EX series special type heads Reprocessing Instruction
2 Pages

Preview
Page 1
Reprocessing Instruction in accordance with ISO17664
EX series special type heads 1 Preparation 20sec.
Cleaning Wipes
2 Cleaning, Disinfection and Drying 38°C
30sec.
3.5L/min
30sec.
30sec.
30sec.
38°C
3.5L/min
Manual 60sec.
1-3min
38°C
Neutral enzymatic detergent for medical purposes
NOTICE It is recommended to cover the ultrasonic bath during operation.
500lx
WL-Adapter 02/G (ALPRO)
WL-Adapter U-1 + U-1a (ALPRO)
WL-clean (ALPRO)
Minuten Wipes (ALPRO)
WL-Adapter 02/G (ALPRO)
WL-Adapter 02/G (ALPRO)
WL-dry (ALPRO) WL-cid WL-Adapter (ALPRO) U-1 + U-1a (ALPRO)
WL-Adapter U-1 + U-1a (ALPRO)
WL-Blow (ALPRO)
Automated ISO 15883-1 500lx Minuten Wipes (ALPRO)
・Automated cleaning should only be performed on products that are Washer-disinfector safe. Refer to the products Operation Manual to check compatibility. ・After cleaning remove any remaining moisture using a dry cloth or compressed air ensuring it is dry internally and externally. ・Cover the product with a cloth when drying with compressed air to prevent water aerosol from being produced. ・Use a nationally approved disinfectant with proven bactericidal, fungicidal, and virucidal properties.
3 Lubrication Manual cleaning and lubrication REF
CHUCK (Only for FG Bur type)
Z020010
INTERIOR
REF REF
Z019090
2~3sec.
Y900084
NSKPANASPRAYPlus
REF
Z182100
Automated lubrication CHUCK (Only for FG Bur type)
INTERIOR
NSK iCare
NSK iCare+ iCare Care3 Plus
4 Packaging, Sterilizing and Drying
ISO 11607-1 Use a sterilization pouch that conforms to ISO 11607-1. Seal the pouch. Maximum temperature for NSK instruments is 135°C.
Type
Class N
Class B
Temperature
132°C
134°C
FullCycleTime
≥15min
3-18min
DryingTime
≥30min
≥30min
Steam sterilization
CAUTION
Perform steam sterilization using the conditions above. Use a steam sterilizer to perform sterilization.
Plasma
EO
5 Storage
OM-DC0915EN 000