally turned to access different monitoring screens and to highlight menu options, and pressed to accept or change selections. C.O. Trend Screen
TURN
PUSH
Rebreathing Curves Screen
PUSH
CO2/SpO2 Screen (No Options)
Flow/Pressure Screen (No Options)
Numerics Screen (No Options)
Rev. 05
NICO User’s Manual
11
KNOB selectable Monitoring Screens
KNOB Selectable Respiratory Screens
The following Respiratory Screens are available in the monitoring mode only when activated. Press the MENU key and select RESP SCREENS from the SELECT A SCREEN menu by turning and then pressing the KNOB. From the ADD RESPIRATORY SCREENS menu, highlight and then select which screens will appear in the base monitoring mode by turning and then pressing the KNOB. When enabled, selected screens can be displayed by turning the KNOB while viewing any monitoring screen.
Respiratory Numerics TURN
Flow Volume Loop and Pressure Volume Loop
Single Breath CO2 Waveform
VCO2/MValv Trend
12
NICO User’s Manual
Rev. 05
MENU key Screen Displays
MENU key Screen Displays Press the MENU key to activate the SELECT A SCREEN menu and illuminate the key’s green icon. Press the key again to return to the previously displayed screen. From the SELECT A SCREEN menu turn the KNOB to highlight the screen you wish to display. Press the MENU key or the KNOB to display that selected screen.
1 PUSH
3
PUSH
2 TURN
Rev. 05
NICO User’s Manual
13
DATA ENTRY key Screen Displays
DATA ENTRY key Screen Displays Press the DATA ENTRY key to activate the DATA ENTRY screen and illuminate the key’s green icon. Press the key again to return to the previously displayed screen. From the DATA ENTRY screen, you can enter patient information including height, weight and respiratory gas mixture, and access the ABG DATA ENTRY screen. (See “Entering Patient Data” on page 28 for details.)
1 PUSH
2
3 TURN
PUSH
4
5 TURN TO ADJUST
14
NICO User’s Manual
PUSH TO ACCEPT
Rev. 05
DATA ENTRY key Screen Displays
ABG Data Entry Screens
From the DATA ENTRY screen, select ABG DATA ENTRY. Turn and press the KNOB to enter PaCO2, PaO2, and Hemoglobin entry screens. (See “Entering Patient Data” on page 28 for details.)
1
PUSH
3
2 TURN
PUSH
4
5 TURN TO SELECT
Rev. 05
PUSH TO ACCEPT
NICO User’s Manual
15